EU MDR + the Belgian layer
The Medical Devices Act of 22 December 2020 and the FAMHP frame the national layer — one agency that designates notified bodies, runs market surveillance, and still operates its own portal registration.
Field guide · PDF
No Belgian CE mark — EU MDR applies directly. Belgium adds one of Europe's densest national layers: FAMHP portal registration that continues alongside EUDAMED, labelling in French, Dutch and German, an annual contribution on Belgian turnover, and a separate RIZIV/INAMI reimbursement gate.
Sourced from the FAMHP, RIZIV/INAMI, the Medical Devices Act of 22 December 2020, and verified 2026 publications.
Enter your work email and we'll send the PDF straight to your inbox.
What's inside
Every figure in this guide traces to a named source — the FAMHP, RIZIV/INAMI, the 2020 Medical Devices Act, and verified 2026 publications.
The Medical Devices Act of 22 December 2020 and the FAMHP frame the national layer — one agency that designates notified bodies, runs market surveillance, and still operates its own portal registration.
FAMHP portal registration for Belgian operators continues next to mandatory EUDAMED registration — one does not discharge the other. Distributors and exporters register too, and every operator designates a materiovigilance contact point.
For implants and invasive devices, the RIZIV/INAMI list — in many categories nominative, brand by brand — decides whether compulsory insurance pays. Registration alone does not put you on it.
Complete guide
Every section of the PDF on one page. Use the form above to keep a copy.
Last updated: October 2026. The EU Medical Device Regulation (MDR 2017/745) and the In Vitro Diagnostic Regulation (IVDR 2017/746) apply directly in Belgium. The Belgian national layer consists mainly of the Medical Devices Act of 22 December 2020, the IVD Act of 15 June 2022, their implementing Royal Decrees (notably 12 May 2021 for the MDR, 18 May 2021 on clinical investigations, and 13 and 25 September 2022 for the IVDR), supervision by the Federal Agency for Medicines and Health Products (FAMHP — in Dutch FAGG, in French AFMPS), a continuing FAMHP portal registration for Belgian economic operators, and a separate reimbursement system run by RIZIV/INAMI (the National Institute for Health and Disability Insurance). Always confirm critical steps against current FAMHP, RIZIV/INAMI and eHealth guidance before filing.
Belgium is a small, dense and disproportionately important European medtech market. It has about 11.7 million inhabitants, a compulsory health-insurance system, a strong hospital sector concentrated around university hospitals such as UZ Leuven, UZ Gent, UCLouvain Saint-Luc, Erasmus Hospital in Brussels and UZ Antwerp, and a central geographic position that has made it a distribution hub for the Benelux and for northern France and western Germany. Osborne Clarke, summarising MedTech Europe figures, has described Belgium as a top-10 European country for both medical devices and IVDs. It is also home to a Belgian-designated notified body — SGS Belgium NV (NB 1639, Antwerp) — designated under both the MDR and, from 23 April 2025, the IVDR.
Regulatory reality check up front: Belgium is not a separate certification regime from the EU. There is no "Belgian CE mark" and no Belgian pre-market product approval on top of CE marking. What Belgium adds — and what foreign manufacturers most often underestimate — is a national layer that sits around and after CE marking, and it is denser than in most member states:
This guide covers the EU layer briefly (it is the same as in Germany, France, Italy, Spain or the Netherlands) and the Belgian layer in depth, including the 2026 transition to mandatory EUDAMED, the continuing FAMHP portal duty, and the RIZIV/INAMI reimbursement route that most generic "sell into the EU" guides omit entirely.
Belgium transposed the MDR and IVDR later and in more pieces than most member states, and legacy pre-MDR decrees still coexist with the new framework. The core instruments are:
Two structural points matter for planning. First, Belgium did not abolish its national portal when EUDAMED became mandatory: Osborne Clarke's analysis of the 2022 decrees states expressly that "local registration requirements on the FAMHP portal remain unchanged by the decree." A Belgian manufacturer, authorised representative, importer, distributor or exporter therefore works in two systems — EUDAMED for the EU duties, the FAMHP portal for the Belgian operator registration — not one. Second, health competences in Belgium are split between the federal state (which runs device regulation, compulsory insurance and hospital financing norms) and the Communities/Regions (prevention, some care organisation). Device registration itself is federal — FAMHP and RIZIV/INAMI — so a foreign manufacturer does not file three regional registrations; the split matters commercially (for example in prevention programmes and some hospital investment), not for CE or FAMHP purposes.
| Body | Role |
|---|---|
| FAMHP / FAGG / AFMPS — Federal Agency for Medicines and Health Products | Belgian competent authority for medical devices and IVDs: designates and supervises Belgian notified bodies, market surveillance and post-market surveillance, vigilance (materiovigilance), clinical investigations and performance studies, Article 59 derogations, Free Sale Certificates (certificates of free sale / export certificates), FAMHP portal registration of economic operators, and enforcement including suspension of an operator's activities for EUDAMED non-confirmation. |
| FPS Public Health, Food Chain Safety and Environment | The federal ministry above the FAMHP; owns health legislation and hospital policy. |
| RIZIV / INAMI — National Institute for Health and Disability Insurance | Runs compulsory health insurance reimbursement: the List of implants and invasive medical devices and the nominative list, the nomenclature of reimbursable services, agreements/conventions with providers, and — historically with FAMHP and the eHealth platform — the financing level (M3) of the mHealthBelgium pyramid. Not a device regulator, but decisive for whether an implant or invasive device is paid for. |
| eHealth platform | Federal platform for secure health-data exchange; owner of the ICT criteria at level M2 of the mHealth pyramid (identification, authentication, interoperability, therapeutic relationship and informed consent). |
| KCE — Belgian Health Care Knowledge Centre | Federal health-technology assessment body; produces HTA reports that inform RIZIV/INAMI and ministerial decisions. Not a filing office. |
| Ethics Committees (ECs) + Act of 7 May 2004 on experiments on the human person | Review clinical investigations and performance studies alongside FAMHP authorisation/notification, under the MDR/IVDR decrees and the human-experimentation framework. |
| SGS Belgium NV (NB 1639) | The Belgian-designated notified body, based in Antwerp, designated by the FAMHP/European Commission under the MDR for all types of devices including Annex XVI products without a medical purpose, and under the IVDR from 23 April 2025. Any EU-designated notified body can be used for the Belgian market; it does not have to be Belgian. |
| beMedTech / Agoria | Sector federations. beMedTech is the Belgian medtech industry federation; together with Agoria it runs the mHealthBelgium platform — since the end of 2023 without the former public partnership with RIZIV/INAMI, according to sector reporting . |
| Mdeon | Non-profit ethics platform for healthcare professionals and industry; operates the prior-visa procedure for sponsorship of HCP participation in scientific events under Article 10 of the Medicines Act. |
| betransparent.be | The central Sunshine Act transparency register, managed under FAMHP supervision; annual publication of premiums and benefits (Section 9). |
Belgium uses the MDR classification system directly — four risk classes, 22 classification rules in Annex VIII MDR:
| Class | Risk Level | Conformity Assessment Route | Examples |
|---|---|---|---|
| Class I | Low risk | Self-declaration (Annex II+III technical documentation; EU Declaration of Conformity; no notified body — except Class Is [sterile], Im [measuring function], Ir [reusable surgical instruments]) | Adhesive bandages, spectacles, wheelchairs, non-sterile examination gloves |
| Class IIa | Low–medium risk | Notified body required (Annex IX chapters I+III, or Annex II+III + Annex XI) | Dental fillings, surgical clamps, hearing aids, ultrasound devices |
| Class IIb | Medium–high risk | Notified body required | Anaesthesia machines, infusion pumps, ventilators, surgical lasers |
| Class III | High risk | Notified body required, strictest scrutiny (Annex IX, including clinical evaluation consultation for certain devices) | Cardiac catheters, devices in direct contact with the heart, central circulatory system or central nervous system, pacemakers, drug-eluting stents |
For IVDs, the IVDR classes A–D apply (A lowest risk — specimen containers, prepared selective media; B medium — pregnancy tests, cholesterol self-tests; C medium-high — blood-glucose self-tests, genetic testing; D highest — hepatitis B blood-donor screening, HIV diagnostic blood tests), with notified-body involvement for Classes B, C and D and for sterile Class A devices.
Classification matters in Belgium for three practical reasons beyond the conformity route itself. First, it drives the reimbursement category if the device is an implant or an invasive device: RIZIV/INAMI's List works with categories A–H that are about how a device is paid (individual price on the nominative list, generic tariff, contract, restricted clinical application), not about MDR risk — teams that confuse the two "classifications" build the wrong dossier (Section 9). Second, classification and device type drive the Belgian study route: which clinical investigations and performance studies need prior FAMHP authorisation plus ethics review, and which are notification-only, is set out in the 2021 and 2022 decrees by study and device type (Section 5). Third, classification drives post-market reporting: PSUR duties apply from Class IIa upwards, and implantables trigger the implant-card and traceability duties that Belgian hospitals operate through their vigilance contact points (Section 11).
Special MDR categories also apply in Belgium: custom-made devices (Annex XIII statement instead of CE marking; no UDI), systems and procedure packs (Article 22 MDR statement; Article 29(2) has its own EUDAMED registration logic, expressly recalled in the Belgian decree), in-house devices manufactured and used only within a health institution under Article 5(5) MDR — for which the Belgian performance-studies decree created a specific authorisation procedure for IVD studies — and Annex XVI products without a medical purpose (non-corrective contact lenses, dermal fillers and similar products).
An incorrect classification is not a clerical error: it invalidates the conformity route, can mean a missing notified body, and can result in FAMHP enforcement. Resolve borderline questions (device vs. medicinal product vs. cosmetic vs. biocidal product vs. food supplement) early — the FAMHP offers a national scientific-technical advice (STA) route for qualification of borderline products.
Belgium imposes no separate Belgian representative requirement on top of the MDR for a foreign manufacturer that simply sells into Belgium through others. Keep the EU roles and the Belgian establishment facts separate — the Belgian consequences attach to whoever is established in Belgium, in whatever role.
PRRC: The Person Responsible for Regulatory Compliance (Art. 15 MDR) requirement applies unchanged: a relevant degree plus one year of professional experience, or four years of experience without a degree; the AR must also have a PRRC permanently and continuously at its disposal. Micro and small manufacturers may have the PRRC permanently and continuously at their disposal without employing them.
No physical office is required to sell into Belgium. A foreign manufacturer does not need a Belgian office, branch or warehouse merely to register and sell devices in Belgium. It needs: an EU AR (if non-EU), an importer/distributor chain that meets Articles 13–14, trilingual labelling, EUDAMED registration and — where an operator in its chain is Belgian-established — that operator's FAMHP portal registration. A Belgian office becomes regulatory-relevant mainly if that establishment itself is the manufacturer, AR, importer, distributor or exporter, in which case the corresponding MDR duties plus the Belgian portal, VCP, Sunshine Act and annual-contribution duties (Sections 4, 8 and 9) attach to it directly.
This is the section most generic "sell into the EU" guides get wrong for Belgium, because Belgium kept its national system when the EU system became mandatory.
The FAMHP portal (continuing, no end date). Belgian economic operators register with the FAMHP through its portal — the registration the 2022 decrees expressly left "unchanged." In substance it covers:
A FAMHP portal registration is an administrative registration, not an approval: it does not certify conformity, and the FAMHP does not "approve" an ordinary CE-marked product through it. Conversely, the absence of a required portal registration is a compliance defect the FAMHP can act on even for a perfectly CE-marked device.
What changed on 28 May 2026 (EUDAMED). With the first four EUDAMED modules becoming mandatory EU-wide, Belgium applies the same EU timetable as every other member state:
What Belgium did NOT add. There is no Belgian equivalent of France's separate ANSM declaration industry or Italy's Repertorio per-product registration fee for ordinary devices, and no Belgian "product approval certificate" on top of CE marking. The Belgian extras are: operator registration on the FAMHP portal, distributor/exporter registration for traceability, the VCP system, trilingual information, the turnover-based FAMHP contribution, and — commercially decisive but legally separate — the RIZIV/INAMI reimbursement lists.
Practical consequence for planning. Budget for two registration systems in 2026 and after — EUDAMED (EU-wide, mandatory, no fee) plus the FAMHP portal for every Belgian-established operator in your chain — plus a data-confirmation calendar (1-year / 2-year EUDAMED confirmations, enforced by suspension), plus, if you sell implants or invasive devices, a RIZIV/INAMI list project that runs on its own evidence and calendar logic (Section 9). Do not assume an EUDAMED registration replaces the FAMHP portal, and do not assume a FAMHP portal registration replaces EUDAMED. In Belgium, in 2026, both statements are false.
Step 1 — Confirm the product is a medical device and classify it under Annex VIII MDR (or IVDR Annex VIII for IVDs). Resolve borderline questions early, using the FAMHP scientific-technical advice route where the qualification is genuinely in doubt; for device–drug combinations, expect the notified body to run the ancillary-substance consultation procedure.
Step 2 — Establish the quality management system aligned to MDR Art. 10(9), normally evidenced by ISO 13485 certification. Notified bodies audit the QMS as part of conformity assessment; it is a prerequisite, not a parallel task.
Step 3 — Compile the technical documentation (Annex II) and clinical evaluation (Annex XIV), including the post-market surveillance plan and PMCF plan where applicable. For IVDs, compile the performance evaluation under the IVDR.
Step 4 — Complete conformity assessment with a notified body (all classes above I, and Is/Im/Ir; IVD Classes B, C, D and sterile A). The notified body does not need to be Belgian — any body designated for your device codes in NANDO qualifies — but Belgium gives you a real local option: SGS Belgium NV (NB 1639) in Antwerp, designated under the MDR for all device types including Annex XVI products, and under the IVDR since April 2025. Verify scope in NANDO, not on the body's marketing page.
Step 5 — Draw up the EU Declaration of Conformity (Annex IV) and affix the CE marking (with the notified body's number where a body was involved). There is no separate Belgian marketing authorisation for an ordinary device: the Belgian system is CE marking plus registrations, policed by the FAMHP — not a Belgian product approval.
Step 6 — Assign UDI (Art. 27 MDR: Basic UDI-DI, UDI-DI, UDI-PI) and prepare UDI carriers for labels, on the staggered MDR timetable by class. Custom-made devices are exempt from UDI. Belgian hospitals check UDI data at goods-in and in their implant traceability workflows.
Step 7 — Appoint the EU Authorized Representative (if you are established outside the EU) and execute the Art. 11 mandate — before registrations, because the AR is a registrant in EUDAMED. Decide deliberately whether the AR sits in Belgium (with the Belgian portal and FAMHP-facing consequences described in Section 3) or elsewhere in the EU.
Step 8 — Decide the Belgian establishment question. Will any company established in Belgium be the manufacturer, AR, importer, distributor or exporter? If yes, that establishment needs Crossroads Bank for Enterprises (KBO/BCE) registration, its own EUDAMED actor registration/SRN where applicable, FAMHP portal registration, a VCP function, and working Article 10/11/13/14 systems — plus Sunshine Act registration on betransparent.be and the turnover-based FAMHP contribution workflow if it places devices on the Belgian market (Sections 8–9). If a Belgian health institution will manufacture devices for its own exclusive use, test the Article 5(5) in-house conditions instead of assuming a full manufacturer route.
Step 9 — Register the actors and the device in EUDAMED. Obtain the SRN for manufacturer, AR and importer through actor registration (validated by the competent authority of the member state of establishment — for Belgian establishments, the FAMHP), then register the device/UDI data. Mandatory since 28 May 2026 for the first four modules; new devices must be registered before placement on the market. Diary the Belgian-enforced data-confirmation duty: confirm accuracy within one year of first submission and every two years thereafter, on pain of FAMHP suspension.
Step 10 — Complete the FAMHP portal registration for every Belgian operator in the chain. Manufacturer, AR/importer and distributor/exporter registrations as applicable (Sections 3–4), with the distributor/exporter product information for traceability. Confirm the live portal form for your operator type at filing — the FAMHP reorganises its web guidance by operator and device type, and older third-party summaries (including CD-ROM-era clinical-investigation instructions still circulating online) are not a reliable map of the current portal.
Step 11 — Prepare the Belgian labelling. Translate labels and IFU into French, Dutch and German as the default (Section 6). Build the professional-user English position, if you intend to rely on it, as a documented, device-specific decision — the user can demand a national language, so your distributor must be able to supply it.
Step 12 — If you will run a clinical investigation or performance study in Belgium, file through the FAMHP/EC route before starting (see the box below). Do not treat a hospital's local approval as a substitute.
Step 13 — Set up vigilance, post-market surveillance and supply-continuity workflows before first sale: PMS system and reporting calendar (Section 11), a serious-incident reporting workflow to the FAMHP, a named VCP on the Belgian side (manufacturer/AR contact and the distributor's and hospitals' contact points), and the Article 10a supply-interruption workflow (notify the FAMHP at least six months in advance, absent exceptional circumstances).
Step 14 — Build the reimbursement and commercial route (separate from registration). Registration makes sale lawful; it does not create a payer. If the device is an implant or an invasive device, work out its position on the RIZIV/INAMI List — generic tariff, nominative-list application, contract or restricted clinical application — before you price the Belgian launch (Section 9). For hospital sales generally: almost all Belgian hospitals are subject to public procurement law, tenders run through the federal e-Procurement platform and, above EU thresholds, TED; expect selection and award criteria, ESPD documentation and Dutch/French tender documents. For a health app, decide whether you are entering the mHealthBelgium pyramid and at which level (Section 9).
> Clinical investigations and performance studies in Belgium (FAMHP + ethics committee). Under the Royal Decree of 18 May 2021, clinical investigations for conformity purposes (MDR Articles 62 and 74(2)) require FAMHP authorisation and a favourable ethics committee opinion before they start in Belgium; PMCF investigations under Article 74(1) that involve additional invasive or burdensome procedures, and Article 82 investigations, follow the notification/ethics routes set out in the decree. For IVDs, the Royal Decree of 25 September 2022 draws the same distinction: performance studies involving invasive sample-taking, interventional clinical performance studies, studies involving additional invasive procedures or other risks, and companion-diagnostic studies (Art. 58(1) and 58(2), first sentence, IVDR) require prior FAMHP authorisation plus EC review; PMPF studies with additional invasive or burdensome procedures and companion-diagnostic studies using only left-over samples are notification plus EC review (for left-over samples, evidence that informed consent was obtained at collection must be supplied). Substantial modifications are, as a rule, subject to FAMHP notification and EC review. The FAMHP can require modification of a non-compliant study, suspend or terminate it, or revoke an authorisation — after hearing the sponsor. Informed-consent and patient documents must be in the patient's language — French, Dutch or German. Fees: FAMHP financing documents have listed the fee for notification of a commercial clinical investigation of a medical device at €8,982, revised to €10,436.42 in the 2019 financing-law fee tables.
What does NOT exist in Belgium: a FAMHP "approval certificate" for an ordinary CE-marked device, a Belgian CE mark, and a per-product national marketing register on top of EUDAMED of the Italian Repertorio type. The proof of lawful Belgian market access is: CE certificate + EU Declaration of Conformity + EUDAMED registrations + FAMHP portal registration for the Belgian operators in the chain (+ a RIZIV/INAMI list position if you want the insurance to pay for an implant or invasive device). Anyone selling you a Belgian "product approval" beyond this is selling you a filing service, not an approval.
Belgium is, with Finland, one of the two EU markets where nationwide distribution in principle means three national languages for patient-facing device information — and the Belgian rule is written into the Act itself, not left to guidance.
Practical rule for planning: budget and produce full French and Dutch labels and IFU, plus German, for anything that reaches a patient, a lay user, an implant card, a field safety notice or a Belgian authority. English-only supply is defensible only for genuinely professional-only devices, as a documented position, with the national-language version producible on demand. Belgian hospitals operate bilingually in Brussels and monolingually by region elsewhere; procurement and vigilance teams will reject single-language patient material at vendor qualification even where a legal argument exists for professional material.
Electronic IFU (eIFU) follows the EU eIFU regulation framework; the language duty is unchanged by the medium.
| Activity | Typical timeline |
|---|---|
| CE certification via notified body (Class IIa, new submission) | 12–18 months as a realistic planning figure; notified-body capacity remains the critical path EU-wide |
| CE certification (Class III) | 18–24+ months, including the clinical evaluation consultation procedure where it applies |
| Class I self-declaration route | 2–4 months once technical documentation and trilingual labelling are complete |
| EUDAMED actor registration / SRN issuance | Weeks, after competent-authority verification (for Belgian establishments, FAMHP validation). Mandatory from 28 May 2026 for the first four modules |
| EUDAMED device registration (new device, from 28 May 2026) | Must be completed before placing the first unit on the market; allow weeks for data preparation and validation, not days |
| Registration of pre-28 May 2026 devices in EUDAMED | Within 12 months of the 27 November 2025 notice — plan for late November 2026 ; the FAMHP portal registration does not discharge this duty |
| EUDAMED data confirmation (Belgian-enforced) | Accuracy confirmed no later than 1 year after first submission, then every 2 years; FAMHP may suspend activities until a confirmed breach is cured |
| FAMHP portal registration (Belgian operators) | No statutory processing time verified in current public FAMHP guidance consulted for this guide ; treat it as a project of weeks including KBO/BCE and mandate paperwork, and complete it before first Belgian supply through that operator |
| Free Sale Certificate / export certificate (FAMHP) | Allow several weeks for a complete application (declaration of conformity, CE certificate where applicable, ISO 13485 certificate where applicable) |
| Clinical investigation (MDR Art. 62/74.2) | MDR Art. 66–74 timetable (validation ~10 days; assessment within 45 days, extendable), routed in Belgium as FAMHP authorisation + ethics committee review under the 18 May 2021 decree, plus local feasibility at each centre |
| RIZIV/INAMI nominative-list application (implants/invasive devices) | An evidence project, not a filing: dossier built on the PICO-structured application form (population, intervention, comparator, outcome) introduced in 2023, commission review and ministerial/Royal Decree steps for list changes. Allow 6–18 months as a planning range |
Transition deadlines for legacy (MDD/AIMDD) devices (Regulation (EU) 2023/607) — the same EU-wide dates apply in Belgium:
Eligibility conditions: continued directive compliance, no significant change in design or intended purpose, no unacceptable risk, an MDR-compliant QMS in place by 26 May 2024, a formal notified-body application by 26 May 2024, and a written notified-body agreement by 26 September 2024. IVDR legacy deadlines (Regulation (EU) 2024/1860) are staggered by class: Class D by end-2027, Class C by end-2028, Class B and sterile Class A by end-2029.
Article 59 derogations — a Belgian route worth knowing. Belgium has made comparatively active use of MDR Article 59 derogations (authorising, on duly justified request and in the interest of public health or patient safety, the placing on the Belgian market of a device that has not completed the conformity procedure), a route that gained popularity during Covid-19 and continues to be used where notified-body bottlenecks bite. In Belgium only the manufacturer, the AR or a person duly mandated by them can trigger the procedure, and the implementing decree sets minimum information requirements — in practice the FAMHP's standard and patient-specific application forms demand considerably more. A derogation is Belgian-territory only, time- and device-specific, and not a substitute for CE marking; treat it as an exceptional bridge, documented as such.
Belgium's national filing fees for ordinary devices are modest; the structural Belgian cost that foreign teams miss is the turnover-based annual FAMHP contribution, on top of certification, investigation, translation and reimbursement-dossier costs.
Ethics-committee and per-centre feasibility fees are additional and set locally; budget per centre, not per country.
Currency: all official fees in euros (€).
Legal market access is only the first gate in Belgium. The second gate is how the compulsory insurance pays, and it works completely differently for implants/invasive devices, for other devices and services, and for health apps.
Implants and invasive devices: the RIZIV/INAMI List. Belgium reimburses implants and invasive medical devices through the List of implants and invasive medical devices maintained by RIZIV/INAMI. The List is organised in chapters by clinical domain — among them ophthalmology (A), neurosurgery (B), ENT (C), urology and nephrology (D), abdominal and digestive surgery (E), thoracic and cardiac surgery (F), vascular surgery (G), gynaecology (H), pulmonology/respiratory (I), plastic and reconstructive surgery (J), general surgery (K) and orthopaedics and traumatology (L) — and every device falls into a reimbursement category A–H that determines how it is paid:
| Category | Meaning (per RIZIV/INAMI / MTRC summaries) |
|---|---|
| A | Reimbursed on the basis of an individual price (tariff set per brand) — registration on the nominative list is required |
| B | Reimbursed at an individual price, no nominative-list registration required |
| C | Reimbursed at an individual price only if registered on the nominative list |
| D | Reimbursed at a generic tariff (same tariff regardless of brand), no nominative-list registration |
| E | Reimbursed at a generic tariff only if registered on the nominative list |
| F | Reimbursed on the basis of the selling price, VAT included |
| G | Reimbursed under a Restricted Clinical Application (ACL) — coverage-with-evidence-development arrangement |
| H | Reimbursed under a contract with RIZIV/INAMI |
Practical consequence: for a Category A, C or E device, no nominative-list registration means no reimbursement, however perfect the CE marking. The application form for inclusion, updated in October 2023, is structured on the PICO principle (population, intervention, comparator, outcome) precisely to force an evidence-based comparison against the current standard of care — prepare clinical and socio-economic evidence, a comparator story and a price justification, not a regulatory file. List updates are frequent and technical (for example, in October 2026 RIZIV/INAMI published new material codes for intracoronary imaging catheters, radial EBUS probes and custom-made external aortic support implants, effective 1 November 2026) — a supplier must track code and tariff changes for its own devices as a routine regulatory task.
Other devices and IVD tests: nomenclature and hospital budgets. Devices that are not implants/invasive devices are generally paid inside hospital financing or through the nomenclature of reimbursable services where a service code covers them. IVD laboratory testing, for example, is reimbursed through the nomenclature (usually at 25% of the tariff for individual tests, with exceptions such as certain genetic tests at 100%), supplemented by lump sums per admission, per day of stay and per prescription, plus patient co-payments; specific RIZIV/INAMI projects exist for areas such as NGS and companion diagnostics. Practical consequence mirrors the Dutch one: for most non-implant hospital devices the commercial question is "get specified by clinicians and fit the hospital's budget," while for a genuinely new service or test the gating item is a nomenclature or agreement step at RIZIV/INAMI — an evidence project, not a filing.
Health apps: the mHealthBelgium validation pyramid. Belgium was one of the first European countries to build a structured route for health apps, the validation pyramid on the mHealthBelgium platform (created in 2018; run by beMedTech and Agoria, originally with FAMHP, the eHealth platform and RIZIV/INAMI each owning a level):
Interactions with healthcare professionals — two Belgian statutes, not one code. First, Article 10 of the Medicines Act of 25 March 1964, extended to medical devices, prohibits offering or accepting premiums or benefits in connection with the supply, prescription, dispensing, administration or use of devices, subject to narrow exceptions (benefits of negligible value, and hospitality at scientific events under conditions). Sponsorship of an HCP's participation in a scientific event generally requires a prior Mdeon visa. The FAMHP has specifically criticised hospital tender practices that demand unrelated benefits (training, software, R&D prestige projects) as award factors — benefits with no link to the tendered supplies sit badly with Article 10, on both sides of the table. Second, the Sunshine Act: every pharmaceutical and medical-device company — Belgian or foreign, even with no product on the Belgian market — must document and publish annually on betransparent.be the premiums and benefits granted directly or indirectly to Belgian HCPs, healthcare organisations and patient organisations (service and consultancy fees, event-participation and event-organisation contributions, donations and grants, and R&D amounts in aggregate). Publication takes place by the end of June for the preceding calendar year and the data remain public for three years; a company that granted nothing must still register and file a nil declaration. Non-compliance carries civil fines reported at €1,600 to €120,000. Structure every Belgian HCP payment to be disclosable, document fair-market value, and diary the June deadline — a US- or UK-style speaker programme transplanted unchanged into Belgium is a compliance incident waiting for a date.
Advertising. For devices, Belgian rules (CMS and Chambers guides) are lighter than for medicines but have hard edges: advertising to the general public for active implantable devices is banned, and implantable devices more broadly may be advertised only to healthcare professionals; non-implantable devices may be advertised to the public and to professionals subject to the general misleading-advertising rules of the Code of Economic Law and to MDR Article 7 (no misleading claims about intended purpose, safety or performance); and advertising must not mention the statutory insurance contribution/reimbursement for the device. There is no pre-clearance system for device advertising; the risk is post-hoc enforcement and competitor complaint.
Do I need a Belgian authorized representative? No. You need an EU Authorized Representative established in any member state (MDR Art. 11) if you are established outside the EU. Choosing an AR established in Belgium is a commercial and logistical decision — it can help with the FAMHP, a Belgian notified body and Belgian distributors, but it also makes that AR a Belgian operator for FAMHP portal and vigilance purposes. Decide it deliberately (Section 3).
Is there a Belgian registration on top of CE marking? Yes — two administrative ones, neither of which is an approval: EUDAMED registration (mandatory EU-wide from 28 May 2026; before placement for new devices, by late November 2026 for devices already on the market) and the continuing FAMHP portal registration for Belgian-established manufacturers, authorised representatives, importers, distributors and exporters, which the 2022 Belgian decrees expressly left unchanged. Distributors and exporters register with product information for traceability.
I registered my Belgian distributor with the FAMHP years ago. Am I done? No. The FAMHP registration does not discharge EUDAMED: actors and devices must be in EUDAMED on the 2026 timetable, and EUDAMED data must be re-confirmed within one year and then every two years — in Belgium, on pain of FAMHP suspension of activities. Keep both registrations current.
Does Class I need Belgian registration? In EUDAMED: yes, like every device. With the FAMHP: a Belgian-established Class I manufacturer follows the FAMHP's Class I manufacturer route on the portal, and any Belgian distributor/exporter of the device registers as above. There is no notified body for ordinary Class I (only for Is/Im/Ir), but "self-declared" does not mean "unregistered" in Belgium.
Do I need a Belgian importer licence or a Belgian office? Neither, as a free-standing licence or office. There is no separate Belgian "importer licence" for ordinary devices; a Belgian importer needs KBO/BCE registration, EUDAMED actor registration, FAMHP portal registration and Article 13 systems. No office is required merely to sell into Belgium — but whichever Belgian-established company acts as importer/distributor carries the Belgian duties described in this guide.
How much does Belgian registration cost? EUDAMED: €0. No per-product FAMHP portal fee has been verified in current public guidance. The structural Belgian charge is the annual contribution on previous-year Belgian device turnover. A Free Sale/export certificate was listed at €117.23 (revised from €159) and a commercial clinical-investigation notification at €10,436.42 (revised from €8,982), both to be verified at application. Real costs are notified-body fees (roughly €10,000–€50,000+ by class and complexity, market-reported), EU AR services, ethics/centre fees if you run studies in Belgium, trilingual translations, and the RIZIV/INAMI dossier if you need a list position.
How long does the whole process take? For a new Class IIa device starting from no CE marking: 12–18 months for notified-body certification dominates everything else; EUDAMED and FAMHP portal registrations are weeks of data work once the certificate and UDI data exist, and EUDAMED must be finished before first placement. For an already CE-marked device entering Belgium: EUDAMED + FAMHP portal registration, trilingual labelling and importer/distributor set-up are typically a 1–3 month project — plus, commercially, 6–18 months to a RIZIV/INAMI list position (where needed) and first meaningful hospital revenue.
Can I use a Belgian notified body if I am not Belgian? Yes. Notified bodies serve the whole EU market. Belgium designates SGS Belgium NV (NB 1639) in Antwerp, under the MDR for all device types including Annex XVI products and under the IVDR since 23 April 2025 — but any EU-designated body with the right scope codes in NANDO can certify for the Belgian market. Choose on scope, queue and sector experience, not on flag.
Do labels and IFUs have to be in French, Dutch and German? For user/patient information on devices made available in Belgium: yes — Article 9(1) of the Medical Devices Act requires the Annex I Section 23 information in the three national languages. The sole exception is professional-only devices, where English is permitted — but the professional user may demand the information in the national language of his or her choice, so the national-language version must exist and be suppliable on demand.
How are implants and invasive devices paid for? Through the RIZIV/INAMI List of implants and invasive medical devices, by chapter (ophthalmology to orthopaedics) and category A–H. Categories A, C and E require registration on the nominative list (brand-level, individual or generic tariff as applicable); B and D pay without it; F pays on selling price including VAT; G is a restricted-clinical-application (coverage-with-evidence) route; H is a contract with RIZIV/INAMI. The list application is PICO-structured and evidence-based — it is a market-access dossier, not a registration formality.
My product is a health app. What is different in Belgium? You enter the mHealthBelgium validation pyramid: M1 — CE-marked as a medical device, GDPR-compliant, notified in the FAMHP database; M2 — plus the eHealth platform's ICT criteria (identification/authentication, interoperability, data protection, therapeutic relationship, informed consent); M3 — financing, but only inside a defined care process/convention (e.g. telemonitoring in chronic heart failure), first temporarily (M3-) and definitively (M3+) once socio-economic value is proven. Belgium pays for the care trajectory, not for the app as a standalone prescription.
What is the single most common Belgian mistake in 2026? Treating Belgium as "just another EUDAMED country." The FAMHP portal survived EUDAMED, distributor/exporter registration is a separate Belgian duty, EUDAMED data must be re-confirmed on a Belgian-enforced clock with suspension behind it, patient information is trilingual by statute, and an implant without a RIZIV/INAMI list position has no insured market. Five Belgian facts — all of them must be in the same project plan.
Sources consulted for this guide (October 2026): FAMHP (famhp.be) — Manufacturers pages (Class I, custom-made, systems and procedure packs, Class IIa/IIb/III, IVDs, clinical investigations, certificates of free sale, exceptional use, fees based on turnover in Belgium) and Distributors, importers and exporters pages (distribution-export, updated 23 September 2026), and FAMHP financing-law fee tables (clinical-investigation notification €8,982 → €10,436.42; export certificate €159 → €117.23); Osborne Clarke (3 November 2022, updated 2026) on the Belgian MDR/IVDR decrees of October 2022 — VCP decree amendment, EUDAMED data-confirmation sanctions and FAMHP suspension, Article 59 derogation procedure, performance-studies decree, and the statement that local FAMHP portal registration remains unchanged; Crowell & Moring / Lexology on the Belgian MedTech landscape — Royal Decree of 12 May 2021 implementing the Act of 22 December 2020, materiovigilance contact-point tasks under the Royal Decree of 15 November 2017, EUDAMED confirmation and derogation clarification; ICLG Digital Health 2026 (Belgium) — Medical Devices Act of 22 December 2020, IVD Act of 15 June 2022, Royal Decree of 13 September 2022, FAMHP post-market surveillance role, and the December 2025 Commission proposal to revise the MDR/IVDR; European Observatory on Health Systems and Policies (Belgium country profile) — CE marking and notified-body supervision by the FAMHP, distributor/exporter registration with the FAMHP for traceability (Official Journal 2013/2017/2019), hospital and distributor vigilance contact points, and the 2019 distribution liberalisation; AbroadLink and European Commission language tables on Article 9(1) of the Belgian Medical Devices Act (French, Dutch and German for user/patient information; English permitted for professional-only devices, with the user's right to demand a national language); SGS — SGS Belgium NV (NB 1639) MDR designation for all device types including Annex XVI products, IVDR designation on 23 April 2025, and its conformity-assessment process (EU QMS certificate for all devices; EU Technical Documentation Assessment for Class III and implantable Class IIb); European Commission / EUDAMED roll-out sources (Decision (EU) 2025/2371; mandatory use of Actor, UDI/Device, Notified Bodies & Certificates and Market Surveillance modules from 28 May 2026; registration of pre-May 2026 devices within twelve months of the 27 November 2025 notice); MTRC market-access summaries for Belgium — RIZIV/INAMI List of implants and invasive medical devices (categories A–H, chapters A–L, nominative list), the PICO-structured list application form (October 2023), IVD nomenclature reimbursement (25% tariff plus lump sums), the October 2026 list update (intracoronary imaging, radial EBUS, external aortic support), and the mHealthBelgium validation pyramid (M1/M2/M3, care-process financing); mHealthBelgium / Osborne Clarke / sector reporting on the pyramid's governance (beMedTech and Agoria; FAMHP for M1, eHealth platform for M2, RIZIV/INAMI for M3; platform continued by the federations after 2023); Belgian Sunshine Act sources (Act of 18 December 2016, Royal Decree of 14 June 2017, betransparent.be FAQ and legal commentary — annual disclosure including nil declarations, foreign companies in scope, publication by end of June, data public for three years, fines €1,600–€120,000) and the FAMHP annual-contribution provision (Act of 18 December 2016, Article 34: maximum 0.4026092% of previous-year Belgian device turnover, minimum €500 per distributor); CMS and Chambers advertising guides for Belgium (ban on public advertising of active implantable devices, implantables advertised only to HCPs, no mention of insurance contribution, Code of Economic Law misleading-advertising rules) and Article 10 Medicines Act / Mdeon commentary (inducement prohibition extended to devices, prior visa for event sponsorship, FAMHP criticism of unrelated benefits in hospital tenders); US International Trade Administration country commercial guide for Belgium (public procurement thresholds and tender publication, local-partner recommendation). Items marked "to be verified" (to be verified) were not confirmed against a primary source at the time of writing and must be checked before filing.
This guide is market-access intelligence, not regulatory or legal advice. FAMHP practice, contribution rates, and RIZIV/INAMI list rules change — professional review is required before business use.
Market Access Strategy Session
A Market Access Strategy Session maps your Belgium route — FAMHP portal and EUDAMED registration, trilingual labelling, the turnover contribution, and the RIZIV/INAMI list strategy.
Professional review is required before business use.
Back to ghmap.io Questions? Talk to our team