Austria Medical Device Registration2026 Field Guide

No Austrian CE mark — EU MDR applies directly. Austria runs two registers in parallel: EUDAMED, validated by the BASG, and the GÖG Medical Devices Register (MPR), where Austrian distributors and custom-made manufacturers must register before they start. Add a German-language rule and an annual levy that also catches foreign direct sellers.

registers in parallel: EUDAMED + MPR
2
annual levy declaration deadline
30 June
of GDP spent on health (2024)
11.7%

Sourced from the BASG, GÖG, MPG 2021 and Statistics Austria, and verified 2026 publications.

Download the guide

Enter your work email and we'll send the PDF straight to your inbox.

Professional review is required before business use. See our privacy policy.

What's inside

Nineteen pages, zero fluff

Every figure in this guide traces to a named source — the BASG, GÖG, MPG 2021, Statistics Austria, and verified 2026 publications.

EU MDR + the Austrian layer

MPG 2021 frames the national layer: the BASG is the competent authority for surveillance, vigilance and classification decisions, with AGES as its operating arm — and under the Act's sanction system even a deliberate attempt is punishable.

EUDAMED + the GÖG register

Actor registration in EUDAMED is mandatory since 28 May 2026 — but distributors and custom-made device manufacturers established in Austria must also register in the GÖG-run MPR before commencing activity. One registration does not replace the other.

German language + the annual levy

Accompanying information must be in German; English suffices only for exclusively professional users. And anyone supplying devices to Austrian end users — wherever based — files a levy declaration with the BASG by 30 June each year.

Complete guide

The guide in full

Every section of the PDF on one page. Use the form above to keep a copy.

Overview

Last updated: November 2026. The EU Medical Device Regulation (MDR 2017/745) and the In Vitro Diagnostic Regulation (IVDR 2017/746) apply directly in Austria. The Austrian national layer consists mainly of the Medical Devices Act 2021 (Medizinproduktegesetz 2021 — MPG 2021, BGBl. I No. 122/2021, as last amended by BGBl. I No. 102/2024), the Medical Device Notification Ordinance 2024 (Medizinproduktemeldeverordnung 2024, BGBl. II No. 23/2024, in force 1 May 2024), the Medical Device Operator Ordinance (Medizinproduktebetreiberverordnung — MPBV), the Medical Device Levy Ordinance (Medizinprodukteabgabeverordnung), the BASG Fee Regulation (Gebührenverordnung) and the Electronic Submission Ordinance 2011 (EEVO 2011). Always confirm critical steps against current BASG and Gesundheit Österreich GmbH (GÖG) guidance before filing.

Austria is a small market with outsized purchasing power: a resident population of roughly 9.1–9.2 million, and current health expenditure of EUR 57.8 billion in 2024, equal to 11.7% of GDP — fourth among the 38 OECD countries, with the public sector accounting for EUR 43.8 billion (75.7%) of that spending (source: Statistics Austria press release 14 097-016/26, 29 January 2026). Public expenditure on State Health Fund (Landesgesundheitsfonds) hospitals alone was about EUR 17.7 billion in 2024. The hospital landscape is concentrated in a handful of public operating groups — above all the Wiener Gesundheitsverbund in Vienna — plus three university hospitals (Vienna, Graz, Innsbruck) and a meaningful private sector. For a foreign manufacturer, Austria is rarely a first EU market, but it is a demanding reference market: German-language documentation, formal procurement, and a regulator that reads the file.

Regulatory reality check up front: Austria is not a separate certification regime from the EU. There is no "Austrian CE mark" and the Federal Office for Safety in Health Care (Bundesamt für Sicherheit im Gesundheitswesen — BASG) does not approve medical devices. The BASG states this expressly in its own FAQ: the manufacturer carries out the conformity assessment procedure on its own responsibility (involving a notified body for higher risk classes), and the BASG is responsible for market surveillance and vigilance after devices are made available. What Austria adds — and what foreign manufacturers most often underestimate — is a national layer with six specific features:

  1. 1. Two registration systems running in parallel — EUDAMED actor and device registration (mandatory since 28 May 2026, validated for Austrian-established operators by the BASG) plus the Austrian Medical Devices Register (Österreichisches Medizinprodukteregister, MPR) run by Gesundheit Österreich GmbH (GÖG), in which manufacturers of custom-made devices and distributors (Händler) established in Austria must register before starting their activity under Section 37 MPG 2021 and the Medizinproduktemeldeverordnung 2024. One registration does not replace the other for the categories the MPR covers.
  2. 2. A German-language rule with a professional-user carve-out — under Section 7 MPG 2021, devices may be supplied to users and patients only if the accompanying information is in German; if use is limited exclusively to professional users, English is sufficient. Field Safety Notices must be in German, and an EU Declaration of Conformity must be provided in German on request. This is more permissive than Sweden's rule for professional devices, and stricter than many manufacturers assume for patient-facing products.
  3. 3. An annual medical-device levy (Medizinprodukteabgabe) that is easy to miss — every natural or legal person, regardless of where their business is based, who supplies medical devices to end users in Austria must file a levy declaration for the previous year with the BASG by 30 June of the following year and pay the flat-rate levy under the Medical Device Levy Ordinance. It is a flat amount by highest device category supplied — not a turnover tax — and it catches foreign direct sellers as well as Austrian distributors.
  4. 4. An Austrian notified body that now covers both regulations — QMD Services GmbH (Notified Body 2962), a Quality Austria subsidiary based in Vienna and Linz, designated under the IVDR on 23 December 2022 and under the MDR on 14 May 2024 — one of only eleven bodies in the EU covering both regulations at the time of its MDR designation. Using it is a choice, not a requirement; any EU notified body may be used.
  5. 5. A wide surveillance net — MPG 2021 monitoring and serious-incident reporting duties extend beyond manufacturers and hospitals to persons and businesses that professionally use devices even occasionally, expressly including businesses such as beauty salons and fitness centres that use devices within the MDR's scope (for example liposuction/lipolysis equipment). The catalogue of sanctionable offences is extensive, and under MPG 2021 even a deliberate attempt to commit an offence is punishable.
  6. 6. No standardised device reimbursement — there is no single Austrian decision that "lists" a device for payment. Inside hospitals, devices are funded inside the LKF activity-based hospital financing system (Austria's DRG-type model); outside hospitals, reimbursement of aids and appliances depends on an arrangement with the individual social-insurance carrier involved, and patients generally contribute a co-payment. Device prices themselves are not statutorily regulated and can be freely set by the manufacturer — the commercial gate is procurement and the LKF / social-insurance logic, not a price approval.

A note on institutions, because Austria splits responsibilities in a way that confuses first-time entrants: the BASG is the competent authority designated in Section 2 MPG 2021 — market surveillance, vigilance, clinical investigations, derogations, qualification/classification decisions, notified-body designation and monitoring, and validation of Austrian EUDAMED actor registrations. The Austrian Agency for Health and Food Safety (AGES) — in practice its medical-market surveillance arm (AGES Medizinmarktaufsicht / AGES MEA) — is the BASG's operating/auxiliary body that effectively interacts with market participants. The BASG sits under the federal ministry responsible for health (Bundesministerium für Soziales, Gesundheit, Pflege und Konsumentenschutz — BMSGPK). GÖG (Gesundheit Österreich GmbH), the national public-health institute wholly owned by the Republic, runs the Austrian Medical Devices Register and also maintains implant registers (including pacemaker, implantable defibrillator and loop-recorder registers, and registers for other implant categories). The ministry itself keeps framework competence, including the annual LKF catalogue process described in Section 9.

1. The Regulatory Framework: EU MDR + Austrian National Layer

1.1 The EU layer (directly applicable)

  • Regulation (EU) 2017/745 (MDR) — applicable since 26 May 2021: classification (Annex VIII), General Safety and Performance Requirements (Annex I), conformity assessment (Article 52), CE marking, UDI (Articles 27–28), EUDAMED registration (Articles 29–31), clinical evaluation and investigations (Articles 61–82), post-market surveillance and vigilance (Articles 83–92), the prohibition of misleading claims (Article 7), and the supply-interruption notification duty in Article 10a.
  • Regulation (EU) 2017/746 (IVDR) — applicable since 26 May 2022, with staggered transition for legacy IVDs.
  • Regulation (EU) 2023/607, as amended by Regulation (EU) 2024/1860 — extended MDR/IVDR transition for legacy devices (to end-2027 / end-2028 for devices, and end-2027 to end-2029 by IVD class), subject to the conditions in Article 120 MDR / Article 110 IVDR, including a formal application to a notified body and a signed written agreement within the deadlines those regulations set.
  • EUDAMED mandatory modules — since 28 May 2026, use of the EUDAMED Actor, Device (UDI/Device) and Certificate modules is mandatory. Devices placed on the market before 28 May 2026 that continue to be placed on the market must complete device registration in EUDAMED by 28 November 2026 at the latest, and notified bodies must have all certificates registered in EUDAMED by 28 May 2027 (source: BASG, "Registration as actor in EUDAMED", basg.gv.at, crawled 1 November 2026).

1.2 The Austrian layer

InstrumentWhat it does
Medizinproduktegesetz 2021 (MPG 2021), BGBl. I No. 122/2021, as amended (last amendment cited in the consolidated text: BGBl. I No. 102/2024)The main Austrian statute supplementing the MDR/IVDR: designates the BASG as competent authority (Section 2), the German-language rule (Section 7), in-house devices (Section 9), qualification/classification applications (Section 10), clinical investigations and ethics committees, the GÖG registration duty for custom-made manufacturers and distributors (Section 37), reporting duties (Section 40), medical device consultants (Medizinprodukteberater), operator/use rules, advertising (Sections 70–77) and the sanction system
Medizinproduktemeldeverordnung 2024 (Medical Device Notification Ordinance 2024), BGBl. II No. 23/2024, in force 1 May 2024Re-regulates the notification duties of custom-made manufacturers, distributors and testing/inspection/certification bodies (PIZ) to GÖG; product categories are now recorded using the European Medical Device Nomenclature (EMDN); registered data (except personal data) are published in the MPR public search
Medizinproduktebetreiberverordnung (MPBV) (Medical Device Operator Ordinance)Rules for setting up, operating, using and maintaining devices in healthcare institutions: incoming inspection before first use, recurring safety inspections, metrological controls, the device file (Gerätedatei), and the role of the hospital technical safety officer (Technischer Sicherheitsbeauftragter)
Medizinprodukteabgabeverordnung (Medical Device Levy Ordinance / "Medical Device Tax Ordinance" in BASG's English rendering)The annual flat-rate levy on persons supplying devices to end users in Austria — declaration and payment to the BASG by 30 June of the following year (Section 8)
Medizinprodukte-Abgabeverordnung / dispensing rulesWhich devices may be dispensed by whom (pharmacies, specialist trade, direct supply) — relevant for consumer-channel products
BASG Gebührenverordnung (Fee Regulation)Fees for BASG official acts — for devices, see in particular Section X of the Annex (qualification/classification and related device procedures), plus the clinical-investigation sections (Section 8)
Elektronische Einreichverordnung 2011 (EEVO 2011)Applications, notifications and reports for clinical investigations / performance studies to the BASG must be submitted exclusively electronically
Bundesvergabegesetz 2018 (BVergG 2018)Federal Public Procurement Act — governs how public hospitals and hospital groups buy devices (Section 9)
Gesetz gegen den unlauteren Wettbewerb (UWG)General unfair-competition law, applicable cumulatively alongside MPG 2021 advertising rules
Allgemeines Sozialversicherungsgesetz (ASVG) and carrier statutesFramework for social-insurance benefits outside hospitals, including aids/appliances (Heilbehelfe/Hilfsmittel) supplied through arrangements with the carriers (Section 9)

1.3 Who does what

  • Manufacturer — full MDR Article 10 duties. If established in Austria: EUDAMED actor registration validated by the BASG, and — if it makes custom-made devices — additional registration in the GÖG register (MPR) before starting activity.
  • Authorised representative (AR) — mandatory for non-EU manufacturers (MDR Article 11). May be established anywhere in the EU; Austria imposes no rule that the AR be Austrian. An Austria-based manufacturer or AR is the only applicant who may request a BASG qualification/classification decision under Section 10 MPG 2021.
  • Importer — MDR Article 13 duties; EUDAMED actor registration (mandatory from 28 May 2026). The importer verifies CE marking, the EU Declaration of Conformity, the manufacturer's identity and AR designation, labelling and UDI, and that the device is registered in EUDAMED.
  • Distributor (Händler) — MDR Article 14 duties, plus an Austrian speciality: distributors established in Austria must register with GÖG in the MPR before commencing their activity (Section 37 MPG 2021, Medizinproduktemeldeverordnung 2024), declaring firm name/address and contact, type of activity and product categories. Exemptions exist for defined low-risk trade under Section 37(2) MPG 2021 / the ordinance on freely tradable devices.
  • Custom-made device manufacturer — MDR Annex XIII statement per device (no CE marking of the device itself in the ordinary route), EUDAMED actor considerations for Class III custom-made implantables, and MPR registration with GÖG including the name/contact of the Person Responsible for Regulatory Compliance (PRRC).
  • System/procedure pack producer (SPPP) — EUDAMED actor registration; a non-EU SPPP whose packs are first placed on the Union market in Austria is validated by the BASG.
  • Healthcare institutions and professional users — MPBV operator duties, and Section 40 MPG 2021 reporting duties (Section 10). In hospitals, serious-incident reports go uniformly through the medical director (ärztlicher Leiter), except in imminent danger.
  • Testing, inspection and certification bodies (PIZ) — also subject to MPR notification under the 2024 ordinance.

2. Device Classification

Austria applies the MDR classification rules (Annex VIII) unchanged — there is no Austrian reclassification layer.

ClassExamplesConformity route in brief
I (incl. Is sterile, Im measuring, Ir reusable surgical)Non-invasive low-risk devices, dressings, examination glovesSelf-declaration for plain Class I; notified body for the sterile/measuring/reusable aspects of Is/Im/Ir
IIaDental materials, short-term invasive devices, many diagnostic and monitoring devicesNotified body (Annex IX or Annex X + XI)
IIbLong-term invasive devices, most implants not in Class III, active therapeutic devices with potentially hazardous energy exchangeNotified body, fuller technical-documentation review
IIIAbsorbable implants, devices incorporating a medicinal substance, heart valves, breast implants, surgical meshes in defined usesNotified body with design examination; expert-panel / clinical-evaluation consultation for defined categories
IVDs (IVDR)Classes A–D under IVDR Annex VIIIA: self-declare (notified body for sterile A); B–D: notified body, D with the heaviest scrutiny, including EU reference-laboratory involvement where applicable

Software is classified under MDR Rule 11 as in the rest of the EU. Two Austrian practical notes:

  1. 1. No separate Austrian SaMD / telemedicine code yet. Professional commentary on Austria consistently notes that neither telemedicine devices nor software/AI medical devices are comprehensively regulated by a dedicated Austrian regime — the MDR analysis (and, for AI, the EU AI Act layering onto MDR conformity for high-risk AI) is the operative test. Do not assume an Austrian "digital health fast track" exists for devices; Austria's DiGA-style reimbursement pathway does not exist for medical-device software.
  2. 2. Borderline and classification help exists — but only from inside Austria. The manufacturer is responsible for correct qualification and classification. Under Section 10 MPG 2021, a manufacturer or authorised representative legally based in Austria may apply to the BASG for a formal qualification/classification decision (form F_I235; fees per Section X of the BASG Fee Regulation). A foreign manufacturer without an Austrian-established AR cannot file this application directly — structure the AR choice accordingly if a borderline question (device vs. medicinal product, device vs. cosmetic/biocide, Annex XVI product without a medical purpose) is genuinely open. For the medicinal-product borderline, the BASG also runs a delineation procedure under Section 1(3b) of the Medicines Act (AMG), with an expert opinion and a formal decision; an informal email enquiry is possible but produces no formal decision and is charged at an hourly rate under the Fee Regulation.

Misclassification risk sits with the manufacturer: a device wrongly self-declared as Class I is a non-compliant device on the market, exposed to BASG market-surveillance measures and the MPG 2021 sanction system.

3. The EU Authorized Representative, Importer and Distributor in Austria

A non-EU manufacturer (for example a Turkish, US, UK or Chinese maker) selling into Austria needs the standard MDR chain, with Austrian specifics at each link:

  1. 1. EU Authorised Representative — appointed by written mandate (MDR Article 11), established anywhere in the Union. The AR verifies the EU Declaration of Conformity and technical documentation, keeps documentation available for the BASG, cooperates on corrective actions, and is jointly liable for defective devices on the same basis as the manufacturer. Austrian choice-of-AR consequences: an Austria-based AR (a) is validated by the BASG in EUDAMED, (b) can file Section 10 qualification/classification applications, and (c) can request a Certificate of Free Sale from the BASG for CE-marked devices exported outside the EU — manufacturers/ARs based in Austria request the CFS from the BASG (BASG guidance). If your export markets ask for an Austrian/EU free-sale certificate, an Austrian AR is operationally useful.
  2. 2. Importer — the first EU operator making a third-country device available. EUDAMED-registered, with Article 13 verification, storage/transport and complaint-register duties. The importer indicates its name and address on the device, its packaging or an accompanying document. If the importer supplies devices directly to end users in Austria, analyse the medical-device levy (Section 8) — the levy follows the supply to the end user, not the manufacturer's seat.
  3. 3. Distributor — Article 14 verification and traceability duties, plus the Austrian GÖG registration duty for Austria-established distributors (Section 1.3). Put the registration in the distribution agreement as a condition precedent to first supply: an unregistered Austrian distributor is a defect in your Austrian market file and in theirs, and hospital customers' supplier checks can surface it. A distributor that translates, relabels or repackages also steps into MDR Article 16 duties (with notified-body certification of that activity where required) — document who does the German labelling (Section 6) and in which legal capacity.
  4. 4. Medical device consultants (Medizinprodukteberater) — MPG 2021 regulates the persons who professionally inform healthcare professionals about devices or instruct in their use (the Austrian analogue of the German Medizinprodukteberater). Sales representatives visiting Austrian hospitals must meet the statutory expertise/reliability requirements, and the company must be able to evidence this. Field teams hired for Austria need this checked before the first hospital visit, not after a complaint.
  5. 5. Trade licence (Gewerbeberechtigung) — trading in medical devices in Austria is a regulated commercial activity in practice: registration documentation for the Austrian register has historically required product data and evidence of the professional/trade licence. An Austrian distributor should hold the appropriate trade licence for medical-device trade; verify the current WKO (Austrian Federal Economic Chamber) position for your exact activity before appointing.

PRRC: the Person Responsible for Regulatory Compliance (MDR Article 15) requirement is EU-wide; micro and small enterprises may use an external PRRC permanently and continuously at their disposal. For GÖG registration, custom-made manufacturers must name the PRRC (name, telephone, email) — have the PRRC contact data ready before filing.

4. The Austrian Registration Architecture: EUDAMED + GÖG Register — Two Systems, Not One

4.1 Register in EUDAMED (mandatory from 28 May 2026)

Per BASG guidance (crawled 1 November 2026):

  • Actors obliged to register before placing on the market: manufacturers, authorised representatives, importers and system/procedure pack producers — for all devices other than custom-made devices (with the Class III custom-made implantable exception below).
  • Austrian validation: for economic operators established in Austria (and non-EU SPPPs first placing packs on the Union market in Austria), the BASG checks the registration request for completeness and correctness and validates it. Incomplete requests trigger a formal improvement request under Article 13(3) of the General Administrative Procedure Act (AVG); if the missing information is not submitted in time, the request is formally rejected and a new request must be filed. BASG processing is in order of receipt and timing "cannot be predicted" — the BASG expressly recommends early registration and warns that a late application is the operator's own risk.
  • SRN / Actor ID: manufacturers, ARs and importers receive a Single Registration Number (SRN); other actors receive an Actor ID. The SRN must appear on the EU Declaration of Conformity and in notified-body conformity-assessment documents. Device registration is only possible with an SRN/Actor ID.
  • Devices: register devices in the UDI/Device module before placing on the market. Devices already on the market before 28 May 2026 and continuing afterwards must be registered by 28 November 2026. Certificates issued after 28 May 2026 are entered by the notified body; all certificates must be in EUDAMED by 28 May 2027 — which is why manufacturers must be actor-registered in time for their notified body to enter certificate data.
  • Maintenance: update data within one week of a change (Article 31(4) MDR / Article 28(4) IVDR); confirm accuracy no later than one year after registration and every two years thereafter.
  • Class III custom-made implantable manufacturers must actor-register in EUDAMED so the notified body can enter certificate details — in addition to GÖG registration below.

4.2 Register with GÖG in the Austrian Medical Devices Register (MPR)

  • Who: (a) manufacturers of custom-made devices and (b) distributors of medical devices/IVDs — in each case before commencing the activity (Section 37 MPG 2021) — and (c) testing/inspection/certification bodies (PIZ).
  • Where: the register is run by Gesundheit Österreich GmbH and is accessible at medizinprodukteregister.at (application portal and public search at medizinprodukteregister.at/registersuche). Contact for registration questions: medizinprodukte@goeg.at (GÖG information letter, May 2024).
  • What a custom-made manufacturer declares: firm name or name, address, contact person with telephone/email and website if any; name, telephone and email of the PRRC; type of activity; product name/trade name or, if unavailable, generic name; and product code under the EMDN nomenclature (Medizinproduktemeldeverordnung 2024, Section 3(1)).
  • What a distributor declares: firm name or name, address, contact person with telephone/email and website if any; type of activity; and product categories (Section 3(2)).
  • Publication: since the 2024 ordinance, the declared data — except personal data — are published and searchable in the public register search. Assume hospital procurement teams and competitors can and do look you up.
  • Transition already run: operators newly registering from 1 May 2024 file online under the new ordinance; operators already registered had to supplement missing data by 30 April 2025. A registration taken out under the old ordinance that was never supplemented is a live compliance gap — check an existing Austrian partner's entry, do not assume it is current.
  • Fees: registration in the Austrian register has historically been free of charge as an online database (GÖG/legacy register guidance and secondary registration summaries). The Austrian state cost for ordinary market access is not the register — it is the BASG levy and the case fees in Section 8.
  • Implant registers: separately, GÖG sets up and maintains registers for pacemakers, implantable defibrillators and loop recorders, and implant registers for active implantable devices and for soft-tissue, cardiovascular, neurological and orthopaedic implants. If your device is an implant in a registered category, ask at launch what register data flows the Austrian hospitals will expect from you — this is a post-market data duty in practice, discovered late by many implant makers.

4.3 What registration is not

Registration — in EUDAMED or in the MPR — is a notification/surveillance act. It is not an assessment, not an approval, and not evidence that the BASG or GÖG has reviewed the device. The BASG does not approve devices at all (Section 1). Do not write "approved by BASG" in Austrian marketing material: besides being untrue, it collides with MDR Article 7 (misleading claims) and MPG 2021 advertising rules (Section 9.4).

5. Step-by-Step: Placing a Medical Device on the Austrian Market

Step 1 — Confirm the product is a medical device (or IVD) and classify it. Apply MDR Article 2 and Annex VIII / IVDR Annex VIII. Screen Austrian borderlines early: products near the medicinal-product line (BASG delineation under the AMG), Annex XVI products without a medical purpose (aesthetic equipment — remember Austria's surveillance expressly reaches beauty/fitness operators), and software (Rule 11; no Austrian NMI-style separate category, but the AI Act layer for high-risk AI). If the borderline is real and you have (or will appoint) an Austria-based AR/manufacturer entity, price in a Section 10 MPG 2021 qualification/classification application (form F_I235) rather than launching on an assumption.

Step 2 — Build the MDR conformity file. QMS (ISO 13485-aligned), technical documentation (Annexes II–III), GSPR checklist (Annex I), risk management (ISO 14971), clinical evaluation (Article 61, Annex XIV) or performance evaluation for IVDs, PMS plan, Basic UDI-DI and UDI-DI assignment, and the PRRC appointment. For Class IIa and above, and for Is/Im/Ir, engage a notified body early — capacity, not the BASG, is the usual bottleneck. QMD Services (NB 2962) is the Austrian option; any EU notified body is legally equivalent.

Step 3 — Fix the operator chain. Appoint the AR (non-EU manufacturers) by written mandate; confirm who is importer of record for Austria/EU; document distributor roles; decide who performs German translation/relabelling and in which MDR capacity (Article 16 consequences); check the Austrian distributor's trade licence and Medizinprodukteberater coverage for field staff.

Step 4 — Register the actors. Complete EUDAMED actor registration and obtain SRNs — early, because BASG validation timing for Austrian-established operators is not predictable and a notified body cannot enter your certificate without your actor registration. Complete GÖG/MPR registration for any Austrian custom-made manufacturing or distributorship before the activity starts, with EMDN product categories prepared in advance (GÖG/WKO guidance recommends preparing your product list before starting the online filing).

Step 5 — German-language package. Prepare labelling and instructions for use under Section 7 MPG 2021 (Section 6): German for anything reaching users/patients; English only where use is genuinely limited exclusively to professional users — and document that limitation, because a "professional" device that reaches a patient or a lay carer loses the carve-out. Prepare the German Field Safety Notice template at the same time; you cannot draft it calmly during an FSCA.

Step 6 — Conformity assessment and CE marking. Complete the applicable Annex route, draw up the EU Declaration of Conformity (with SRN), affix CE marking (with the notified-body number where involved) and UDI. Keep the DoC available in German on request (Section 7 MPG 2021).

Step 7 — Register the devices. Register devices in the EUDAMED UDI/device module before placing on the Austrian market (legacy devices by 28 November 2026). Custom-made device manufacturers complete the Annex XIII statement per device and keep it available for the BASG.

Step 8 — Set up the Austrian money obligations before first supply. Calendar the medical-device levy: whoever in your chain supplies Austrian end users must file the annual declaration and pay by 30 June of the following year (Section 8). Decide contractually who bears it — Austrian distributors will assume the manufacturer has priced it; foreign direct sellers discover they are the liable person themselves. Diarise BASG case fees for any qualification, derogation, clinical-investigation or free-sale applications you plan.

Step 9 — Clinical investigations in Austria (if applicable). Applications, notifications and reports to the BASG are electronic only (EEVO 2011). BASG guidance requires the final favourable ethics-committee vote to be included in the initial application — submitting to the ethics committee in parallel during the BASG procedure is expressly not recommended, because the final vote usually cannot be delivered within the MDR timelines and the application is then rejected. On receipt of complete/corrected documents the BASG completes validation within 5 days; for Article 70(7)(a) MDR / Article 66(7)(a) IVDR investigations the study may start on validation confirmation, for Article 70(7)(b) cases the 45-day scientific/regulatory review runs from there (source: BASG, "Clinical investigations with medical devices", basg.gv.at). MPG 2021 adds Austrian specifics, including ethics-committee composition and defined exclusions of certain study-subject groups (for example Section 25 MPG 2021 for persons in residential care institutions, per professional commentary).

Step 10 — Commercial setup: hospital groups, tenders and distribution. Map the buyers before promising revenue (Section 9): Wiener Gesundheitsverbund (Vienna), the state hospital operators (among them KAGes in Styria, OÖ Gesundheitsholding/Oberösterreich, NÖ Landesgesundheitsagentur in Lower Austria, SALK in Salzburg, Tirol Kliniken, Gesundheit Burgenland, KABEG in Carinthia and the Vorarlberg hospital operator), the university hospitals, religious-order and private groups, and central purchasing via Bundesbeschaffung GmbH (BBG) where the category is bought centrally. Public tenders run under BVergG 2018 and are published through Austrian tender platforms and TED. Prepare German tender documentation — in Austrian public procurement, formal defects exclude bids without a merits review.

Step 11 — Post-market system live before first sale. PMS/PMCF running; vigilance reporting lines to the BASG tested (Section 10) — including the current email route for manufacturer reports and German FSN/customer-list attachments for FSCA; supply-interruption (Article 10a) notification process using the BASG form; implant-register data flows if applicable; complaint handling aligned with Austrian hospitals' own Section 40 reporting (they report to the BASG and inform you — expect to hear about Austrian incidents from the authority as well as from the customer).

6. Language, Labelling and Information Supplied with the Device

Austria's rule (Section 7 MPG 2021) is one sentence with a consequential exception:

  • Default: German. Medical devices may be provided to users and patients only if the information accompanying them is provided in German. Labelling, instructions for use, implant cards and patient information for Austria are a German deliverable.
  • Professional-only exception: English sufficient. If the use of the device is limited exclusively to professional users, information in English is sufficient. This is a genuine Austrian advantage for capital equipment and professional systems — but it is a factual limitation, not a labelling choice. If the same device is also supplied to patients, home-care users or lay carers, the German duty applies to that supply.
  • On request: Declaration of Conformity in German. On a user's or patient's request, the EU Declaration of Conformity must be provided in German. The BASG may also request additional conformity evidence in German.
  • Field Safety Notices: German. FSNs under Article 89(8) MDR must be provided in German; for an FSCA the BASG expects the FSN in German plus the customer list together with the FSCA report form.
  • Notified-body documents: audit, assessment and inspection reports of the notified body may be in German or English.
  • UDI and symbols: standard MDR UDI rules apply; harmonised symbols reduce translation volume but do not remove the IFU duty where it applies.
  • Tender reality: even where English is legally sufficient for the device information, Austrian hospital tenders, training, service and safety documentation are in practice conducted in German, and hospital operators' purchasing terms (for example the Wiener Gesundheitsverbund General Terms and Conditions of Purchase, version valid from 3 March 2025) require products to comply with all relevant legal and technical regulations. Budget German commercial documentation as well as regulatory translation.

7. Timelines

StepRealistic timing
Class I self-declaration (file complete, German IFU ready where required)Weeks — the gating items are usually translation, UDI data and EUDAMED data quality, not the declaration itself. Secondary registration summaries cite about 4–6 weeks for an Austrian Class I registration process
Notified-body conformity assessment, Class IIaCommonly 6–12 months including queue time (planning range — varies by body, scope and file quality; QMD Services has publicly cited 12–18 months as a normal certification span and ~9 months for its first IVDR certificate — treat body-specific quotes as the only reliable figure)
Notified-body assessment, Class IIb / IIICommonly 12–24 months; EU-wide capacity constraints persist
EUDAMED actor registration (Austrian-established operator)BASG processing is in order of receipt and expressly not predictable; file early — an incomplete request triggers an AVG improvement request and possible formal rejection, restarting the clock
GÖG / MPR registration (custom-made manufacturer, distributor, PIZ)Online filing before commencing activity; prepare the product/EMDN list in advance. No statutory approval clock — registration is notification
Legacy devices into EUDAMEDBy 28 November 2026
All notified-body certificates into EUDAMEDBy 28 May 2027 (notified body's duty — but it requires your actor registration to exist)
Clinical investigation, AustriaEthics-committee vote before the BASG initial application; BASG validation within 5 days of complete documents; Article 70(7)(b) MDR review period 45 days thereafter. Add ethics-committee lead time — it is the critical path
BASG qualification/classification decision (Section 10 MPG 2021)No planning figure in BASG public guidance relied on for this guide
Austrian tender cyclesNot a registration step — allow 6–18 months from market readiness to first meaningful public-sector revenue; framework agreements run in multi-year cycles
Medical-device levyDeclaration and payment for year N by 30 June of year N+1 — an annual recurring deadline, not a one-off

Transition deadlines for legacy devices (Regulations (EU) 2023/607 and 2024/1860) — the EU-wide dates apply in Austria unchanged: MDR legacy devices may remain on the market until 31 December 2027 (Class III and Class IIb implantable, with the stated exceptions) or 31 December 2028 (other Class IIb, Class IIa and Class I sterile/measuring), and IVDR legacy devices until 31 December 2027 / 2028 / 2029 by class, in each case only if the Article 120 MDR / Article 110 IVDR conditions (including timely notified-body application and written agreement) were met.

8. Fees and Costs

Austria's state-fee model is different from most EU markets: registration is cheap or free; the recurring levy and the case fees are the real state costs, and notified-body fees dominate everything for Class IIa and above.

8.1 Registration fees

ItemFee
EUDAMED actor / device registrationNo EU fee; BASG validation of Austrian requests is part of the authority's procedure
GÖG Austrian Medical Devices Register (MPR) registrationHistorically free of charge (online register)

8.2 The annual medical-device levy (Medizinprodukteabgabe) — the fee everyone misses

Per the BASG's Medical Device Levy Ordinance guidance and its manufacturer page:

  • Who pays: every natural or legal person — regardless of where their business is based — who provides/supplies medical devices to end users in Austria. "End user" means anyone who does not sell or pass on the device for payment — a consumer using it personally, or a user applying it in medical treatment. In a typical import model the liable person is whoever makes that final supply in Austria; in a direct-to-end-user cross-border model it can be the foreign seller. Allocate it expressly in the distribution agreement.
  • How much: a flat-rate annual amount set in the Annex to the Levy Ordinance by device category. If a liable person supplies devices of different classes / IVDs / implantable devices, the amounts are not added together — only the highest applicable flat-rate amount is payable.
  • Threshold: the levy is payable only if the levy amount is less than 1% of the person's sales revenue from medical devices — a de-minimis design so that an occasional seller (the BASG's own example is the corner shop that occasionally sells a thermometer or plaster) is not captured. Do not read this backwards: for any real device business the levy will be well under 1% of device revenue and therefore payable.
  • When: a declaration for the previous fee year must be submitted to the BASG in accordance with the Levy Ordinance no later than 30 June of the following year, and the levy paid by that deadline.
  • Exact flat-rate amounts:

8.3 BASG case fees (Gebührenverordnung)

ItemFee basis
Qualification / classification application (Section 10 MPG 2021, form F_I235)Per the current BASG Fee Regulation, Section X of the Annex
Derogation from conformity assessment (MDR Article 59 / IVDR Article 54)Per the Fee Regulation; contact the BASG in advance by email (medizinprodukte@basg.gv.at) — derogations are for health-protection emergencies, not for early marketing ahead of a pending assessment
Clinical investigations / performance studiesSeparate fee items in the Fee Regulation (clinical-investigation sections); combined device+medicinal-product submissions have specific apportionment rules in the Regulation
Certificate of Free Sale (for Austria-based manufacturer/AR, CE-marked devices for non-EU export)Per the Fee Regulation
Informal borderline enquiry (AMG delineation)Charged at an hourly rate under Section 8(2) of the Fee Regulation; the formal delineation procedure is a chargeable official act with expert opinion and decision
Notified-body conformity assessmentCommercial fees; the dominant cost for Class IIa+ — obtain quotes (including from QMD Services, NB 2962). Not a state fee

Not charged by Austria: no per-device MDR registration fee for ordinary CE-marked devices, no manufacturing or import licence fee for devices as such, and no advertising licence or fee (CMS advertising guide for Austria). Non-Austrian EU operators selling through an Austrian chain do not pay a separate Austrian "market entry" fee — their Austrian state-cost exposure is the levy analysis above (which follows the end-user supply) and any case fees they themselves trigger.

9. Reimbursement, Procurement and Advertising — the Austrian Commercial Reality

9.1 Hospitals: the LKF system pays, inside a fixed envelope

Austria finances public (State Health Fund) hospitals through Leistungsorientierte Krankenanstaltenfinanzierung (LKF) — an activity-based, DRG-type system in place since 1997 with two pillars: a nationally uniform core area and a steering area adapted to hospital-care levels and the federal states. Cases are grouped into leistungsorientierte Diagnosefallgruppen (LDF), scored by points that determine payment to the hospital. New — usually costly — medical services using devices enter the system through the hospital individual services catalogue (MEL — Katalog medizinischer Einzelleistungen), a positive list of roughly 2,000 services that is updated annually in a process coordinated by the federal health ministry; public hospitals submit proposals for inclusion or price changes, and proposals are assessed on clinical-benefit evidence — a preliminary rejection is the usual recommendation where benefit evidence is absent or inconclusive, in which case a service may be coded for documentation only (XN code) without changing payment (source: analysis of Austrian hospital individual-services catalogue decisions, BMC Health Services Research / Springer, covering 2008–2020).

Practical consequences for a device maker:

  • CE marking does not create LKF money. If your device enables a service that has no MEL position, or is used inside an LDF whose points were calculated without your device's cost, the hospital funds it from an envelope calculated for something else. Your Austrian evidence dossier must answer the LKF question — which code, which points, what cost offset — in the hospital's language, literally and economically.
  • Timing is annual. The catalogue is updated once a year; a missed cycle is a year, not a quarter. Engage the hospitals that would propose your service well before the submission window — at use).
  • Reimbursement is not standardised (Wolf Theiss, Medical Devices in CEE & SEE, Austria chapter): inside hospitals it differs by hospital operator type; the LKF logic above is the public-hospital core, private operators negotiate their own terms.

9.2 Outside hospitals: social insurance, carrier by carrier

For devices used outside hospitals (aids and appliances — Heilbehelfe/Hilfsmittel — home-use devices, defined consumables), reimbursement depends on the social-insurance carrier involved. Austria's carriers were consolidated from 22 to 5 in the 2020 Social Insurance Organisation reform; the largest is the Österreichische Gesundheitskasse (ÖGK), covering the large majority of insured persons, alongside the carriers for the self-employed/farmers (SVS), public employees/railways/mining (BVAEB) and others. The working rules from GÖG's pricing-and-reimbursement system description:

  • Prices are not statutorily regulated — manufacturers set device prices freely.
  • Reimbursement requires an arrangement with the sickness fund(s). A manufacturer/distributor that wants a device reimbursed must reach an arrangement with the carrier; because of the federal/carrier structure, national coverage can mean more than one arrangement.
  • Patients generally contribute a co-payment toward the cost of the device (professional sources, Wolf Theiss Austria chapter).
  • There is no public overview of device prices in Austria — competitive price intelligence is tender-by-tender and contract-by-contract, not a published list.

9.3 Procurement: BVergG 2018, hospital groups and the BBG

Public hospitals buy under the Federal Public Procurement Act 2018 (BVergG 2018). Expect published tenders (Austrian tender platforms and TED), framework agreements, and strict formalism. Structural features to plan around:

  • Concentrated buyers. Vienna's Wiener Gesundheitsverbund operates the capital's municipal hospitals and is routinely among the largest single device buyers in the country; the state operators and the three university hospitals (AKH Wien / Medical University of Vienna, Graz, Innsbruck) follow the same public-procurement logic. Purchasing terms are published and non-trivial — for example, the Vienna group's purchasing terms require (medical) products to comply with the state of the art and all relevant legal, official and technical regulations.
  • Central purchasing. The Bundesbeschaffung GmbH (BBG) is the federal central purchasing body; where a category is bought through BBG frameworks, that framework — not the individual hospital — is your route to a large share of public volume.
  • German documentation. Tender documents, bids, service and training documentation are in German in practice. A formally defective or late bid is excluded without a merits review — use Austrian procurement support for the first bid.
  • Service coverage is scored. Austria is geographically compact but operationally Alpine; national service coverage from Vienna/Linz/Graz with credible response times is a recurring award factor for equipment categories, and consolidation among hospital groups is lengthening tender processes and favouring suppliers that can offer multi-site framework terms.

9.4 Advertising and promotion (MPG 2021, Sections 70–77)

Device advertising remains nationally regulated in Austria, layered under MDR Article 7's EU-wide ban on misleading claims:

  • Consumer advertising is prohibited for devices that are (Section 71 MPG 2021): subject to prescription; intended to be used only by healthcare professionals; or to be used by the consumer only in connection with medical or dental treatment.
  • Permitted consumer advertising is content-controlled (Section 73 MPG 2021): it must not imply the treatment's effect is superior to another device's, must not be directed primarily at children, and must not imply that medical treatment is unnecessary (among other prohibitions, including restrictions on pictorial representations of changes in the human body — CMS Austria advertising guide). It must contain the name of the device, a short description of its intended purpose, essential information on its application and further prescribed particulars.
  • Professional advertising must be accurate, current, verifiable and complete enough for the recipient to form their own judgement; inducement rules and the Austrian Criminal Code's anti-corruption provisions apply to benefits for healthcare professionals, and the AUSTROMED industry code of conduct frames industry practice (it does not replace the statute).
  • No advertising licence or fee is required — compliance is policed after the fact by the BASG/authorities and by competitors under the UWG, which is an active route in Austria: a competitor complaint can hurt more than an inspection.
  • Never claim BASG "approval" (there is none) and never market beyond the CE-marked intended purpose; keep German marketing claims word-consistent with the German IFU — inconsistency between promotional German and regulatory German is a recurring surveillance and litigation trigger.

9.5 Consumer and pharmacy channels

Devices sold direct to consumers (pharmacies, specialist retail, e-commerce) follow the same MDR + Section 7 language rules plus the Austrian dispensing rules for who may supply which devices. Distance selling is possible within the MDR framework; the seller's MDR role (usually distributor, sometimes importer) determines its duties, and an Austria-established seller must check the GÖG distributor-registration duty and the levy before the first sale, not at year-end.

10. Post-Market Surveillance, Vigilance and Market Surveillance in Austria

  • PMS system (MDR Articles 83–86): PMS plan and PMS report for Class I; PSUR for Class IIa (at least every two years) and for IIb/III (annually for Class III and implantable IIb) — unchanged EU rules, checked in BASG surveillance.
  • Serious-incident reporting by the manufacturer (Article 87): within the MDR deadlines (2 days for a serious public-health threat, 10 days for death or serious deterioration, 15 days as the standard). Austrian channel note: BASG guidance states that because EUDAMED was not fully operational, manufacturer notifications are made by email to medizinprodukte@basg.gv.at using the European reporting forms, per MDCG 2021-1 and Article 2 EEVO 2011; for an FSCA, the customer list and the Field Safety Notice in German must be submitted in addition to the FSCA form (source: BASG, "Manufacturer and representative", basg.gv.at, crawled October 2026). The EUDAMED modules made mandatory on 28 May 2026 were Actor, Device, Certificates and Market Surveillance — the vigilance module was not in that first mandatory set.
  • Austrian-side reporting — broader than the MDR (Section 40 MPG 2021): health professionals, persons who professionally/commercially use or operate devices, heads of testing/inspection/certification bodies, assemblers of systems/procedure packs, persons who repack devices, and hospitals' technical safety officers must report information on every serious incident that becomes known to them in their professional activity immediately (unverzüglich) to the BASG, sharing all observations and data relevant to device safety — and must also inform the manufacturer or its AR. In hospitals, reports are made uniformly through the medical director, except where there is imminent danger. Suspected falsified devices must be reported immediately. Devices involved in incidents must not be destroyed — they are subject to root-cause analysis. Design your complaint handling for early, factual Austrian reports arriving from both the hospital and the authority.
  • Trend reporting (Article 88), FSCA and FSN: as in the EU generally, with the Austrian specifics above (German FSN, customer list to the BASG).
  • Supply-interruption notification (Article 10a MDR/IVDR): a manufacturer expecting an interruption in supply must report it to the BASG using the BASG form, state the reasons, and inform the affected economic operators, healthcare institutions and healthcare professionals it supplies directly — at least six months in advance (source: BASG, "Medical devices vigilance", basg.gv.at).
  • Operator duties in healthcare institutions (MPBV): incoming inspection before first use at the place of operation for listed devices (and others designated by the technical safety officer), recurring safety inspections (sicherheitstechnische Kontrollen), metrological controls, maintenance per manufacturer instructions, staff instruction, and a maintained device file. The hospital technical safety officer is a statutory expert role, involved in procurement planning for medical-technical equipment — for equipment sales, this person is a technical gatekeeper, not an administrator. Selling into Austria without a service/inspection concept that fits MPBV duties is selling half a product.
  • Market surveillance: the BASG supervises manufacturers and ARs (Articles 93 MDR / 88 IVDR, Section 2 MPG 2021) and the wider MPG 2021 population — including commercial users outside healthcare (beauty salons, fitness centres) and even occasional professional users for reporting purposes. It can inspect, demand documentation (including in German), require corrective action and prohibit non-compliant devices. Sanctions: MPG 2021 carries a comprehensive catalogue of administrative penalties, and deliberate attempts are also punishable (CMS, "Austria: What's new in medical device law?"). Assume the file you would show the BASG is the file your Austrian distributor and your largest hospital customer will also eventually see — consistency across the three is an Austrian survival skill.
  • In-house (healthcare-institution) devices: institutions manufacturing and using devices internally work under MDR Article 5(5) as supplemented by Section 9 MPG 2021, under which the health ministry may prohibit the in-house manufacture/use of specific devices — relevant where a university hospital is both customer and in-house maker.
  • Records: keep technical documentation and registration data for the MDR periods (10 years; 15 years for implantables). Deregistration does not end the documentation duty, and ceasing Austrian sales does not cancel a levy declaration already due for the year of supply.

11. Common Mistakes Foreign Manufacturers Make in Austria

  1. 1. Assuming EUDAMED registration replaced Austrian registration. It did not. Austrian distributors and custom-made manufacturers still register with GÖG in the MPR under Section 37 MPG 2021 / the 2024 Notification Ordinance — and since 2024 their data are published. Appointing an unregistered Austrian distributor is a defect on day one.
  2. 2. Assuming the BASG approves devices — and marketing that assumption. The BASG expressly does not approve devices; it surveils the market. "BASG-approved" in Austrian material is a misleading claim under MDR Article 7 and MPG 2021, and competitors enforce the UWG actively.
  3. 3. Missing the medical-device levy. The levy follows whoever supplies Austrian end users, regardless of the supplier's seat, with a declaration and payment deadline of 30 June for the previous year. Foreign direct sellers and e-commerce operators are the classic defaulters; Austrian distributors assume it has been priced and say nothing.
  4. 4. Reading the professional-user English exception as a general English permission. Section 7 MPG 2021 allows English only where use is limited exclusively to professional users. The first home-care patient, lay carer or consumer sale on the same SKU ends the exception for that supply — and FSNs are German regardless.
  5. 5. Filing a clinical investigation in parallel with ethics review. The BASG requires the final favourable ethics vote in the initial application and expressly advises against parallel submission. A parallel filing is a rejection strategy with fees attached.
  6. 6. Leaving the Austrian distributor's paperwork unexamined. Trade licence, MPR registration (with current EMDN categories and supplemented post-2024 data), Article 16 status if it relabels/translates, Medizinprodukteberater coverage for its field staff — check all four before signing, because hospital supplier audits will check them after.
  7. 7. Treating the LKF/MEL question as a post-launch problem. A device whose service has no MEL position is funded from an envelope calculated for other services. The annual catalogue cycle means a late evidence strategy costs a full year.
  8. 8. Quoting one social-insurance arrangement as "Austria covered". Outside hospitals, reimbursement is carrier-by-carrier. ÖGK is the largest carrier, not the only one; national claims need the carrier map checked.
  9. 9. Ignoring the technical safety officer and MPBV. Austrian hospitals must incoming-inspect, periodically safety-check and document your device. If your service concept does not deliver the inspections, intervals and documentation the operator's file needs, the device fails operationally even if it wins clinically.
  10. 10. Destroying or returning incident devices before analysis. Austrian law requires devices involved in incidents to be preserved for root-cause analysis, and Austrian hospitals report serious incidents to the BASG immediately and inform you in parallel. A logistics process that auto-returns the suspect unit abroad breaks the investigation and the relationship at once.
  11. 11. Under-preparing the BASG actor validation. For Austrian-established operators, an incomplete EUDAMED actor request triggers a formal AVG improvement request and can end in formal rejection — restarting a queue whose timing the BASG expressly does not predict. File complete, file early.
  12. 12. Assuming "small market, light touch". Austria's surveillance population expressly includes occasional professional users and non-healthcare businesses, its offence catalogue is extensive, and attempts are punishable. The Austrian layer is thin in pages and heavy in enforcement reach.

12. Frequently Asked Questions

Does Austria require a separate national approval on top of CE marking? No. There is no Austrian marketing authorisation for medical devices and the BASG does not approve CE marking. Market access = MDR/IVDR conformity + CE marking + the registration, language, levy and post-market duties in this guide.

Who is the Austrian regulator for medical devices? The BASG (Bundesamt für Sicherheit im Gesundheitswesen — Federal Office for Safety in Health Care), designated in Section 2 MPG 2021, operating with AGES (Austrian Agency for Health and Food Safety / AGES Medizinmarktaufsicht) as its operating body, under the federal health ministry. GÖG runs the Austrian Medical Devices Register and implant registers — it is a register operator, not the market-surveillance authority.

Do I need an Austrian authorised representative? No — the AR must be established in the EU, not necessarily in Austria. But an Austria-based AR can file BASG qualification/classification applications (Section 10 MPG 2021, form F_I235), request Certificates of Free Sale from the BASG, and is validated by the BASG in EUDAMED. Choose the AR's seat with those functions in mind.

I am an EU manufacturer already registered in my home country. Do I register again in Austria? Not as a manufacturer — your EUDAMED actor registration covers you EU-wide. Austrian duties that still apply to you: the Section 7 language rule for Austrian supply, vigilance for Austrian incidents, the levy analysis if you supply Austrian end users directly, and GÖG registration only if you act in Austria as a distributor or custom-made manufacturer in your own right.

My Austrian distributor says it is "registered". Which register should it be in? The GÖG Austrian Medical Devices Register (MPR) at medizinprodukteregister.at — distributors established in Austria register there before commencing activity (Section 37 MPG 2021), declaring firm/contact data, activity type and product categories (EMDN). Since the Medizinproduktemeldeverordnung 2024 the data are public — verify the entry yourself in the public search, including that pre-2024 registrations were supplemented by 30 April 2025.

What does Austria charge per year? There is no per-device registration fee for ordinary MDR/IVDR devices and GÖG registration has historically been free. The recurring state cost is the medical-device levy: a flat annual amount by highest device category supplied to Austrian end users, declared and paid to the BASG by 30 June of the following year, payable only where the amount is under 1% of the liable person's device sales revenue (in practice: payable for any real device business). Exact flat-rate amounts must be taken from the current Annex to the Levy Ordinance. BASG case fees (qualification/classification, derogations, clinical investigations, free-sale certificates) are in the BASG Fee Regulation.

Must the IFU be in German even for hospital professional devices? Not necessarily — Austria is more permissive here than several member states. Under Section 7 MPG 2021, German is the default for information supplied to users and patients, but where use is limited exclusively to professional users, English is sufficient. The exception is factual: any supply to patients or lay users needs German, Field Safety Notices are always German, and the Declaration of Conformity must be provided in German on request.

Is there an Austrian notified body? Yes — QMD Services GmbH, Notified Body 2962 (Vienna and Linz; a Quality Austria subsidiary), designated under the IVDR on 23 December 2022 and under the MDR on 14 May 2024, and described at its MDR designation as one of only eleven EU bodies covering both regulations. You may use any EU notified body; the choice is commercial.

How do hospitals in Austria actually buy devices? Public hospitals buy under the Federal Public Procurement Act (BVergG 2018) — tenders and framework agreements run by the hospital operating groups (largest: Wiener Gesundheitsverbund) and, for centrally bought categories, the federal purchasing body BBG. They are paid through the LKF activity-based financing system; a service's position in the MEL individual-services catalogue determines whether your device's economics work inside the hospital's envelope. Registration and CE marking make you eligible to sell; tenders and LKF fit make you revenue.

Is there national reimbursement for devices like there is for medicines? No standardised device reimbursement exists. Inside hospitals: LKF/MEL logic, differing in detail by operator. Outside hospitals: an arrangement with the individual social-insurance carrier (largest: ÖGK), with patient co-payments. Device prices are not statutorily regulated and are set freely by manufacturers.

What happens if we run a clinical investigation in Austria? Electronic submission only (EEVO 2011) to the BASG, with the final favourable ethics-committee vote included in the initial application. BASG validation completes within 5 days of complete documents; Article 70(7)(a) MDR investigations may start on validation confirmation, Article 70(7)(b) investigations after the 45-day review. MPG 2021 adds national rules on ethics committees and study subjects — confirm the current sections for your protocol before filing.

How are incidents reported in Austria? Manufacturers report serious incidents/FSCA to the BASG within MDR deadlines — currently by email to medizinprodukte@basg.gv.at on the European forms (plus German FSN and customer list for FSCA) per BASG guidance, pending the EUDAMED vigilance module. Separately, Austrian professionals, operators, pack assemblers, repackers and hospital technical safety officers must report serious incidents to the BASG immediately under Section 40 MPG 2021 and inform the manufacturer/AR; hospitals report through their medical director. Suspected falsified devices are reported immediately, and incident devices must be preserved for root-cause analysis.

Can I sell devices online to Austrian consumers? Yes, within the MDR framework — with German user information, the correct MDR role allocation, the Austrian dispensing rules, and two Austrian traps checked before the first sale: whether you are the levy-liable supplier to Austrian end users, and whether any Austria-established selling entity triggers GÖG distributor registration. Consumer advertising is prohibited for prescription, professional-only and treatment-linked devices (Section 71 MPG 2021).

What are the penalties for non-compliance? The BASG can inspect, demand documentation, require corrective action and prohibit devices; MPG 2021 provides a comprehensive catalogue of administrative penalties and makes even deliberate attempts punishable. Unpaid levy and fee obligations are pursued as public charges. The practical penalty in Austria is often commercial as well as legal: a surveillance finding is visible to the concentrated hospital-group buyers who dominate the market.

We make custom-made devices. What is different in Austria? Three things: the MDR Annex XIII statement regime (no ordinary CE route for the device); GÖG/MPR registration before starting activity, including PRRC name/contact and EMDN product codes; and, for Class III custom-made implantables, EUDAMED actor registration as well, so the notified body can enter the certificate.

Sources & scope

  • basg.gv.at, BASG — market surveillance, EUDAMED actor validation, the annual medical-device levy.
  • goeg.at, Gesundheit Österreich GmbH (GÖG) — the Austrian Medical Devices Register (MPR) and implant registers.
  • eur-lex.europa.eu, Regulation (EU) 2017/745 (MDR), Regulation (EU) 2023/607.

This guide is market-access intelligence, not regulatory or legal advice. BASG and GÖG practice, levy rates and LKF catalogues change — professional review is required before business use.