Bulgaria Medical Device Registration2026 Field Guide

No Bulgarian CE mark — EU MDR applies directly. Bulgaria adds a class-dependent national filing on top: Class I devices are registered with the BDA within 14 days of placing on the market, Class II devices go through a mandatory BDA notification, and wholesale trading needs a BDA permit. Add a Bulgarian-language rule written into the Act, and a payment system split between NHIF clinical pathways and the Ordinance No. 7 list.

Class I registration deadline after placing
14 days
of GDP spent on health (2023)
7.9%
hospitals (NSI, end-2024)
319

Sourced from the BDA, the Medical Devices Act, the NSI and the OECD, and verified 2026 publications.

Download the guide

Enter your work email and we'll send the PDF straight to your inbox.

Professional review is required before business use. See our privacy policy.

What's inside

Eighteen pages, zero fluff

Every figure in this guide traces to a named source — the BDA, the Medical Devices Act, Ordinance No. 7 of 2021, the NSI, the OECD, and verified 2026 publications.

EU MDR + the Bulgarian layer

The Medical Devices Act (ZMI, 2007) frames the national layer: the Bulgarian Drug Agency (BDA) controls placing on the market, putting into service and trade, issues clinical-investigation and wholesale permits, and can suspend, prohibit and withdraw non-conforming devices. A CE file alone completes none of the Bulgarian steps.

EUDAMED + BDA filing + wholesale permit

Actor registration in EUDAMED is mandatory since 28 May 2026 — but Bulgaria adds a class-dependent national filing: Class I registration with the BDA no later than 14 days after placing on the market, a mandatory notification for Class II devices, and a BDA wholesale-trade permit decided within 30 days for anyone trading wholesale.

Bulgarian language + NHIF access

Article 10(3) of the Act requires the instruction for use also in Bulgarian, and the BDA's notification review tests labels and IFUs. And hospital revenue is decided inside NHIF clinical pathways and public tenders, while devices payable with public funds follow the Ordinance No. 7 list — in euro since 1 January 2026.

Complete guide

The guide in full

Every section of the PDF on one page. Use the form above to keep a copy.

Overview

Last updated: November 2026. The EU Medical Device Regulation (MDR, Regulation (EU) 2017/745) and the In Vitro Diagnostic Regulation (IVDR, Regulation (EU) 2017/746) apply directly in Bulgaria. The Bulgarian national layer is built around one framework statute — the Medical Devices Act (Zakon za meditsinskite izdeliya — ZMI, promulgated in State Gazette No. 46 of 12 June 2007, as amended, most recently cited amendment State Gazette No. 41 of 10 May 2024) — and the ordinances adopted under it, in particular the Ordinance on the essential requirements and the conformity assessment procedures for medical devices (adopted under Article 18 of the Medical Devices Act), Ordinance No. 7 of 31 March 2021, which governs the list of medical devices that may be paid for with public funds through the National Health Insurance Fund (NHIF), and Ordinance No. 9 of 1 December 2015 on health technology assessment. Around that core sit the Health Act, the Health Insurance Act (which governs public funding of devices), the Public Procurement Act (which governs how public hospitals buy), the Tariff of Fees Collected under the Medical Devices Act (the BDA fee tariff), and — since 1 January 2026 — the euro framework: Bulgaria adopted the euro as its official currency on 1 January 2026 at the irrevocable conversion rate of EUR 1 = BGN 1.95583, and the lev ceased to be legal tender on 1 February 2026 (source: European Commission, "Bulgaria and the euro", economy-finance.ec.europa.eu). Always confirm critical steps against current Bulgarian Drug Agency (BDA) guidance and the consolidated texts published in the State Gazette (Darzhaven vestnik) before filing: Bulgarian device guidance is issued largely through the Act, its ordinances and BDA practice, fee amounts originally set in lev are being restated in euro, and the current consolidated wording at the date of filing governs.

Bulgaria is a hospital-heavy, import-dependent market with low health spending by EU standards and an unusually dense hospital network for its population. The population was 6,445,481 in 2024 (source: OECD / European Observatory on Health Systems and Policies, Bulgaria: Country Health Profile 2025, demographic table, data from Eurostat). Bulgaria devoted 7.9% of GDP to health in 2023, well below the EU average of 10.0%, and public funding covered only 63% of health expenditure against an EU average of 80%, with nearly all of the remainder paid out of pocket (source: Bulgaria: Country Health Profile 2025, OECD Data Explorer and Eurostat data for 2023). Health expenditure per capita was about half the EU average after adjusting for differences in cost of living (source: Bulgaria: Country Health Profile 2025). Capacity is hospital-centred to a degree unmatched in the EU: at 31 December 2024 Bulgaria had 319 hospitals with 53,806 beds — 203 public and 116 private — equal to 870.9 hospital beds per 100,000 population (source: National Statistical Institute (NSI), "In-patient, out-patient and other health care establishments as of 31.12.2024"); the OECD records 8.2 hospital beds per 1,000 population, almost double the OECD average of 4.3, with hospitals accounting for 37% of health spending in 2022 (source: OECD Reviews of Health Systems: Bulgaria 2026). The physician workforce is comparatively dense: 30,149 physicians at the end of 2024, or 46.8 per 10,000 population (source: NSI, as above). The device sector is supplied predominantly by imports — published analysis puts the market above EUR 300 million with imports above 90% (source: peer-reviewed review, "Medical devices in Bulgaria", 2025 — an analytical estimate, not an official statistic). For a foreign manufacturer, the EU conformity file is necessary but not sufficient: the Bulgarian layer decides who must register or notify what to the BDA, who may trade wholesale and on what permit, and the payment layer — NHIF clinical pathways, the Ordinance No. 7 list and public procurement — decides whether the launch earns money.

Regulatory reality check up front: Bulgaria is not a separate certification regime from the EU. There is no "Bulgarian CE mark" and the Bulgarian Drug Agency (Izpalnitelna agentsiya po lekarstvata — BDA) does not approve medical devices in the sense of a pre-market product authorisation: the manufacturer carries out the conformity assessment under the MDR or IVDR, involving a notified body for the higher risk classes, and affixes the CE mark on its own responsibility. The BDA is the competent authority for the placing on the market, putting into service and trade of medical devices: it controls conformity, evaluates conformity-assessment matters, issues permits for clinical investigations and for wholesale trade in medical devices, and its Executive Director may suspend or prohibit the placing on the market or putting into service of non-conforming devices and order their withdrawal (source: CMS, Legislative frameworks to combat illicit trade in medical products — Bulgaria). What Bulgaria adds — and what foreign manufacturers most often underestimate — is a national layer with six specific features:

  1. 1. A registration and notification system that is class-dependent, not uniform. Persons placing Class I medical devices or custom-made active implantable devices on the market must apply for registration with the BDA no later than 14 days after the product is placed on the market and/or put into operation in Bulgaria (source: Legal 500, Pharmaceutical & Medical Devices Comparative Guide — Bulgaria). CE-marked Class IIa, IIb and III devices already placed on the market in the EU are not subject to prior registration or a prior permit before placement in Bulgaria — but a mandatory notification procedure to the BDA applies to Class II devices, under which the BDA verifies the conformity documents, the validity of certificates and the compliance of labels and instructions for use (source: sector guidance on the Class IIa/IIb notification procedure, IV Law Firm). One rule does not fit all classes: map the Bulgarian filing by class before the launch, not after it.
  2. 2. Wholesale trade is a permitted activity, decided in 30 days. The BDA is competent to issue permits for wholesale trade in medical devices; manufacturers and importers may wholesale only the products included in their manufacturing or import permit, and wholesalers may not retail (source: Legal 500, as above). The permit procedure starts with an application to the BDA, and within 30 days of submission the Executive Director issues the permit or a reasoned refusal (source: sector guidance on the wholesale permit procedure, IV Law Firm). A distribution model that is lawful in another member state does not authorise wholesale trading in Bulgaria — the permit attaches to the Bulgarian operator and its premises, storage and transport arrangements.
  3. 3. A Bulgarian-language rule written into the Act itself. Article 10(3) of the Medical Devices Act provides that the instruction for use shall also be written in Bulgarian (source: Medical Devices Act, English text published by the BDA, bda.bg). EU language-requirement compilations record the Bulgarian position as Bulgarian for instructions for use and labelling, with label content in Bulgarian where the device is not intended for professional use only. Plan the Bulgarian pack as a launch deliverable, not a post-launch translation project (Section 6).
  4. 4. A euro changeover that rewrites every lev figure in the file. Bulgaria joined the euro area on 1 January 2026 as its 21st member; the euro and the lev circulated side by side until the end of January 2026 and from 1 February 2026 the lev is no longer legal tender (sources: European Commission, as above; Deutsche Bundesbank, "Bulgaria introduces the euro"). Dual display of prices in lev and euro, compulsory from 8 August 2025, applies until 8 August 2026 (source: European Commission, as above). Every fee, price, tender value and contract amount originally denominated in lev — including the BDA tariff — must be read in its euro restatement at the fixed rate of EUR 1 = BGN 1.95583, and pre-2026 lev figures quoted in older guidance are historical, not current (Section 8).
  5. 5. A payer structure split between clinical pathways and a dedicated list. The National Health Insurance Fund (NHIF/NZOK) is the central payer: hospital care is financed through clinical pathways, while devices payable with public funds outside that logic are governed by the list under Ordinance No. 7 of 31 March 2021, which also grounds the electronic database of those devices (source: peer-reviewed review, as above). A CE file opens neither track (Section 10).
  6. 6. Sanctions with real numbers, and no device price regulation for private sales. Placing or using a device without the required conformity assessment can incur fines of approximately EUR 5,000 to EUR 10,000 (source: CMS, DTx Legal Navigator — Bulgaria), while prices of privately paid devices are not specifically regulated and remain contractual (source: CMS, as above). The public price is decided by NHIF and procurement logic, not by a price approval.

A note on institutions: the BDA is the specialised state regulatory body under the Minister of Health — for devices it controls placing on the market, putting into service and trade, and issues clinical-investigation and wholesale permits (source: CMS, as above). The Ministry of Health holds legislative competence and organises public tenders in its sphere; the NHIF is financial — it reimburses devices to insured persons and finances hospital care (sources: Legal 500, as above). The National Council on Prices and Reimbursement registers and regulates medicinal-product prices — a medicines machinery; do not import it into a device plan. The NSI publishes the hospital and workforce statistics behind any serious Bulgarian market sizing. No domestic notified body for the mainstream MDR/IVDR scopes could be verified from the sources used — verify the live position in NANDO/EUDAMED; any EU notified body may be used.

1. The Regulatory Framework: EU MDR + Bulgarian National Layer

1.1 The EU layer (directly applicable)

  • Regulation (EU) 2017/745 (MDR) — applicable since 26 May 2021: classification (Annex VIII), General Safety and Performance Requirements (Annex I), conformity assessment (Article 52), CE marking, UDI (Articles 27–28), EUDAMED registration (Articles 29–31), clinical evaluation and investigations (Articles 61–82), post-market surveillance and vigilance (Articles 83–92), the prohibition of misleading claims (Article 7), and the supply-interruption notification duty in Article 10a.
  • Regulation (EU) 2017/746 (IVDR) — applicable since 26 May 2022, with staggered transition for legacy IVDs.
  • Regulation (EU) 2023/607, as amended by Regulation (EU) 2024/1860 — extended MDR/IVDR transition for legacy devices (to end-2027 / end-2028 for devices, and end-2027 to end-2029 by IVD class), subject to the conditions in Article 120 MDR / Article 110 IVDR, including a formal application to a notified body and a signed written agreement within the deadlines those regulations set.
  • EUDAMED mandatory modules — since 28 May 2026, use of the EUDAMED Actor, UDI/Device and Certificate modules is mandatory across the EU, Bulgaria included. Devices placed on the market before 28 May 2026 that continue to be placed on the market must complete device registration in EUDAMED within the transitional window ending in late November 2026, and notified bodies must have their certificates registered in EUDAMED by 28 May 2027 (sources: European Commission, EUDAMED overview; Commission Decision (EU) 2025/2371 — confirm the live timetable at filing).

1.2 The Bulgarian layer

Instrument What it does

--- ---

InstrumentWhat it does
Medical Devices Act (ZMI), State Gazette No. 46/12 June 2007, as amended (last amendment cited in the consolidated text: SG No. 41/10 May 2024)The main Bulgarian statute for devices: regulates the terms and procedures for placing devices on the market and putting them into service, the designation of conformity-assessment bodies, clinical investigations, and trade in devices; Article 10 places responsibility for placing on the market on the manufacturer, requires a written authorised representative where the manufacturer is established outside the EU/EEA, and provides in Article 10(3) that the instruction for use shall also be written in Bulgarian; manufacturers or their authorised representatives must ensure safe installation and guaranteed maintenance where the device's character requires it (Article 17).
Ordinance on the essential requirements and conformity assessment procedures (under Article 18 of the Medical Devices Act)The implementing ordinance on essential requirements and the procedures for assessing conformity with them, including the software provision cited at Article 39a for devices incorporating or consisting of software.
Tariff of Fees Collected under the Medical Devices ActThe BDA fee schedule for device procedures, including the wholesale-trade permit. Amounts were historically set in lev; since 1 January 2026 they fall to be read and paid in euro at the fixed conversion rate — take every amount from the live euro tariff at filing (Section 8).
Ordinance No. 7 of 31 March 2021 (Ministry of Health)Regulates the list of medical devices that can be paid for with public funds through the NHIF and creates the legal basis for the electronic database of those devices — the core text for the outpatient/listed-device payment track (Section 10).
Ordinance No. 9 of 1 December 2015 (Ministry of Health)Sets the conditions and procedures for conducting health technology assessment (HTA) in Bulgaria — relevant where a device or procedure seeks a public-funding position that turns on assessed value.
Ordinance No. 25 of 10 November 2008 (Ministry of Health)Regulates the emergency cases in which devices may be put into service under special circumstances (life-threatening situations) where alternatives with valid CE marking are lacking — an exception regime, not a launch route.
Health Act and Health Insurance ActFrame the healthcare system and the public funding of medicines and medical devices, including the NHIF's financing of hospital care through clinical pathways (Section 10).
Public Procurement ActGoverns public procurement procedures for public expenditure on products, including hospital purchasing of devices (Section 10).
Euro Adoption framework (Council approval of 8 July 2025; conversion rate EUR 1 = BGN 1.95583)Bulgaria adopted the euro on 1 January 2026; the lev ceased to be legal tender on 1 February 2026; dual price display applies until 8 August 2026 (source: European Commission, "Bulgaria and the euro").

1.3 Who does what

The BDA controls the placing on the market, putting into service and trade of devices, evaluates conformity-related matters within its competence, issues permits for clinical investigations and for wholesale trade in devices, and supervises advertising of medicines — for devices, no specific licence, approval or fee is required to advertise to either the general public or healthcare professionals under the framework summarised by CMS (source: CMS, Expert Guide to Advertising of Medicines and Medical Devices — Bulgaria). The BDA Executive Director may order the suspension or prohibition of placing on the market or putting into service, and the withdrawal, of non-conforming devices that may endanger the health and safety of patients, professionals or third parties (source: CMS, anti-counterfeiting guide — Bulgaria). The Ministry of Health drafts legislation, adopts the ordinances in Section 1.2 and organises tenders in its sphere; the NHIF finances and reimburses; the National Council on Prices and Reimbursement is a medicines body whose device relevance is routinely overstated by entrants — there is no equivalent device price-registration step for privately paid devices (source: CMS, DTx Legal Navigator — Bulgaria). Verify any domestic notified-body designation live in NANDO before relying on it.

2. Device Classification

Classification in Bulgaria is EU classification. Medical devices are classified into Classes I, IIa, IIb and III under the 22 rules of MDR Annex VIII, taking account of intended purpose, invasiveness, duration of contact and the vulnerability of the body structures involved; IVDs are classified into Classes A, B, C and D under the seven rules of IVDR Annex VIII. There is no Bulgarian re-classification and no national class list: a device that is Class IIb under the MDR is Class IIb in Sofia. The practical classification questions in Bulgaria are therefore the EU questions — software qualification and Rule 11, substance-based devices under Rule 21, borderline products — with two Bulgarian overlays.

The first overlay is the class-dependent Bulgarian filing. The national registration and notification duties described in Section 4 turn on the MDR class: Class I (and custom-made active implantable) devices carry a BDA registration duty within 14 days after placing on the market; Class II devices carry the BDA notification procedure; CE-marked higher-class devices already on the EU market carry neither a prior registration nor a prior permit for placement itself. Classify the device for the EU file, then map the class to the Bulgarian filing — the two exercises are linked but not identical, and running only the first is the classic Bulgarian omission.

The second overlay is software and the national ordinance. Bulgarian implementing law addresses software expressly: Article 39a of the Ordinance on essential requirements deals with devices incorporating software and software that is itself a medical device, requiring validation in accordance with the state of the art, taking account of lifecycle development, risk management, verification and validation (source: CMS, DTx Legal Navigator — Bulgaria). A software-as-a-medical-device launch into Bulgaria should therefore evidence the validation file against both MDR Rule 11 and the national ordinance text.

MDR classTypical conformity routeNotified body?Bulgaria note
------------
Class I (non-sterile, non-measuring, non-reusable surgical)Self-declaration, MDR Annexes II–III documentationNoRegistration with the BDA no later than 14 days after the device is placed on the market and/or put into operation in Bulgaria (source: Legal 500 — Bulgaria). Instructions for use in Bulgarian (Act, Article 10(3)).
Class I sterile / measuring / reusable surgical (Is, Im, Ir)Annex IX or XI with limited notified-body involvementYes, limited scopeSterile/measuring claims must match the certificate exactly in EUDAMED, the Bulgarian notification file and tender data. Confirm the class-specific Bulgarian filing position at launch — to be verified against live BDA practice for the Is/Im/Ir sub-classes.
Class IIaAnnex IX (or Annex XI route)YesMandatory notification procedure to the BDA: application with device and manufacturer data, EU conformity documentation; the BDA verifies documents, certificate validity and label/IFU compliance (source: sector guidance, IV Law Firm).
Class IIbAnnex IX (or Annex X + XI route)YesSame Class II notification procedure. Includes many implantable and energy-delivering devices; installation and maintenance duties under Article 17 of the Act follow the installed base.
Class IIIAnnex IX with design examination scrutinyYesCE-marked Class III devices already placed on the EU market are not subject to prior Bulgarian registration or a prior placement permit (source: sector guidance, IV Law Firm) — but wholesale trading still needs the Section 3 permit, and vigilance exposure is highest.
IVD Class A–D (IVDR)Self-declaration (A, non-sterile) up to full Annex IX with EU reference laboratory involvement (D)By classThe same operator and language analysis applies; confirm the IVD-specific Bulgarian filing position at filing — to be verified against the live BDA procedure for the class in question.

Legacy devices deserve a warning. Devices under the Regulation (EU) 2023/607 transition may be placed in Bulgaria only while every condition holds, with the EUDAMED timetable (Section 4) running in parallel. A lapsed notified-body agreement or a significant change ends transition cover EU-wide — and Bulgarian tenders increasingly test MDR certificates or documented transition status alongside the bidder's permit and Bulgarian documentation.

3. The Authorised Representative, Importer and Distributor in Bulgaria

Bulgarian law works with the MDR's economic-operator chain and adds two Bulgarian elements at the national end of it: the BDA registration/notification duties of Section 4 and the BDA wholesale-trade permit for the trading activity itself. The roles must not be blurred: the same company may hold several roles, but each role carries its own duties, its own EUDAMED registration where the MDR requires one and, in Bulgaria, its own place in the permit analysis.

3.1 The authorised representative

A manufacturer established outside the EU/EEA must authorise in writing an authorised representative (AR): under Article 10(2) of the Medical Devices Act, where the manufacturer is not established in a Member State or an EEA state, the AR's particulars are additionally specified on the packaging and in the instructions for use of devices imported from third countries into the EU/EEA (source: Medical Devices Act, English text, BDA). At EU level the AR is designated by written mandate (MDR Article 11), verifies the declaration of conformity and technical documentation, keeps the documentation available, cooperates with competent authorities and registers as an actor in EUDAMED, validated by the competent authority of its member state of establishment — for a Bulgarian-established AR, the BDA. An AR mandate does not itself confer the right to trade wholesale in Bulgaria: an AR that also distributes needs the Section 3.4 permit for that activity. A non-EU manufacturer cannot register as an EUDAMED actor until its AR is registered, so the AR appointment is on the critical path for every Bulgarian launch from outside the EU.

3.2 The importer

The importer — the EU-established person who places a device from a third country on the EU market (MDR Article 2(33)) — verifies CE marking, labelling, UDI and EUDAMED registration, checks storage and transport conditions and registers as an actor in EUDAMED (MDR Articles 13 and 31). Where the importer operates in Bulgaria, three consequences follow. First, manufacturers and importers may wholesale only products included in their manufacturing/import permit position (source: Legal 500 — Bulgaria). Second, the importer is the BDA's first point of contact for the Class II notification file, surveillance and vigilance follow-up, and needs Bulgarian-language capability and a functioning traceability file, not merely a warehouse. Third, manufacturers or their authorised representatives must ensure safe installation according to the instructions for use and guarantee maintenance where the device's character requires it (Article 17, Medical Devices Act) — evidence the arrangement in the launch file, not after the first delivery.

3.3 The distributor

Distributors — the supply-chain operators who make devices available without importing them from a third country (MDR Article 2(34)) — carry the Article 14 verification duties: CE marking, declaration of conformity, labelling and language, UDI, storage and transport conditions, and a bar on supplying products they believe to be non-compliant. Distributors do not register as actors in EUDAMED. In Bulgaria, however, the trading activity itself is permitted: wholesale trade in medical devices is subject to a permit issued by the BDA (Section 3.4), and retail sale of devices runs through pharmacies and drugstores under the national trade framework (source: Legal 500 — Bulgaria). Two practical consequences follow. First, sequencing: the wholesale permit belongs before wholesale trading commences. Second, channel discipline: a wholesale permit is not a retail authorisation — wholesalers may not retail products (source: Legal 500, as above) — so a direct-to-patient or own-shop model must be re-checked against the Bulgarian trade framework before it is switched on.

3.4 The wholesale-trade permit

The permit regime, in the structure recorded by sector guidance on the Medical Devices Act procedure (source: IV Law Firm, "Marketing Authorization for wholesale of medical devices"), runs as follows. A permit for wholesale trade in medical devices in Bulgaria may be issued to individuals and to trading companies. The procedure starts with an application to the BDA by the person seeking the permit or their authorised representative, accompanied by the supporting documentation — including a declaration that the applicant owns or has means of transport ensuring proper storage during distribution and transportation, the address of any premises for the storage of medical devices in Bulgaria, and proof of the paid fee. Within 30 days of submission, the Executive Director of the BDA issues the permit or, where the requirements are not met, a reasoned refusal. A fee is payable under the Tariff of Fees Collected under the Medical Devices Act; the amount recorded in pre-euro sector guidance is BGN 5,000 — since 1 January 2026 that figure falls to be read in euro at the fixed rate (EUR 1 = BGN 1.95583, i.e. approximately EUR 2,556), and the current euro amount must be taken from the live BDA tariff at filing (Section 8). The permit attaches to the operator and the products and arrangements it covers: manufacturers and importers wholesale only products included in their permit position, so a scope question (new product groups, new premises, new storage or transport arrangements) is a filing question, not an assumption. Collect the permit evidence before first shipment, not after first invoice.

A direct-sales model needs the same care: a foreign manufacturer's own Bulgarian subsidiary is the operator that must hold the wholesale permit, with its own premises and transport declarations. The permit does not travel with the goods — a logistics provider's permit does not cover the seller's own wholesale activity.

4. The Registration Architecture: EUDAMED + BDA Registration/Notification + Wholesale Permit

Bulgaria runs a three-element architecture, and the elements answer different questions. EUDAMED carries the EU actor and device registration for manufacturers, authorised representatives and importers. The BDA registration and notification procedures carry the national record of what is placed on the Bulgarian market, class by class. The wholesale permit authorises the Bulgarian operator's trading activity. None of the three replaces either of the others.

4.1 Register in EUDAMED (mandatory from 28 May 2026)

Actor registration comes first: the manufacturer (with its AR, where applicable) and the importer register and receive their Single Registration Numbers (SRNs), validated by the competent authority of the member state of establishment — for Bulgarian-established actors, the BDA. Devices are then registered in the UDI/Device module with their Basic UDI-DI, UDI-DI and EMDN nomenclature before being placed on the market. Devices placed on the market before 28 May 2026 that continue to be placed must complete registration within the transitional window ending in late November 2026, and notified-body certificates follow the timetable in Section 1.1 (sources: European Commission, EUDAMED overview; Commission Decision (EU) 2025/2371). Keep the identifiers identical across EUDAMED, the BDA notification file, the wholesale permit file, the Bulgarian labelling and any NHIF or tender submission: name, model, reference and UDI data that differ between systems are the single most common self-inflicted delay in Bulgarian launches.

4.2 The BDA registration and notification procedures

The national procedures, in the structure recorded by the comparative and sector guidance used for this guide (sources: Legal 500 — Bulgaria; IV Law Firm, Class IIa/IIb notification guidance), run as follows:

  • Class I and custom-made active implantable devices — registration. Persons placing these devices on the market apply for registration with the BDA no later than 14 days after the product is placed on the market and/or put into operation in Bulgaria. The registration application is submitted to the Executive Director of the BDA with the device and manufacturer data set: commercial-register standing, device name, type and model, intended purpose and description (in Bulgarian and English in the notification practice described below), the conformity documentation and declaration, the instructions for use in Bulgarian, and proof of the paid tariff fee (source: registration-dossier summaries of the Medical Devices Act procedure — confirm the live data set with the BDA at filing).
  • Class IIa and IIb devices — notification. CE-marked Class II devices already placed on the market in the EU are not subject to prior registration or a prior permit before placement in Bulgaria, but a mandatory notification to the BDA applies. The procedure starts with an application form to the BDA providing detailed data on the device and its manufacturer, the wholesale distributor's details, and device data including the device name, a brief description of the device's designation in Bulgarian and English, its composition, trade name and catalogue number, accompanied by documents including the EU certificate of conformity. On submission, the BDA reviews whether the device holds the necessary documents and authorisations for distribution, whether the certificates are valid, and whether the labels and instructions for use comply with the legal requirements; once the European and Bulgarian requirements are satisfied, distribution in Bulgaria may proceed, and inconsistencies identified by the BDA must be resolved (source: IV Law Firm, as above).
  • Class III and other CE-marked higher-class devices — no prior placement registration. Class IIa, IIb and III devices manufactured outside Bulgaria have historically been exempt from BDA registration as such, and sector guidance confirms that CE-marked Class IIa/IIb devices already on the EU market need no prior registration or placement permit — with the Class II notification above as the operative national step (source: IV Law Firm, as above). Do not generalise that position into "no Bulgarian filing": the notification, the language pack, EUDAMED and the wholesale permit analyses all still apply, class by class and operator by operator.

The filing discipline is evidential: for every device, the operator should be able to produce the conformity file reference, the chain's EUDAMED SRNs, its wholesale permit where it trades wholesale, and the Bulgarian-language supply file. The live data sets, forms and channels are the BDA's — to be verified at filing (Section 8).

4.3 What registration is not

Neither EUDAMED registration, nor a BDA registration or notification, nor the wholesale permit is a product approval, a quality endorsement or a reimbursement decision. The BDA does not certify devices; notified bodies assess conformity for the higher classes and the manufacturer declares it for all classes. The permit authorises the operator's wholesale activity; the filings document who is placing what, so that surveillance, vigilance and — through the separate NHIF and procurement tracks — payment can function. A plan that treats a registration receipt as the finish line skips the language and payment steps that gate revenue (Sections 5, 6 and 10).

5. Step-by-Step: Placing a Medical Device on the Bulgarian Market

The sequence below assumes a CE-marked device from a manufacturer inside or outside the EU, sold into Bulgaria through a Bulgarian distribution structure — the most common foreign-entrant pattern. Steps 1–3 are EU steps with Bulgarian checkpoints; steps 4–8 are the Bulgarian layer; steps 9–10 decide whether the launch earns money.

  • Step 1 — Confirm classification, conformity status and the Bulgarian filing map. Fix the MDR or IVDR class (Annex VIII) and the conformity route; for legacy devices, confirm every Regulation (EU) 2023/607 transition condition. In parallel, map the class to the Bulgarian filing (Class I registration within 14 days; Class II notification; no prior placement registration for CE-marked higher classes) and map the operators to the wholesale permit analysis (Sections 2–4).
  • Step 2 — Close the EU file. EU declaration of conformity, CE marking, UDI assignment (Basic UDI-DI and UDI-DI), technical documentation, clinical evaluation, risk management and PMS plan in place; notified-body certificates where applicable. Freeze designations, models and references before filing anywhere.
  • Step 3 — Appoint and verify the operator chain. Written AR mandate for non-EU manufacturers, with the AR's particulars on the packaging and instructions for use of third-country imports as Article 10(2) of the Act requires; importer and distributor agreements that allocate the MDR Article 13–14 duties, the Bulgarian permit and notification duties operator by operator, Bulgarian-language duties and vigilance cooperation. Verify each Bulgarian partner's existing wholesale permit — products and activities covered — rather than assuming it.
  • Step 4 — File for the wholesale permit (or verify the existing one). For a new Bulgarian operator, submit the application and supporting documentation (including the transport/storage declaration and premises address) to the BDA and run the procedure to issuance; the Executive Director decides within 30 days of submission. For an existing partner, collect the permit evidence and confirm the launch products are inside its scope; a scope gap is a new filing, sequenced before first trading.
  • Step 5 — Register the actors in EUDAMED. Manufacturer, AR and importer obtain SRNs; Bulgarian-established actors are validated by the BDA. Do this before device registration, so the same SRNs and identifiers flow through every system.
  • Step 6 — Register the devices in EUDAMED and complete the BDA class filing. Basic UDI-DI, UDI-DI, EMDN code, risk class and certificate data entered before placing on the market; legacy devices diarised against the late-November 2026 transitional deadline. Then complete the Bulgarian step for the class: the Class I registration within 14 days after placing on the market, or the Class II notification with the bilingual device description, the EU conformity certificate and the Bulgarian labels and instructions, resolved to the BDA's satisfaction before distribution proceeds.
  • Step 7 — Complete the Bulgarian language pack. Bulgarian instructions for use across the range (Act, Article 10(3)); Bulgarian label content, in particular for devices not intended for professional use only; Bulgarian tender and service documentation. Freeze the Bulgarian texts under change control — they are regulatory documents, not marketing copy (Section 6).
  • Step 8 — Clear the channel. For hospital products, confirm the buyer's procurement route (institutional tender under the Public Procurement Act, framework arrangement, Ministry of Health procedure) and the bidder's eligibility and permit scope. For retail-facing products, confirm the pharmacy/drugstore channel position; a wholesale permit does not authorise retail. For distance selling, re-check the model against the Bulgarian trade framework before the channel opens — to be verified for the specific model at filing.
  • Step 9 — Open the payment track. For products that can be paid for with public funds outside hospital budgets, prepare the Ordinance No. 7 list analysis in parallel with the regulatory track. For hospital products, map the clinical-pathway economics and the tender calendar of the target hospitals (Section 10).
  • Step 10 — Stand up vigilance, servicing and in-operation readiness. Bulgarian-language complaint intake, incident escalation meeting the MDR Article 87 clocks (Section 9), field-safety-notice distribution in Bulgarian, and the installation and maintenance arrangement that Article 17 of the Act requires manufacturers or their authorised representatives to guarantee for devices that need it.

Clinical investigations follow their own track: MDR Articles 62–82, with the BDA issuing permits for conducting clinical investigations of devices in Bulgaria (source: Legal 500 — Bulgaria) alongside the competent ethics-committee structure. Sponsors not established in the EU need a legal representative established in the Union (MDR Article 62(2)). Authority and ethics fees are to be verified in the live tariff at submission (Section 8).

6. Language, Labelling and Information Supplied with the Device

The rule is anchored in the Act itself: under Article 10(3) of the Medical Devices Act, the instruction for use shall also be written in Bulgarian (source: Medical Devices Act, English text, BDA). Bulgaria is a single-language market in which Bulgarian — written in Cyrillic — is the only official language, and EU language-requirement compilations record the Bulgarian position as Bulgarian for instructions for use and labelling, with the label content in Bulgarian where the device is not intended for professional use only. The practical reach of the rule is wider than the sentence suggests:

  • Instructions for use are Bulgarian, full stop. The Act's wording admits no professional-user carve-out for the IFU itself: plan a Bulgarian IFU for every device placed on the Bulgarian market, professional and lay alike. Do not plan an English-only pack for Bulgaria on the strength of another member state's professional-user exception.
  • Labels follow the same logic, with the recorded professional-use nuance. Label information should be in Bulgarian; the compilation position records Bulgarian label content in particular where the device is not intended for professional use only. Apply any narrower position product by product, on a documented basis in the regulatory file — never as a range-wide English-only policy adopted on a sales team's preference. The BDA's Class II notification review expressly tests whether labels and instructions for use comply with the legal requirements (source: IV Law Firm), so the language pack is examined, not merely filed.
  • The notification file itself is bilingual in part. The Class II notification practice records a brief description of the device's designation in Bulgarian and English, alongside the trade name and catalogue number (source: IV Law Firm). Prepare the bilingual device description as part of the filing data set, consistently with EUDAMED and the pack.
  • Tender and service documentation is Bulgarian in practice. Public procurement runs in Bulgarian, and a language defect in a tender file is a formal defect with exclusion risk. Plan Bulgarian operating, maintenance and training documentation as part of the bid, not as a post-award translation project — and remember the Article 17 installation and maintenance guarantee, which is performed in Bulgarian hospitals, in Bulgarian.
  • Symbols stay, text translates. MDR harmonised symbols, UDI carriers and CE marking follow the EU rules; the accompanying text is Bulgarian as above. Translations must be controlled versions of the assessed IFU, updated in the same change cycle. Cyrillic typesetting belongs in the production plan, not in a distributor's inbox the week before launch.
  • Registration data do not replace the pack. EUDAMED entries and BDA filings are authority-facing; they do not discharge the duty to supply Bulgarian information with the device itself. Surveillance tests the physical and digital information supplied with the product, not the register entries.

7. Timelines

Bulgaria publishes no single end-to-end "registration timeline", because most of the clock sits in EU procedures, in the class-dependent BDA filings and in payment tracks rather than in a Bulgarian product approval. The honest timeline is a stack of separate clocks:

PhaseClockSource / status
---------
Notified-body conformity assessment (Classes Is/Im/Ir, IIa, IIb, III; IVD B–D)Months to well over a year by class and body capacity; the dominant critical path for higher classesEU procedure; no Bulgaria-specific statutory clock.
Class I / custom-made active implantable registration (BDA)Application no later than 14 days after the device is placed on the market and/or put into operation in BulgariaLegal 500, Pharmaceutical & Medical Devices Comparative Guide — Bulgaria.
Class II notification (BDA)No prior-placement clock: notification procedure with BDA review of documents, certificate validity and label/IFU compliance; distribution proceeds once the European and Bulgarian requirements are satisfiedSector guidance, IV Law Firm. A BDA decision period for the notification is not stated in the sources used — to be verified at filing.
Wholesale-trade permit (BDA)Executive Director issues the permit or a reasoned refusal within 30 days of submission of the applicationSector guidance on the Medical Devices Act wholesale procedure, IV Law Firm.
EUDAMED actor registration (SRN)Administrative; allow days to a few weeks including validation by the BDA for Bulgarian-established actorsEuropean Commission, EUDAMED framework. Mandatory use since 28 May 2026.
EUDAMED device registrationBefore placing on the market; legacy devices continuing on the market by late November 2026Commission Decision (EU) 2025/2371 timetable.
Bulgarian language packManufacturer-controlled; typically runs parallel to Steps 4–6Section 6. Allow translation, Cyrillic proofreading and print/digital production time in the launch plan.
Public tender cycle (hospital devices)Set by the contracting authority's notice under the Public Procurement Act; framework arrangements run in multi-year cyclesPublic Procurement Act. Missing a framework window can cost a full cycle — diarise expiries per category.
NHIF list / clinical-pathway trackSeparate administrative and budgetary tracks; plan in months, in parallel with registration, not after itOrdinance No. 7 of 2021 (list) and the Health Insurance Act framework (pathways). Current list content, values and procedural periods: to be verified at filing (Section 10).

Two planning rules follow. First, never quote a single "Bulgaria approval time": quote the stack — the notified-body phase is the long pole for higher classes, the 30-day wholesale decision and class filings are short but strictly sequenced, and the 14-day Class I clock runs after placing on the market, making it a diarised deadline rather than a launch gate. Second, start the payment track with the regulatory track: in Bulgaria the payment clock usually decides the launch date that matters.

8. Fees and Costs

This section separates what could be verified from a named source from what could not. ghmap.io does not publish fee figures it cannot source, and Bulgaria's euro changeover makes pre-2026 lev figures historical by definition: every amount below must be re-confirmed against the BDA's live euro tariff at the date of filing.

ItemAmount / positionStatus
---------
EUDAMED actor and device registration (EU level)No EU feeVerified — EUDAMED registration is free of charge (European Commission).
Wholesale-trade permit (BDA)BGN 5,000 under the Tariff of Fees Collected under the Medical Devices Act, as recorded in pre-euro sector guidance; at the fixed conversion rate (EUR 1 = BGN 1.95583) that equals approximately EUR 2,556Amount sourced from sector guidance on the wholesale procedure (IV Law Firm). Current euro amount: to be verified in the live BDA tariff — do not budget the lev figure, or its conversion, as if it were the current schedule.
Class I registration / Class II notification (BDA)A fee in the amount fixed by the tariff accompanies the registration filing; the notification file likewise sits inside the tariff frameworkDuty to pay a tariff fee verified in structure (registration-dossier summaries of the Medical Devices Act procedure). Current euro amounts by class and procedure: to be verified in the live tariff at filing — none is restated here.
Clinical investigation permit (BDA) and ethics reviewAuthority and ethics fees apply within the tariff and committee frameworksTo be verified — take the figures from the current tariff and the ethics committee's schedule at submission.
Notified-body conformity assessmentContractual; by class, scope and bodyQuotation-based. The single largest regulatory cost for higher classes.
Non-compliance exposure (not a fee)Placing a device on the market or using it without the required conformity assessment can incur fines of approximately EUR 5,000 to EUR 10,000Sourced from CMS, DTx Legal Navigator — Bulgaria (accessed for this guide). Re-confirm the current sanction ranges in the consolidated Act at filing.
Private-market device pricingNo specific price regulation for devices paid by private parties; pricing is contractualVerified position (CMS, DTx Legal Navigator — Bulgaria). Public-funds payment follows the NHIF logic in Section 10, not a price approval.
Imports from outside the EUCustoms duty by tariff classification, plus Bulgarian VAT on importation and domestic supply at the rate applicable to the specific productProduct- and date-specific VAT treatment: to be verified with Bulgarian tax advice at filing; this guide states no VAT rate it could not verify as current for the product in question.

The cost lesson is that the visible tariffs are the small part. The material costs are the notified-body programme, the Bulgarian (Cyrillic) language pack, the permitted operator's premises, storage and transport arrangements, tender participation, and the working capital of a system where public funding covers only 63% of health expenditure (source: Bulgaria: Country Health Profile 2025). Confirm every tariff from the live euro schedule.

9. Vigilance, Post-Market Surveillance and Incident Reporting

Bulgaria applies the MDR vigilance system without a parallel national clock. Manufacturers must report serious incidents and field safety corrective actions to the competent authority under MDR Article 87: within 15 days of becoming aware of a serious incident as the standard deadline; within 10 days where the incident led to death or an unanticipated serious deterioration in health; within 2 days of becoming aware of a serious public health threat; and, for incidents similar to previously reported ones showing a significant increase in frequency or severity, under the trend-reporting duty of Article 88. Periodic safety update reports (PSURs) follow Article 86 by class, and post-market surveillance plans and reports follow Articles 83–85. Until the EUDAMED vigilance module becomes mandatory, reports for Bulgaria are filed through the national channel to the BDA in the manner its current guidance prescribes; once the module is mandatory, the same clocks run into EUDAMED. The deadlines do not change with the channel — diarise the clocks, not the portal.

Bulgarian specifics to build into the PMS system:

  • The BDA can remove devices from the market. The Executive Director of the BDA may issue an order suspending or prohibiting the placing on the market or putting into service of devices, and ordering their withdrawal, where they do not conform to the regulatory requirements and may endanger the health and safety of patients, medical professionals or third parties (source: CMS, anti-counterfeiting guide — Bulgaria). A vigilance failure in Bulgaria can therefore interrupt the right to trade, not merely attract a fine.
  • Bulgarian-language field safety notices. Issue the Bulgarian version in the same release cycle as the source notice, distribute it through the channel that actually reaches the installed base, and evidence receipt.
  • Cooperation is contractual. Bulgarian distribution agreements should name the escalation contact, the maximum internal forwarding time in hours — the Article 87 clocks leave no room for a weekly complaints batch — and the Bulgarian-language intake route.
  • Installation and maintenance are vigilance infrastructure. The Article 17 duty — safe installation according to the instructions for use and guaranteed maintenance, owed by manufacturers or their authorised representatives — means the servicing record of the installed base is part of the Bulgarian vigilance file in practice. Repeated in-operation findings on the same model are trend data; hold them where the PMS system can see them.
  • Sanctions frame the risk. Non-compliance carries administrative sanctions, including the fine range of approximately EUR 5,000 to EUR 10,000 recorded for placing or using a device without conformity assessment (source: CMS, DTx Legal Navigator — Bulgaria). Keep the permit, notification and vigilance files consistent: surveillance tests exactly that chain.

10. Market Access in Practice: Procurement and Reimbursement

10.1 Who buys, and how

Bulgaria's delivery system is dominated by hospitals to an extent the OECD flags as structural: 8.2 hospital beds per 1,000 population against an OECD average of 4.3, hospital discharge rates of 325 per 1,000 population in 2022 against an OECD average of 129, and hospitals taking 37% of health spending while ambulatory providers took 15% (source: OECD Reviews of Health Systems: Bulgaria 2026). The NSI counts 319 hospitals (203 public, 116 private) and 2,289 out-patient establishments at the end of 2024 (source: NSI, as above). Buyer behaviour runs through three layers. First, public hospitals buy through public procurement under the Public Procurement Act (source: Legal 500 — Bulgaria): specifications are formal, Bulgarian documentation defects can exclude, and premium positioning must be translated into evaluable advantages. Second, the Ministry of Health organises tenders in its own sphere, including centralised procedures for defined categories. Third, private hospitals — 116 hospitals holding 30.4% of beds at the end of 2024 (source: NSI) — buy commercially, on distributor relationships and service credibility. Map which layer buys your category before you price it.

10.2 Reimbursement: two different worlds

Hospital devices are, as a rule, not reimbursed item by item. Hospital care is financed by the NHIF substantially through clinical pathways — defined bundles of hospital activity paid per case — so the device is bought from the provider's pathway-financed budget, which is why Section 10.1, not a reimbursement list, is the hospital revenue map. There is no separate "device reimbursement application" that unlocks hospital sales; there is a tender, a framework or a direct institutional purchase, and a specification to win. Pathway values and the device cost they must absorb decide the realistic price corridor — obtain the current pathway framework at filing (to be verified per category).

Devices paid with public funds outside the hospital budget are the listed world. Ordinance No. 7 of 31 March 2021 regulates the list of medical devices that can be paid for with public funds through the NHIF, and creates the legal basis for the electronic database of those devices (source: Ministry of Health, via the peer-reviewed review cited in the Introduction). Inclusion is therefore a list-and-conditions question: which device groups the current list covers, at what payment values and under what supply conditions, decides the outpatient and community business case. Health technology assessment under Ordinance No. 9 of 2015 frames the evidence conversation where a funding position turns on assessed value. Because the list and its values are updated through ministerial action, the current content is to be verified against the live Ordinance No. 7 list at filing. For products outside the list and outside hospital budgets, the remaining routes are private institutional purchase and out-of-pocket purchase by patients — a materially larger route in Bulgaria than in most of the EU, given that public funding covers 63% of health expenditure and nearly all the rest is out of pocket (source: Bulgaria: Country Health Profile 2025) — but neither replaces the list or pathway analysis in a launch plan.

10.3 The payer landscape in one paragraph

One device authority (the BDA), one central payer (the NHIF — clinical pathways for hospital care, the Ordinance No. 7 list for listed devices), one rule-maker (the Ministry of Health) and one procurement statute. At 7.9% of GDP with per-capita spending about half the EU average, budgets are contested — but the hospital network is dense and, since 1 January 2026, every contract and tender runs in euro. ghmap.io's feasibility-first sequence fits: verify demand, the pathway or list position and the permit scope before building distribution.

11. Common Mistakes

  1. 1. Treating CE marking as the whole Bulgarian entry. CE conformity opens the EU file; it does not complete the class-dependent BDA registration or notification, does not permit anyone to trade wholesale in Bulgaria, and does not put a device on the NHIF list. The Bulgarian layer is separate and sequenced. 2. Missing the 14-day Class I clock. Class I (and custom-made active implantable) registration is due no later than 14 days after placing on the market — a deadline that is easy to miss precisely because it falls after launch day, when the team's attention has moved on. Diarise it before the first shipment. 3. Assuming CE-marked higher classes need nothing at all. Class II devices carry the mandatory BDA notification, in which the Agency tests documents, certificate validity and label/IFU compliance. "No prior registration" is not "no Bulgarian filing" — and the wholesale permit analysis applies at every class. 4. Trading wholesale on a partner's assumed permit. Wholesale trade needs the BDA permit, decided within 30 days on the operator's own application, premises and transport declarations; manufacturers and importers wholesale only products inside their permit position. Verify each partner's permit scope and collect the evidence before first shipment.
  2. 5. Budgeting from a lev figure. Every pre-2026 lev amount — including the BGN 5,000 wholesale fee in older guidance — is historical since the euro changeover of 1 January 2026. Convert nothing by assumption: take amounts from the live euro tariff at the fixed rate of EUR 1 = BGN 1.95583 and confirm them with the BDA. 6. Launching with an English instructions pack. Article 10(3) of the Medical Devices Act requires the instruction for use also in Bulgarian, and the BDA's notification review tests labels and IFUs. An English-only pack is a surveillance and notification finding, and in tenders a formal defect. 7. Confusing the medicines machinery with the device machinery. The National Council on Prices and Reimbursement regulates medicinal-product prices; privately paid device prices are not regulated, and publicly paid devices follow the NHIF list and pathway logic. A launch plan built on a "device price registration" step that does not exist will wait forever. 8. Expecting hospital reimbursement that does not exist. Ordinary hospital devices are not reimbursed item by item; they are bought inside clinical-pathway-financed budgets through Public Procurement Act procedures. The revenue map is the tender and the pathway economics, not a reimbursement application. 9. Ignoring the Article 17 installation and maintenance duty. For devices whose character requires it, manufacturers or their authorised representatives must ensure safe installation according to the instructions for use and guarantee maintenance in Bulgaria. Selling installed equipment without a Bulgarian servicing arrangement is a compliance gap, not a commercial choice. 10. Treating EUDAMED as the whole registration. EUDAMED actor and device registration does not discharge the BDA class filing or the wholesale permit — and neither of those discharges EUDAMED. Bulgaria runs three elements, not one.

12. Frequently Asked Questions

Q: Is there a Bulgarian CE mark, or a Bulgarian device approval? No. The MDR and IVDR apply directly; conformity is assessed under the EU procedures, with a notified body for the higher classes, and the manufacturer affixes the CE mark. The BDA registers and receives notifications class by class, permits wholesale trade and clinical investigations, and surveils the market; it does not approve devices before marketing in the sense of a product authorisation (Introduction; Section 1).

Q: Who is the Bulgarian competent authority for medical devices? The Bulgarian Drug Agency (BDA), the specialised state regulatory body under the Minister of Health. For devices it controls placing on the market, putting into service and trade, issues wholesale-trade and clinical-investigation permits, and its Executive Director may suspend, prohibit and withdraw non-conforming devices (Sections 1 and 9).

Q: What must be filed with the BDA, and when? It depends on the class. Class I and custom-made active implantable devices: registration no later than 14 days after placing on the market and/or putting into operation in Bulgaria. Class IIa and IIb devices: the mandatory notification procedure, with the BDA reviewing conformity documents, certificate validity and label/IFU compliance. CE-marked higher-class devices already on the EU market: no prior placement registration or permit — but EUDAMED registration, the Bulgarian language pack and the wholesale permit analysis still apply (Sections 2 and 4).

Q: Who needs a wholesale-trade permit, and how long does it take? Any operator trading wholesale in medical devices in Bulgaria, on its own application to the BDA with premises and transport/storage declarations and the tariff fee; the Executive Director issues the permit or a reasoned refusal within 30 days of submission. Manufacturers and importers may wholesale only products included in their permit position, and wholesalers may not retail (Section 3.4).

Q: Does EUDAMED registration replace the BDA filing or the wholesale permit? No. They are parallel elements with different owners and different content: EUDAMED carries EU actor and device registration (validated for Bulgarian actors by the BDA), the BDA procedures carry the national class-by-class record, and the wholesale permit authorises the Bulgarian operator's trading activity. A Bulgarian launch normally needs all three analyses (Section 4).

Q: Must labels and instructions be in Bulgarian? Yes. Article 10(3) of the Medical Devices Act requires the instruction for use also in Bulgarian; label information should be in Bulgarian, in particular for devices not intended for professional use only, and the BDA's notification review tests label and IFU compliance. Tender and service documentation is Bulgarian in practice (Section 6).

Q: What does the wholesale permit cost? Sector guidance records BGN 5,000 under the Tariff of Fees Collected under the Medical Devices Act — a pre-euro figure. Since 1 January 2026 Bulgaria uses the euro at the fixed rate of EUR 1 = BGN 1.95583 (the lev figure converts to approximately EUR 2,556); take the current amount from the live BDA euro tariff at filing. Class registration and notification fees likewise sit in the tariff and are not restated in this guide (Section 8).

Q: How are hospital devices paid for in Bulgaria? Generally not by itemised reimbursement. The NHIF finances hospital care substantially through clinical pathways, and public hospitals buy devices under the Public Procurement Act from pathway-financed budgets. Devices payable with public funds outside that logic follow the list under Ordinance No. 7 of 31 March 2021, maintained with an electronic database (Section 10).

Q: Can I sell medical devices online or direct to patients in Bulgaria? Wholesale permit holders may not retail; retail sale runs through pharmacies and drugstores. A distance-selling model from another member state must be re-checked against the Bulgarian trade and permit position before it opens — to be verified for the specific model (Section 3.3).

Q: Which notified bodies are in Bulgaria? Designations change and no domestic body for the mainstream MDR/IVDR scopes could be verified from the sources used for this guide — verify the current position in NANDO/EUDAMED before contracting. Any EU notified body may be used for a device sold in Bulgaria (Section 1.3).

Sources & scope

  • bda.bg, Bulgarian Drug Agency (BDA) — market control, class registration and notification, wholesale-trade permits and the fee tariff.
  • nhif.bg, National Health Insurance Fund (NHIF) — clinical-pathway financing of hospital care and payment for listed devices under Ordinance No. 7 of 2021.
  • eur-lex.europa.eu, Regulation (EU) 2017/745 (MDR), Regulation (EU) 2017/746 (IVDR), Regulation (EU) 2023/607.

This guide is market-access intelligence, not regulatory or legal advice. BDA practice, tariff amounts in euro, the Ordinance No. 7 list and clinical-pathway values change — professional review is required before business use.

Market Access Strategy Session

Bulgaria on your roadmap? Start with the full picture.

A Market Access Strategy Session maps your Bulgaria route — the BDA class filing, the wholesale-trade permit, Bulgarian labelling, and NHIF clinical-pathway and Ordinance No. 7 list access.

Professional review is required before business use.