Greece Medical Device Registration2026 Field Guide

No Greek CE mark — EU MDR applies directly. Greece runs two systems in parallel: EUDAMED, validated by EOF, and EOF's own national notification. Then the payment layer decides: EKAPY central procurement with a price-observatory ceiling, EOPYY reimbursement with clawback exposure, and a statutory rebate on public-hospital device sales.

hospitals counted in 2024 (ELSTAT)
266
of GDP spent on health (2023)
8.4%
of hospitals are private clinics (2024)
53.0%

Sourced from EOF, ELSTAT and the OECD/European Commission Country Health Profile 2025, and verified 2026 publications.

Download the guide

Enter your work email and we'll send the PDF straight to your inbox.

Professional review is required before business use. See our privacy policy.

What's inside

Twenty-one pages, zero fluff

Every figure in this guide traces to a named source — EOF, Regulation (EU) 2017/745 and Regulation (EU) 2017/746, the Government Gazette instruments, and verified 2026 legal and trade publications.

EU MDR + the Greek layer

Greece has no new MDR execution law: the MDR applies directly, but Joint Ministerial Decisions 130648/2009 and 130644/2009 still frame national practice, and EOF — established by Law no. 1316/1983 — remains the competent authority for notification, surveillance and vigilance.

EUDAMED + the EOF notification

Every device placed on the Greek market is notified to EOF — identification data, labelling and instructions for use — through the National Electronic Medical Device Registry, alongside mandatory EUDAMED registration. One system does not replace the other, and the Greek chain has to be able to evidence both.

Registry, clawback + the hospital rebate

Devices reimbursed by EOPYY outside hospital care enter through the EKAPY–EOPYY registry and reimbursement lists, with clawback exposure when a category overspends its budget. Public-hospital sales carry a statutory rebate under Law 5129/2024 — 0%, 3% or 5% of quarterly turnover per hospital — on top of tender pricing.

Complete guide

The guide in full

Every section of the PDF on one page. Use the form above to keep a copy.

Overview

Last updated: November 2026. Greece is a full EU member state, so the EU Medical Device Regulation (MDR, Regulation (EU) 2017/745) and the In Vitro Diagnostic Regulation (IVDR, Regulation (EU) 2017/746) apply directly in Greece. The Greek national layer is built around the National Organization for Medicines (EOF), the older Joint Ministerial Decisions that transposed the former directives, and the Greek rules on notification, language, pricing, reimbursement and public procurement that sit on top of the MDR. Always confirm critical steps, the current EOF registry channel and current fee amounts against EOF guidance in force at filing before you file. Where this guide cannot verify a current figure from a primary source, it says so and marks it to be verified.

Greece is a market of roughly ten and a half million people, served by a National Health System (ESY) with more than 120 public hospitals organised under seven Regional Health Authorities (YPE), a single dominant public payer — EOPYY, the National Organisation for the Provision of Healthcare Services — and a large private hospital and clinic sector concentrated in Athens, Thessaloniki and the other regional centres. For a foreign manufacturer, Greece is not entered on population alone. It is entered because three systems have to line up at the same time: the EU conformity system that puts the CE mark on the device, the Greek notification and language system administered by EOF that makes the device lawfully present on the Greek market, and the Greek payment system — EOPYY reimbursement for devices supplied to insured patients outside hospital, the price observatory and EKAPY central procurement for devices supplied into public hospitals — that decides whether anyone actually pays for it. A device can be perfectly CE-marked and still not sell in Greece if the EOF notification is missing, if the Greek-language labelling set is missing, or if the price and rebate logic has never been worked through.

Regulatory reality check up front: Greece is not a separate certification regime from the EU. There is no "Greek CE mark" and EOF — Ethnikos Organismos Farmakon, the National Organization for Medicines — does not grant a pre-market approval for medical devices in the way a non-EU regulator grants a registration certificate. EOF is the Greek competent authority for medicines and medical devices, operating under the Ministry of Health. It was established by Law 1316/1983. It receives the national notification of devices placed on the Greek market through its National Electronic Medical Device Registry, supervises the Greek market, handles vigilance and field safety corrective actions, authorises clinical investigations together with the National Ethics Committee, and supervises distributors and other economic operators established in Greece. The manufacturer remains responsible for compliance; a conforming device bears the CE mark and can then be traded across the EU/EEA. What Greece adds — and what foreign manufacturers most often underestimate — is a national layer with six specific features:

1. Notification to EOF is a real Greek duty, not a formality left over from the directives. Under the Greek medical devices framework — historically Joint Ministerial Decision DY8d/ 130648/2009 on medical devices, DY8d/130644/2009 on active implantable devices and DY8d/ 3607/892/2001 on IVDs — any medical device placed on the Greek market is subject to a notification obligation to EOF. Manufacturers or their authorised representatives submit to EOF all data allowing identification of the device, together with its labelling and its instructions for use, when the device is placed on the market and put into service in Greece. Notification is effected through EOF's National Electronic Medical Device Registry. A registration procedure in that registry also applies to IVDs, to Class I devices and to custom-made devices. This Greek notification sits alongside, not instead of, EUDAMED registration. Do not assume that an EUDAMED record alone closes the Greek file, and do not assume that an old EOF notification under the directives closes the MDR file. Doğrulanacak: the exact current EOF registry workflow, data fields and whether a given device family is filed device-by-device or by device group, at filing — EOF has been aligning its national registry practice with mandatory EUDAMED, and the division of work between the two systems must be confirmed for your category in 2026.

2. Greek is the language of the file. Labelling and instructions for use supplied in Greece must be in Greek. Language tables for EU device documentation list Greece as "Greek — mandatory", without the professional-use English allowance that some northern markets operate in practice. A file that passed in an English-speaking market will fail here on language alone, exactly as in Czechia, Poland or Portugal. Field safety notices reaching Greek users must also be in Greek.

3. The legacy Greek decisions are still cited everywhere — read them as history, not as the current conformity route. A large share of English-language material on Greece still describes the world of Directive 93/42/EEC, Directive 90/385/EEC and Directive 98/79/EC as transposed by the 2001 and 2009 Joint Ministerial Decisions. Those decisions explain where the EOF notification, the clinical-investigation procedure and the advertising rules come from, and parts of them remain the national procedural layer. The conformity assessment itself, however, is now governed directly by the MDR and IVDR, which have applied since 26 May 2021 and 26 May 2022 respectively. Greek sources in 2025–2026 also cite a newer Joint Ministerial Decision D3(a) 52922/2025 in the life-sciences framework. Doğrulanacak: the precise scope of JMD D3(a) 52922/2025 for medical devices and which provisions of the 2009 decisions it amends or replaces, at filing.

4. For reimbursed devices, Greece runs a registry, pricing and clawback system that is commercially decisive. Devices reimbursed by EOPYY for insured patients — notably consumables and devices supplied outside inpatient hospital care — enter the EOPYY reimbursement world through the EKAPY–EOPYY registry and pricing framework, are subject to EOPYY's reimbursement lists and conditions, and have historically been subject to a clawback mechanism when EOPYY expenditure in a category exceeds its budget: suppliers return a proportion of the excess. The Greek Council of State ruled in 2026 on challenges to the 2017 and 2018 clawbacks — annulling the 2017 clawback on the ground that the registry system was inadequately regulated at that time, and upholding the 2018 clawback for parties who had registered in that year.

The practical lesson is not the litigation; it is that in Greece the registry entry, the price basis (import/wholesale versus retail) and the clawback exposure are part of the product's economics from day one, not an afterthought.

5. For public-hospital sales, Greece now adds a statutory rebate on top of tender pricing.

Law 5129/2024, Article 72, introduced a rebate on sales of medical devices and reagents procured by public hospitals, detailed by Ministerial Decision 3263/11-3-2025. The rebate applies to sales through EKAPY or other contracting authorities and to direct sales to public hospitals, calculated per supplier, per hospital, per quarter on invoice value excluding VAT, on a scale reported as 0% for quarterly turnover up to €10,000, 3% from €10,000 to €30,000, and 5% above €30,000, payable after the quarter on notification. Exceptions include capital medical equipment for prevention and diagnosis or for detecting and measuring body structure or function, and tender sales where a saving of more than 5% against the budgeted tender value was achieved.

The rebate applies to invoices issued after 11 March 2025. Price your Greek hospital business net of this rebate, or the margin in the business plan will not survive the first quarter.

6. There is a central purchasing authority, but Greece remains a fragmented hospital market.

EKAPY — the National Central Health Procurement Authority, established by Law 4472/2017 and supervised by the Ministry of Health, plans central purchases, maintains the electronic health procurement platform with registers of products, suppliers and prices, and runs central tenders whose framework agreements and dynamic purchasing systems hospitals then call off. Everything EKAPY has not centralised is bought by the individual hospital under its health region, in the national electronic procurement system ESIDIS, recorded in KHMDIS, and published in TED above EU thresholds. For medical supplies and devices a price observatory records award prices per item, and tender documents often set the observatory price as the maximum acceptable bid — a higher bid is rejected. There is therefore no single Greek door: there is a central door for centralised categories and more than 120 hospital doors for the rest.

A note on institutions: the Ministry of Health sets policy. EOF, headquartered in Athens (Mesogeion Avenue, Cholargos), is the competent authority, the vigilance authority and the operator of the national device registry. EOPYY is the payer for insured services and reimbursed devices. EKAPY is the central procurement authority. The National Ethics Committee gives the ethics opinion required for clinical investigations, alongside EOF's scientific assessment. The seven YPE (Regional Health Authorities) supervise the public hospitals in their regions.

1. The Regulatory Framework: EU MDR/IVDR (Directly Applicable) + Greek National Layer

1.1 The EU layer (applies directly in Greece — no transposition step)

  • Regulation (EU) 2017/745 (MDR) — applicable since 26 May 2021: classification (Article 51 and Annex VIII, 22 rules), General Safety and Performance Requirements (GSPR, Annex I), conformity assessment (Article 52), CE marking (Article 20), UDI (Articles 27–28), EUDAMED registration (Articles 29–31), clinical evaluation and investigations (Articles 61–82), post-market surveillance and vigilance (Articles 83–92), the ban on misleading claims (Article 7), and the supply-interruption notification duty in Article 10a (applicable since 10 January 2025, introduced by Regulation (EU) 2024/1860).
  • Regulation (EU) 2017/746 (IVDR) — applicable since 26 May 2022, with staggered transition for legacy IVDs by class (to end-2027, end-2028 and end-2029 depending on class, subject to the Article 110 conditions).
  • Regulation (EU) 2023/607, as amended by Regulation (EU) 2024/1860 — extended MDR transition for eligible legacy devices to 31 December 2027 for Class III and Class IIb implantable devices (with stated exceptions) and to 31 December 2028 for other Class IIb, Class IIa and Class I devices that require a notified body, subject to the Article 120 conditions, including a timely formal application to a notified body and a signed written agreement. A legacy device that does not meet those conditions may not be placed on the Greek market on the strength of an old directive certificate.
  • Annex XVI products without a medical purpose — the MDR common specifications (Commission Implementing Regulation (EU) 2022/2346) apply to listed non-medical products such as coloured contact lenses, cosmetic implants and fillers, liposuction equipment and certain lasers and light-based equipment.
  • EUDAMED mandatory modules — following the Commission's functionality decisions, use of the Actor Registration, UDI/Device Registration, Notified Bodies & Certificates, and Market Surveillance modules became mandatory EU-wide in 2026, including in Greece. Devices first placed on the market under the mandatory regime must be registered in EUDAMED before first placement; devices already on the market that continue to be placed on the market must complete device registration within the transitional window set by the Commission decision. EOF validates EUDAMED actor applications for Greek-established manufacturers, authorised representatives, importers and system/procedure pack producers. Doğrulanacak: the exact EUDAMED transitional deadline that applies to your device's first-placement date, at filing.

1.2 The Greek layer

InstrumentWhat it does
Law 1316/1983Establishes EOF, its organisation and its competence for medicines, medical devices and other health products.
Joint Ministerial Decision DY8d/130648/2009 (Government Gazette B' 2198/2.10.2009)The historic "Medical Devices" decision transposing Directive 93/42/EEC: the source of the EOF notification duty (notably Article 14 in the sources describing it), the data, labelling and IFU to be submitted, the national registry procedure, and the Greek clinical-investigation and advertising provisions. Read together with the MDR today.
Joint Ministerial Decision DY8d/130644/2009Active implantable medical devices — transposition of Directive 90/385/EEC. Historic layer for AIMDs, now governed on conformity by the MDR.
Joint Ministerial DecisionIn vitro diagnostic medical devices — transposition of Directive 98/79/EC, read with
DY8d/3607/892/2001EOF Circular 62701/2003 on IVD registration in the sources describing the older regime. Conformity is now governed by the IVDR.
EOF Circular No. 42353/2011 and Article 15 of the Medical Devices MDClinical investigations of non-CE-marked devices, or CE-marked devices tested for new uses outside their label: application to EOF, with the National Ethics Committee opinion.
EOF Circular No. 55220/2009; EOF Circular No. 18158/2010; the SEP CodeHealth-product and device advertising and industry-conduct layer in Greece. Device advertising is governed by the Medical Devices MD and the AIMD decision in combination with EOF circulars and the Greek industry code (SEP, the local equivalent of the MedTech Europe / EUCOMED code tradition). Doğrulanacak: the current EOF advertising circular and code edition at campaign launch.
Law 4472/2017Establishes EKAPY, the National Central Health Procurement Authority, and the central health procurement framework, including its product, supplier and price registers.
Law 5129/2024, Article 72, and Ministerial Decision 3263/11-3-2025The public-hospital device rebate: scope, the 0% / 3% / 5% quarterly scale by turnover per hospital, the capital-equipment and >5%-saving tender exceptions, quarterly calculation on invoices excluding VAT, and payment deadlines after each quarter (for the first quarter, by 31 May in the published summary). Applies to invoices issued after 11 March 2025.
EOPYY reimbursement framework and the EKAPY–EOPYY RegistryThe payer layer for devices reimbursed outside inpatient care: registry entry as a precondition of lawful clawback and reimbursement, reimbursement conditions and lists, and the clawback mechanism itself, as tested in the Council of State rulings on the 2017 and 2018 clawbacks delivered in 2026. Doğrulanacak: the current EOPYY list, price basis and clawback percentage methodology for your category, at entry — these are set and revised by ministerial and EOPYY decisions and must never be taken from a secondary summary.
Greek public procurement law (Law 4412/2016, as amended) and the ESIDIS / KHMDIS systemsHow public hospitals and EKAPY buy: electronic procedures in ESIDIS above the national threshold (reported as €30,000, with direct awards below and all hospital tenders above €1,000 recorded in KHMDIS), publication in TED above the EU threshold (reported as €216,000 for supplies in the current period — to be verified at tender, EU thresholds are revised every two years), the price observatory as a ceiling in many device tenders, and the 60-day payment term that applies to hospital payments rather than the general 30-day term.

1.3 Who does what

  • Manufacturer — full MDR Article 10 duties. If established in Greece: EUDAMED actor registration validated by EOF, and EOF notification of devices placed on the Greek market. If established elsewhere in the EU and supplying Greece: no second Greek marketing authorisation, but the Greek notification, language and payment layers in this guide still have to be worked through with the Greek chain.
  • Authorised representative (AR) — mandatory for manufacturers established outside the EU (MDR Article 11), established anywhere in the EU. Greece imposes no rule that the AR be Greek — but a Greek-established AR is validated by EOF in EUDAMED, and in the historic Greek framework it is the manufacturer or its authorised representative who submits the EOF notification.
  • Importer — MDR Article 13 duties; the first EU operator making a third-country device available.

A Greek-established importer registers in EUDAMED, validated by EOF, verifies CE marking, labelling, UDI and the EU Declaration of Conformity, and carries Greek traceability and complaint duties.

  • Distributor — MDR Article 14 duties, plus the Greek reality that a distributor operating in Greece must be a properly established Greek economic operator able to be supervised by EOF, to hold the device data, Greek labelling and IFU for the devices it supplies, and to execute a field safety corrective action in Greek without delay. Vet a Greek distributor on its EOF-facing status, its EKAPY/EOPYY registry experience for your category and its tender record, before signing — not after the first hospital asks for the notification evidence.
  • Custom-made device manufacturers — in the Greek framework described in the Legal 500 and Lexology sources, manufacturers of Class I and made-to-order (custom-made) devices selling in Greece under their own name are registered in the EOF register, and custom-made devices follow the MDR Annex XIII statement procedure rather than CE marking. Do not label a mass-produced device adapted to a professional user's requirements as "custom-made" to use this route.
  • Health institutions that manufacture and use devices in-house (MDR Article 5(5)) — subject to the MDR conditions for in-house manufacture and use, documentation and safety reporting.

Doğrulanacak: EOF's current notification and documentation practice for in-house devices at implementation.

2. Device Classification

Greece applies the MDR classification rules (Annex VIII) unchanged — there is no Greek reclassification layer. Classification is by intended purpose and inherent risk, across 22 rules in four groups (non-invasive Rules 1–4, invasive Rules 5–8, active Rules 9–13, special Rules 14–22). Where more than one rule applies, the rule resulting in the higher class prevails. Accessories are classified in their own right, software is classified under Rule 11, and the classification rationale must be documented in the technical documentation. The European Commission's classification guidance summarises the four classes as I, IIa, IIb and III depending on risk during use, with the detailed principles in Article 51 and Annex VIII and the conformity procedures in Article 52.

ClassTypical examplesConformity route in brief
I (incl. Is sterile, Im measuring, Ir reusable surgical)Non-invasive low-risk devices, simple dressings, non-sterile examination gloves, wheelchairs, simple instrumentsSelf-declaration for plain Class I (Article 52(7)); notified body involvement limited to the sterile, measuring or reusable-surgical aspect for Is/Im/Ir
IIaShort-term invasive devices, hearing aids, dental materials, many monitoring and diagnostic devices, most decision-support software as the starting point under Rule 11Notified body (Annex IX QMS + technical-documentation assessment on at least one representative device per category, or Annex X + XI route); typical end-to-end timelines in current EU practice of around 12–18 months
IIbLong-term invasive devices, infusion pumps, ventilators, most implants not in Class III, active therapeutic devices with potentially hazardous energy exchangeNotified body, with technical-documentation assessment per generic device group; commonly around 12–18 months, longer for implantables
IIIHeart valves, implantable pacemakers and neurological devices, absorbable implants, breast implants, devices incorporating an ancillary medicinal substance, defined surgical meshesNotified body with full technical-documentation / design examination; Summary of Safety and Clinical Performance (SSCP) publicly accessible via EUDAMED; expert-panel or clinical-evaluation consultation and, for ancillary medicinal substances, competent-authority/EMA consultation where applicable; commonly 18–24 months
IVDs (IVDR)Classes A–D under IVDR Annex VIII (Rules 1–7)A: self-declare (notified body for sterile A); B–D: notified body, with the heaviest scrutiny, EU reference laboratory involvement and common-specification compliance for Class D

Two Greek practical notes:

1. Do not import an old MDD class into a Greek filing. Software (Rule 11), substance-based devices, nanomaterial devices and reusable surgical instruments in particular moved up under the MDR. EOF, a notified body, EOPYY checking a reimbursement file or a hospital buyer checking your EUDAMED and EOF record will work from the MDR class — and the class in the Greek notification must match it.

2. Greece has no widely listed MDR notified body of its own for a general device scope. A manufacturer may use any EU-designated notified body whose NANDO scope covers the device and the route; there is no rule requiring a Greek body for the Greek market. Doğrulanacak: the current NANDO list at application — do not assume any Greek-designated body covers your device codes, and obtain the chosen body's own current lead time in writing.

Software deserves a separate line for Greece: MDR Rule 11 puts most software that provides information used for diagnostic or therapeutic decisions at Class IIa or higher, and higher where the decision could cause death or serious deterioration of health. Classify the claim you actually make in the Greek IFU — the Greek text is the text Greek users, EOF and a Greek court will read.

3. The EU Authorised Representative, Importer and Distributor in Greece

A non-EU manufacturer (for example a Turkish, US, UK or Chinese maker — note that Great Britain is a third country for this purpose, while Northern Ireland follows EU goods rules under the Windsor Framework) selling into Greece needs the standard MDR chain, with Greek specifics at each link:

1. EU Authorised Representative — appointed by written mandate (MDR Article 11), established anywhere in the EU. The AR verifies that the EU Declaration of Conformity and technical documentation have been drawn up, keeps documentation available for EOF and other competent authorities for at least 10 years after the last device is placed on the market (15 years for implantables), cooperates on corrective actions, and is jointly liable for defective devices on the same basis as the manufacturer where the manufacturer has not complied with its obligations. If the AR is established in Greece, EOF is its validating authority in EUDAMED.

2. Importer — the first EU operator making a third-country device available, with Article 13 verification (CE marking, labelling, UDI, EU Declaration of Conformity, manufacturer and AR identification), storage and transport conditions that do not jeopardise compliance, and complaint-register and traceability duties. A Greek-established importer registers in EUDAMED, validated by EOF.

3. Distributor — Article 14 verification and traceability duties, plus the Greek layer: the distributor is the operator EOF, EKAPY, EOPYY and the hospitals will actually deal with day to day. It must be able to evidence the device's EOF notification and EUDAMED registration, supply the Greek labelling and IFU set, register products and prices in the EKAPY/EOPYY systems where the category requires it, bid in ESIDIS in Greek, and carry the rebate and clawback mechanics in its pricing. This is why distributor selection in Greece is a regulatory and payer act, not only a commercial one.

4. Servicing partner — if your device needs installation, maintenance, calibration, safety checks or repair in Greece, put a competent Greek service chain in place before the first tender. Greek hospital tenders routinely score or require local servicing, spare-parts and response-time capability, and a foreign manufacturer with no Greek service answer loses on that ground alone, however strong the CE file is.

5. Relabelling / repackaging trap (MDR Article 16) — a distributor or importer that translates, relabels or repackages a device for Greece can step into manufacturer-level duties for that activity, including notified-body certification of its quality system for the activity where required. Greek translation of labelling and IFU should therefore be done under the manufacturer's control wherever possible, with the translator working from the controlled master file — not improvised by the local partner.

6. PRRC — the Person Responsible for Regulatory Compliance (MDR Article 15) requirement is EU-wide. Micro and small enterprises may rely on an external PRRC who is permanently and continuously at their disposal.

7. Customs and tax border — Greece is inside the EU customs union and uses the euro, so a shipment from a German or Italian warehouse to an Athens hospital is an intra-EU supply, not an import. Goods arriving from outside the EU are imported through Greek customs, with import VAT (standard Greek VAT is 24%, with reduced rates for defined categories — to be verified: the current VAT treatment for your device category and the duty position for your commodity code, at shipment) and any duty accounted for at import by commodity code.

Free movement point: a device that is CE-marked and lawfully marketed in another EU country can be brought into Greece with no additional Greek marketing authorisation. The Greek work is therefore not a second certification; it is (a) the EOF national notification, (b) the Greek-language set, (c) a compliant Greek distribution and servicing chain, and (d) the EOPYY / EKAPY / price-observatory / rebate work that makes the device payable and biddable in Greece.

4. The Greek Registration Architecture: EUDAMED + EOF National Registry (+ the Payment Registries)

Greece currently has two regulatory registration layers, plus payment layers that foreign teams routinely conflate with them. Keep them separate in your launch plan.

4.1 Layer 1 — EUDAMED (EU-wide, mandatory in 2026)

1. Actor registration — manufacturer, AR, importer and system/procedure pack producers register and receive a Single Registration Number (SRN) after validation by the competent authority of their place of establishment — EOF for Greek-established operators. The non-EU manufacturer links to its AR in EUDAMED; that linking can only be completed once the AR has its SRN.

2. Device registration — in the UDI/Device module, at Basic UDI-DI / UDI-DI level with the European Medical Device Nomenclature (EMDN). Receiving an SRN does not register the device: actor registration and device registration are separate acts, and both must be complete before first placement under the mandatory regime.

3. Certificates and vigilance modules — notified body certificates are recorded in EUDAMED, and market surveillance and vigilance reporting increasingly run through EUDAMED alongside the national EOF channel. Keep both channels in your vigilance procedure until EOF confirms a national channel is closed for your report type.

4.2 Layer 2 — EOF National Electronic Medical Device Registry (Greek-specific)

Under the Greek framework described in Section 1.2, manufacturers or their authorised representatives notify EOF of devices placed on the Greek market, submitting all data allowing identification of the device together with its labelling and instructions for use, at the time the device is placed on the market and put into service in Greece. Notification is through EOF's National Electronic Medical Device Registry. The sources describe a registry procedure for medical devices generally, and specifically for IVDs, for Class I devices and for custom-made devices.

Practical consequences:

  • File the Greek notification as a Greek step in the launch plan, with its own owner, its own data set (device identification, labelling, IFU in Greek) and its own completion evidence — do not bury it inside "EUDAMED done".
  • Keep the Greek record current. A new model, a changed intended purpose, a changed Greek IFU or a new device family supplied through the same distributor is a notification-update event, not just a new price-list line.
  • Never write "approved by EOF" in Greek marketing material. Notification and registration are identification and surveillance processes — not an EOF assessment of safety or performance, and not an endorsement. The claim collides with MDR Article 7 (misleading claims), the Greek advertising layer and Greek consumer law.

Doğrulanacak: the current EOF registry portal, account and credential process, data fields, any per-device or per-family grouping practice, and any current administrative fee for the notification, at filing. No reliable current published EOF tariff for the ordinary device notification was verified for this guide, so no figure is stated in Section 7 as a verified tariff.

4.3 Payment layers (not marketing authorisations)

  • EKAPY–EOPYY Registry (reimbursed devices): for devices reimbursed by EOPYY outside inpatient care, registry entry and the correct price basis are the gateway to reimbursement and the precondition the Council of State treated as decisive for lawful clawback. Without it, the device is not in the payer system, however clean its CE and EOF file is.
  • EKAPY central procurement registers: EKAPY maintains registers of products, suppliers and prices for central procurement. For centralised categories, being correctly recorded there is the gateway to framework agreements and call-offs.
  • Price observatory: for hospital supplies and devices, recorded award prices per item act in many tenders as the maximum acceptable price. Entering Greece at a price above the observatory price for the item is, in those tenders, entering at a price that will be rejected.

5. Step-by-Step: Placing a Medical Device on the Greek Market

Step 1 — Confirm the product is a medical device (or IVD) and classify it. Apply MDR Article 2 and Annex VIII / IVDR Annex VIII. Screen Greece-relevant borderlines early: Annex XVI aesthetic products, software (Rule 11), substance-based devices, and products near the medicinal-product line (EOF regulates both medicines and devices, so a borderline argument will be read by an authority that owns both sides of the line).

Step 2 — Build the MDR conformity file. Quality management system (ISO 13485-aligned; ISO 13485 certificates are typically issued for a three-year cycle with annual surveillance), technical documentation (Annexes II–III), GSPR checklist (Annex I), risk management (ISO 14971), clinical evaluation (Article 61, Annex XIV) or performance evaluation for IVDs, post-market surveillance plan, Basic UDI-DI and UDI-DI assignment, and the PRRC appointment.

Step 3 — Decide, at the outset, which payment track the device will live on. EOPYY-reimbursed outpatient/consumable track (registry, reimbursement conditions, clawback exposure), public-hospital track (EKAPY central or hospital tender, price observatory ceiling, Law 5129/2024 rebate), or purely private/commercial track (private hospitals, clinics and retail, where EOPYY and EKAPY do not pay but EOF and language duties still apply in full). This decision drives the evidence, pricing and partner work in Steps 7–10 and cannot be retrofitted cheaply.

Step 4 — Fix the operator chain. Appoint the AR (non-EU manufacturers) by written mandate; confirm who the importer is; select the Greek distributor and verify its establishment, EOF-facing status and category experience before contract signature; confirm who will perform servicing and instruction in Greece; document who performs any Greek relabelling or repackaging and whether Article 16 duties are triggered. Remember the sequencing rule: the AR must have its SRN before the non-EU manufacturer can link to it in EUDAMED.

Step 5 — Conformity assessment. Plain Class I: self-declare. Is/Im/Ir, IIa, IIb, III and IVD Classes B–D: engage an EU-designated notified body with the right NANDO scope (Section 2). EU MDR certificates are issued for a maximum of five years, with annual surveillance and recertification. Notified-body capacity — not EOF — is the usual bottleneck, so apply early, and never lodge a parallel application for the same procedure with two notified bodies.

Step 6 — Register the actors, then the devices, in EUDAMED. Actor registration and SRN for manufacturer, AR, importer and system/procedure pack producers (EOF-validated for Greek-established operators); device registration in the UDI/Device module before first placement under the mandatory regime. Keep the information current thereafter.

Step 7 — Complete the Greek national layer. EOF notification through the National Electronic Medical Device Registry, with the device identification data, labelling and IFU (Section 4.2). Confirm at filing whether EOF requires any additional national step for your device type — in particular for Class I, custom-made and IVD devices, where the Greek sources describe a specific registry procedure.

Step 8 — Greek labelling and IFU. Produce the language set in Section 6 before the first Greek delivery: Greek label, Greek IFU, Greek implant-card content where applicable, and Greek field-safety material ready to issue. Allow real translation and independent review time — this step, not the authority queue, is what most often slips in a Greek launch.

Step 9 — Payment readiness. Reimbursed track: confirm the EKAPY–EOPYY registry entry, the reimbursement conditions and the price basis for the category, and model the clawback exposure in the price. Hospital track: confirm whether the category is EKAPY-centralised or hospital-bought, check the price observatory position for the item, map the target hospitals' and health regions' procurement calendars, and rebuild the price net of the Law 5129/2024 rebate (0% / 3% / 5% by quarterly turnover per hospital). Private track: confirm the private groups' own listing and health-technology assessment processes — they are commercial, but they will still ask for the CE, EUDAMED and EOF evidence.

Step 10 — Logistics and tax setup for the first shipment. Intra-EU supply: no customs declaration, normal EU VAT treatment in euros. Third-country supply: importer of record, customs declaration, import VAT (and any duty) accounted for — agree in writing who carries this and under which Incoterm.

Step 11 — Commercial entry: tender or private channel. Map the buyer: EKAPY frameworks for centralised categories, individual ESY hospitals and their YPE health regions buying in ESIDIS for the rest (notices recorded in KHMDIS and, above EU thresholds, in TED), the EOPYY-contracted provider channel for reimbursed devices, and the private hospital groups and clinics. Monitor by CPV code — health purchases concentrate in CPV division 33, notably 33100000 (medical equipment), 33140000 (medical consumables) and 33190000 (miscellaneous medical devices).

Step 12 — Only if relevant: clinical investigation in Greece. Applications for a clinical investigation of a device that does not bear the CE mark, or that is to be tested for new uses not covered by its CE mark, are submitted to EOF, under Article 15 of the Medical Devices MD and EOF Circular 42353/2011 in the framework described by the Greek sources, within the MDR Articles 70–82 timetable. A positive scientific assessment by EOF and a positive ethical assessment by the National Ethics Committee are both required; in case of a negative ethics opinion, EOF rejects the application, and EOF informs the sponsor of the outcome. Participant-facing documents, summaries and labelling/IFU for the investigation must be in Greek. Doğrulanacak: the current EOF clinical-investigation application channel, form and fee at application — do not take a fee or a fixed-day promise from a secondary summary. A CE-marked device used within its label does not need a clinical-investigation authorisation; confirm your pathway against current EOF guidance before submission.

Step 13 — Only if relevant: derogation and custom-made routes. The MDR Article 59 derogation route (competent-authority authorisation of a specific device that has not completed conformity assessment, in the interest of public health or patient safety) is administered nationally by EOF. A derogation is an exception for a defined situation, not a launch strategy. Custom-made devices follow the Annex XIII statement route and the Greek registry procedure for custom-made devices — they do not bear a CE mark for that statement route.

Step 14 — Post-market system live before first sale. PMS/PMCF (or PMPF for IVDs), vigilance reporting to EOF (and through EUDAMED where the module applies) within MDR deadlines, FSCA/ FSN handling in Greek, Article 10a supply-shortage notification readiness, servicing records with your Greek service partner, complaint handling with the Greek chain, and document retention (10 years; 15 for implantables) — see Section 9.

6. Language Requirements: the Greek Layer in Detail

Greece is a Greek-language market for regulatory purposes. The MDR requires device information in the language(s) determined by the member state where the device is made available;

Greece's determination is Greek, listed as mandatory in EU language-requirement tables, with no general professional-use English allowance stated in those tables.

InformationRequired language
Labelling and instructions for use — professional useGreek
Labelling and instructions for use — lay users (patients and other consumers)Greek — the version the patient actually receives must be comprehensible to that patient in Greek
Implant card / implantable-device information for the patient (MDR Article 18)Greek, covering device identification, precautions and interference, expected service life and follow-up, and safe-use information
Documents filed in the EOF notification (device identification, labelling, IFU)Greek labelling and IFU as submitted to EOF; confirm at filing which supporting documents EOF accepts in English for your procedure — to be verified
Clinical-investigation fileParticipant information, summaries and device labelling/IFU in Greek; application and supporting file language to be confirmed document-by-document with EOF at submission — to be verified
EU Declaration of Conformity and technical documentationKeep the controlled master in your QMS language; be ready to provide Greek versions of user-facing and authority-requested documents — do not assume an English-only file will be accepted for a Greek procedure
Field safety notices reaching Greek usersGreek, issued promptly through the Greek chain
Tender documents and bids in ESIDIS / EKAPY proceduresGreek — tender documentation is issued in Greek and deadlines are hard

Practical points:

  • Translation is a conformity activity, not a marketing task. The Greek IFU is part of the device as supplied in Greece. A mistranslated warning, a sizing table that diverges from the master, or a Greek label that omits a master-file symbol explanation is a nonconformity, and it will be read against the master file in any EOF inspection or incident investigation.

Use a translator and an independent reviewer working in Greek with medical-device experience, and lock the approved Greek text under document control. Do not let a distributor improvise a Greek text outside the manufacturer's controlled file — apart from the conformity risk, it can trigger the Article 16 trap in Section 3.

  • English is a working convenience, not a right. EOF publishes selected information in English and its staff work with English in EU procedures, but the registry, tender documentation, reimbursement lists and user-facing information are Greek. Plan on Greek; treat any English acceptance as procedure-specific, to be confirmed for that procedure.

7. Timelines and Fees

7.1 Timelines (indicative — the conformity route, not EOF, sets most of the clock)

StepTypical timeNotes
Plain Class I self-declaration (file, EU Declaration of Conformity, EUDAMED registration)~4–8 weeksIndicative range assuming the conformity file is ready; Greek translation and review are usually the critical path
EOF national notificationFiled when the device is placed on the market and put into service in Greece; allow several weeks including registry account setup, preparing the Greek data set and internal reviewDoğrulanacak: EOF's current processing and validation practice at filing — no reliable published fixed-day standard was verified for this guide, so do not quote one to management as a commitment
EUDAMED actor validation by EOFAllow several weeksData readiness and the AR-linking sequence (Step 4) are what usually slip, not the authority queue alone
Notified-body certification, Class IIaCommonly ~12–18 months end-to-end in current EU practiceObtain the notified body's own current lead time in writing at application
Notified-body certification, Class IIb / IIICommonly ~12–24+ monthsClass III design examination and, where applicable, expert-panel / ancillary-substance consultation extend the clock
IVDR Class C / D certificationCommonly ~12–24+ monthsIVDR notified-body capacity remains tight EU-wide
Clinical investigation (EOF + National Ethics Committee)Within the MDR Article 70 timetable (validation and assessment phases totalling roughly 45–60+ days depending on completeness and clock stops)Both EOF scientific assessment and National Ethics Committee opinion are required; sequence ethics and competent-authority work together; to be verified at application
EKAPY / hospital tender cycleProcedure windows are set per tender, typically several weeks to a few months from notice to award, with supply contracts and frameworks then running for their term (commonly 1–4 years)Notices in ESIDIS/KHMDIS and TED; a missed window in a key hospital or a centralised framework usually means waiting for the next competition in that category
Hospital payment term60 days for hospital payments, rather than the general 30-day public-sector term, in the procurement summariesModel Greek public-hospital receivables on 60 days as the legal term — and stress-test the plan for longer actual collection, given the scale of hospital overdue debt reported in Greece

Certificates issued under the MDR are valid for a maximum of 5 years, with annual surveillance and recertification; the EU Declaration of Conformity and technical documentation must be kept for at least 10 years after the last device is placed on the market (15 years for implantable devices).

7.2 Fees and costs

Greece's fee picture must be stated plainly rather than filled with an invented tariff:

Cost itemAmount / positionNotes
EUDAMED registrationNo EUDAMED filing feeCosts are internal data-preparation and maintenance time
EOF national notificationNo current published EOF tariff for the ordinary device notification was verified in a primary EOF source for this guide.Doğrulanacak — important: confirm, before budgeting, whether any administrative fee under current EOF practice applies to your specific notification, clinical investigation or other request, at filing. Do not present any secondary-source "EOF registration fee" figure to a board as a current verified tariff without that confirmation
Clinical investigation / performance studyFee position to be confirmed with EOF at applicationDoğrulanacak: current EOF amount and the National Ethics Committee's current review fee for your investigation type
EKAPY–EOPYY registry / reimbursement entryNo per-device listing fee was verified in the sources used for this guide; the economic content of the system is the reimbursement price basis and the clawback exposure, not a filing feeDoğrulanacak: whether any charge applies to your registry or pricing request, and the current price and clawback methodology for your category, at entry
Public-hospital rebate (Law 5129/2024)0% / 3% / 5% of quarterly invoice value excluding VAT, per supplier per hospital, by quarterly turnover band (up to €10,000 / €10,000– €30,000 / above €30,000), with the capital-equipment and >5%-tender-saving exceptions in Section 1.2This is not a filing fee — it is a recurring deduction from Greek public-hospital revenue. It must be in the price model, not discovered in the first quarterly notification. Applies to invoices issued after 11 March 2025
Notified-body conformity assessmentCommercial fees — the real cost centre. As indicative market ranges only (not official tariffs): initial MDR certification for a single device family commonly runs into the tens of thousands of euros (QMS audit day rates plus technical-documentation review), with annual surveillance audits and 5-year recertification on top; complex Class III files cost materially moreObtain a written quotation from the notified body for your device scope
Greek translation (labelling, IFU, implant card, FSN templates)Commercial cost, per-word/per-document plus independent reviewBudget Greek medical translation + reviewer as a conformity line
Import VAT / duty (third-country goods only)Import VAT on the customs value, plus any duty by commodity code — normally recoverable as input VAT for a VAT-registered Greek importer, a cash-flow cost otherwiseDoğrulanacak: current VAT rate treatment for your category (standard 24% vs a reduced rate) and code-specific duty at shipment
Greek distributor / AR / servicing servicesCommercial feesAR mandate fees, distributor margin and Greek servicing capacity are negotiated — they are operating costs, not authority fees

Budgeting rule for Greece: the authority-fee line is small and, for the ordinary notification, unverified — do not invent it. The real Greek budget lines are notified-body certification, controlled Greek translation, the operator chain (AR, importer, a Greek distributor and service partner), and the payer economics: the price-observatory ceiling, the Law 5129/2024 hospital rebate and, on the reimbursed track, clawback exposure. A budget that shows a large "EOF registration fee" and no translation, rebate or clawback line has the Greek cost structure backwards.

8. Selling After Registration: Reimbursement, Procurement and Channels

8.1 Three payment tracks — keep them apart

Track A — EOPYY-reimbursed devices (outpatient / supplied to insured patients). EOPYY is the single public payer for insured healthcare services. For devices and consumables it reimburses outside inpatient hospital care, the gateway is inclusion in the reimbursement system through the EKAPY–EOPYY registry, at the correct (import/wholesale) price basis, under the reimbursement conditions and lists in force for the category. Expenditure in a category is monitored against its budget, and overspend has historically triggered clawback from suppliers. The 2026 Council of State rulings draw the line precisely: 2017 clawback annulled where the registry system was not adequately regulated and products had not been properly brought into it; 2018 clawback upheld for parties who had registered and thereby accepted the framework. Register properly, price on the correct basis, and reserve for clawback in the category forecast. Doğrulanacak: the current reimbursement list position, conditions and clawback percentage for your exact category at entry.

Track B — Public hospitals (ESY). More than 120 hospitals under seven YPE regions buy either through EKAPY central frameworks and dynamic purchasing systems (for centralised categories) or through their own tenders in ESIDIS, recorded in KHMDIS and published in TED above EU thresholds, with CPV division 33 as the monitoring vocabulary. Tender files commonly require, in the procurement summaries: CE marking under Regulation (EU) 2017/745, registration in the national device register, and ISO 13485 for manufacturers. The price observatory price for the item is often the maximum acceptable bid. Net price must then survive the Law 5129/2024 rebate in Section 7.2, and cash collection must be modelled on the 60-day hospital payment term with a stress case beyond it.

Track C — Private sector. Private hospital groups, clinics, diagnostic centres, pharmacies and direct professional sales. No EOPYY or EKAPY payment, no public rebate — but full EOF notification, Greek language, vigilance and advertising duties, and each group's own commercial listing process.

For many foreign manufacturers this is the fastest first revenue in Greece, while the public tracks are built in parallel — provided the private price does not destroy the later public price-observatory position. Set the Greek price architecture once, across all three tracks, at entry.

8.2 Advertising and professional interaction

Device advertising in Greece is governed by the Medical Devices MD and the AIMD decision in combination with EOF Circular 18158/2010 and the Greek industry (SEP) code, on top of MDR Article 7's ban on misleading claims. Keep Greek promotional claims inside the CE-marked intended purpose as stated in the Greek IFU, never claim EOF "approval", and run gifts, hospitality, sponsorship and healthcare-professional interaction through the Greek code as well as your home code.

Doğrulanacak: the current EOF advertising circular and SEP code edition at campaign launch.

9. Post-Market Obligations in Greece

Registration is the start of the Greek obligations, not the end of them:

  • Post-market surveillance (MDR Articles 83–86): a PMS system and plan for every device, a PMS report for Class I and a Periodic Safety Update Report (PSUR) for Class IIa, IIb and III, updated at the MDR frequencies, plus post-market clinical follow-up (PMCF) — or performance follow-up (PMPF) for IVDs — feeding back into the clinical evaluation and risk file.
  • Vigilance (Articles 87–92): serious incidents and field safety corrective actions reported to EOF (and through EUDAMED where the applicable module requires it) within the MDR deadlines — immediately for a serious public-health threat, within 2 days; within 10 days for death or unanticipated serious deterioration; within 15 days for other serious incidents. Trend reporting for non-serious incidents and statistically significant increases.
  • Field safety notices: issued in Greek to Greek users, through the Greek distributor chain, with evidence the notice reached the affected hospitals and users — a notice sent only to a distributor's inbox is not a completed Greek FSCA.
  • Supply-interruption notification (Article 10a): manufacturers must notify the competent authority and the economic operators and health institutions they supply directly of an anticipated interruption or discontinuation that could result in serious harm or a risk of serious harm, in principle at least six months in advance, except in exceptional circumstances.
  • Traceability and UDI: UDI carriers on labels and packaging at the MDR phase-in dates for the class, UDI and device data kept current in EUDAMED, and implant cards for implantable devices (Article 18) in Greek.
  • Distributor and importer duties continue after sale: verification on receipt, storage and transport conditions, complaint forwarding, and cooperation on corrective actions — audit your Greek chain against Articles 13 and 14, not only against its sales targets.
  • Servicing and maintenance records: for devices in Greek hospitals, maintenance, calibration and safety-check records held with the Greek service partner, available to the hospital and, on request, to EOF.
  • Document retention: technical documentation, EU Declaration of Conformity and certificates available for at least 10 years after the last device is placed on the market, 15 years for implantables.
  • Rebate and clawback administration: quarterly Law 5129/2024 rebate calculation and payment per hospital on notification (first-quarter deadline reported as 31 May, with the following quarters on the ministerial decision's calendar — to be verified: the current year's exact quarterly deadlines), and EOPYY clawback settlement on the reimbursed track. Put both in the Greek finance calendar, with an owner — a missed rebate deadline is a compliance failure, not a late payment.

10. Common Mistakes in Greece

1. Treating EUDAMED as the whole Greek file. EUDAMED registration does not replace the EOF national notification. File both, evidence both.

2. Treating EOF notification as an "approval" — in the plan or in marketing. It is a notification and registry entry. "Approved by EOF" is untrue and unlawful as a claim.

3. Launching with an English IFU and a promise to translate later. Greek labelling and IFU are mandatory. The Greek set is a pre-launch conformity item, not a post-launch localisation project.

4. Building the launch on the old directive decisions alone. The 2001/2009 Joint Ministerial Decisions explain the Greek procedural layer; the conformity route, classification, UDI, EUDAMED and PMS duties are MDR/IVDR duties now. A plan that cites only DY8d/130648/2009 will miss them.

5. Choosing a Greek distributor on sales coverage alone. The distributor carries the EOF-facing, EKAPY/EOPYY, ESIDIS tender, servicing and FSCA execution for your device in Greece.

Vet its regulatory and payer capability with the same rigour as its hospital relationships.

6. Pricing Greece off the home-market price list. The price observatory ceiling, the 0%/3%/5% hospital rebate and, on the reimbursed track, clawback exposure and the correct price basis decide the real Greek net price. Model all three before quoting the first hospital.

7. Registering a reimbursed product on the wrong price basis or outside the registry, and assuming clawback cannot reach you. The Council of State litigation shows the opposite structure: proper registry entry is what makes the system lawful and predictable; improvisation is what creates annulment fights and unbudgeted returns.

8. Missing the central-versus-hospital split. A category EKAPY has centralised is won or lost in the central framework competition, not hospital by hospital — and the reverse is also true. Confirm which track your category is on before building a hospital-by-hospital sales plan.

9. Ignoring the Article 16 trap in Greek translation. A distributor translating and relabelling outside the manufacturer's control can acquire manufacturer-level duties — and the manufacturer acquires a Greek text it does not control. Translate under document control.

10. Forgetting Cyprus is not a Greek filing shortcut. Cyprus uses Greek as an official language and has its own competent authority and national requirements. A Greek EOF notification does not cover Cyprus, and a Cypriot filing does not cover Greece — even where the same Greek-language IFU can be reused after a separate check.

11. Frequently Asked Questions

Q: Who regulates medical devices in Greece? A: EOF — the National Organization for Medicines (Ethnikos Organismos Farmakon), a public entity under the Ministry of Health, established by Law 1316/1983. EOF is the competent authority for notification, market surveillance, vigilance and clinical investigations. EOPYY is the payer, and EKAPY is the central procurement authority — neither is the regulator.

Q: Do I need a separate Greek approval if my device is CE-marked? A: No separate Greek marketing authorisation or Greek CE mark exists. A CE-marked device can be traded across the EU.

Greece adds a national layer: EOF notification through its National Electronic Medical Device Registry, Greek labelling and IFU, a compliant Greek operator chain, and the payer/procurement work in Section 8.

Q: Is CE marking mandatory in Greece? A: Yes. Devices placed on the Greek market must comply with the EU MDR (or IVDR for IVDs) and bear the CE mark under the applicable conformity route, except for the defined exceptions such as custom-made devices on the Annex XIII statement route and Article 59 derogations.

Q: What are the device classes in Greece? A: The MDR classes, unchanged: Class I (including Is, Im, Ir), IIa, IIb and III under Annex VIII's 22 rules, and Classes A–D for IVDs under the IVDR.

Classification is by intended purpose; where more than one rule applies, the higher class prevails.

Q: Do non-EU manufacturers need a local representative in Greece specifically? A: They need an EU Authorised Representative established anywhere in the EU (MDR Article 11) — Greece does not require the AR to be Greek. They also need, in practice, a Greek importer/distributor chain for the Greek notification execution, Greek language set, tenders and servicing. A Greek-established AR or importer is validated by EOF in EUDAMED.

Q: What documents are required for the Greek file? A: The MDR core set — EU Declaration of Conformity, CE certificate (notified body, where the class requires it), technical documentation summary, clinical evaluation, risk management file, ISO 13485 / QMS evidence, UDI assignment, EUDAMED actor and device registration — plus the Greek set: device identification data, labelling and IFU in Greek for the EOF notification, implant-card content in Greek where applicable, and, for the payment tracks, the EKAPY–EOPYY registry and pricing data or the EKAPY/tender evidence set for the category.

Q: How long does Greek market entry take? A: For a plain Class I device with a ready file, the conformity, EUDAMED and EOF work is typically a matter of weeks, with Greek translation usually on the critical path. For Class IIa, IIb, III and IVD B–D devices, the notified-body route dominates: commonly around 12–18 months for IIa/IIb and 18–24+ months for Class III in current EU practice.

MDR certificates are valid for a maximum of five years. No verified fixed-day EOF processing standard is stated in this guide — see Section 7.1.

Q: What does Greek registration cost? A: EUDAMED registration has no filing fee. No current published EOF tariff for the ordinary device notification was verified for this guide, so no figure is presented as verified — confirm any EOF fee for your specific procedure at filing (to be verified).

The real costs are notified-body certification (typically tens of thousands of euros for an initial device family), controlled Greek translation, the operator chain, and the payer economics: the Law 5129/2024 hospital rebate at 0%/3%/5% of quarterly turnover per hospital and, on the reimbursed track, clawback exposure.

Q: Can I sell to Greek public hospitals without a tender? A: Public hospitals buy through EKAPY central frameworks for centralised categories or their own procedures in ESIDIS; direct awards exist only below the national threshold and within the procurement law's conditions. Assume a competitive procedure, a price-observatory ceiling and Greek-language tender documents as the default, and build the tender calendar into the launch plan.

Q: Does Greece reimburse medical devices? A: Through EOPYY, for defined devices and consumables supplied to insured patients outside inpatient hospital care, via the EKAPY–EOPYY registry, reimbursement lists and conditions — with clawback exposure when category expenditure exceeds budget. Hospital devices are paid through the hospital procurement track, not the EOPYY outpatient track. Confirm your category's current list position and conditions at entry (to be verified).

Q: I already sell in another EU country. What is left to do for Greece? A: Four things: (1) EOF national notification with the Greek data set; (2) the Greek labelling/IFU set under document control;

(3) a Greek distributor and servicing chain vetted for EOF, tender and FSCA execution; and (4) the payment work — EKAPY–EOPYY registry and clawback modelling if reimbursed, EKAPY/ESIDIS tender readiness, price-observatory check and rebate-net pricing if hospital-supplied.

Sources & scope

  • eof.gr, EOF (National Organisation for Medicines) — device notification and the national electronic registry, market surveillance and vigilance, and the e-fees (e-Paravolo) pages.
  • statistics.gr, ELSTAT — Hospital Census 2024; and the OECD/European Commission State of Health in the EU — Greece Country Health Profile 2025.
  • eur-lex.europa.eu, Regulation (EU) 2017/745 (MDR), Regulation (EU) 2023/607.

This guide is market-access intelligence, not regulatory or legal advice. EOF practice, the transitional position of the 2009 national framework, fee amounts and EKAPY programmes change — professional review is required before business use.