EU MDR + the Finnish layer
Act 719/2021 frames the national layer: Fimea is the competent authority for supervision, vigilance and the register of serious incidents — and it does not approve devices before marketing.
Field guide · PDF
No Finnish CE mark — EU MDR applies directly. Finland runs two registers in parallel: EUDAMED for manufacturers, authorised representatives and importers, and Fimea's national CERE register for notifiable distributors, custom-made manufacturers and health institutions. Add a safe-use language rule in Finnish and Swedish, and county-level buying.
Sourced from Fimea, the Medical Devices Act 719/2021 on Finlex and the OECD Country Health Profile 2025, and verified 2026 publications.
Enter your work email and we'll send the PDF straight to your inbox.
What's inside
Every figure in this guide traces to a named source — Fimea, the Medical Devices Act 719/2021, the OECD Country Health Profile 2025, and verified 2026 publications.
Act 719/2021 frames the national layer: Fimea is the competent authority for supervision, vigilance and the register of serious incidents — and it does not approve devices before marketing.
Manufacturers, authorised representatives and importers established in Finland register in EUDAMED — but notifiable distributors, custom-made manufacturers and health institutions manufacturing for their own use register in Fimea's national CERE register. Which register applies depends on the operator's role.
Device information may be in Finnish, Swedish or English — but safe-use information and patient instructions must be in Finnish and Swedish. And the commercial gate is procurement by the 21 wellbeing services counties, Helsinki and the HUS Group, not a reimbursement listing.
Complete guide
Every section of the PDF on one page. Use the form above to keep a copy.
Last updated: November 2026. Finland is a member state of the European Union, so the EU Medical Device Regulation (MDR, Regulation (EU) 2017/745) and the In Vitro Diagnostic Regulation (IVDR, Regulation (EU) 2017/746) apply directly in Finland — there is no separate Finnish transposition of the Regulations themselves. The Finnish national layer consists mainly of the Medical Devices Act (laki lääkinnällisistä laitteista, Act 719/2021, in force 19 July 2021), the Act on Certain Medical Devices Specified in EU Directives (Act 629/2010, which continues to govern the legacy directive devices during the transition periods), Fimea Administrative Regulation 2/2021 on registration notifications, the Decree of the Ministry of Social Affairs and Health on fees chargeable by the Finnish Medicines Agency, and the Act on Public Contracts (1397/2016) on the procurement side. The competent authority is the Finnish Medicines Agency — Fimea (in Finnish: Lääkealan turvallisuus- ja kehittämiskeskus) — which operates under the Ministry of Social Affairs and Health. Always confirm critical steps against current Fimea guidance and the consolidated texts on finlex.fi before filing.
Finland is a small market with high purchasing power, near-universal public coverage and an unusually strong domestic health-technology industry. The resident population was 5,603,851 in 2024 (source: OECD / European Observatory on Health Systems and Policies, Country Health Profile 2025: Finland). In 2023 Finland spent EUR 3,996 per capita on health (adjusted for purchasing power), equal to 10.5% of GDP — slightly above the EU average of 10.0% — with public sources accounting for 81% of current health expenditure, the highest public share in two decades, and out-of-pocket payments making up about 14% (source: OECD / European Observatory, Country Health Profile 2025: Finland, based mainly on official statistics provided to Eurostat and the OECD). Since 1 January 2023, publicly funded health and social services have been organised by 21 self-governing wellbeing services counties (hyvinvointialueet), plus the City of Helsinki, which organises services in the capital, and the HUS Group, which is responsible for demanding specialised healthcare in the Uusimaa region. For a foreign manufacturer, Finland is rarely a first EU market, but it is a demanding reference market: two official languages, a national register running in parallel to EUDAMED, and county-level procurement that rewards suppliers who arrive with Finnish-language documentation ready.
Regulatory reality check up front: Finland is not a separate certification regime from the EU. There is no "Finnish CE mark" and Fimea does not approve medical devices before they are marketed. The manufacturer carries out the conformity assessment procedure on its own responsibility, involving a notified body for the higher risk classes, and Fimea is responsible for supervision, market surveillance and vigilance after devices are made available. What Finland adds — and what foreign manufacturers most often underestimate — is a national layer with six specific features:
A note on institutions, because Finland splits responsibilities in a way that confuses first-time entrants: Fimea is the competent authority under section 37 of the Medical Devices Act 719/2021 — general steering and supervision of medical devices, registration (CERE), market surveillance, vigilance and the register of serious incidents, clinical investigations and performance studies, derogations, and designation and supervision of Finnish notified bodies. The Ministry of Social Affairs and Health (Sosiaali- ja terveysministeriö, STM) sits above Fimea and issues the fee decree. The National Supervisory Authority for Welfare and Health (Valvira) supervises healthcare units and professionals — it is not the device register authority, a confusion older secondary sources still repeat. The wellbeing services counties, Helsinki and the HUS Group are buyers and providers, not regulators. The Finnish Institute for Health and Welfare (THL) produces health statistics and registers but does not register devices. Keep those lanes separate: registration questions go to Fimea, tender questions go to the counties, and statistics questions go to THL and Statistics Finland.
| Instrument | What it does |
|---|---|
| Medical Devices Act 719/2021 (laki lääkinnällisistä laitteista) | The main Finnish statute complementing the MDR and IVDR, in force 19 July 2021: language requirements (section 5), clinical investigations and performance studies (Chapter 3), health-institution manufacturing (sections 27–31), Fimea's supervisory powers (Chapter 5, section 37), the register of serious incidents (section 48), the notification/registration obligation (section 49), the supervision fee (section 50), derogations (section 58), and sanctions |
| Act on Certain Medical Devices Specified in EU Directives 629/2010 | The predecessor act, renamed and kept partly in force for devices still governed by the old directives (AIMD/MDD/IVDD) during the MDR/IVDR transition periods; listed provisions of Act 719/2021 also apply to those legacy devices |
| Fimea Administrative Regulation 2/2021 | Fimea's binding regulation specifying in detail who notifies what, to which register (CERE or EUDAMED), with what information, and by which procedure (source: Fimea, "Device and operator registrations") |
| Decree of the Ministry of Social Affairs and Health on fees chargeable by Fimea | Sets the first registration fee, the annual supervision-fee amounts and the processing fees for derogations, certificates of free sale, clinical investigations and other chargeable Fimea actions; the current decree text and amounts must be checked before budgeting (amounts: to be verified against the current decree — see Section 8) |
| Act on Public Contracts 1397/2016 (laki julkisista hankinnoista ja käyttöoikeussopimuksista) | Governs how wellbeing services counties, Helsinki, the HUS Group and other public buyers tender device purchases; notices are published in Hilma and, above the EU thresholds, forwarded to TED |
| Act on Organising Healthcare and Social Welfare Services 612/2021 and the wellbeing services counties legislation | The 2023 reform framework: 21 wellbeing services counties plus Helsinki organise health and social services from 1 January 2023, funded mainly by central government funding; the HUS Group carries demanding specialised care in Uusimaa (source: Ministry of Social Affairs and Health, stm.fi) |
Finland applies the MDR classification rules (Annex VIII) unchanged — there is no Finnish reclassification layer. Fimea is the authority that decides borderline classification questions; product classification decisions are chargeable and, on Fimea's published figures for medicines-side classification work, individual decisions can take months — treat any borderline Finland question (device vs. medicinal product, cosmetic, food supplement or biocide) as a Fimea decision to request early, not as an internal memo (source: Fimea, "Principles of classification", fimea.fi — the page describes Fimea's classification competence; device-specific timelines: to be verified with Fimea).
| Class | Examples | Conformity route in brief |
|---|---|---|
| I (incl. Is sterile, Im measuring, Ir reusable surgical) | Non-invasive low-risk devices, dressings, examination gloves | Self-declaration for plain Class I; notified body for the sterile / measuring / reusable aspects of Is / Im / Ir |
| IIa | Dental materials, short-term invasive devices, many diagnostic and monitoring devices | Notified body (Annex IX or Annex X + XI) |
| IIb | Long-term invasive devices, most implants not in Class III, active therapeutic devices with potentially hazardous energy exchange | Notified body, fuller technical-documentation review |
| III | Absorbable implants, devices incorporating a medicinal substance, heart valves, breast implants, surgical meshes in defined uses | Notified body with design examination; expert-panel / clinical-evaluation consultation for defined categories; notified body also for Class III custom-made implantable devices |
| IVDs (IVDR) | Classes A–D under IVDR Annex VIII | A: self-declare (notified body for sterile A); B–D: notified body, D with the heaviest scrutiny, including EU reference-laboratory involvement where applicable |
Two Finland-specific classification notes. First, products without an intended medical purpose listed in MDR Annex XVI (for example certain aesthetic equipment) are inside the regime in Finland as in the rest of the EU — and their manufacturers established in Finland carry a specific CERE notification by form (source: Fimea, CERE guidance), so a beauty-channel product can still create a Finnish registration duty. Second, software is classified under MDR Rule 11 as elsewhere in the EU, and the Finnish language rule in Section 6 extends to the information software presents to its user: plan Finnish- and Swedish-language safe-use content as part of classification and labelling work, not as an afterthought. Digital health solutions sold to the wellbeing services counties are additionally commonly assessed with the national Digi-HTA framework coordinated by FinCCHTA (the Finnish Coordinating Center for Health Technology Assessment) — Digi-HTA is an assessment framework used by purchasers, not a Fimea approval, and its current role for your product category is to be verified with the county or FinCCHTA before you rely on it.
A manufacturer established outside the EU must mandate an authorised representative established in the EU before placing a device on the Finnish market (MDR Article 11). The mandate must cover at least the Article 11(3) tasks: verifying that the EU Declaration of Conformity and technical documentation have been drawn up, keeping the documentation available, cooperating with Fimea on corrective actions, and informing the manufacturer of complaints and vigilance reports. The AR's name and address go on the label and in EUDAMED, and an AR established in Finland registers as an actor in EUDAMED and is validated by Fimea.
Practical points for Finland. First, an AR established anywhere in the EU serves Finland; there is no Finnish-AR requirement, and equally a Finland-established AR serves the entire EU market — relevant for US, UK and Swiss manufacturers choosing an AR seat. Second, documentation duties have a Finnish-language dimension: under section 5 of Act 719/2021, the conformity-assessment documentation referred to in MDR Article 52(12) / IVDR Article 48(12) must be available in Finnish, Swedish or English, and Fimea may, where safety requires, order the manufacturer and the AR to submit information and documentation in Finnish or Swedish free of charge — budget translation capacity into the AR mandate. Third, the AR is jointly and severally liable with the manufacturer for defective devices on the same basis where the manufacturer has not complied with its obligations (MDR Article 11(5)) — Finnish AR mandates should price and insure that exposure exactly as mandates elsewhere in the EU do.
Where a device is imported into the EU through Finland, the importer carries the full MDR Article 13 verification chain before making the device available: the device is CE marked, the EU Declaration of Conformity has been drawn up, the manufacturer is identified and has designated an AR, the device is labelled in accordance with the MDR and accompanied by the required information — in Finland, that means checking the Finnish/Swedish safe-use information, not just any EU language version — and the device is registered in EUDAMED where registration is required. The importer adds its own name and address to the label or accompanying documents without obscuring the manufacturer's information, keeps a copy of the Declaration of Conformity and certificates for ten years (fifteen for implantable devices), maintains a register of complaints and non-conforming devices, and cooperates with Fimea on corrective actions.
Finland-specific importer notes. A Finland-established importer registers as an actor in EUDAMED and must link itself to each manufacturer it imports from; where the manufacturer is not yet in EUDAMED, Fimea's transitional route is the Importer's Linked Manufacturers Form E in Finnish or Swedish (source: Fimea, EUDAMED guidance). Devices do not require a separate Finnish import licence — importing is a registration-and-duties regime, not a licensing regime — but do not confuse the device position with medicines, where wholesale trade is licensed by Fimea under the Medicines Act. Finland is inside the EU customs union, so EEA goods move without the customs step that surprises sellers in Norway or Switzerland; VAT and logistics remain commercial matters, not device-law matters.
Distributors carry MDR Article 14 duties: verifying CE marking, the Declaration of Conformity, the accompanying information in the languages Finland requires, importer identification where applicable, and UDI assignment; not supplying devices they have reason to believe are non-conforming; ensuring storage and transport conditions maintain conformity; keeping traceability records; and forwarding complaints and incident information to the manufacturer, AR and importer. Fimea has published a dedicated guideline for distributors of medical devices (revised 16 June 2025), which is the practical starting text for any Finnish distribution model (source: Fimea, Guidelines for distributors of medical devices).
On top of Article 14, Finland adds a notification duty that has no equivalent for distributors in most member states. A distributor is liable to file a notification to Fimea if it makes a medical device available on the Finnish market to retailers, healthcare and social welfare operators and other professional users, or if it makes available in Finland a self-testing IVD or a device containing human tissue or a human blood or plasma derivative that it has imported into Finland. A liable distributor must register as an operator in Fimea's CERE register before making devices available. Three qualifications matter in practice:
In 2025, the supervision fee for a distributor liable to file a notification is EUR 500 (source: Fimea, distributor guideline). Build the CERE operator entry, the Article 14 evidence pack and Finnish/Swedish complaint-handling capacity before the first county tender, not after the first question from a procurement team.
Manufacturers and authorised representatives must have a PRRC meeting MDR Article 15. Micro and small enterprises may contract the PRRC externally. Finnish law adds no separate PRRC registration for ordinary devices, but the PRRC's details sit inside the EUDAMED actor registration, and Fimea will treat the PRRC as the operational counterpart in surveillance and vigilance matters. If your PRRC does not work in Finnish or Swedish, ensure the function has language support for Fimea correspondence and for the Finnish- and Swedish-language field safety work described in Section 9.
Finland's registration architecture is the single most distinctive feature of its national layer, and Fimea states it plainly: there are two registers — the national CERE register maintained by Fimea, and EUDAMED maintained by the European Commission — and the Medical Devices Act 719/2021 (section 49) together with Fimea Administrative Regulation 2/2021 decide which operators notify which register (source: Fimea, "Device and operator registrations").
| Operator (established in Finland) | Register | Route |
|---|---|---|
| Manufacturer | EUDAMED | Actor module, validated by Fimea; devices in the UDI/Device module |
| Authorised representative | EUDAMED | Actor module, validated by Fimea |
| Importer | EUDAMED | Actor module; importer links itself to each manufacturer (Form E fallback in Finnish/Swedish if the manufacturer is not yet registered) |
| System / procedure pack producer or steriliser thereof | EUDAMED | Actor module |
| Custom-made device manufacturer, where a notified body was used (Class III implantable route) | EUDAMED | Actor module |
| Custom-made device manufacturer (other cases) | CERE | E-submissions via Suomi.fi identification |
| Health institution self-manufacturer (MDR Article 5(5) / Act sections 27–31) | CERE | E-submissions; all self-manufactured devices in a single notification |
| Distributor supplying retailers, healthcare/social welfare operators or other professional users in Finland | CERE | E-submissions for operator notification; device notifications suspended until further notice |
| Sterilisation service provider | CERE | Notification form |
| Manufacturer of an MDR Annex XVI device without a medical purpose | CERE | Notification form |
| Distributor (incl. established abroad) supplying a self-testing IVD or a human-tissue / blood-derivative device it imported, or a foreign distributor supplying Finnish professional users | CERE | Notification form |
Legacy actors are included: the EUDAMED obligation also covers Finnish-established legacy actors under MDR Article 120 and IVDR Article 110 (source: Fimea, EUDAMED guidance). EMDN nomenclature is mandatory inside EUDAMED, while in CERE its use is voluntary; UDI is mandatory in EUDAMED but is not required for operators who register only in CERE (source: Fimea, EUDAMED guidance).
Fimea describes registration as a two-stage process (source: Fimea, "About medical device registration", fimea.fi):
Practical access notes. CERE e-submissions run through Suomi.fi identification: use requires a Finnish personal identity code and strong identification (online banking codes, a mobile certificate or a chip identity card), and acting for a company requires either an official Trade Register role (for example managing director) or a Suomi.fi mandate named "Management of information related to medical devices" granted by the organisation (source: Fimea, CERE guidance). A foreign group setting up a Finnish distributor should therefore sequence company formation, Trade Register entries and Suomi.fi mandates before it plans the CERE filing date — the filing itself is short; the access rights are the long pole. Form-route notifications are sent to Fimea's device registration mailbox as Fimea directs on its forms pages, unsigned, with all mandatory fields completed.
Neither a CERE entry nor an EUDAMED entry is an approval, a listing decision or a quality endorsement. Fimea does not examine the device before registration in the way a notified body examines a file; registration records the operator and the device for supervision. Marketing material that presents a Finnish registration number as a Fimea approval misstates the system and collides with MDR Article 7's prohibition of misleading claims.
The sequence below assumes a CE-marked (or to-be-CE-marked) device sold into Finland through a Finnish importer/distributor structure, which is the model most foreign manufacturers use. Adjust steps 4–5 if you establish a Finnish manufacturer or AR instead.
Finland's language regime is set by section 5 of the Medical Devices Act 719/2021 and explained in Fimea's distributor guideline. It is more flexible than Sweden's or Norway's at the base level — and stricter than most entrants expect at the safety level, because Finland has two official languages and both are required.
The information and documents referred to in MDR Article 10(11) and Annex I section 23 (label and instructions for use) and IVDR Article 10(10) must be in Finnish, Swedish or English, unless the information is provided in the form of internationally recognised symbols. A manufacturer may therefore prepare instructions for use in Finnish, Swedish or English as its base version (sources: Medical Devices Act 719/2021, section 5; Fimea, Guidelines for distributors of medical devices).
The override is where launches fail. Information necessary for the safe use of the device must be available in Finnish and in Swedish. Fimea's distributor guideline makes the two concrete applications explicit: (1) information allowing safe use of the device, and (2) instructions for use for devices intended to be used by patients and other consumers, must be provided in both Finnish and Swedish. The manufacturer's risk assessment determines which information is necessary for safe use — but that judgement is the manufacturer's to defend, and the conservative reading is the working rule: anything a user needs to use the device without harm ships in both languages, for professional and lay devices alike. Do not read the English base rule as an English-only permission for patient-facing products: it is not one.
Worked contrast from Fimea's own guideline logic: a laboratory analyser used only by laboratory professionals may carry its general IFU in English, but the safe-use information identified by the risk assessment must exist in Finnish and Swedish; a home-use self-test sold to consumers needs its instructions for use in Finnish and Swedish in full. Symbols can replace text only where they are internationally recognised symbols within the meaning of section 5.
MDR Article 7 prohibits misleading claims in Finland as elsewhere, and Finnish consumer-protection law applies in parallel to consumer marketing. The industry code of Sailab – MedTech Finland (based on the MedTech Europe code) governs member marketing practice. A CERE or EUDAMED entry, a Finnish reference customer, or a Digi-HTA assessment must never be presented as a Fimea approval or endorsement.
Finland publishes few statutory processing clocks for device registration, because registration is a notification system, not an authorisation system. The honest planning table separates what is sourced from what must be verified for your file.
| Step | Indicative timing | Basis |
|---|---|---|
| EUDAMED actor registration (Finnish actors) | Plan in weeks, not days; validation by Fimea follows the EUDAMED request | EUDAMED process; no Finland-specific statutory clock published in the Fimea guidance relied on — to be verified for your filing |
| CERE operator notification by form | Fimea states it emails the operator reference number within 2 weeks of receiving the form | Fimea, CERE guidance (fimea.fi) |
| CERE e-submissions (operator + any device data currently collected) | Electronic filing; the practical lead time is Suomi.fi access and mandates, not the form | Fimea, CERE guidance |
| Distributor device notifications to CERE | Suspended until further notice — no current filing deadline | Fimea, distributor notification notice |
| Notified-body conformity assessment (higher classes) | Commonly the longest element of the whole route; SGS publishes 12–18 months or longer for MDR certification as a general market indication (secondary source — your body's queue and scope decide; to be verified per file) | SGS Finland, MDR certification services page |
| Clinical investigation / performance study start in Finland | Ethics-committee opinion must precede the Fimea application; Fimea processing time varies by study type — reserve generous time and contact Fimea during preparation | Fimea, investigations and derogation guidance |
| Derogation (MDR Article 59 / IVDR Article 54; Act section 58) | Processing time varies; Fimea advises reserving enough time and contacting it while preparing the application | Fimea, "Derogation" (fimea.fi) |
| Legacy-device EUDAMED registration transition | Complete during 2026 planning horizon; verify the exact end date in the EUDAMED Information Centre | EUDAMED mandatory-use framework from 28 May 2026 |
| County framework agreements | Multi-year cycles set by each wellbeing services county / HUS procurement calendar; missing a cycle, not Fimea, is usually the real delay | Hilma / county procurement practice (Section 10) |
Two planning conclusions. First, for a device that is already CE marked, the Finnish critical path is almost never Fimea — it is the language package, the Suomi.fi/CERE access chain, and the county tender calendar. Second, any timeline a consultant gives you for a notified-body certificate or a Fimea study decision that is not tied to a named body's or Fimea's current published position should be treated as an estimate, not a commitment.
Finnish device fees rest on two instruments: section 50 of the Medical Devices Act 719/2021 (the supervision fee) and the Decree of the Ministry of Social Affairs and Health on fees chargeable by Fimea (first registration fee and processing fees). Fimea states the structure clearly; the full current amount table sits in the decree and Fimea's fees pages, and this guide does not reproduce amounts it could not verify against that current decree text.
| Item | Amount / basis | Status |
|---|---|---|
| First registration fee for an operator | Charged on first registration (CERE or EUDAMED route); amount set by the current Fimea fees decree | Structure: Fimea, "Device and operator registrations". Amount: to be verified against the current decree before budgeting |
| Annual supervision fee — distributor liable to file a notification (2025) | EUR 500 | Sourced: Fimea, Guidelines for distributors of medical devices (revised June 2025); the same EUR 500 figure was published for 2023 (Fimea notice) |
| Annual supervision fee — other operator roles (manufacturer, AR, importer, custom-made, self-manufacturer) | Operator-specific; charged per registered role; determined according to the number of devices notified by the operator (Act 719/2021, section 50) | Basis sourced from Finlex text of section 50; per-role amounts: to be verified against the current fees decree |
| Multiple roles | Each registered role is invoiced separately — e.g. distributor + manufacturer = two supervision fees | Sourced: Fimea, supervision-fee notice for 2024 collection |
| Fee determination date | The fee follows the registered information at year-end; an operator that has notified termination (via CERE or EUDAMED) by 31 December is not charged for the following cycle on that role | Sourced: Fimea, supervision-fee notice (2024 collection, describing the standing section 50 mechanism) |
| Derogation application (Act section 58) | Processing fee per the fees decree | Structure sourced: Fimea, "Derogation". Amount: to be verified |
| Certificate of free sale / registration certificate | Charged per the fees decree; copies ordered at the same time are cheaper; an express-delivery surcharge applies (delivery under 2 weeks) | Sourced (structure): Fimea, "Certificate of free sale". Amounts: to be verified |
| Clinical investigation / performance study processing | In principle, a fee is charged for notifications and applications to Fimea | Structure sourced: Fimea, "Notifications and applications". Amounts: to be verified |
Budget the Finnish launch on total cost, not on the fee table: notified-body fees for the conformity route (the largest single item for Class IIa and above); Finnish AND Swedish translation of the safe-use and patient package, maintained under change control; AR fees if the manufacturer is outside the EU; the distributor's Article 14 quality system and CERE administration, including Suomi.fi mandate management; vigilance and field-safety execution in two languages; and tender participation costs across county frameworks. Device prices themselves are not approved or capped by a Finnish authority for ordinary devices — pricing is a commercial and procurement matter (Section 10), which is precisely why the tender calendar, not a price filing, decides Finnish revenue timing.
The MDR/IVDR post-market system applies in full: PMS plans and reports (MDR Articles 84–86), PSURs where required, trend reporting (Article 88), serious-incident reporting (Article 87), field safety corrective actions and field safety notices (Article 89), and the Article 10a duty to notify foreseeable supply interruptions. In Finland the competent-authority address for that system is Fimea, and Act 719/2021 adds the national machinery: Chapter 5 supervision powers, and section 48, under which Fimea maintains a register of serious incidents (source: Medical Devices Act 719/2021, Finlex).
Manufacturers report serious incidents (MDR/IVDR devices) and incidents (legacy directive devices) to Fimea using the European Manufacturer Incident Report (MIR) form. Fimea notes that the European Commission has published MIR form version 7.3.1 and that manufacturers must implement it no later than May 2026 (source: Fimea, "Notifications and applications", fimea.fi). Until the EUDAMED vigilance module becomes the mandatory channel, national reporting to Fimea remains the operative Finnish route — verify the current channel on Fimea's incident pages before each filing, because the EUDAMED transition changes addresses, not duties. FSCAs are notified to Fimea, and the field safety notice that reaches Finnish users follows the safe-use language rule: Finnish and Swedish (Section 6.3). Periodic summary reporting and trend reports follow the MDR timelines; nothing in the Finnish layer shortens them.
Finland formalises the user side. Professional users report incidents involving devices to Fimea — through their organisation's Fimea electronic interface or on Fimea's PDF form — and in urgent cases the report may first be made by phone, with the written report following without delay (source: Fimea, "Notifications and applications" — Fimea publishes the urgent incident phone route on that page). Healthcare units' internal device-incident systems therefore feed a national register, and a manufacturer selling into Finnish hospitals should expect incident information to flow both ways: build the distributor's and the county customer's incident contacts into your PMS procedure, in Finnish and Swedish, before first delivery.
Chapter 3 of Act 719/2021, together with MDR Chapter VI / IVDR Chapter VI, governs investigations run in Finland. The Finnish sequence has one feature sponsors miss: a sponsor cannot submit a clinical-investigation application (or a substantial modification) to Fimea before the competent ethics committee has given its opinion on the investigation — for combination studies the ethical evaluation is carried out by the National Committee on Medical Research Ethics (Tukija) (sources: Fimea, "Investigations with devices" and "Combination studies", fimea.fi). Performance studies follow the IVDR split: an application to Fimea where the study involves surgically invasive sampling for study purposes, interventional elements that may influence treatment decisions, or additional invasive procedures or risks — and for defined companion-diagnostic situations; a notification where the study uses leftover samples with a companion diagnostic (Act section 23) or evaluates a CE-marked device within its intended purpose with additional invasive or burdensome procedures (source: Fimea, "Performance studies of IVD devices", fimea.fi). Safety reporting during investigations follows the MDR Article 80 / IVDR Article 76 framework to Fimea until EUDAMED reporting takes over. Investigation and study fees follow the Fimea fees decree (amounts: to be verified — Section 8).
Where a specific device's use is in the interest of public health or patient safety, Fimea may grant a temporary derogation from conformity-assessment procedures under MDR Article 59 / IVDR Article 54 and section 58 of Act 719/2021. Fimea advises contacting it while preparing the application and reserving enough time for processing, which varies (source: Fimea, "Derogation"). A derogation is a case-specific safety decision, not a fast-track registration route, and it does not waive the language, vigilance or procurement layers around the device.
The 2023 health and social services reform moved organising responsibility from municipalities to 21 self-governing wellbeing services counties (hyvinvointialueet). The City of Helsinki continues to organise services in the capital outside the county structure, and the HUS Group is responsible for demanding specialised healthcare duties laid down by law in Uusimaa. The counties are funded mainly by central government funding and do not, as yet, have the right to levy taxes (source: Ministry of Social Affairs and Health, "Wellbeing services counties", stm.fi). The five university-hospital catchments (Helsinki/HUS, Turku, Tampere, Kuopio, Oulu) form the cooperation-area backbone for the most demanding care. Private providers supplement public services, and occupational healthcare — employer-funded and near-universal for employees — is a parallel channel for working-age ambulatory care that device sellers routinely overlook.
Public device buying follows the Act on Public Contracts 1397/2016. Contract notices are published in Hilma, the national electronic procurement notice service maintained under the Ministry of Finance / Hansel, and notices above the EU thresholds are forwarded to TED (Tenders Electronic Daily). Hansel Ltd, the state-owned central purchasing body, runs joint procurement frameworks and expert services that public buyers — including wellbeing services counties — can use, and it maintains Hilma; most device-category buying is nonetheless organised by the counties themselves, by joint county arrangements, or by the HUS Group for its area (sources: Ministry of Economic Affairs and Employment, "Key operators"; Hansel published material — the device-category split for your product is to be verified against live Hilma notices). Framework agreements run in multi-year cycles, tender documentation is in Finnish (and, where relevant, Swedish), and award criteria combine price with quality, service and lifecycle terms. There is no device equivalent of a national reimbursement application to file before tendering: if your category is inside a live framework, the framework is the market until it reopens.
Finland reimburses medicines through a statutory system (Kela-administered, with prices and reimbursement status confirmed for medicines by the Pharmaceuticals Pricing Board). Ordinary medical devices have no equivalent national listing or price-confirmation step. Devices supplied as part of public care are funded inside the county, Helsinki and HUS budgets; medical aids and assistive devices are arranged through the public service system for the individual user under the applicable healthcare and disability-services rules rather than through a manufacturer-facing reimbursement register. The practical consequences for a manufacturer: (1) there is no Finnish "reimbursement dossier" that unlocks the hospital market — the unlock is winning supply positions in county procurement; (2) price is negotiated inside tenders and frameworks, not approved in advance; (3) for novel digital or high-cost technologies, expect county-level health-technology assessment practice — including FinCCHTA-coordinated Digi-HTA assessment for digital health solutions — to shape adoption discussions even though it is not a statutory gate (current practice for your category: to be verified with the target counties). Private and occupational-health channels follow commercial terms, with Kela reimbursing patients for defined private-care costs under the Health Insurance Act — a patient-side mechanism, not a manufacturer listing.
Register correctly (EUDAMED + CERE by role), finish the bilingual safety package, then work the procurement map county by county: identify which counties and frameworks cover your category, diarise the Hilma notice flow for at least one full cycle before you need revenue, and treat HUS and the university-hospital counties as reference sites whose clinicians other counties listen to. Finland rewards suppliers who arrive early, documented and bilingual — and it is unforgiving of suppliers who discover the county structure after they have signed an exclusive distributor measured on national sales that no single distributor controls.
Does Finland have its own medical device approval? No. The EU MDR and IVDR apply directly, the manufacturer (with a notified body where required) completes conformity assessment and CE marking, and Fimea supervises the market. There is no Finnish CE mark and no pre-market Fimea approval for ordinary devices.
Who is the Finnish competent authority for medical devices? The Finnish Medicines Agency, Fimea, under section 37 of the Medical Devices Act 719/2021. Valvira supervises healthcare units and professionals; it is not the device registration authority.
What is CERE, and who registers there? CERE is Fimea's national register of medical device operators and their devices. Custom-made device manufacturers (except the notified-body Class III implantable route), health institutions manufacturing devices for their own use, notifiable distributors (including foreign distributors supplying Finnish professional users), sterilisation service providers and Annex XVI manufacturers register in CERE — by e-submissions or by form, depending on the category (Fimea, CERE guidance).
Who registers in EUDAMED instead? Finland-established manufacturers, authorised representatives, importers, system/procedure-pack producers and their sterilisers, and custom-made manufacturers who used a notified body — plus legacy actors under the transition rules. Fimea validates Finnish actor registrations.
My device is already CE marked elsewhere in the EU. What is left for Finland? Three things: the operator registrations that match your Finnish structure (EUDAMED for manufacturer/AR/importer; CERE for the notifiable distributor), the Finnish/Swedish safe-use and patient language package, and county procurement access. The CE mark itself travels; the national layer does not.
What languages must the label and instructions be in? Base information may be in Finnish, Swedish or English (or internationally recognised symbols). Information necessary for safe use, and instructions for devices intended for patients and other consumers, must be in Finnish and Swedish (Act 719/2021, section 5).
How much does Finnish registration cost? A first registration fee applies, plus an annual supervision fee per registered role, determined by the number of devices notified (Act section 50; Fimea fees decree). The one amount this guide states from a current Fimea source is the notifiable distributor's supervision fee: EUR 500 in 2025. All other amounts must be verified against the current fees decree before budgeting — see Section 8.
We sell only through a distributor. Does the manufacturer have Finnish duties? Yes. The manufacturer's MDR duties, EUDAMED actor and device registration, Article 10a supply-interruption notifications and vigilance duties remain the manufacturer's, wherever it is established. The Finnish distributor adds its own Article 14 duties and its CERE operator notification.
Are distributor device lists filed in CERE today? No — Fimea has suspended distributors' device notifications until further notice. The operator notification is filed now, and the suspension should be monitored: it is a Fimea decision that can be lifted when its e-service is ready.
Is there a Finnish notified body? Yes. Finland-designated bodies include Eurofins Electric & Electronics Finland Oy (0537, MDR and IVDR) and SGS Fimko Oy (0598, MDR, with IVDR designation announced in October 2025); Fimea lists three notified bodies established in Finland. Any EU notified body may be used; verify scope in NANDO.
How are devices paid for in Finland? Through the budgets of the 21 wellbeing services counties, the City of Helsinki and the HUS Group, via public procurement under the Act on Public Contracts (notices in Hilma and TED). There is no national device reimbursement listing or price approval for ordinary devices.
Can single-use devices be reprocessed in Finland? The MDR leaves reprocessing of single-use devices to national choice (Article 17). This guide did not verify a current Finnish permission position from a primary Fimea source — treat reprocessing models as to be verified with Fimea before any reliance.
We want to run a study in Finland. Where do we start? With the ethics committee: its opinion must precede the Fimea application for a clinical investigation. Chapter 3 of Act 719/2021 and Fimea's investigations pages set the Finnish sequence for investigations and IVD performance studies.
This guide is market-access intelligence, not regulatory or legal advice. Fimea practice, fee-decree amounts and county frameworks change — professional review is required before business use.
Market Access Strategy Session
A Market Access Strategy Session maps your Finland route — Fimea and EUDAMED validation, CERE registration, the Finnish and Swedish language package, the per-role supervision fee, and wellbeing services county tender cycles.
Professional review is required before business use.