Iceland Medical Device Registration2026 Field Guide

No Icelandic CE mark — the MDR applies in Iceland through the EEA Agreement. Iceland adds its own layer on top: an IMA distributor registration that is immediate but fee-bearing, no national device register at all (devices live in EUDAMED), a published device tariff in krónur, an Icelandic-first language rule with a class-dependent English and Nordic opening, and a single payer for assistive devices, Iceland Health, whose grants only flow through contracted suppliers.

people living in Iceland on 1 January 2025 (Statistics Iceland)
389,444
spent on health per person (PPP) — OECD average USD 5,967 (OECD, Health at a Glance 2025)
USD 6,770
life expectancy at birth, 1.3 years above the OECD average (OECD, Health at a Glance 2025)
82.4 years

Sourced from the IMA, Ísland.is, the Act on Medical Devices No. 132/2020, the IMA fee tariff, the European Commission language-requirements overview, Statistics Iceland and the OECD, and verified 2026 publications.

Download the guide

Enter your work email and we'll send the PDF straight to your inbox.

Professional review is required before business use. See our privacy policy.

What's inside

Seventeen pages, zero fluff

Every figure in this guide traces to a named source — the IMA, Ísland.is, the Act on Medical Devices No. 132/2020, the IMA fee tariff, the European Commission, Statistics Iceland and the OECD — and every figure that could not be verified is flagged as such.

EU MDR via the EEA + the Icelandic layer

The MDR and IVDR have the force of law in Iceland through the EEA Agreement and the Act on Medical Devices No. 132/2020, supervised by the Icelandic Medicines Agency (Lyfjastofnun, IMA) under the Ministry of Health — the competent authority in the sense of Article 101 of the MDR and Article 96 of the IVDR. Iceland adds no second conformity assessment and no national device class.

EUDAMED for devices, the IMA for distributors

The IMA does not maintain a register of medical devices — devices are registered via EUDAMED, and manufacturers, importers and authorised representatives register as EUDAMED actors. What Iceland registers nationally is the distributor: distributors marketing devices in Iceland register with the IMA under Article 28(3) of the Act, distribution may begin on submission, and the Agency invoices ISK 53,400 under its published device tariff in force from 1 January 2026.

Icelandic-first rule + Iceland Health grants

Icelandic is the default language of the label and instructions for use — English or a Nordic language other than Finnish is allowed for Class I and IIa lay information, Icelandic or English for professionals, and the implant card must be in Icelandic. Payment for assistive devices runs through Iceland Health (Sjúkratryggingar Íslands), whose grants are conditional on purchase from a contracted company, while Landspítali, the only university hospital, anchors public hospital purchasing under the Act on Public Procurement No. 120/2016.

Complete guide

The guide in full

Every section of the PDF on one page. Use the form above to keep a copy.

Overview

Last updated: November 2026. Iceland is a member of the European Economic Area (EEA), not of the European Union, and the EU Medical Device Regulation (MDR, Regulation (EU) 2017/745) and the In Vitro Diagnostic Medical Devices Regulation (IVDR, Regulation (EU) 2017/746) have the force of law in Iceland through the Act on Medical Devices No. 132/2020 of 8 December 2020, which gives effect to the two regulations as adopted into the EEA Agreement. The competent authority is the Icelandic Medicines Agency (Lyfjastofnun, IMA — lyfjastofnun.is), acting under the Ministry of Health. On top of the EEA layer, Iceland runs a genuinely national set of rules: distributors who market devices in Iceland must register with the IMA under Article 28 of the Act, devices themselves are registered in EUDAMED and not in any Icelandic device register, the IMA charges for its device supervision under a published tariff (gjaldskrá) set under Article 39 of the Act — the tariff in force from 1 January 2026 is reproduced in Section 8 — and the language rule in Articles 12 and 19 of the Act, detailed in Regulation No. 630/2022, makes Icelandic the default language of the label and the instructions for use, with a defined opening to English and the other Nordic languages for lower-risk and professional devices. On the paying side, Iceland Health (Sjúkratryggingar Íslands) administers the statutory health insurance and the grants for assistive devices (hjálpartæki), and almost all hospital care is delivered by public institutions, above all Landspítali. Always confirm critical steps against current IMA guidance before filing.

There were 389,444 people living in Iceland on 1 January 2025, an increase of 5,718 (1.5%) on the year before, and 138,772 of them lived in Reykjavík (source: Statistics Iceland, The population on 1 January 2025, March 2025). Iceland spends USD 6,770 per person per year on health (PPP) — above the OECD average of USD 5,967 — while health spending equals 9.0% of GDP, close to the OECD average of 9.3% (source: OECD, Health at a Glance 2025: Iceland). Life expectancy was 82.4 years, 1.3 years above the OECD average, and 84% of health spending was covered by mandatory prepayment, against an OECD average of 75% (source: OECD, Health at a Glance 2025: Iceland). For a foreign manufacturer, Iceland is the smallest EEA device market that still runs the full EEA rulebook: one competent authority, one dominant public payer for assistive technology, one university hospital anchoring tertiary care — and a language spoken by fewer than 400,000 people that your labelling team has probably never budgeted for. Its value is not volume; it is completeness. A team that treats Iceland as "EEA, therefore CE mark, therefore done" will be stopped by a distributor registration it never filed, an Icelandic instruction leaflet it never commissioned, and a payer contract it never knew existed.

Because Iceland reads as "Nordic, English-speaking, EEA member", entrants miss five things. First, the MDR and IVDR apply as incorporated by decisions of the EEA Joint Committee — the Act states that the two regulations, as adopted in the EEA Agreement with the adaptations from Decisions No. 288/2019 and No. 90/2020, have the force of law in Iceland (source: Act on Medical Devices No. 132/2020, Article 2). Second, there is no Icelandic device register: the IMA does not maintain a registration of medical devices, devices are registered via EUDAMED, and what Iceland registers nationally is people — above all distributors (source: IMA, Registration of distributors; Ísland.is). Third, the distributor registration is fee-bearing but immediate: distribution may begin on submission, and the IMA invoices afterwards (source: Ísland.is). Fourth, Icelandic is the default language, with class-dependent exceptions — English or a Nordic language other than Finnish for Class I and IIa lay information, Icelandic or English for professionals, Icelandic for the implant card (source: Act, Articles 12 and 19; Commission language table, Rev. 3, Iceland row). Fifth, assistive-device payment runs through Iceland Health's grant system, conditional on purchase from a contracted company — a device outside the contract and product-list system is, for insured patients, commercially invisible (source: IBA Survey — Iceland; Ísland.is, January 2025).

Regulatory reality check up front: Iceland is not a separate certification regime from the EEA. There is no "Icelandic CE mark", and the IMA does not approve devices in the sense of granting a medicines-style marketing authorisation. What Iceland adds is a national layer with six specific features:

  1. 1. One act that carries the whole EEA framework — the Act on Medical Devices No. 132/2020 of 8 December 2020 gives the MDR and the IVDR the force of law in Iceland as adopted in the EEA Agreement, covers the full device life cycle from production to use, and makes the IMA the competent authority in the sense of Article 101 of the MDR and Article 96 of the IVDR, under the Minister of Health (source: Act on Medical Devices No. 132/2020, Articles 1–5). Where a question arises whether a product is a medical device at all, the IMA decides (source: Article 3; IMA, About Medical Devices).
  2. 2. A national register of operators, not of devices — the IMA maintains a register of the parties operating enterprises in Iceland which manufacture devices or are responsible for their marketing, while devices themselves are registered in EUDAMED (source: IMA, About Medical Devices; IMA, Registration of distributors). Manufacturers, importers and authorised representatives register as actors in EUDAMED; distributors responsible for marketing devices in Iceland register with the IMA itself, under Article 28(3) of the Act, and pay for it under the tariff (source: Ísland.is, Registration of medical device distributors).
  3. 3. A distributor registration that is immediate but not free — filed through the Ísland.is service portal or, if the web form cannot be used, by email to lyfjastofnun@lyfjastofnun.is with a defined data set (name, ID number, address, contact persons, and the brand list with a contact person per brand; Section 3.3 lists it in full) (source: Ísland.is). Processing is immediate — distribution may begin on submission — and the IMA registers the distributor as an entity subject to surveillance and invoices the fee under its tariff: ISK 53,400 for registration and ISK 26,700 for maintenance registration (source: Ísland.is; IMA fee tariff, Article 2).
  4. 4. An Icelandic-first language rule with class-dependent exceptions — the general rule is that instructions for use are in Icelandic, and where patients themselves operate the device the guidelines must be in Icelandic; under Article 12 of the Act, lay-user information must be in Icelandic, except that for Class I and IIa devices it may be in English or another Nordic language except Finnish; for professional users it may be in Icelandic or English; the implant card must be in Icelandic under Article 19, with an IMA-practice fallback to English or another Nordic language except Finnish where Icelandic is not possible (source: Act, Articles 12 and 19; Regulation No. 630/2022; Commission language table, Rev. 3, Iceland row; IMA, About Medical Devices).
  5. 5. A published, itemised device tariff — unlike many small EEA states, Iceland publishes exactly what the authority charges: a tariff set under Article 39 of the Act, in force from 1 January 2026 and replacing tariff No. 1555/2023, covering free sale certificates (ISK 53,400 for five copies), distributor registration (ISK 53,400), supervision (ISK 133,500 to ISK 774,300 by type), derogation assessments (ISK 200,250 or ISK 293,700 by class), clinical investigation fees (assessment ISK 1,161,450) and an expert hourly rate of ISK 26,700 (source: IMA fee tariff, Stjórnartíðindi B-deild, 30 December 2025). Section 8 reproduces the items. Fees are invoiced with a 30-day due date and are enforceable, with default interest on late payment (source: the same tariff, Article 10).
  6. 6. One payer for assistive technology and one anchor hospital — Iceland Health (Sjúkratryggingar Íslands) administers the statutory health insurance covering everyone who has legally resided in Iceland for six months, and contributes to the cost of the devices listed in the Regulation on grants for assistive devices, on condition that the device is purchased from a contracted company; where no contract exists, bids are sought or price surveys conducted (source: IBA Survey — Iceland). Its framework tender for wheelchairs and walking aids put contracts with seven companies in place to 30 November 2026, with a published product list (source: Ísland.is, Iceland Health news, 13 January 2025). On the hospital side, Landspítali is the country's only tertiary and university hospital, with around 25,000 inpatient stays and 407,000 outpatient visits a year, 624 beds and around 4,500 full-time-equivalent staff recorded for 2019 (source: Ministry of Health, December 2021). Winning Iceland is one authority file, one distributor registration, one payer contract system and one anchor hospital — small numbers, no hiding places.

A note on institutions. The Ministry of Health (Heilbrigðisráðuneytið) holds health policy and supervises the Act. The IMA (Lyfjastofnun) is the device competent authority: it registers distributors, keeps the register of responsible parties, runs post-market surveillance and an inspection programme, assesses derogation applications and clinical investigations, issues free sale certificates and charges for this work under the tariff (source: IMA, About Medical Devices; IMA fee tariff, 2025). Iceland Health (Sjúkratryggingar Íslands) is the insurance institution: it administers health-insurance rights, negotiates and pays for healthcare services, and runs the assistive-device grants and framework contracts (source: IBA Survey — Iceland; Ísland.is, Iceland Health). Public procurement runs under the Act on Public Procurement No. 120/2016; the central purchasing function formerly exercised by Ríkiskaup passed to the Financial Management Authority (Fjársýslan, FMA) on 1 August 2024, when Ríkiskaup was decommissioned (source: Ísland.is, 2024). This guide keeps those roles separate, because most Iceland errors are a filing, a grant application or a pitch sent to the wrong one of them.

A note on terminology and money. Iceland sources use the MDR terms manufacturer, authorised representative, importer and distributor; the authorised representative is viðurkenndur fulltrúi in Icelandic. Fees in this guide are in Icelandic krónur (ISK) as published in the tariff; no euro conversion is offered, because the tariff is set in krónur and exchange rates move. Where a figure could not be verified against a named official source, this guide marks the point to be verified rather than smoothing it over. No figure in this guide is a price for any service of ghmap.io, and nothing here is a promise of an outcome or a substitute for professional advice.

1. The Regulatory Framework: EU MDR via the EEA + the Icelandic National Layer

1.1 The two EU regulations apply in Iceland through the EEA Agreement — Iceland does not re-run them

Iceland is not an EU Member State. The MDR and the IVDR reach Iceland through the EEA Agreement, which extends the EU internal market for goods to Iceland, Liechtenstein and Norway, and through the Act on Medical Devices No. 132/2020, whose Article 2 gives the two regulations — as adopted in the EEA Agreement, with the adaptations from EEA Joint Committee Decisions No. 288/2019 and No. 90/2020 — the force of law in Iceland (source: Act on Medical Devices No. 132/2020, Article 2). Classification, conformity assessment, clinical and performance evaluation, the EU declaration of conformity, CE marking, UDI, economic-operator obligations and vigilance therefore run on the EU texts as incorporated, read with MDCG guidance. No Icelandic authority repeats a notified body's conformity assessment, and no Icelandic record replaces CE marking. The IMA's registration page states the consequence plainly: the IMA does not maintain a registration of medical devices; devices should be registered via EUDAMED (source: IMA, Registration of distributors). An Icelandic distributor registration is an administrative, traceability and market-surveillance matter for a device that is already lawfully CE marked — it is not a second conformity decision, and it must never be described as an Icelandic "approval" of safety or performance.

The transition architecture is the EEA one: the extended MDR transitional provisions under Regulation (EU) 2023/607, the staggered IVDR transition, and the phased mandatory use of EUDAMED under Regulation (EU) 2024/1860, each as incorporated into the EEA Agreement. Because EEA incorporation can lag the EU's application dates, any date that depends on a transition deadline or on whether a given EUDAMED module is mandatory for an Iceland-established operator at the moment of filing is to be verified at the time of filing. In practice the EUDAMED file and the Icelandic file must agree: the same operator and distributor data must read identically in both.

1.2 The Icelandic layer: the Act on Medical Devices No. 132/2020 and its regulations

On top of the EEA layer, Iceland's national device law is the Act on Medical Devices No. 132/2020 of 8 December 2020. Its stated aim is to ensure the quality and safety of medical devices, with public safety as the guiding principle, and it covers production, importing, distribution, sale, placing on the market, market surveillance, maintenance and use of devices, including the Annex XVI devices without an intended medical purpose (source: Act on Medical Devices No. 132/2020, Articles 1 and 3). The Minister of Health exercises supervision under the Act, and the IMA is the competent authority in the sense of Article 101 of the MDR and Article 96 of the IVDR, charged with applying the regulations and the Act and with monitoring devices under them; the IMA also sets its own rules of procedure, for example for examining clinical investigation applications (source: Act on Medical Devices No. 132/2020, Article 5).

Three features of the national layer matter most in practice. First, the language rules are national law, not guidance: Articles 12 and 19 of the Act set the language of the information accompanying devices and of implant cards, and the Regulation on instructions for use accompanying medical devices No. 630/2022 details the rules for laypersons and professionals (source: Act on Medical Devices No. 132/2020; Regulation No. 630/2022, as cited in the European Commission's MDR language requirements table, Rev. 3). Section 6 works through the matrix. Second, the fee power is national and exercised: Article 39 of the Act is the legal basis of the IMA's device tariff, and Article 28(3) is the basis of the distributor-registration fee (source: IMA fee tariff, Articles 2 and 11). Third, the supervision net is national and fee-bearing: the Act gives the IMA supervision of the use of devices (Article 16), of their maintenance (Article 17) and of the operators in the supply chain (Article 34), and the tariff prices each of those supervision types — an Icelandic operator can be invoiced for being inspected, which is a budgeting fact most entrants discover late (source: Act on Medical Devices No. 132/2020; IMA fee tariff, Articles 3–5).

One structural point for teams coming from the medicines side: medicines in Iceland are governed by the separate Medicinal Products Act No. 100/2020, with its own marketing-authorisation tariff (currently No. 350/2026). Medicines fees and routes are not device rules and must never be budgeted as device costs; the device tariff is the separate instrument reproduced in Section 8 (source: Stjórnartíðindi, tariff No. 350/2026; IMA fee tariff, 2025).

ElementEEA levelIceland level
Legal basisMDR (EU) 2017/745; IVDR (EU) 2017/746, as incorporated into the EEA Agreement (Joint Committee Decisions No. 288/2019 and No. 90/2020)Act on Medical Devices No. 132/2020 of 8 December 2020; Regulation on instructions for use No. 630/2022
Conformity assessmentManufacturer + notified body where the class requires itNot repeated in Iceland
Device authorityEUDAMED; competent authorities networkMinistry of Health — Icelandic Medicines Agency (Lyfjastofnun, IMA), competent authority under MDR Article 101 / IVDR Article 96
Medicines (separate regime)EMA networkMedicinal Products Act No. 100/2020 and its own tariff — not device rules
Actor registrationEUDAMED actor module, Single Registration NumberManufacturers, importers and representatives register in EUDAMED; the IMA keeps the national register of responsible parties
Device registrationEUDAMED UDI / device module as modules become mandatoryNo Icelandic device register — the IMA does not register devices; devices go via EUDAMED
Distributor registrationNo EEA-level distributor register under the MDRDistributors marketing devices in Iceland register with the IMA (Act Article 28(3)); distribution may begin on submission; fee ISK 53,400
LanguageLanguage determined by the state where the device is made availableIcelandic default; English or a Nordic language other than Finnish for Class I and IIa lay information; Icelandic or English for professionals; implant card in Icelandic (Act Articles 12 and 19)
FeesNone at EEA levelPublished IMA device tariff under Act Article 39, in force from 1 January 2026 (Section 8)
PayerNone at EEA levelIceland Health (Sjúkratryggingar Íslands) — statutory insurance; assistive-device grants conditional on purchase from a contracted company
Public buyersNone at EEA levelLandspítali, the only university hospital (624 beds, ~25,000 inpatient stays, 2019); regional hospitals incl. Akureyri
ProcurementDirective 2014/24/EU principles as incorporated in the EEA AgreementAct on Public Procurement No. 120/2016; central purchasing with the Financial Management Authority (Fjársýslan) since 1 August 2024, formerly Ríkiskaup; national portal and TED above EEA thresholds

1.3 What the IMA is — and what it is not

The Icelandic Medicines Agency (Lyfjastofnun, IMA) is Iceland's device competent authority, based in Reykjavík and reached through lyfjastofnun.is and Ísland.is. Its recorded device functions are the registration of distributors, the register of parties that manufacture devices in Iceland or are responsible for their marketing, post-market surveillance and vigilance, an inspection programme covering correct maintenance and use, the assessment of derogation and clinical investigation applications, and the issue of certificates such as free sale certificates (source: IMA, About Medical Devices; IMA fee tariff, Articles 1 and 6). The IMA can require information from regulated parties, take samples and perform tests to prevent patient harm, and importers, manufacturers, sellers, owners and users are expected to provide the assistance and information requested (source: IMA, About Medical Devices).

Three consequences follow. First, the IMA is a supervisor, not a promoter: a distributor registration is not an endorsement that can be quoted in marketing. Second, the IMA is small, like the market it supervises: the tariff prices expert time by the hour (ISK 26,700 under Article 7), and an Icelandic filing is a conversation with a handful of named officials, not a ticket in a large portal queue — conduct and accuracy are diligence items. Third, the IMA works inside the EEA network, exchanging incident and faulty-device reports with the other competent authorities (source: IMA, About Medical Devices).

1.4 The market in numbers

Iceland's device market is the EEA's smallest by population and one of its richest per person: 389,444 people on 1 January 2025, health spending of USD 6,770 per person (PPP) and 9.0% of GDP, life expectancy of 82.4 years and 84% of spending covered by mandatory prepayment (sources: Statistics Iceland, 2025; OECD, Health at a Glance 2025: Iceland). Care is overwhelmingly public, and the hospital map is short: Landspítali as the only university and tertiary hospital, Akureyri Hospital in the north, and a small set of regional health institutions (source: Ministry of Health, December 2021). There is no Icelandic device-market size figure from an official source that this guide can stand behind: any market-size number quoted by a third party is to be verified against its methodology before use in a board paper. What can be said from official data is structural — a market of under 400,000 people, one anchor hospital, one insurance institution for assistive technology, and per-capita health spending above the OECD average. Iceland rewards a complete, correct, small file; it does not reward scale assumptions.

2. Device Classification

2.1 Classification is the EEA classification — Iceland adds no class of its own

Classification in Iceland is the MDR and IVDR classification, because the two regulations have the force of law under Article 2 of the Act. Medical devices are classified in Classes I, IIa, IIb and III under MDR Article 51 and Annex VIII; in vitro diagnostics in Classes A, B, C and D under IVDR Article 47 and Annex VIII. Iceland adds no national class and no national positive list: the class your notified body has certified (or, for Class I and Class A, the class you have self-declared) is the class in Iceland. Two Icelandic specifics do attach to the class. The language opening in Article 12 of the Act is class-dependent: the English/Nordic alternative for lay-user information exists only for Class I and IIa devices, so the class decides the translation budget (source: Act on Medical Devices No. 132/2020, Article 12; European Commission, MDR language requirements table, Rev. 3, Iceland row). And the IMA's derogation-assessment fee is class-dependent: ISK 200,250 for Class IIa and non-invasive Class IIb and III devices, ISK 293,700 for the Class IIb and III band including implantables, long-term invasive devices and IVDs (source: IMA fee tariff, Articles 3.3 and 3.4). Where a genuine doubt exists whether a product is a medical device at all, the IMA resolves the question (source: Act on Medical Devices No. 132/2020, Article 3; IMA, About Medical Devices).

2.2 The classification table for Iceland planning

Class (MDR)Conformity route (EEA)Iceland-specific consequence
Class I (incl. Is, Im, Ir)Self-declaration; notified body only for sterile, measuring or reusable surgical featuresLay information may be in English or a Nordic language other than Finnish instead of Icelandic (Act Article 12); devices in EUDAMED, not with the IMA
Class IIaNotified bodySame Class I/IIa language opening for lay information (Act Article 12); derogation assessment fee ISK 200,250
Class IIbNotified bodyLay information in Icelandic; derogation fee ISK 200,250 if non-invasive, ISK 293,700 if the invasive/implantable band applies
Class IIINotified body, incl. clinical evaluation consultation where applicableLay information in Icelandic; implant card in Icelandic (Act Article 19); derogation assessment fee ISK 293,700
IVD Class ASelf-declaration (sterile A: notified body)Investigation pre-screening priced in the lower band (ISK 200,250)
IVD Classes B, C, DNotified bodyPre-screening in the same ISK 200,250 band; investigation assessment ISK 1,161,450 for all classes

The practical reading: classification work done for the EEA carries into Iceland unchanged, and the Icelandic deltas are downstream of the class — language, derogation fees and investigation fees. A borderline question is one of the few classification matters that is genuinely Icelandic, because the Act assigns the decision to the IMA; ask early, in writing.

3. Authorised Representative, Importer and Distributor

3.1 The authorised representative (viðurkenndur fulltrúi)

A manufacturer established outside the EEA must appoint a single authorised representative established in the EEA under MDR Article 11 (IVDR Article 11), and that rule applies to Iceland exactly as in the EU. The representative does not have to be established in Iceland: an EEA-established representative serves Iceland as it serves every EEA state. Representatives register as actors in EUDAMED and receive a Single Registration Number; the IMA's registration pages direct manufacturers, importers and authorised representatives to EUDAMED (source: IMA, Registration of distributors; IMA, Medical Device Registration). Two Iceland specifics follow. First, a representative established in Iceland is inside the IMA's supervision net: market surveillance of manufacturers and representatives is priced at ISK 133,500 and their quality-management supervision at ISK 774,300 (source: IMA fee tariff, Articles 5.3 and 5.4). Second, the IMA register of responsible parties covers parties operating enterprises in Iceland that manufacture devices or are responsible for their marketing, so an Iceland-established representative should expect to appear in it (source: IMA, About Medical Devices). Whether the IMA requires any separate national notification beyond the EUDAMED actor record is to be verified at the time of appointment — the published pages direct representatives to EUDAMED, and this guide does not invent a second filing.

3.2 The importer

The importer is the EEA-established operator that places a device from a third country on the EEA market, with the duties of MDR Article 13 (IVDR Article 13). For Iceland the question has a geographic edge: goods arrive by sea or air freight into Reykjavík-area logistics, and the Iceland-established importer is the operator the IMA can inspect, invoice and hold to account. Importers register as actors in EUDAMED (source: IMA, Registration of distributors). The tariff prices importer supervision expressly: market surveillance at ISK 133,500, quality-management supervision at ISK 387,150 (source: IMA fee tariff, Articles 5.1 and 5.2). A manufacturer selling into Iceland through an EEA importer established elsewhere should confirm who in its chain is the importer for Icelandic placing on the market; a mismatch between the EUDAMED record and the party actually importing is the kind of small-market discrepancy a small authority notices.

3.3 The distributor — the one genuinely Icelandic registration

The distributor is where Iceland's national layer bites. Distributors who are responsible for the marketing of medical devices in Iceland must register with the IMA, and the registration is subject to the applicable fees (source: IMA, Registration of distributors). The legal basis is Article 28(3) of the Act, and the tariff defines the distributor as every natural or legal person in the supply chain, other than the manufacturer or the importer, that makes a device available on the market up to the point of putting it into service (source: IMA fee tariff, Article 2). The registration is filed through the Ísland.is service portal; where the web form cannot be used, it is sent by email to lyfjastofnun@lyfjastofnun.is with the distributor's name, ID number if available, address and postcode, the contact person and their email address, the telephone number, and the list of brand names to be marketed, with a contact person for each brand (source: Ísland.is, Registration of medical device distributors).

Three features define the step. First, timing: processing is immediate, and a distributor can begin distributing as soon as the registration is submitted — there is no Icelandic waiting period and no approval letter to chase (source: Ísland.is). Second, money: the Agency registers distributors as entities subject to surveillance and invoices the registration under its fee schedule — ISK 53,400, and ISK 26,700 for maintenance registration (source: Ísland.is; IMA fee tariff, Article 2). The periodicity of the maintenance charge is not stated on the service page and is to be verified with the IMA when the first invoice arrives. Third, scope discipline: the same page states the mirror-image rule for devices — despite maintaining a register of distributors, the IMA does not maintain a register of medical devices (source: Ísland.is). A manufacturer that asks its distributor "are our devices registered in Iceland?" is asking the wrong question twice over: the distributor is registered, the devices are in EUDAMED.

The manufacturer's diligence on an Icelandic distributor therefore has four items: evidence of the IMA registration (with the brand list covering your brands), the distributor's EUDAMED actor status where it also imports, its vigilance intake and Icelandic-language complaint handling, and its standing with Iceland Health's contracted-supplier system for assistive technology. In a market this small, the distributor is usually also the service engineer, the tender respondent and the payer's counterparty; choose accordingly.

4. Registration Architecture

Iceland's registration architecture has three layers, and keeping them separate is most of the battle. The first is EUDAMED: manufacturers, authorised representatives and importers register as actors and receive Single Registration Numbers, and devices are registered in the UDI/device module as mandatory use phases in under Regulation (EU) 2024/1860 as incorporated into the EEA Agreement (source: IMA, Registration of distributors). The second is the IMA's national registers of people: the distributor register under Article 28(3) of the Act, and the register of parties operating enterprises in Iceland which manufacture devices or are responsible for their marketing (source: IMA, About Medical Devices; Ísland.is). The third is not a register at all but a payer list: Iceland Health's contracted companies and product lists for assistive devices, which decide whether an insured patient can obtain the device with a grant (source: Ísland.is, Iceland Health news, January 2025; IBA Survey — Iceland).

What does not exist is equally important. There is no Icelandic device registration certificate, no national device number to print on a label, no renewal cycle for devices, and no device fee per SKU: the tariff's charges attach to operators, certificates and supervision events, not to device listings (source: IMA fee tariff, Articles 1–7). Custom-made devices follow the MDR Article 29 and Annex XIII route as incorporated in the EEA; whether the IMA expects any national notification from an Iceland-established custom-made manufacturer beyond the EUDAMED actor record is to be verified. Clinical investigations sit outside this architecture as their own track: before starting one, the sponsor must determine whether it is subject to application, notification or neither to the IMA (source: IMA, Medical Devices; IMA fee tariff, Article 6). The ethics-review step Icelandic health research ordinarily requires in addition is to be verified for the specific design before the sponsor commits to a site timeline.

5. Step-by-Step Process

For a foreign manufacturer entering through a local distributor, the sequence is:

  • Step 1 — Confirm the EEA file is complete. Valid conformity assessment for the class, EU declaration of conformity, CE marking, UDI assignment and technical documentation under the MDR or IVDR as incorporated in the EEA Agreement. Iceland adds no second conformity step; a gap in the EEA file cannot be cured by any Icelandic filing.
  • Step 2 — Fix the operator chain. Confirm the EEA authorised representative (for non-EEA manufacturers), identify the importer for Icelandic placing on the market, and select the distributor. Check that each party's role matches the MDR definitions — a party that imports is not a distributor for registration purposes (source: IMA fee tariff, Article 2).
  • Step 3 — Register the actors in EUDAMED. Manufacturer, authorised representative and importer actor registrations with Single Registration Numbers, and device registration in the UDI/device module as its mandatory use applies (source: IMA, Registration of distributors).
  • Step 4 — Register the distributor with the IMA. File through Ísland.is or by email to lyfjastofnun@lyfjastofnun.is with the defined data set, including the brand list with a contact person per brand. Distribution may begin on submission; keep the submission evidence (source: Ísland.is).
  • Step 5 — Pay the registration invoice. The IMA invoices the registration under the tariff — ISK 53,400, maintenance registration ISK 26,700 — with a 30-day due date; agree in the distribution agreement who bears the fee and the future supervision fees under tariff Articles 3–5 (source: IMA fee tariff, Articles 2 and 10).
  • Step 6 — Close the language file. Icelandic label and IFU where required, the Class I/IIa decision documented where the English/Nordic opening is used, implant card in Icelandic, and the declaration of conformity and field safety notice versions ready (Act Articles 12 and 19). Section 6 gives the matrix.
  • Step 7 — Wire vigilance before the first shipment. Named vigilance contact at the distributor, Icelandic-language complaint intake, and the reporting clocks of Section 9 agreed; everyone who sells, owns or professionally uses devices in Iceland must report adverse incidents to the IMA (source: IMA, About Medical Devices).
  • Step 8 — Work the payer and buyer files in parallel. For assistive technology, establish whether the device falls under the Regulation on grants for assistive devices and whether Iceland Health buys it under a framework contract, by individual bid or by price survey; for a hospital device, map Landspítali and the regional hospitals and watch the tender channels under the Act on Public Procurement No. 120/2016 (source: IBA Survey — Iceland).
  • Step 9 — If you need paper for a third market, order the free sale certificate: ISK 53,400 for five copies, ISK 26,700 per extra copy, other certificates ISK 40,050 per copy (source: IMA fee tariff, Article 1).
  • Step 10 — Diarise the maintenance points. Distributor registration updates when brands, contacts or addresses change (filed the same way), EUDAMED data maintenance, tariff invoices, and the payer-side framework calendar (source: Ísland.is).

Two deviations are common. A manufacturer with its own Icelandic entity becomes its own importer or distributor and takes the registrations — and the supervision-fee exposure — in-house. And a sponsor running a clinical investigation leaves this track for the IMA investigation track priced in Section 8 (source: IMA, Medical Devices).

6. Language and Labelling

6.1 The rule

Iceland determines its device language rule in Articles 12 and 19 of the Act, detailed by Regulation No. 630/2022, and the European Commission's MDR language requirements table (Rev. 3, August 2025) records the Iceland position as follows (source: Commission language table, Rev. 3, Iceland row, citing the Act and Regulation No. 630/2022):

DocumentLay user / patientProfessional user
Label and instructions for use (MDR Article 10(11), Annex I §23)Icelandic — but for Class I and IIa devices the information is allowed to be in English or a Nordic language other than Finnish (Act Article 12)Icelandic or English (Act Article 12)
Implant card (MDR Article 18)Icelandic (Act Article 19)Icelandic (Act Article 19)
EU declaration of conformity (MDR Article 19)Icelandic or EnglishIcelandic or English
Field safety notice (MDR Article 89(8))Icelandic or EnglishIcelandic or English
Graphic user interface (e.g. apps)Icelandic — for Class I and IIa, allowed to be in English or a Nordic language other than FinnishIcelandic or English

The IMA's own summary matches: the general rule is that instructions for use shall be in Icelandic, and in cases where patients themselves operate the device, the guidelines should be in Icelandic (source: IMA, Medical Devices; IMA, About Medical Devices). For implantables the Agency adds a practical softening the table does not show: implant cards handed out by health institutions should normally be in Icelandic, but when that is not possible they may be in English or another Nordic language except Finnish — and the same applies to the supporting documents (the patient information leaflet and the summary of safety and clinical performance) published under MDR Article 18 (source: IMA, About Medical Devices). Treat that softening as practice for cards, not as permission to plan English-first: the Act's text says Icelandic, and the card is the document the patient keeps for life.

6.2 Labelling practice in a very small language

Icelandic is spoken by under 400,000 people, which creates three practical consequences no other EEA market combines. First, translation supply is thin: medical Icelandic translation is a specialist service with slower turnaround and higher unit costs than the large EU languages, and it belongs in the launch plan's critical path, not its final week. Second, the class decision is a money decision: a Class I or IIa consumer device can lawfully reach lay users in English or in Danish, Swedish or Norwegian, while a Class IIb or III patient device cannot, so the classification file and the translation budget must be settled together (source: Act Article 12). Third, mixed-language packs are normal in Iceland because volumes are small: an Iceland-specific over-sticker or an Icelandic insert leaflet inside a Nordic pack is the standard commercial answer, provided the MDR's legibility and indelibility requirements are met and the distributor verifies the pack before release under MDR Article 14 (source: IMA fee tariff, Articles 3 and 5). One boundary is absolute: the "Nordic language" opening excludes Finnish, expressly, in both the Act-based table and the IMA's implant-card practice (source: Commission language table, Rev. 3, Iceland row; IMA, About Medical Devices). A Finnish IFU is not a Nordic-language solution in Iceland.

7. Timelines

Iceland publishes one hard timeline and otherwise runs on EEA clocks. The hard one is the distributor registration: processing is immediate and distribution can begin on submission, so the step adds days of preparation — assembling the brand list and contact data — and no waiting period (source: Ísland.is). The IMA invoices the registration fee after submission, with a 30-day payment term (source: IMA fee tariff, Article 10).

Everything else is a range to be managed rather than a date to be quoted. EUDAMED actor registration follows the EUDAMED process with the competent authority's validation step, and no Iceland-specific service standard was located in the sources verified for this guide — any promised number of days is to be verified with the IMA (source: IMA, Registration of distributors). Free sale certificates are issued on request against the tariff fees; no published issuance clock was located, to be verified at ordering (source: IMA fee tariff, Article 1). Derogation and investigation assessments carry the Section 8 fees but no published Icelandic decision clock; the MDR's own periods govern the investigation track as incorporated in the EEA Agreement, and the concrete handling time is to be verified for the specific file (source: IMA fee tariff, Articles 3 and 6). On the commercial side, the true timeline is set by two calendars that are not the authority's: Icelandic translation turnaround, and Iceland Health's framework-tender cycle — the current wheelchair and walking-aid contracts run to 30 November 2026, and missing a framework window can cost a full contract period (source: Ísland.is, January 2025). A realistic launch plan therefore sequences: EEA file complete, distributor registered (immediate), language file closed (the long pole), payer route identified — and quotes a date only for the steps Iceland actually clocks.

8. Fees and Costs

8.1 Official fees: a published tariff, in krónur

Iceland is one of the few small EEA markets that publishes its device fees in full. The tariff for the IMA's supervision of medical devices (gjaldskrá vegna eftirlits Lyfjastofnunar með lækningatækjum) is set by the Minister of Health on the Agency's proposals under Article 39 of the Act, entered into force on 1 January 2026, and replaced tariff No. 1555/2023 (source: IMA fee tariff, Article 11; Stjórnartíðindi B-deild, 30 December 2025). The device-relevant items are:

Item (tariff article)Fee (ISK)
Free sale certificate (FSC) — five copies (1.1)53,400
FSC / IVD free sale certificate — each additional copy beyond five (1.2)26,700
Other certificates — per copy (1.3)40,050
Distributor registration fee (2.1)53,400
Distributor maintenance registration (2.2)26,700
Supervision of use — device supervision (3.1)160,200
Supervision of use — management-system supervision (3.2)387,150
Derogation assessment — Class IIa, and non-invasive Class IIb and III (3.3)200,250
Derogation assessment — Class IIb, III incl. implantable, long-term invasive and IVD (3.4)293,700
Other derogations (3.5)200,250
Supervision of maintenance — device (4.1)200,250
Supervision of maintenance — management system (4.2)387,150
Market surveillance — distributors and importers (5.1)133,500
Quality management supervision — distributors and importers (5.2)387,150
Market surveillance — manufacturers and authorised representatives (5.3)133,500
Quality management supervision — manufacturers and authorised representatives (5.4)774,300
Clinical/performance investigation — pre-screening, lower band: Class I, non-invasive IIa/IIb, IVD A (6.1)200,250
Clinical/performance investigation — pre-screening, higher band: IIa, IIb, III incl. implantable/long-term invasive, IVD B, C, D (6.2)200,250
Clinical investigation application assessment — all classes (6.3)1,161,450
Clinical investigation approved by the IMA (6.4)387,150
Amendment to a clinical investigation (6.5)106,800
Expert hourly rate for supervision and services under the Act (7)26,700 per hour

Read the tariff the way the IMA will apply it. Registration is cheap and certain: ISK 53,400 once, ISK 26,700 for maintenance registration, invoiced after an immediate filing. Supervision is the variable cost: where an inspection turns out to be extensive, the IMA may charge the hourly rate on top of the fixed fee, travel and subsistence costs are charged on top, and an invoice is issued when the inspection report is available (source: IMA fee tariff, Articles 3, 7 and 8). Clinical investigations are the expensive track by design: pre-screening ISK 200,250, assessment ISK 1,161,450 for all classes, ISK 387,150 for an approved investigation and ISK 106,800 per amendment, with hourly top-up if the assessment costs more than the fee covers — the applicant must be told of the additional cost and given 14 days to withdraw instead of paying it; investigation fees are non-refundable if the application is refused or withdrawn, though the IMA may waive or reduce the assessment fee in exceptional cases (source: IMA fee tariff, Article 6). All tariff invoices fall due 30 days from issue, default interest applies to late payment, the fees are enforceable, and enforcement follow-up work is itself charged at the hourly rate (source: IMA fee tariff, Articles 9 and 10).

8.2 The real cost stack

The official fees are the small part of an Iceland budget. The full stack for a typical distributor-led entry is: the registration and maintenance fees above; Icelandic translation of the label, IFU and patient documents (Section 6) at small-language rates; the EUDAMED and notified-body costs already carried in the EEA file; freight to an island market; the distributor's costs of tendering into Iceland Health frameworks and hospital tenders; and a contingency for a fee-bearing supervision event under tariff Articles 3–5. What must not appear in the stack: any per-device Icelandic registration fee (none exists), any medicines tariff figure from the Medicinal Products Act No. 100/2020 regime (a different law and tariff), and any "Iceland approval fee" quoted by an intermediary — the IMA does not sell approvals, it registers distributors and charges for supervision (source: IMA fee tariff, Articles 1–11; Ísland.is). The periodicity of the ISK 26,700 maintenance registration — annual or event-driven — is to be verified with the IMA; budget it as recurring until the Agency confirms otherwise.

9. Vigilance and Post-Market Surveillance

Iceland runs the MDR vigilance system with a national reporting duty broader than the manufacturer's. The IMA states that manufacturers or companies responsible for devices in Iceland must carry out post-market surveillance and report adverse events to the Agency — and that everyone who sells, owns or professionally uses devices must report adverse incidents (source: IMA, About Medical Devices). The reportable events are the MDR set: incidents that caused death or serious harm, and incidents that might cause death or serious harm (source: IMA, About Medical Devices).

For the manufacturer, the IMA publishes the reporting clocks: a serious public-health risk is reported immediately or within 2 days because of special circumstances; death or serious harm, immediately or within 10 days; otherwise, immediately or within 30 days — and a manufacturer that delays must be able to rationalise the delay (source: IMA, About Medical Devices). These mirror the MDR Article 87 periods; as EUDAMED's vigilance module becomes mandatory, the filing channel changes but the clocks do not, and the channel in force at the moment of an incident is to be verified against current IMA guidance. Field safety notices run through the same system, in Icelandic or English (source: Commission language table, Rev. 3, Iceland row), and the distributor's registration data — a named contact per brand — is exactly the data the IMA will use to reach the market fast, one more reason the brand list must be current (source: Ísland.is).

Post-market surveillance beyond incidents is the MDR system — PMS plan, PMS report for Class I, periodic safety update report for higher classes, trend reporting — read with the IMA's inspection programme, under which the Agency monitors that devices are maintained and used correctly (source: IMA, About Medical Devices). The supervision of use and of maintenance is fee-bearing under tariff Articles 3 and 4, so a sloppy installed-base record is not only a compliance risk; it is an invoiced event waiting to happen. Finally, the IMA cooperates with the other EEA competent authorities and exchanges incident reports with them, so an Icelandic incident is visible to the network and a network incident lands in Iceland (source: IMA, About Medical Devices). Plan Icelandic vigilance as EEA vigilance with an Icelandic front door: local intake in Icelandic, clocks as published, one named owner.

10. Procurement and Reimbursement

10.1 Iceland Health: how assistive devices get paid for

Iceland's statutory health insurance is administered by Iceland Health (Sjúkratryggingar Íslands) under the Ministry of Health, and covers all individuals who have legally resided in Iceland for six months; the institution negotiates and pays for healthcare services and covers part or all of the cost for the insured (source: IBA Survey — Iceland). For devices, the operative instrument is the Regulation on grants for assistive devices, which defines which devices Iceland Health contributes to and the extent and conditions of coverage (source: IBA Survey — Iceland). The decisive commercial rule is the contracting condition: a grant is conditional on the device being purchased from a specific contracted company; where Iceland Health has concluded no contracts, bids are sought for individual devices or price surveys are conducted, and grants are awarded based on that price (source: IBA Survey — Iceland).

In practice this is a framework-agreement market. Iceland Health tenders framework contracts for assistive-device categories and publishes the resulting product list: the current framework for manual and electric wheelchairs and walking aids took effect on 1 December 2024 and runs to 30 November 2026, the number of contracted companies rose from four to seven, and wheelchairs were consolidated into one contract (source: Ísland.is, Iceland Health news, 13 January 2025). Suppliers invoice Iceland Health electronically through its aid-accounts web service (source: Ísland.is, Iceland Health). For a manufacturer of assistive technology, the sequence is: confirm the category sits inside the assistive-devices regulation, identify whether a framework contract exists and when it re-opens, and be inside the contracted-supplier set — through your own Icelandic company or a contracted distributor — before marketing spend begins. A device outside the contracted set can still be sold privately, but in a country of 389,444 people with near-universal public coverage, the private residue is small (source: Statistics Iceland, 2025; OECD, 2025). Whether a given category is inside the regulation, and at what grant percentage, is to be verified against the current text and Iceland Health's product lists before a revenue line is booked.

Hospital devices follow the other money channel: they are paid from the budgets of the public hospitals that use them — the sale is the reimbursement. That makes Section 10.2 the whole game for hospital product lines.

10.2 Hospitals and public procurement

Iceland's hospital map is the shortest in this guide series. Landspítali — the National University Hospital of Iceland in Reykjavík — is the main hospital for the whole country and its only tertiary and university hospital: around 25,000 inpatient stays and 407,000 outpatient visits per year, 624 beds, 21 operating theatres and around 4,500 full-time-equivalent employees recorded for 2019, with an annual budget of around ISK 78 billion (source: Ministry of Health, December 2021). Around 70% of Icelandic children are born there (source: Landspítali). The second hospital is Akureyri Hospital (Sjúkrahúsið á Akureyri) in the north; the rest of the map is a small set of regional health institutions and health centres. There are no private hospitals of consequence: hospital care is a public system, and access to it is a public-procurement and clinical-relationship exercise, not a private sales exercise.

Public purchases run under the Act on Public Procurement No. 120/2016: procurements above the national thresholds are tendered nationally, and those above the EEA thresholds must be tendered throughout the EEA, which means publication in the EU's Tenders Electronic Daily (TED) as well as the national channels (source: Act No. 120/2016, Articles 2 and 23). The central purchasing function sits with the Financial Management Authority (Fjársýslan, FMA), which on 1 August 2024 assumed all responsibilities of Ríkiskaup, the former Central Public Procurement office, which was officially decommissioned (source: Ísland.is, Financial Management Authority news, 2024). Any market plan that still names Ríkiskaup as the live counterparty is working from pre-August 2024 sources — a small detail that dates a proposal instantly in Reykjavík. Before buying, public bodies check whether the product is covered by a central contract; if it is, purchase follows the contract's terms, by direct purchase or mini-competition (source: Ísland.is, Buying for the first time).

For a device manufacturer the practical shape of hospital access is: one anchor account (Landspítali) whose departments set the national standard for tertiary care, one northern account (Akureyri), published and therefore monitorable tender windows, and framework or central contracts where they exist. Evaluation and first use at Landspítali carry weight out of proportion to the unit count, because the same small specialist community staffs the national service; conversely, a lost framework or a failed first evaluation has nowhere else to hide. Volumes will be small — price the Icelandic hospital file as a reference market, not as a volume market.

11. Common Mistakes

  • Assuming a CE mark finishes Iceland. The EEA file is necessary, but the Icelandic layer — distributor registration, language, payer contracting — is what makes the device sellable and paid.
  • Looking for an Icelandic device register. There is none: the IMA does not maintain one; devices are registered via EUDAMED, and what Iceland registers is the distributor (source: IMA, Registration of distributors).
  • Letting the distributor launch unregistered. Distributors responsible for marketing devices in Iceland must register with the IMA under Article 28(3); the step is immediate, but skipping it leaves the brand outside the register of entities subject to surveillance (source: Ísland.is).
  • Budgeting a per-device Icelandic fee. No such fee exists; the tariff charges operators, certificates, supervision events and investigations — ISK 53,400 for distributor registration, ISK 26,700 for maintenance registration (source: IMA fee tariff, Articles 1–7).
  • Quoting medicines fees as device fees. The Medicinal Products Act No. 100/2020 tariff (currently No. 350/2026) prices medicines authorisations; devices have their own tariff under Article 39 of the Medical Devices Act. Mixing the two destroys a board paper's credibility with anyone who checks.
  • Shipping the English IFU for a patient-operated Class IIb device. The English/Nordic opening covers Class I and IIa lay information only; above that, Icelandic is required (source: Act Article 12; IMA, About Medical Devices).
  • Offering a Finnish IFU as the "Nordic" version. The opening expressly excludes Finnish, for labels, for interfaces and in the IMA's implant-card practice (source: Commission language table, Rev. 3, Iceland row).
  • Forgetting the implant card language. The card must be in Icelandic under Article 19; the IMA's English/Nordic fallback applies where Icelandic is not possible, not as a default (source: IMA).
  • Pitching Ríkiskaup. The central purchasing office was decommissioned on 1 August 2024 and its tasks sit with the Financial Management Authority (Fjársýslan); an outdated counterparty name marks the proposal as recycled (source: Ísland.is, 2024).
  • Selling assistive devices outside the contract system. Iceland Health grants are conditional on purchase from a contracted company; a non-contracted route leaves insured patients without the grant and the manufacturer outside the product list (source: IBA Survey — Iceland).
  • Treating Landspítali as one account among many. It is the country's only university and tertiary hospital; its evaluation outcome effectively sets the national position for tertiary categories (source: Ministry of Health, 2021).
  • Leaving vigilance to the distributor by default. The reporting duty reaches everyone who sells, owns or professionally uses devices, and the manufacturer's clocks are 2, 10 and 30 days by severity; an unwired local intake misses the first clock before the manufacturer hears of the event (source: IMA, About Medical Devices).

12. Frequently Asked Questions

Is Iceland in the EU for medical device purposes? Iceland is an EEA member, not an EU member. The MDR and IVDR have the force of law in Iceland through the Act on Medical Devices No. 132/2020, as adopted in the EEA Agreement by Joint Committee Decisions No. 288/2019 and No. 90/2020, so a CE-marked device lawfully placed on the EEA market is recognised in Iceland, with no second assessment (source: Act, Article 2).

Who is the Icelandic competent authority for medical devices? The Icelandic Medicines Agency (Lyfjastofnun, IMA), under the Ministry of Health — the competent authority in the sense of Article 101 of the MDR and Article 96 of the IVDR (source: Act, Article 5).

Do I register my device with the IMA? No. The IMA does not maintain a registration of medical devices; devices are registered via EUDAMED. What Iceland registers nationally is operators — above all distributors (source: IMA, Registration of distributors; Ísland.is).

What must my Icelandic distributor do before selling? Register with the IMA under Article 28(3) of the Act, via Ísland.is or by email to lyfjastofnun@lyfjastofnun.is, providing its name, ID number, address, contact person and email, telephone number and the list of brand names it will market, with a contact person per brand. Processing is immediate and distribution may begin on submission; the IMA invoices ISK 53,400 under its tariff (source: Ísland.is; IMA fee tariff, Article 2).

What does the IMA charge, and where is it published? In the tariff for the supervision of medical devices, set under Article 39 of the Act and in force from 1 January 2026: distributor registration ISK 53,400, maintenance registration ISK 26,700; free sale certificate ISK 53,400 for five copies; operator supervision from ISK 133,500 to ISK 774,300; investigation assessment ISK 1,161,450; expert time ISK 26,700 per hour. Invoices fall due in 30 days (source: IMA fee tariff, Articles 1–11).

Does my authorised representative have to be in Iceland? No. The representative must be established in the EEA; an EEA-established representative covers Iceland. Manufacturers, importers and authorised representatives register as actors in EUDAMED (source: IMA, Registration of distributors). An Iceland-established representative falls inside the IMA's fee-bearing supervision (source: IMA fee tariff, Article 5).

Which language must the label and IFU be in? Icelandic by default. For Class I and IIa devices, lay-user information may instead be in English or a Nordic language other than Finnish; for professional users, Icelandic or English; the implant card must be in Icelandic, with an IMA-practice fallback to English or another Nordic language except Finnish where Icelandic is not possible (source: Act Articles 12 and 19; Regulation No. 630/2022; Commission language table, Rev. 3; IMA, About Medical Devices).

How are incidents reported in Iceland, and how fast? To the IMA, by the manufacturer and by everyone who sells, owns or professionally uses devices. The manufacturer's clocks: serious public-health risk immediately or within 2 days; death or serious harm within 10 days; otherwise within 30 days (source: IMA, About Medical Devices).

Can I get a derogation to supply a non-CE-marked device? The IMA assesses applications to use devices that do not meet the requirements in the interest of public health or patient safety; the tariff prices the assessment at ISK 200,250 for Class IIa and non-invasive Class IIb/III, and ISK 293,700 for the band including implantables and long-term invasive devices (source: IMA fee tariff, Articles 3.3–3.5). Derogations are exceptional and case-by-case; they are not a launch strategy.

How do I run a clinical investigation in Iceland? The sponsor must first determine whether the investigation is subject to application, notification or neither to the IMA. The tariff prices pre-screening at ISK 200,250, assessment at ISK 1,161,450 for all classes, ISK 387,150 for an approved investigation and ISK 106,800 per amendment (source: IMA, Medical Devices; IMA fee tariff, Article 6). The additional ethics-review step for the specific study is to be verified before committing site timelines.

How are assistive devices paid for? Through Iceland Health grants under the Regulation on grants for assistive devices — conditional on purchase from a contracted company, typically under a tendered framework agreement with a published product list; the wheelchair and walking-aid framework runs to 30 November 2026 with seven companies (source: IBA Survey — Iceland; Ísland.is, January 2025).

Who buys hospital devices in Iceland? Public hospitals, above all Landspítali — the only university hospital, with 624 beds and around 25,000 inpatient stays a year — and Akureyri Hospital, buying under the Act on Public Procurement No. 120/2016, with the central purchasing function at the Financial Management Authority (Fjársýslan) since 1 August 2024 (source: Ministry of Health, 2021; Ísland.is, 2024).

How big is the Icelandic device market? 389,444 people on 1 January 2025, health spending of USD 6,770 per person (PPP) and 9.0% of GDP (sources: Statistics Iceland; OECD, 2025). No official device-market size figure was located; any third-party number is to be verified before use.

Sources & scope

  • Act on Medical Devices No. 132/2020 (Government of Iceland) — force of law of the MDR and IVDR as adopted in the EEA Agreement, IMA as competent authority, distributor registration (Article 28(3)), language rules (Articles 12 and 19), fee basis (Article 39); IMA fee tariff for the supervision of medical devices, Stjórnartíðindi, in force 1 January 2026.
  • Icelandic Medicines Agency (ima.is), About Medical Devices and Registration of distributors; Ísland.is, Registration of medical device distributors — no IMA device register, devices via EUDAMED, immediate distributor registration, vigilance duties and reporting clocks; European Commission, language requirements for manufacturers under the MDR (Rev. 3, August 2025) — Iceland row citing the Act and Regulation No. 630/2022.
  • Statistics Iceland, population on 1 January 2025; OECD, Health at a Glance 2025: Iceland; Ministry of Health, The future development of Landspítali's services (2021); IBA, Healthcare Financing and Reimbursement Survey — Iceland, on the Regulation on grants for assistive devices and the contracted-company condition; Act on Public Procurement No. 120/2016; Ísland.is, Financial Management Authority news (Fjársýslan assumed the Ríkiskaup functions on 1 August 2024).

This guide is market-access intelligence, not regulatory or legal advice. Iceland authority practice, Iceland Health contract criteria and procurement details change — professional review is required before business use.

Market Access Strategy Session

Iceland on your roadmap? Start with the full picture.

A Market Access Strategy Session maps your Iceland route — EUDAMED actor and device registration, the IMA distributor registration and tariff exposure, Icelandic-first labelling, Iceland Health framework contracting for assistive devices, and hospital access through Landspítali and public tenders under the Act on Public Procurement No. 120/2016.