Step 1 — Confirm the product is a medical device (or IVD) and classify it. Apply MDR Article 2 and Annex VIII / IVDR Annex VIII. Screen Czechia-relevant borderlines early: Annex XVI aesthetic products, software (Rule 11), substance-based devices, and products near the medicinal-product line (SÚKL regulates both medicines and devices, so a borderline argument will be read by an authority that owns both sides of the line).
Step 2 — Build the MDR conformity file. QMS (ISO 13485-aligned), technical documentation (Annexes II– III), GSPR checklist (Annex I), risk management (ISO 14971), clinical evaluation (Article 61, Annex XIV) or performance evaluation for IVDs, PMS plan, Basic UDI-DI and UDI-DI assignment, and the PRRC appointment.
Step 3 — Decide, at the outset, which payment track the device will live on. Voucher/outpatient (categorisation + monthly catalogue), hospital/ZUM and CZ-DRG (hospital budgets and tenders), or purely private/commercial. This decision drives the evidence, pricing and partner work in Steps 7–9 and cannot be retrofitted cheaply — a voucher product launched without its categorisation work done has no payer.
Step 4 — Fix the operator chain. Appoint the AR (non-EU manufacturers) by written mandate; confirm who the importer is; select the Czech distributor and verify its ISZP notification status and the device list declared in it before contract signature; confirm who will perform servicing and instruction in Czechia and that that person is ISZP-notified and competent for the device; document who performs any Czech relabelling/ repackaging and whether Article 16 duties are triggered. Remember the sequencing rule: the AR must have its SRN before the non-EU manufacturer can link to it in EUDAMED.
Step 5 — Conformity assessment. Plain Class I: self-declare. Is/Im/Ir, IIa, IIb, III and IVD Classes B–D: engage an EU-designated notified body with the right NANDO scope (Section 2). Notified-body capacity — not SÚKL — is the usual bottleneck, so apply early, and never lodge a parallel application for the same procedure with two notified bodies.
Step 6 — Register the actors, then the devices, in EUDAMED. Actor registration and SRN for manufacturer, AR, importer and system/procedure pack producers (SÚKL-validated for Czech-established operators); device registration in the UDI/Device module before first placement for devices placed from 28 May 2026 (and by 28 November 2026 for devices already on the market that continue to be sold). Keep the information current thereafter.
Step 7 — Complete the Czech national layer. Distributor ISZP notification (before operations begin, with your devices on its declared list — name, intended purpose, risk class, Basic UDI-DI); servicing-person ISZP notification; custom-made manufacturer ISZP notification if that is genuinely your case. If you are an EU-established manufacturer supplying directly with no Czech distributor, record the SÚKL position that no operation notification is due from you — and make sure no one in the chain files an unnecessary one, or omits a necessary distributor filing in reliance on yours.
Step 8 — Czech labelling and IFU. Produce the language set in Section 6 before the first Czech delivery:
Czech label, Czech IFU, Czech implant-card content where applicable, and Czech field-safety material ready to issue. Allow real translation and review time — this step, not the authority queue, is what most often slips in a Czech launch.
Step 9 — Payment readiness. Voucher track: confirm the Appendix 3 category, the reimbursement limit and the prescription/indication/volume conditions for the category; complete the supplier's classification notification to SÚKL; verify the device's appearance in the monthly catalogue before promising insurers' payment to a dispensing partner. Hospital track: map the target hospitals' procurement calendars, the CZ-DRG / ZUM treatment of your category, and each hospital's internal assessment process. Doğrulanacak: the current notification form, data elements and catalogue cut-off dates for your category at filing — the catalogue is monthly, so a missed cut-off costs a month, not a quarter.
Step 10 — Logistics and tax setup for the first shipment. Intra-EU supply: no customs declaration, normal EU VAT treatment in CZK. Third-country supply: importer of record, customs declaration, import VAT (and any duty) accounted for — agree in writing who carries this and under which Incoterm.
Step 11 — Commercial entry: hospital tender or dispensing channel. Map the buyer: faculty and regional hospitals buying under the ZZVZ (notices in the VVZ / Zakázky GOV and TED), the pharmacy and dispensing-point channel for voucher devices, and the private clinic channel. Monitor by CPV code (Section 8.2).
Step 12 — Only if relevant: clinical investigation in Czechia. Applications go to SÚKL with the opinion of the competent ethics committee — ECRIN's Czech country data records that the ethics committee's approval is required as part of the submission to the competent authority (in practice: ethics first / in parallel, with the EC document in the CA file), that approval is explicit (written), and that the general authorisation timespan is about 60 days within the MDR Article 70 timetable, with substantial-modification timelines of roughly 30–38 days.
Participant-facing documents, summaries and labelling/IFU for the investigation must be in Czech. Fees are charged under SÚKL guideline UST-29 (with UST-24 on cost reimbursement, waivers and refunds); non-commercial sponsors pay the fee and then seek a waiver under that framework — to be verified: the current UST-29 amount for your investigation type at application. A CE-marked device used within its label is exempt from clinical-investigation authorisation/notification; confirm your pathway against Act 375/2022 Section 20 and SÚKL guidance before submission.
Step 13 — Only if relevant: derogation. Separately from ordinary notification, the MDR Article 59 derogation route (competent-authority authorisation of a specific device that has not completed conformity assessment, in the interest of public health or patient safety) is administered nationally. A derogation is an exception for a defined situation, not a launch strategy. Doğrulanacak: the current Czech derogation application channel and form at request.
Step 14 — Post-market system live before first sale. PMS/PMCF (or PMPF for IVDs), vigilance reporting to SÚKL within MDR deadlines, FSCA/FSN handling, Article 10a supply-shortage notification readiness, servicing and safety-technical-control records with your Czech service partner, complaint handling with the Czech chain, and document retention (10 years; 15 for implantables) — see Section 10.