Czechia Medical Device Registration2026 Field Guide

No Czech CE mark — EU MDR applies directly. Czechia runs two systems in parallel: EUDAMED for manufacturers, authorised representatives and importers, and SÚKL's national ISZP, where Czech distributors notify their operation and device list before they start. Add a Czech-language rule and a separate categorisation track that decides reimbursement.

systems in parallel: EUDAMED + ISZP
2
four EUDAMED modules became mandatory
28 May 2026
SÚKL reimbursement catalogue for voucher devices
Monthly

Sourced from SÚKL, Act No. 375/2022 Coll., Act No. 48/1997 Coll. and Regulation (EU) 2017/745, and verified 2026 publications.

Download the guide

Enter your work email and we'll send the PDF straight to your inbox.

Professional review is required before business use. See our privacy policy.

What's inside

Twenty-one pages, zero fluff

Every figure in this guide traces to a named source — SÚKL, Act No. 375/2022 Coll., Act No. 48/1997 Coll., Regulation (EU) 2017/745 and Decree No. 377/2022 Coll., and verified 2026 publications.

EU MDR + the Czech layer

Act No. 375/2022 Coll. consolidated the national layer in a single statute, with Decree No. 377/2022 Coll. beneath it: SÚKL is the competent authority — it validates Czech operators in EUDAMED, runs the ISZP notification system, and supervises distributors, importers, servicing persons and manufacturers operating in Czechia.

EUDAMED + the ISZP notification

Manufacturers, authorised representatives and importers register in EUDAMED, mandatory since 28 May 2026 — but Czech distributors, servicing persons and custom-made manufacturers must also notify SÚKL in the national ISZP system before commencing operations, distributors with a device-by-device list. One registration does not replace the other.

Czech language + categorisation

Labelling and instructions for use must be in Czech, with no verified general English-only route for professional users. And a prescription (voucher) device is reimbursed only inside SÚKL's categorisation system — the catalogue SÚKL publishes monthly is what health insurers pay against.

Complete guide

The guide in full

Every section of the PDF on one page. Use the form above to keep a copy.

Overview

Last updated: November 2026. Czechia is a full EU member state, so the EU Medical Device Regulation (MDR, Regulation (EU) 2017/745) and the In Vitro Diagnostic Regulation (IVDR, Regulation (EU) 2017/746) apply directly in Czechia. The Czech national layer is Act No. 375/2022 Coll., on Medical Devices and In Vitro Diagnostic Medical Devices, as amended, with implementing Decree No. 377/2022 Coll. Always confirm critical steps, the current ISZP notification channel and current fee amounts against SÚKL guidance in force at filing before you file.

Czechia is a Central European market of roughly ten and a half to eleven million people, with a compulsory public health-insurance system, a strong hospital sector centred on Prague, Brno and the regional university hospitals, and a domestic medical-technology industry that is unusually capable for a market of this size. For a foreign manufacturer, Czechia is rarely entered on population alone. It is entered because the file, the distributor and the reimbursement logic have to work together here in a way that exposes weak European launches early: a device can be perfectly CE-marked and still not sell in Czechia if the Czech-language set is missing, if the Czech distributor has not notified its activity in the national system, or — for devices prescribed on a voucher — if the product has never been placed into the reimbursement categorisation that Czech insurers actually pay against.

Regulatory reality check up front: Czechia is not a separate certification regime from the EU. There is no "Czech CE mark" and SÚKL — Státní ústav pro kontrolu léčiv (the State Institute for Drug Control) — does not approve medical devices before marketing in the way a non-EU regulator grants a registration certificate. SÚKL is the Czech competent authority for medical devices and IVDs. It validates Czech-established actors in EUDAMED, receives the national activity notifications that Czech law still requires through its Medical Device Information System (ISZP), supervises distributors, importers, servicing persons and manufacturers operating in Czechia, handles vigilance and field safety corrective actions, oversees clinical investigations together with ethics committees, supervises device advertising, and administers the monthly reimbursement catalogue for voucher-prescribed devices. The manufacturer remains responsible for compliance; a conforming device bears the CE mark and can then be traded across the EU/EEA. What Czechia adds — and what foreign manufacturers most often underestimate — is a national layer with six specific features:

1. A single, modern adaptation act — with older acts still visible in old sources. Act No. 375/2022 Coll. is the current framework. It replaced the split regime that preceded it (Act No. 89/2021 Coll. for medical devices and the remains of Act No. 268/2014 Coll. for IVDs). A large share of English-language material still online cites Act 268/2014 or Act 89/2021, the old Register of Medical Devices (RZPRO) and a 15-day device-notification duty and five-year notification validity that belonged to that older regime. Do not build a 2026 Czech launch plan on those citations. Where this guide cannot verify a current figure from a primary source, it says so and marks it to be verified.

2. The national notification that survives is an activity notification, in ISZP — and it falls mainly on the Czech-side operators. Under Sections 8 and 23 of Act 375/2022, distributors, persons servicing medical devices and manufacturers of custom-made devices must notify SÚKL via ISZP before commencing their operations. Manufacturers, authorised representatives and importers, by contrast, provide actor and device data to EUDAMED before placing a device on the market, under Article 31 MDR/ IVDR — SÚKL states this expressly in its current registration-and-notification guidance. A manufacturer established elsewhere in the EU that supplies the Czech market directly is not subject to the Czech operation-notification duty. The practical consequence is sharp: in most foreign launches, the Czech notification risk sits with your distributor and your service provider, not with you — which is why distributor selection in Czechia is a regulatory act, not only a commercial one.

3. Czech is the language of the file. Labelling and instructions for use supplied in Czechia must be in Czech. SÚKL's own ethics-committee guidance, for example, requires the instructions for use in the Czech language for devices used in clinical trials, and the adaptation act carries language requirements for documentation and information, including the statement that must accompany a custom-made device. A file that passed in an English-speaking market will fail here on language alone, exactly as in Portugal or Poland.

4. Prescribing, dispensing, use and servicing are nationally regulated in detail. Act 375/2022 regulates areas the MDR leaves to the member states — in Part 4 (Sections 11–18) and the surrounding provisions: prescribing and dispensing of devices, their use in healthcare, servicing, professional maintenance and safety technical controls, and the instruction/training (instruktáž) that must be given for defined devices. A Czech hospital does not simply "buy and use" a device; the device enters a regulated use-and-maintenance regime, and suppliers who cannot support servicing and instruction in Czechia lose tenders on that ground alone.

5. For voucher devices, reimbursement is a categorisation system run through SÚKL — not an insurer price list. Since 1 January 2019, following a Constitutional Court judgment that struck down the old insurer-code-list system as non-transparent and non-reviewable, reimbursement of medical devices prescribed on a voucher in outpatient care is governed by the Public Health Insurance Act (Act No. 48/1997 Coll.): devices are placed into reimbursement categories in the categorisation tree in Appendix 3 to that Act, each category carrying a reimbursement limit and prescription, indication and volume conditions set in the Act itself, and SÚKL issues the system-wide reimbursement catalogue monthly, which health insurers must respect. Hospital devices — separately billed material (ZUM, zvlášť účtovaný materiál) such as implants, stimulators, valves, catheters and stents — sit outside the voucher system and are paid inside hospital care, increasingly through the CZ-DRG system. Section 8 separates the two tracks.

6. There is no Czech central purchasing body for devices. Unlike Portugal's SPMS, Czech public hospitals buy for themselves (faculty/university hospitals, regional hospitals and institutes) under the Public Procurement Act (Act No. 134/2016 Coll., the ZZVZ). Notices appear in the Journal of Public Contracts (Věstník veřejných zakázek, VVZ) and on the Zakázky GOV portal, and above EU thresholds in TED.

Tender documentation is in Czech and deadlines are hard. There is no single national catalogue to join; there are dozens of hospital buyers to be ready for, one tender at a time.

A note on institutions and on old information: the Ministry of Health (Ministerstvo zdravotnictví, MZ ČR) sets policy and acts as the appellate body in the voucher-reimbursement system. SÚKL, headquartered at Šrobárova 48, 100 41 Prague 10, is the competent authority, the EUDAMED validating authority for Czech-established operators, the market-surveillance and vigilance authority, the ISZP operator, the advertising supervisor and the publisher of the monthly reimbursement catalogue. Ethics committees established at healthcare facilities (and SÚKL's own ethics committee structures for multicentre work) handle the ethics side of clinical investigations. The Office for the Protection of Competition (ÚOHS) supervises public procurement.

ÚNMZ (the Czech Office for Standards, Metrology and Testing) oversees the notified-body designation framework under Act No. 22/1997 Coll. Older sources that describe only RZPRO, Act 268/2014, or device notifications decided by SÚKL with five-year validity predate Act 375/2022 and mandatory EUDAMED — use them only as history.

1. The Regulatory Framework: EU MDR/IVDR (Directly Applicable) + Czech National Layer

1.1 The EU layer (applies directly in Czechia — no transposition step)

  • Regulation (EU) 2017/745 (MDR) — applicable since 26 May 2021: classification (Annex VIII), General Safety and Performance Requirements (GSPR, Annex I), conformity assessment (Article 52), CE marking (Article 20), UDI (Articles 27–28), EUDAMED registration (Articles 29–31), clinical evaluation and investigations (Articles 61–82), post-market surveillance and vigilance (Articles 83–92), the ban on misleading claims (Article 7), and the supply-interruption notification duty in Article 10a (applicable since 10 January 2025, introduced by Regulation (EU) 2024/1860).
  • Regulation (EU) 2017/746 (IVDR) — applicable since 26 May 2022, with staggered transition for legacy IVDs by class.
  • Regulation (EU) 2023/607, as amended by Regulation (EU) 2024/1860 — extended MDR/IVDR transition for eligible legacy devices (to end-2027 / end-2028 for MDR devices, and end-2027 to end-2029 by IVD class), subject to the Article 120 MDR / Article 110 IVDR conditions, including a timely formal application to a notified body and a signed written agreement. A legacy device that does not meet those conditions may not be placed on the Czech market on the strength of an old directive certificate.
  • Annex XVI products without a medical purpose — the MDR common specifications (Commission Implementing Regulation (EU) 2022/2346) apply to listed non-medical products such as coloured contact lenses, cosmetic implants and fillers, liposuction equipment and certain lasers.
  • EUDAMED mandatory modules — Commission Decision (EU) 2025/2371 of 26 November 2025 (published in the Official Journal on 27 November 2025) confirmed four EUDAMED modules as functional.

Since 28 May 2026, use of the Actor Registration, UDI/Device Registration, Notified Bodies & Certificates, and Market Surveillance modules is mandatory EU-wide, including in Czechia. Devices first placed on the market from 28 May 2026 must be registered in EUDAMED before first placement; devices placed on the market before that date that continue to be placed on the market must complete device registration by 28 November 2026. SÚKL validates EUDAMED applications for Czech-established manufacturers, authorised representatives, importers and system/procedure pack producers. SÚKL's current guidance states the same position in Czech terms: the manufacturer, authorised representative and importer provide actor and device data to EUDAMED before placing the device on the market, in compliance with Article 31 MDR/IVDR.

1.2 The Czech layer

InstrumentWhat it does
Act No. 375/2022 Coll., on Medical Devices and In Vitro Diagnostic Medical Devices, as amendedThe current national framework. An adaptation act for both the MDR and the IVDR: it designates the national authorities (SÚKL in the lead), defines offences and administrative fines for breaches of MDR/IVDR and national duties, carries the distributor/servicing/custom-made notification duty via ISZP (notably Sections 8 and 23), sets language requirements for documentation and information, defines rules for ethics committees and performance studies, provides for central registration and evaluation of serious adverse events and field safety corrective actions, sets procedures for devices posing an unacceptable risk, and regulates the areas the Regulations leave to the state — prescribing, dispensing, use and servicing (Part 4, Sections 11–18, and related provisions). It retained the systematic structure of the Act 89/2021 Coll. it replaced.
Decree No. 377/2022 Coll.Implementing decree for certain provisions of Act 375/2022 — the detail layer beneath the Act (forms, particulars and procedures). Check the current wording at filing; implementing decrees are amended more often than the Act.
Act No. 48/1997 Coll., on Public Health Insurance, as amended (notably the 2019 amendment)Governs reimbursement. Its Appendix 3 categorisation tree defines the reimbursement categories for voucher-prescribed devices, with the reimbursement limit and prescription/indication/ volume conditions for each category set in the Act. SÚKL is the regulating body for this system; the Ministry of Health is the appellate body; SÚKL publishes the reimbursement catalogue monthly. The same Act's hospital-care provisions are the background for ZUM / CZ-DRG payment of inpatient devices.
Act No. 40/1995 Coll., on Advertising Regulation, as amendedContains the dedicated medical-device advertising rules introduced alongside the MDR adaptation (Section 9.3): what counts as device advertising, the separate regimes for professionals and the general public, the gift/inducement ban for professionals, and SÚKL's role as advertising supervisor.
Act No. 134/2016 Coll., on Public Procurement (ZZVZ), as amendedHow public hospitals and other contracting authorities buy devices: procedures, the small-scale contract limits (raised with effect from 3 April 2025 to CZK 3,000,000 excluding VAT for supplies and services and CZK 9,000,000 for works), publication in the VVZ and Zakázky GOV, and ÚOHS supervision (Section 8.2). Doğrulanacak: the current above-threshold values in Government Regulation No. 172/2016 Coll. at tender — EU thresholds are revised every two years, most recently for 2026–2027.
Act No. 22/1997 Coll., on Technical Requirements for ProductsThe background act for conformity assessment and for the designation/authorisation framework under which Czech notified bodies operate.
SÚKL guidelines UST-29 and UST-24SÚKL's fee and cost-reimbursement guidelines: UST-29 covers administrative fees and reimbursement of the costs of expert activities; UST-24 covers reimbursement of costs of expert activities performed on request, waivers and refunds. These — not a tariff in the Act — are where a given procedure's current cost is stated (Section 7.2).

1.3 Who does what

  • Manufacturer — full MDR Article 10 duties. If established in Czechia: EUDAMED actor registration validated by SÚKL. If established elsewhere in the EU and supplying Czechia directly: no Czech operation notification (SÚKL states this expressly). If established in Czechia and making custom-made devices:

ISZP notification duty under Sections 8/23, plus the MDR Annex XIII statement for each device, in Czech where the Act so requires.

  • Authorised representative (AR) — mandatory for manufacturers established outside the EU (MDR Article 11), established anywhere in the EU. Czechia imposes no rule that the AR be Czech — but a Czech-established AR is validated by SÚKL in EUDAMED.
  • Importer — MDR Article 13 duties; the first EU operator making a third-country device available. A Czech-established importer registers in EUDAMED, validated by SÚKL. Note the change from the old regime: under Acts 268/2014 and 89/2021 the importer also carried Czech notification duties; under the current SÚKL guidance the importer's registration channel is EUDAMED. Doğrulanacak: confirm at filing whether any residual ISZP duty applies to your specific importer structure — in particular where the same Czech entity acts as both importer and distributor, because the distributor duty in the next bullet then applies to it in that capacity.
  • Distributor — MDR Article 14 duties, plus the Czech layer that catches foreign suppliers out: a distributor must notify its activity to SÚKL via ISZP before commencing operations (Sections 8 and 23 of Act 375/2022), submitting a list of the devices it supplies — for each device: name, intended purpose, risk class and Basic UDI-DI (if assigned). Two exemptions matter commercially: the duty does not apply to a distributor who supplies exclusively risk class I devices or class A IVDs, and it does not apply to a distributor who supplies exclusively to a user who is not a healthcare provider. A pharmacy-care provider is relieved of part of the data duty (UDI-DI and intended purpose) under the Act's distributor provisions. Vet a Czech distributor on its ISZP notification status before signing, not after the first hospital tender asks for it.
  • Persons servicing devices — servicing (including professional maintenance and periodic safety technical controls) is a notified activity: a servicing person must notify SÚKL via ISZP before commencing operations, on the same Sections 8/23 basis, with a parallel exemption for a person servicing exclusively class I devices or class A IVDs. Servicing must be performed by persons with the professional competence the Act requires, following the manufacturer's instructions — a foreign manufacturer's Czech service partner is therefore part of the regulated chain, and an unnotified service partner is a market-access defect, not an administrative detail.
  • Health institutions that manufacture and use devices in-house (MDR Article 5(5)) — subject to the Act's conditions for in-house manufacture and use, documentation and safety reporting. Doğrulanacak: the current SÚKL notification/documentation practice for in-house devices at implementation.
  • Prescribers and dispensers — devices are prescribed and dispensed under Part 4 of the Act: voucher (poukaz) prescription for the outpatient reimbursement track, dispensing by authorised providers (pharmacies and dispensing points, výdejny zdravotnických prostředků), and defined rules on who may prescribe, dispense and instruct the patient for each device type.

2. Device Classification

Czechia applies the MDR classification rules (Annex VIII) unchanged — there is no Czech reclassification layer.

Classification is by intended purpose and inherent risk, across 22 rules in four groups (non-invasive Rules 1–4, invasive Rules 5–8, active Rules 9–13, special Rules 14–22). Where more than one rule applies, the rule resulting in the higher class prevails. Accessories are classified in their own right, and the classification rationale must be documented in the technical documentation. Czech terminology calls the classes riziková třída (risk class) — the term used in the ISZP notification, where the distributor must state the risk class of every device it supplies.

ClassTypical examplesConformity route in brief
I (incl. Is sterile, Im measuring, Ir reusable surgical)Non-invasive low-risk devices, simple dressings, non-sterile examination gloves, simple instrumentsSelf-declaration for plain Class I; notified body involvement limited to the sterile, measuring or reusable-surgical aspect for Is/Im/Ir
IIaDental materials, short-term invasive devices, many monitoring/diagnostic devices, most decision-support software as the starting point under Rule 11Notified body (Annex IX QMS + technical-documentation assessment on at least one representative device per category, or Annex X + XI route)
IIbLong-term invasive devices, most implants not in Class III, active therapeutic devices with potentially hazardous energy exchangeNotified body, with technical-documentation assessment per generic device group
IIIAbsorbable implants, devices incorporating an ancillary medicinal substance, heart valves, breast implants, defined surgical meshesNotified body with full technical-documentation / design examination; expert-panel or clinical-evaluation consultation and, for ancillary medicinal substances, competent-authority/EMA consultation where applicable
IVDs (IVDR)Classes A–D under IVDR Annex VIIIA: self-declare (notified body for sterile A); B–D: notified body, with the heaviest scrutiny for Class D

Two Czech practical notes:

1. Do not import an old MDD class into a Czech filing. Software (Rule 11), substance-based devices and reusable surgical instruments in particular moved up under the MDR. SÚKL, a notified body, an insurer checking the reimbursement catalogue or a hospital buyer checking your EUDAMED record will work from the MDR class — and the risk class your distributor declares in ISZP must match it, because that declaration is part of the distributor's own notification file.

2. Czechia has designated notified bodies of its own — with limited scopes. The NANDO database lists Czech-designated bodies under the MDR including Český metrologický institut (Czech Metrology Institute, NB 1383) and Institut pro testování a certifikaci, a.s. (ITC Zlín, NB 1023). Their designation scopes cover defined device codes only. A manufacturer may use any EU-designated notified body whose NANDO scope covers the device and the route; there is no rule requiring a Czech body for the Czech market. Doğrulanacak: the current NANDO designation scope of any body you name in a plan, at application — scopes change, and neither Czech body's scope should be assumed to cover your device codes.

Software deserves a separate line for Czechia: MDR Rule 11 puts most software that provides information used for diagnostic or therapeutic decisions at Class IIa or higher, and higher where the decision could cause death or serious deterioration. Classify the claim you actually make in the Czech IFU — the Czech text is the text Czech users, SÚKL and a Czech court will read.

3. The EU Authorised Representative, Importer and Distributor in Czechia

A non-EU manufacturer (for example a Turkish, US, UK or Chinese maker — note that Great Britain is a third country for this purpose, while Northern Ireland follows EU goods rules under the Windsor Framework) selling into Czechia needs the standard MDR chain, with Czech specifics at each link:

1. EU Authorised Representative — appointed by written mandate (MDR Article 11), established anywhere in the EU. The AR verifies that the EU Declaration of Conformity and technical documentation have been drawn up, keeps documentation available for SÚKL and other competent authorities, cooperates on corrective actions, and is jointly liable for defective devices on the same basis as the manufacturer where the manufacturer has not complied with its obligations. If the AR is established in Czechia, SÚKL is its validating authority in EUDAMED.

2. Importer — the first EU operator making a third-country device available, with Article 13 verification (CE marking, labelling, UDI, EU Declaration of Conformity, manufacturer and AR identification), storage/ transport, and complaint-register duties. A Czech-established importer registers in EUDAMED, validated by SÚKL.

3. Distributor — Article 14 verification and traceability duties, plus the Czech ISZP layer: activity notification to SÚKL before operations begin, with the per-device list (name, intended purpose, risk class, Basic UDI-DI if assigned), unless an exemption in Section 1.3 applies. This is the single most-missed Czech step in foreign launches, because teams accustomed to markets where "registration" means the manufacturer's filing assume the distributor needs nothing. In Czechia the distributor's notification is the distributor's own legal duty — but its absence becomes the manufacturer's commercial problem the moment a hospital, an insurer or SÚKL asks for it.

4. Servicing partner — if your device needs installation, maintenance, safety technical controls or repair in Czechia, the person performing that servicing must itself be ISZP-notified (Section 1.3) and professionally competent for the work. Put the service chain in place before the first tender: Czech hospital tenders routinely score or require local servicing and instruction capability, and retrofitting a notified service partner after award is slow.

5. Relabelling / repackaging trap (MDR Article 16) — a distributor or importer that translates, relabels or repackages a device for Czechia can step into manufacturer-level duties for that activity, including notified-body certification of its quality system for the activity where required. Czech translation of labelling/IFU should therefore be done under the manufacturer's control wherever possible, with the translator working from the controlled master file — not improvised by the local partner on its own letterhead.

6. PRRC — the Person Responsible for Regulatory Compliance (MDR Article 15) requirement is EU-wide.

Micro and small enterprises may rely on an external PRRC who is permanently and continuously at their disposal.

7. Customs and tax border — Czechia is inside the EU customs union (currency: Czech koruna, CZK — Czechia has not adopted the euro), so a shipment from a German or Polish warehouse to a Prague hospital is an intra-EU supply, not an import. Goods arriving from outside the EU are imported through Czech customs, with import VAT and any duty accounted for at import by commodity code. Doğrulanacak: the current Czech VAT treatment for your device category — Czechia's post-consolidation VAT structure has a standard rate and a single reduced rate, and medical devices fall on different sides of that line by category — and the duty position for your commodity code, at shipment.

Free movement point: a device that is CE-marked and lawfully marketed in another EU country can be brought into Czechia with no additional Czech marketing authorisation. A manufacturer established in another EU member state that has registered via EUDAMED does not need a second Czech "approval," and does not even owe SÚKL an operation notification. The Czech work is therefore not a second certification; it is (a) a compliant Czech distribution and servicing chain, notified in ISZP where the Act requires it, (b) the Czech-language set, and (c) — for voucher devices — placement in the reimbursement categorisation and the monthly SÚKL catalogue that make the device payable by Czech insurers.

4. The Czech Registration Architecture: EUDAMED + ISZP Activity Notification (+ the Reimbursement Catalogue for Voucher Devices)

Czechia currently has two regulatory layers, plus a third payment layer for one device family. Foreign teams routinely conflate them — and routinely import a fourth, dead layer (RZPRO) from old sources. Keep them separate in your launch plan.

4.1 Layer 1 — EUDAMED (EU-wide, mandatory since 28 May 2026)

1. Actor registration — manufacturer, AR, importer and system/procedure pack producers register and receive a Single Registration Number (SRN) / Actor ID after validation by the competent authority of their place of establishment — SÚKL for Czech-established operators. The non-EU manufacturer links to its AR in EUDAMED; that linking can only be completed once the AR has its SRN.

2. Device registration — in the UDI/Device module, at Basic UDI-DI / UDI-DI level with the EMDN nomenclature. Devices first placed on the market from 28 May 2026 must be registered before first placement; devices already on the market that continue to be sold must be registered by 28 November 2026. An actor registration does not register the device: receiving an SRN does not complete device registration.

3. Who files: SÚKL's current guidance is explicit — the manufacturer, authorised representative and importer provide actor and device data to EUDAMED before placing the device on the market (Article 31 MDR/IVDR). Distributors as such do not register devices in EUDAMED; their Czech duty is Layer 2.

4.2 Layer 2 — ISZP national activity notification (live, Czech-specific)

The Medical Device Information System (ISZP, operated by SÚKL; access and manuals via SÚKL's ISZP pages, historically under niszp.cz / iszp.sukl.cz) is the current national system. It replaced RZPRO: data from the previous RZPRO system has been migrated to ISZP, and RZPRO survives only provisionally for certain transitional activities (including field-safety-notice tracking) pending full EUDAMED implementation. Do not file a new Czech launch in RZPRO on the strength of an old guide.

Under Sections 8 and 23 of Act 375/2022:

  • Who must notify: distributors, persons servicing medical devices, and manufacturers of custom-made devices — before commencing operations.
  • Exemptions: a distributor supplying exclusively class I devices or class A IVDs; a distributor supplying exclusively to a user who is not a healthcare provider; a person servicing exclusively class I devices or class A IVDs.
  • How: in ISZP, via the "New notification of an operation" function (SÚKL publishes step-by-step manuals and a video manual for the Person Module / distributor notification). Access credentials for ISZP must be obtained first — build that lead time into the distributor's start date.
  • What is declared: for a distributor, a list of the medical devices supplied to the market — for each device its name, intended purpose, risk class and Basic UDI-DI (if assigned). Keep this list current: adding a new device family to a Czech distributor's range is a notification-update event, not just a new price-list line.
  • Custom-made trap: a maker established outside Czechia whose products genuinely meet the MDR custom-made definition is assessed under the custom-made rules; a maker whose "custom" products are in fact mass-produced devices adapted to a professional user's requirements is not a custom-made manufacturer for this purpose — SÚKL's guidance states that in that case the operation must be registered as a distributor operation in ISZP and all devices supplied to the Czech market declared. Misclassifying an adapted mass-produced range as "custom-made" is a known Czech filing error.

4.3 Layer 3 (payment, voucher devices only) — Reimbursement categorisation and the monthly SÚKL catalogue For devices prescribed on a voucher in outpatient care, a third layer decides whether the device is paid for at all (detail in Section 8.1): the device must sit inside a reimbursement category in the Appendix 3 categorisation tree to Act 48/1997, the supplier notifies SÚKL of the device's classification into that category (the system works on the notification principle — SÚKL does not classify each device by individual administrative decision, but will open proceedings if it, an insurer or a competitor establishes that a device was classified incorrectly), and the device then appears in the reimbursement catalogue SÚKL issues monthly, which insurers must respect when paying on vouchers. This layer is not a marketing authorisation and does not apply to hospital/ZUM devices — but for a voucher product, skipping it is commercially fatal: a CE-marked, ISZP-clean voucher device that is not in the catalogue is a device the patient pays for entirely out of pocket, against competitors the insurer reimburses.

4.4 Registration is not approval

Notification in ISZP, registration in EUDAMED and listing in the reimbursement catalogue are notification, validation and payment-identification processes — not a SÚKL assessment of safety or performance, and not an endorsement. Never write "approved by SÚKL" in Czech marketing material: besides being untrue, it collides with MDR Article 7 (misleading claims), the Czech Advertising Act and Czech consumer law.

5. Step-by-Step: Placing a Medical Device on the Czech Market

Step 1 — Confirm the product is a medical device (or IVD) and classify it. Apply MDR Article 2 and Annex VIII / IVDR Annex VIII. Screen Czechia-relevant borderlines early: Annex XVI aesthetic products, software (Rule 11), substance-based devices, and products near the medicinal-product line (SÚKL regulates both medicines and devices, so a borderline argument will be read by an authority that owns both sides of the line).

Step 2 — Build the MDR conformity file. QMS (ISO 13485-aligned), technical documentation (Annexes II– III), GSPR checklist (Annex I), risk management (ISO 14971), clinical evaluation (Article 61, Annex XIV) or performance evaluation for IVDs, PMS plan, Basic UDI-DI and UDI-DI assignment, and the PRRC appointment.

Step 3 — Decide, at the outset, which payment track the device will live on. Voucher/outpatient (categorisation + monthly catalogue), hospital/ZUM and CZ-DRG (hospital budgets and tenders), or purely private/commercial. This decision drives the evidence, pricing and partner work in Steps 7–9 and cannot be retrofitted cheaply — a voucher product launched without its categorisation work done has no payer.

Step 4 — Fix the operator chain. Appoint the AR (non-EU manufacturers) by written mandate; confirm who the importer is; select the Czech distributor and verify its ISZP notification status and the device list declared in it before contract signature; confirm who will perform servicing and instruction in Czechia and that that person is ISZP-notified and competent for the device; document who performs any Czech relabelling/ repackaging and whether Article 16 duties are triggered. Remember the sequencing rule: the AR must have its SRN before the non-EU manufacturer can link to it in EUDAMED.

Step 5 — Conformity assessment. Plain Class I: self-declare. Is/Im/Ir, IIa, IIb, III and IVD Classes B–D: engage an EU-designated notified body with the right NANDO scope (Section 2). Notified-body capacity — not SÚKL — is the usual bottleneck, so apply early, and never lodge a parallel application for the same procedure with two notified bodies.

Step 6 — Register the actors, then the devices, in EUDAMED. Actor registration and SRN for manufacturer, AR, importer and system/procedure pack producers (SÚKL-validated for Czech-established operators); device registration in the UDI/Device module before first placement for devices placed from 28 May 2026 (and by 28 November 2026 for devices already on the market that continue to be sold). Keep the information current thereafter.

Step 7 — Complete the Czech national layer. Distributor ISZP notification (before operations begin, with your devices on its declared list — name, intended purpose, risk class, Basic UDI-DI); servicing-person ISZP notification; custom-made manufacturer ISZP notification if that is genuinely your case. If you are an EU-established manufacturer supplying directly with no Czech distributor, record the SÚKL position that no operation notification is due from you — and make sure no one in the chain files an unnecessary one, or omits a necessary distributor filing in reliance on yours.

Step 8 — Czech labelling and IFU. Produce the language set in Section 6 before the first Czech delivery:

Czech label, Czech IFU, Czech implant-card content where applicable, and Czech field-safety material ready to issue. Allow real translation and review time — this step, not the authority queue, is what most often slips in a Czech launch.

Step 9 — Payment readiness. Voucher track: confirm the Appendix 3 category, the reimbursement limit and the prescription/indication/volume conditions for the category; complete the supplier's classification notification to SÚKL; verify the device's appearance in the monthly catalogue before promising insurers' payment to a dispensing partner. Hospital track: map the target hospitals' procurement calendars, the CZ-DRG / ZUM treatment of your category, and each hospital's internal assessment process. Doğrulanacak: the current notification form, data elements and catalogue cut-off dates for your category at filing — the catalogue is monthly, so a missed cut-off costs a month, not a quarter.

Step 10 — Logistics and tax setup for the first shipment. Intra-EU supply: no customs declaration, normal EU VAT treatment in CZK. Third-country supply: importer of record, customs declaration, import VAT (and any duty) accounted for — agree in writing who carries this and under which Incoterm.

Step 11 — Commercial entry: hospital tender or dispensing channel. Map the buyer: faculty and regional hospitals buying under the ZZVZ (notices in the VVZ / Zakázky GOV and TED), the pharmacy and dispensing-point channel for voucher devices, and the private clinic channel. Monitor by CPV code (Section 8.2).

Step 12 — Only if relevant: clinical investigation in Czechia. Applications go to SÚKL with the opinion of the competent ethics committee — ECRIN's Czech country data records that the ethics committee's approval is required as part of the submission to the competent authority (in practice: ethics first / in parallel, with the EC document in the CA file), that approval is explicit (written), and that the general authorisation timespan is about 60 days within the MDR Article 70 timetable, with substantial-modification timelines of roughly 30–38 days.

Participant-facing documents, summaries and labelling/IFU for the investigation must be in Czech. Fees are charged under SÚKL guideline UST-29 (with UST-24 on cost reimbursement, waivers and refunds); non-commercial sponsors pay the fee and then seek a waiver under that framework — to be verified: the current UST-29 amount for your investigation type at application. A CE-marked device used within its label is exempt from clinical-investigation authorisation/notification; confirm your pathway against Act 375/2022 Section 20 and SÚKL guidance before submission.

Step 13 — Only if relevant: derogation. Separately from ordinary notification, the MDR Article 59 derogation route (competent-authority authorisation of a specific device that has not completed conformity assessment, in the interest of public health or patient safety) is administered nationally. A derogation is an exception for a defined situation, not a launch strategy. Doğrulanacak: the current Czech derogation application channel and form at request.

Step 14 — Post-market system live before first sale. PMS/PMCF (or PMPF for IVDs), vigilance reporting to SÚKL within MDR deadlines, FSCA/FSN handling, Article 10a supply-shortage notification readiness, servicing and safety-technical-control records with your Czech service partner, complaint handling with the Czech chain, and document retention (10 years; 15 for implantables) — see Section 10.

6. Language Requirements: the Czech Layer in Detail

Czechia is a Czech-language market for regulatory purposes. The MDR requires device information in the language(s) determined by the member state where the device is made available; Czechia's determination is Czech, carried in Act 375/2022's language provisions for documentation and information.

InformationRequired language
Labelling and instructions for use — professional useCzech
Labelling and instructions for use — lay users (patients and other consumers)Czech — the version the patient actually receives must be comprehensible to that patient in Czech
Implant card / implantable-device information for the patient (MDR Article 18)Czech, covering device identification, precautions/interference, service life and follow-up, and safe-use information
Custom-made device statement (MDR Annex XIII)Czech where Act 375/2022 so requires — failure to attach the statement in Czech is expressly among the offences the current Act was drafted to capture
Documents filed in ISZP (activity notification and device list)Filed electronically in ISZP in Czech (the system, its manuals and the notification fields are Czech); device name, intended purpose, risk class and Basic UDI-DI per device
Clinical-investigation fileApplication language Czech; English accepted in part — ECRIN Czech data records Czech as the application language with English accepted for defined parts, while protocol summaries, participant information and labels must be in Czech. Doğrulanacak: the current SÚKL position document-by-document at submission
EU Declaration of Conformity and technical documentationKeep the controlled master in your QMS language; be ready to provide Czech versions of user-facing and authority-requested documents — do not assume an English-only file will be accepted for a Czech procedure
Field safety notices reaching Czech usersCzech, issued promptly through the Czech chain (and tracked in the national system — field-safety notices were among the functions retained provisionally from RZPRO)

Practical points:

  • Translation is a conformity activity, not a marketing task. The Czech IFU is part of the device as supplied in Czechia. A mistranslated warning, a sizing table that diverges from the master, or a Czech label that omits a master-file symbol explanation is a nonconformity, and it will be read against the master file in any SÚKL inspection or incident investigation.
  • Slovak is not Czech. The two languages are close and mutually intelligible in speech, and suppliers with a Slovak range are tempted to reuse Slovak texts. Regulatory Czech is a separate controlled language version: terminology, diacritics and statutory vocabulary differ. Use a translator/reviewer working in Czech with medical-device experience, and lock the approved Czech text under document control.
  • English is a working convenience, not a right. SÚKL publishes selected guidance in English (its registration-and-notification Q&A included) and its staff work with English in EU procedures, but ISZP, tender documentation, the reimbursement catalogue and user-facing information are Czech. Plan on Czech; treat any English acceptance as procedure-specific, to be confirmed for that procedure.

7. Timelines and Fees

7.1 Timelines (indicative — the conformity route, not SÚKL, sets most of the clock)

StepTypical timeNotes
Plain Class I self-declaration (file, DoC, EUDAMED registration)~4–8 weeksIndicative range assuming the conformity file is ready; Czech translation/review is usually the critical path
ISZP activity notification (distributor / servicing / custom-made)Notification is filed electronically before operations commence; allow several weeks including obtaining ISZP access credentials, preparing the per-device list and internal reviewDoğrulanacak: SÚKL's current processing/ validation practice and credential lead time at filing — no reliable published fixed-day standard was verified for this guide, so do not quote one to management as a commitment
EUDAMED actor validation by SÚKLAllow several weeksData readiness and AR-linking sequence (Step 4) are what usually slip, not the authority queue alone
Notified-body certification, Class IIaCommonly ~9–18 months end-to-end in current EU practiceObtain the notified body's own current lead time in writing at application
Notified-body certification, Class IIb / IIICommonly ~12–24+ monthsClass III design examination and, where applicable, expert-panel / ancillary-substance consultation extend the clock
IVDR Class C / D certificationCommonly ~12–24+ monthsIVDR notified-body capacity remains tight EU-wide
Clinical investigation (SÚKL + ethics committee)General timespan of about 60 days for authorisation (ECRIN Czech country data), within the MDR Article 70 timetable including validation; substantial modifications ~30–38 daysThe ethics-committee opinion is part of the competent-authority file — sequence ethics and CA work together; to be verified at application
Reimbursement catalogue (voucher devices)Monthly cycle — SÚKL issues the catalogue monthly and insurers must respect the current issueA missed monthly cut-off delays first payable dispensing by a month; build the catalogue date into the launch plan
Hospital tender cycle (ZZVZ)Procedure windows are set per tender, typically several weeks to a few months from notice to award, with supply contracts then running for their term (commonly 1–4 years for device frameworks)Notices in the VVZ / Zakázky GOV and TED; a missed window in a key hospital usually means waiting for its next competition in that category

Certificates issued under the MDR are valid for a maximum of 5 years, with annual surveillance and recertification; the EU Declaration of Conformity and technical documentation must be kept for at least 10 years after the last device is placed on the market (15 years for implantable devices).

7.2 Fees and costs

Czechia's fee picture is different from markets with a published per-device registration tariff, and this guide states it plainly rather than inventing a tariff:

Cost itemAmount / positionNotes
EUDAMED registrationNo EUDAMED filing feeCosts are internal data-preparation and maintenance time
ISZP activity notification (distributor / servicing / custom-made)No per-notification administrative fee was verified in a current primary SÚKL source for the ordinary Sections 8/23 activity notification. SÚKL charges for expert activities are governed by guidelines UST-29 (administrative fees; reimbursement of costs of expert activities) and UST-24 (cost reimbursement on request; waivers and refunds) and by the Administrative Fees Act (Act No. 634/2004 Coll.) where it appliesDoğrulanacak — important: confirm, before budgeting, whether any fee under the current UST-29 / Act 634/2004 applies to your specific notification or request at filing. Do not present any secondary-source "SÚKL registration fee" figure to a board as a current verified tariff without that confirmation
Clinical investigation / performance studyFee mandatory, amount set in SÚKL guideline UST-29 for the investigation type; ethics-committee review fees are set by the committees themselves (ECRIN Czech data gives an indicative ~€700–800 for initial ethics approval and ~€200 for an amendment, as committee-level figures, not a SÚKL tariff)Non-commercial/academic sponsors: pay the full SÚKL fee, then seek a waiver under UST-24 — to be verified: current UST-29 amounts and the assigned ethics committee's current fee at application
Reimbursement categorisation (voucher devices)No per-device listing fee was verified in the sources used for this guide; the economic content of the system is the category reimbursement limit in Appendix 3 to Act 48/1997, not a filing feeDoğrulanacak: whether any UST-29 charge applies to your categorisation-related request, and the current Appendix 3 limit and conditions for your category, at filing
Notified-body conformity assessmentCommercial fees — the real cost centre. As indicative market ranges only (not official tariffs): initial MDR certification for a single device family commonly runs into the tens of thousands of euros (QMS audit day rates plus technical-documentation review), with annual surveillance audits and 5-year recertification on top; complex Class III files cost materially more. Published schedules for the Czech Metrology Institute (NB 1383) in comparative 2025–2026 data show day rates in the lower EU range — obtain its own current schedule rather than relying on a comparison tableObtain a written quotation from the notified body for your device scope
Czech translation (labelling, IFU, implant card, FSN templates)Commercial cost, per-word/per-document plus independent reviewBudget Czech medical translation + reviewer; do not reuse a Slovak text without a full Czech revision
Import VAT / duty (third-country goods only)Import VAT on the customs value, plus any duty by commodity code — normally recoverable as input VAT for a VAT-registered Czech importer, a cash-flow cost otherwiseDoğrulanacak: current VAT rate treatment for your category (standard vs reduced rate) and code-specific duty at shipment
Czech distributor / AR / servicing servicesCommercial feesAR mandate fees, distributor margin and notified servicing capacity are negotiated — they are operating costs, not authority fees

Budgeting rule for Czechia: the authority-fee line is small compared with most non-EU markets; the real Czech budget lines are notified-body certification, controlled Czech translation, the operator chain (AR, importer, an ISZP-notified distributor and service partner) and — for voucher devices — the pricing work needed to live inside an Appendix 3 category limit. A budget that shows a large "SÚKL registration fee" and no translation, service-partner or categorisation line has the Czech cost structure backwards.

8. Selling After Registration: Reimbursement, Procurement and Channels

8.1 Two payment tracks — keep them apart

Track A — Voucher devices (outpatient, prescribed on a poukaz). This is the system the 2019 reform rebuilt. Its elements:

  • Legal base: Act No. 48/1997 Coll., on Public Health Insurance, as amended with effect from 1 January 2019, after the Constitutional Court declared the previous insurer-code-list regulation unconstitutional (non-transparent, non-reviewable) and repealed it as of 31 December 2018.
  • Categorisation tree: individual device types are divided into reimbursement categories listed in Appendix 3 to the Act. The Act itself lays down, for each category, a reimbursement limit and the other reimbursement conditions — prescription restrictions, indication restrictions and volume (quantity/ time) restrictions. (The category for glucose meters, with a limit of CZK 435 per piece excluding VAT in the sources describing the reform, is the standard worked example of the pattern: a statutory limit plus prescription, indication and volume conditions. Doğrulanacak: the current Appendix 3 limit and conditions for your category — limits and conditions are amended over time and the 2019-era example figure must not be used as a current price.)
  • Notification principle: devices are not classified by SÚKL in formal proceedings. The manufacturer/ supplier itself notifies SÚKL of the classification of its device into a category. If SÚKL discovers — on its own initiative, or on a complaint by a health insurer or a competitor — that a device was classified incorrectly, it opens formal administrative proceedings on the misclassification. Competitor complaints are a real enforcement channel in this system: a rival already in your category has standing and motive to challenge an aggressive classification.
  • Monthly catalogue: health insurers continue to issue/pay on vouchers, but they must respect the system-wide reimbursement catalogue issued by SÚKL every month. If your device is not correctly in the catalogue, the insurer's payment does not follow, whatever your CE mark says.
  • Scope limit: the system covers devices prescribed in outpatient care and picked up at pharmacies or dispensing sites — diabetic aids and glucose meters, orthoses, wheelchairs, crutches, hearing aids, glasses and contact lenses, incontinence products and similar. It does not cover inpatient devices.

Track B — Hospital devices (inpatient, incl. ZUM). Devices consumed in hospital care — separately billed material (ZUM) such as implants, cardiostimulators, valves, cannulas, catheters, probes, staplers, laparoscopy devices and stents — are paid inside hospital care, not through the voucher catalogue. Historically they were reported against insurer codebooks (the ZUM codebook system); that mechanism has been losing its function as hospital payment has moved toward CZ-DRG case-based payment, in which the device cost increasingly sits inside the case payment rather than being reimbursed item-by-item at a codebook maximum. The commercial consequence: for a hospital device, the buyer conversation is a value-and-budget conversation with the hospital (and its tender), not a listing conversation with SÚKL. Doğrulanacak: the current payment treatment (separately reported ZUM vs inside CZ-DRG) for your exact device category and target hospital type at entry — this boundary is moving and is category-specific.

Private channel: private clinics and out-of-pocket sales buy commercially. Supplier qualification, Czech documentation, servicing and vigilance expectations mirror the public side even though tender and catalogue rules do not apply.

8.2 Public procurement: hospitals, the VVZ and TED

  • Who buys: there is no central purchasing body for devices in Czechia. Faculty (university) hospitals, regional and district hospitals, specialised institutes and other health providers buy as individual contracting authorities under the Public Procurement Act (ZZVZ, Act No. 134/2016 Coll.), supervised by ÚOHS.
  • Where notices appear: all public contract notices are published in the Journal of Public Contracts (VVZ, Věstník veřejných zakázek) at vvz.nipez.cz, administered by the Ministry for Regional Development; notices and most tender documentation are also on the Zakázky GOV portal (zakazky.gov.cz); above-threshold notices are also published in TED. Individual contracting authorities run procedures through certified electronic tools (the national NEN system and commercial profiles of the contracting authority).
  • Thresholds: below the small-scale contract limits — since the amendment effective 3 April 2025, CZK 3,000,000 excluding VAT for supplies and services (raised from CZK 2,000,000) and CZK 9,000,000 for works — the formal ZZVZ award procedure does not have to be used, though transparency, proportionality, equal treatment and non-discrimination still apply. Above the EU thresholds (set in Government Regulation No. 172/2016 Coll. and revised every two years; for orientation, the recent values were about CZK 3,491,000 for central-authority supply/service contracts and about CZK 5,386,000 for other public contracting authorities such as municipalities and many health providers), the full above-threshold regime and TED publication apply. Doğrulanacak: the current threshold values at tender — the 2026–2027 EU revision (Commission Delegated Regulation (EU) 2025/2152) changes the underlying amounts.
  • Aggregation warning: ÚOHS actively enforces the rule that similar supplies must be aggregated when estimating contract value — Czech healthcare case law includes a hospital fined for splitting purchases of similar products across suppliers without a procedure. A "keep each order small" sales strategy aimed at staying under thresholds is a compliance risk for your hospital customer, and hospitals know it.
  • How buying feels in practice: specifications and scoring are fixed by the time the notice is published, all tender documentation is in Czech, electronic submission deadlines are hard, and servicing, instruction and Czech documentation are commonly scored or required. The work that wins Czech tenders happens before publication — clinical evaluation by the hospital's users, trial/sample supply through lawful channels, and service readiness.

8.3 Advertising and promotion

Device advertising in Czechia is specifically regulated — this is not a gap in the law:

  • Legal base: Act No. 40/1995 Coll., on Advertising Regulation, as amended to add dedicated medical-device and IVD advertising rules alongside the MDR adaptation, together with MDR Article 7 and the Civil Code. SÚKL supervises device advertising. No licence, approval or fee is required to advertise — but only a device that can lawfully be placed on the market under the MDR/IVDR may be advertised at all.
  • "Advertising" is defined broadly: promotion of devices, and all forms of information, canvassing and inducement intended to promote the prescription, supply, sale or use of devices and IVDs.
  • To professionals: advertising may be disseminated only through communication channels intended chiefly for professionals (professional publications and programming), must contain the basic information from the instructions for use and enough information for the professional to form their own opinion, and is subject to a ban on giving professionals gifts or other benefits in connection with promotion. Advertising of devices that may be prescribed or dispensed only by authorised persons may target only those persons.
  • To the general public: an advertisement must clearly state that the product is a medical device (or IVD), identify its purpose, and clearly invite a careful reading of the instructions for use. Devices intended, under the manufacturer's instructions, for use only by a healthcare professional, or that may be issued only on a doctor's voucher/request, may not be advertised to the general public.
  • Reimbursed devices: advertising a device reimbursed in full or in part from public health insurance by means of a consumer competition, lottery or similar game tied to the quantity prescribed, dispensed or used is prohibited, and advertising must not refer to specific public figures/authorities in the ways the Act prohibits. Doğrulanacak: the current Advertising Act wording for your device category and audience (professional vs public, reimbursed vs not) before a Czech campaign — the CMS and SÚKL guidance on this area is actively interpreted.

9. Common Mistakes in Czechia

1. Building the launch on the old law. Citing Act 268/2014 or 89/2021, filing in RZPRO, or planning around a 15-day device notification and a five-year SÚKL notification decision — all of that is the pre-375/2022 regime. The current architecture is EUDAMED (manufacturer/AR/importer) + ISZP activity notification (distributor/servicing/custom-made).

2. Assuming CE marking alone completes a Czech launch. Check, in order: the distributor's and service partner's ISZP notifications, the Czech-language set, and (voucher devices) the categorisation and monthly catalogue.

3. Signing a distributor that has not notified — or cannot notify — in ISZP. No ISZP notification (where required), no declared device list with your devices on it, no compliant Czech business. The class I / class A and non-healthcare-provider exemptions are narrow and fact-specific; a distributor supplying class IIa devices to hospitals is squarely inside the duty.

4. Forgetting the service partner is regulated. Servicing, professional maintenance and safety technical controls are a notified, competence-based activity in Czechia. An unnotified or unqualified service arrangement surfaces at the first hospital tender — or the first incident.

5. Launching with English or Slovak labelling. Czech is required for labelling and IFU. Slovak texts are not a substitute, however close the languages are.

6. Letting the distributor translate and relabel on its own. That can trigger MDR Article 16 manufacturer-level duties for the distributor and takes the Czech text out of the manufacturer's document control.

Translate under manufacturer control; let the distributor review for local usability, not originate the controlled text.

7. Treating a voucher device like a hospital device (or the reverse). A voucher product without Appendix 3 categorisation and catalogue presence has no payer; a hospital implant does not go through the voucher system at all, and preparing a catalogue filing for it wastes the launch quarter. Decide the track in Step 3.

8. Self-classifying into a reimbursement category aggressively. The categorisation system runs on supplier notification — and on competitor and insurer complaints that trigger SÚKL proceedings. A classification you cannot defend against the Appendix 3 category definition will be challenged by the rivals who share the category.

9. Quoting a "SÚKL registration fee" — or the 2019 glucose-meter limit — as current fact. The verified position is in Section 7.2: fees live in UST-29/UST-24 and must be confirmed per procedure; category limits live in Appendix 3 as amended and must be confirmed per category. Secondary sources repeat both kinds of figure long after they move.

10. Splitting supply to stay under procurement thresholds. ÚOHS enforces aggregation of similar supplies; hospitals fined for split purchasing will not repeat the pattern with your products. Sell into the tender calendar instead.

11. Advertising like a consumer product. Gift/inducement bans for professionals, channel restrictions, the public-advertising ban for professional-only and voucher-only devices, and the ban on competition/lottery promotion of reimbursed devices are all live Czech rules supervised by SÚKL — a campaign lawful in another member state can breach several of them at once in Czechia.

10. Post-Market Obligations in Czechia

Registration is the start of the Czech compliance load, not the end of it:

  • Post-market surveillance (MDR Articles 83–91): PMS system and plan, PSUR (periodic safety update report) for class IIa and above, PMCF; for IVDs, performance follow-up (PMPF) and the corresponding reports. Trend reporting applies EU-wide.
  • Vigilance: serious incidents and field safety corrective actions are reported within the MDR deadlines (2 / 10 / 15 days by seriousness) — to SÚKL for incidents in Czechia, via the channels in force until the EUDAMED vigilance module becomes mandatory; field safety notices must reach Czech users in Czech, promptly, through the distributor/service chain you notified in ISZP. Act 375/2022 provides for central registration and evaluation of serious adverse events and safety corrective actions at national level, and for SÚKL procedures against devices posing an unacceptable risk — expect SÚKL to be an active counterpart, not a mailbox.
  • Supply interruption (MDR Article 10a): since 10 January 2025, manufacturers must notify the competent authorities (SÚKL for Czechia) and the economic operators / health institutions they directly supply, in advance, of an anticipated interruption or discontinuation of supply of a device where that could result in serious harm or a risk of serious harm to patients or public health.
  • Use-phase duties: Czech healthcare providers' use of devices is regulated (Part 4 of the Act): instructions for use must be available and followed, instruction/training (instruktáž) must be provided for the device categories the Act defines, professional maintenance and periodic safety technical controls must be performed at the manufacturer's prescribed intervals by competent, notified servicing persons, and records kept. Your Czech contracts should allocate who provides instruction and controls, in Czech, for every installed device — providers will ask, because the duty lands on them and they pass it up the chain.
  • Prescribing and dispensing discipline: voucher devices continue to be dispensed only against a valid voucher and within the category's prescription, indication and volume conditions; the monthly catalogue is the operative payment reference — a device whose catalogue entry lapses or is corrected mid-year creates dispensing disputes immediately, so monitor your own entries monthly.
  • Traceability and UDI: UDI assignment and database duties (Articles 27–31 MDR) are live; implantable-device information to the patient (implant card, Article 18) in Czech; Czech providers record UDI for implantables as part of traceability.
  • Retention: technical documentation and the EU Declaration of Conformity kept at least 10 years after the last device is placed on the market (15 years for implantables).
  • Sanctions: Act 375/2022 defines the offences and administrative fines for breaches of the MDR/IVDR and of the national duties in the Act — including failures in notification, language, corrective-action and clinical-investigation duties — precisely because the predecessor act had gaps in its offence coverage.

Doğrulanacak: the current fine ceiling applicable to your specific offence at the time of any incident — do not rely on the CZK 2 million ceiling quoted in commentary on the repealed 2014 act.

11. Frequently Asked Questions

Does SÚKL approve my device before I can sell it in Czechia? No. SÚKL does not grant a pre-market approval or registration certificate for ordinary devices. Conformity is established under the MDR/IVDR (with a notified body where the class requires one), the device is CE-marked, actors and devices are registered in EUDAMED — and the Czech national duties in this guide (ISZP activity notifications in the chain, Czech language, and for voucher devices the reimbursement categorisation) are completed. SÚKL supervises the market; it does not pre-approve the product.

I am a manufacturer in Germany/Poland. Must I notify SÚKL of my activity? No — SÚKL states expressly that a manufacturer established in the EU outside Czechia that supplies its devices directly to the Czech market is not subject to the operation-notification obligation. Your duties are EUDAMED registration (validated by your own member state's authority) and the MDR file. If you appoint a Czech distributor, that distributor has its own ISZP duty.

My Czech distributor only sells class I devices. Must it notify in ISZP? Not under the Sections 8/23 duty, if it supplies exclusively risk class I devices (or class A IVDs) — or exclusively to users who are not healthcare providers. The exemption is lost as soon as the range includes a class IIa device supplied to a healthcare provider. Confirm the distributor's actual range, not its intended range.

Is RZPRO still the system to file in? No. RZPRO is the predecessor system; its data was migrated to ISZP, which is the current system for distributor, servicing and custom-made notifications. RZPRO survives only provisionally for certain transitional functions. An instruction to "file in RZPRO" is a sign the source predates Act 375/2022.

Do I need a Czech authorised representative? You need an EU authorised representative if you are established outside the EU — but it may be established in any EU member state. Czechia does not require the AR to be Czech. If your AR is Czech-established, SÚKL validates it in EUDAMED.

Can I use a Czech notified body? You may use any EU-designated notified body whose NANDO scope covers your device and route. Czechia has its own designated bodies under the MDR — the Czech Metrology Institute (NB 1383) and ITC Zlín (NB 1023) — with defined, limited scopes. Check the current NANDO scope at application; do not assume either body covers your codes, and do not assume you must use a Czech body.

How is my device paid for in Czechia? It depends on the track (Section 8.1). A device prescribed on a voucher in outpatient care is reimbursed under the Appendix 3 categorisation system in Act 48/1997: a statutory category, limit and conditions, supplier notification of classification to SÚKL, and payment against SÚKL's monthly catalogue. A hospital device is paid inside hospital care (historically ZUM separately-billed material; increasingly inside CZ-DRG case payment) and is won through hospital tenders, not catalogue listing.

A purely private sale is commercial.

Is there a central buyer I should register with, like a national framework? No. Czech public hospitals buy individually under the Public Procurement Act. Notices are in the VVZ and on Zakázky GOV, and in TED above EU thresholds. There is no single national device framework or catalogue for hospital devices.

What does the Czech launch cost in authority fees? Very little that is verifiable as a fixed tariff: no EUDAMED fee, and no per-device SÚKL registration fee verified for the ordinary notifications. SÚKL's charges for expert activities sit in guidelines UST-29/UST-24 and must be confirmed per procedure. The real costs are notified-body certification, Czech translation, the operator/service chain, and (voucher track) pricing inside a category limit. See Section 7.2 — and treat any precise "SÚKL fee" quoted by a secondary source as to be verified until SÚKL confirms it.

How long does a Czech launch take? The authority steps (ISZP notification, EUDAMED validation) are measured in weeks. The launch is set by the conformity route (about 9–18 months for class IIa notified-body certification in current EU practice; 12–24+ months for IIb/III), by Czech translation, and by the payment calendar — the monthly catalogue cycle for voucher devices, or the target hospital's tender window for hospital devices.

Can I advertise my device to patients in Czechia? Only within the Advertising Act limits: the device must be lawfully on the market; the advertisement must state it is a medical device, identify its purpose and invite a careful reading of the instructions; and devices for professional use only, or issuable only on a doctor's voucher/ request, may not be advertised to the general public at all. Reimbursed devices may not be promoted through consumer competitions or lotteries tied to quantities. SÚKL supervises this area.

What happens if my distributor never notified in ISZP? The duty and the offence are the distributor's — but the commercial consequences are yours: hospital and insurer checks surface the gap, SÚKL supervision can interrupt supply, and replacing or regularising a distributor mid-launch costs the season. Verify the notification, and the presence of your devices on its declared list, before the first shipment.

Sources & scope

  • sukl.gov.cz, SÚKL (State Institute for Drug Control) — registration and notification, the ISZP system, categorisation and reimbursement.
  • Collection of Laws (Sbírka zákonů) — Act No. 375/2022 Coll. on Medical Devices and Act No. 48/1997 Coll. on Public Health Insurance.
  • eur-lex.europa.eu, Regulation (EU) 2017/745 (MDR), Regulation (EU) 2023/607, Commission Decision (EU) 2025/2371.

This guide is market-access intelligence, not regulatory or legal advice. SÚKL and Ministry of Health practice, ISZP functionality, categorisation lists and fee schedules change — professional review is required before business use.