India CDSCO Medical Device Registration2026 Field Guide

The MD-14 to MD-15 import licence, the mandatory Indian Authorised Agent, the predicate fork, and the five-year retention rule — for a 1.4-billion-patient market.

MD-14 timeline
6–9 mo
Perpetual license
MD-15
Fees per site
$1K–3K

Sourced from CDSCO and the Medical Devices Rules, 2017 — checked November 2026.

Download the guide

Enter your work email and we'll send the PDF straight to your inbox.

Professional review is required before business use. See our privacy policy.

What's inside

Thirteen pages, zero fluff

Every figure in this guide traces to a named source — CDSCO, the Medical Devices Rules 2017, and the sources named in the guide.

The MD-14 → MD-15 route

MD-14 filed on SUGAM leads to the MD-15 license — perpetual validity, 6–9 month typical timelines.

The Indian Authorized Agent

Foreign manufacturers cannot file directly — the agent files, answers queries, and handles vigilance.

Fees, retention and vigilance

Second Schedule fees by class, the no-predicate route (MD-26/MD-27), 15/30-day materiovigilance, and the mistakes that reset applications.

Complete guide

The guide in full

Every section of the PDF on one page. Use the form above to keep a copy.

Overview

Last updated: November 2026. India regulates medical devices — including in vitro diagnostics (IVDs) — nationally under the Drugs and Cosmetics Act, 1940 and the Medical Devices Rules, 2017 (MDR 2017). The national regulatory authority is the Central Drugs Standard Control Organisation (CDSCO), headed by the Drugs Controller General of India (DCGI) under the Ministry of Health and Family Welfare (MoHFW). India is not part of the EU MDR system: a CE mark is supporting evidence, never an Indian authorisation. Since notification S.O. 648(E) of 11 February 2020, all medical devices are regulated under MDR 2017, with mandatory licensing phased in for Class A and B devices from October 2022 and for Class C and D devices from October 2023. Always confirm the current classification list, fee schedule and portal workflow on CDSCO's own sites — cdsco.gov.in, the SUGAM portal and the CDSCO Medical Devices Online portal (cdscomdonline.gov.in) — and, for higher-risk and single-window filings, the National Single Window System (NSWS) before filing.

India is Asia's fourth-largest medical device market after Japan, China and South Korea, and one of the world's top 20. The India Brand Equity Foundation (IBEF), citing Department of Pharmaceuticals data, valued the industry at about Rs. 1,02,660 crore (about US$12 billion) in FY24. Government and industry sources checked for this guide put imports at roughly 70–80% of domestic demand — about US$8.6–8.8 billion in FY2024-25, concentrated in imaging, electro-medical equipment, IVDs, implants and high-end consumables — against exports of about US$4.0–4.1 billion in the same year. That import dependence is precisely why the import-licence gate matters: for a foreign manufacturer, India is not a distributor-led free sale market. You cannot legally import, stock or sell a regulated device without a CDSCO licence held by an Indian entity, and the licence is tied to a specific overseas manufacturing site and specific devices.

Regulatory reality check up front — seven features of the Indian system that foreign manufacturers most often underestimate:

  1. 1. There is no single "CDSCO registration". Indian law issues different instruments by route and risk: a system-generated registration number for Class A non-sterile, non-measuring devices; a manufacturing licence in Form MD-5 (Class A sterile/measuring and Class B) or MD-9 (Class C and D); an import licence in Form MD-15 for all imported classes; and a sale/distribution registration certificate in Form MD-42. Quoting "we are CDSCO registered" without naming the form is meaningless in due diligence.
  2. 2. A foreign manufacturer cannot apply directly. Under Rules 3(f) and 34 of MDR 2017, the Form MD-14 import application must be filed by an Indian Authorised Agent (IAA) — an Indian legal entity holding a manufacturing licence, a wholesale licence, or a Form MD-42 certificate — acting under an apostilled or embassy-authenticated Power of Attorney. The agent holds the licence and carries post-approval duties in India.
  3. 3. Classification is decided by CDSCO, not by you. Under Rule 4, only the Central Licensing Authority (CLA) classifies devices, applying Part I of the First Schedule and publishing dynamic classification lists. Your EU or FDA class is a starting proposal. Since 27 November 2025 CDSCO has operated a Risk Classification Module on the SUGAM portal for novel or innovative intended uses not already listed — to be verified: current module workflow at filing.
  4. 4. The predicate question changes everything. If a substantially equivalent predicate device is already licensed in India, the pathway is a standard MD-14/MD-15 review. If there is no predicate, you fall under Rule 63: application in Form MD-26, permission in Form MD-27, Subject Expert Committee review, and potentially a clinical investigation in India (Form MD-22 → permission MD-23). Budget and timeline diverge sharply at this fork.
  5. 5. Licences are perpetual — retention is not. An MD-15 (and manufacturing licences) remain valid indefinitely provided the retention fee is paid every five years under Rule 37. Late payment attracts 2% per month for up to 90 days; after that the licence is deemed cancelled. "Perpetual" has ended more Indian launches through a missed diary date than any technical rejection.
  6. 6. The portal is the process. Applications are filed online — SUGAM, the CDSCO MD Online portal and, for integrated filings, NSWS. Paper-era forms (old Form 40 / Form 10 references still circulating on the web) are obsolete for devices. A filing that is not correctly built in the portal does not exist.
  7. 7. Price and procurement are regulated second gates. Registration makes a device legal; it does not set its sellable price or win its demand. Certain notified devices (coronary stents and knee implants are the established examples) are under National Pharmaceutical Pricing Authority (NPPA) ceiling prices under the Drugs (Prices Control) Order, and public buyers purchase through tenders and the Government e-Marketplace (GeM). to be verified: whether your exact device category is price-controlled at filing — the list changes by notification.

A note on old information: pre-2020 sources describe only a short list of "notified" devices, a three-year registration certificate and Form 40 filing. All three are out of date for a 2026 filing. The current system covers all devices, uses MDR 2017 forms, and grants perpetual licences with five-year retention.

1. The Regulatory Framework

1.1 Principal instruments

InstrumentWhat it does
Drugs and Cosmetics Act, 1940Parent statute under which MDR 2017 is made; offences and penalties for unlicensed manufacture, import or sale sit here.
Medical Devices Rules, 2017 (effective 1 January 2018)The operative device rules: classification (Chapter II, Rule 4, First Schedule), manufacturing licensing (Chapter V), import licensing (Chapter VI, Rules 34–39), clinical investigation (Chapter VII), labelling (Chapter VI / Rule 46 onwards), testing laboratories, Notified Bodies and post-approval changes. Amended repeatedly since 2020 to bring all devices into scope.
Notification S.O. 648(E), 11 February 2020Extended the device definition so that *all* medical devices fall under MDR 2017, ending the notified-devices-only era.
Second Schedule to MDR 2017Government fee schedule: per-site and per-device fees for manufacturing and import (Section 7.2).
Fifth Schedule / QMS requirementsQuality management system requirements aligned to ISO 13485; Plant Master File and Device Master File are the core dossier expressions of this system.
National Medical Devices Policy, 2023Industrial policy (Department of Pharmaceuticals): domestic manufacturing, PLI incentives, medical device parks, export promotion. Policy, not a licensing instrument — it shapes incentives, not approvals.
Production Linked Incentive (PLI) Scheme for Medical DevicesRs. 3,420 crore outlay, FY23–FY27 in the sources checked: 5% incentive on incremental sales for defined categories (cancer care / radiotherapy, radiology and imaging, anaesthetics and cardio-respiratory including renal care, and implants). Relevant if you localise manufacturing; irrelevant to an import licence.
Legal Metrology Act, 2009 and Packaged Commodities RulesConsumer-pack labelling layer that sits alongside MDR labelling for imported packaged devices: importer name and address, maximum retail price (MRP) inclusive of taxes, country of origin, customer-care details.
Drugs (Prices Control) Order (DPCO) / NPPA notificationsCeiling prices for scheduled / notified devices. Applies to a defined list only — check, do not assume.

India also operates the Materiovigilance Programme of India (MvPI), launched 6 July 2015 and coordinated by the Indian Pharmacopoeia Commission (IPC) as National Coordination Centre, and a network of registered Notified Bodies (accredited through the National Accreditation Board for Certification Bodies) that audit Class A and B manufacturing sites, while CDSCO Medical Device Officers and authorised auditors handle Class C and D scrutiny.

1.2 Who does what

  • Central Drugs Standard Control Organisation (CDSCO) / DCGI — India's National Regulatory Authority. The DCGI is the Central Licensing Authority (CLA) for: all import licences (MD-15), Class C and D manufacturing licences (MD-9/MD-10), test licences, clinical investigation permissions, no-predicate permissions, and classification decisions.
  • State Licensing Authorities (SLA) — state drug controllers / State FDA organisations. The SLA grants Class A (sterile/measuring) and Class B manufacturing licences (MD-5/MD-6) and all sale, stock and distribution licences / certificates (MD-42 via Form MD-41). A national product still needs state-level distribution coverage where it is stocked and sold.
  • Indian Authorised Agent (IAA) — the Indian applicant and licence holder for a foreign manufacturer (Section 3).
  • Notified Bodies — CDSCO-registered bodies (application MD-1, certificate MD-2) that audit manufacturing sites, principally for Class A and B, and support QMS assessment.
  • Medical Device Testing Laboratories (MDTL) — registered laboratories for device testing and evaluation where CDSCO requires performance testing.
  • Indian Pharmacopoeia Commission (IPC) / MvPI — receives and analyses medical device adverse events through Medical Device Adverse Event Monitoring Centres (775 centres as of September 2026 in the updated hospital guidance checked) and forwards recommendations and safety alerts to CDSCO.
  • Government e-Marketplace (GeM) and tender authorities — buyers, not regulators (Section 8).

1.3 The form matrix — learn this table before anything else

RouteApplicationLicence / outputAuthority
Class A non-sterile, non-measuring — manufacture or importOnline registration on CDSCO MD portal (Chapter IIIB self-certification framework, G.S.R. 777(E) in the sources checked)System-generated registration number — no licenceCDSCO portal
Manufacture Class A (sterile/measuring) and Class BMD-3 (loan licence MD-4)MD-5 (loan MD-6)State Licensing Authority
Manufacture Class C and DMD-7 (loan licence MD-8)MD-9 (loan MD-10)Central Licensing Authority
Import, all classesMD-14MD-15Central Licensing Authority
Sale, stock, exhibit or distributionMD-41MD-42 registration certificateState Licensing Authority
Test licence — manufacture for test / evaluation / clinical investigationMD-12MD-13CLA / SLA by class
Test licence — import for test / evaluation / clinical investigationMD-16MD-17Central Licensing Authority
Clinical investigation of investigational deviceMD-22MD-23 permissionCentral Licensing Authority
Clinical performance evaluation of new IVDMD-24MD-25 permissionCentral Licensing Authority
Device with no predicate — permission to import / manufactureMD-26MD-27 permissionCentral Licensing Authority

2. Device Classification

India classifies devices into four risk classes under Rule 4 of MDR 2017, aligned to GHTF/IMDRF principles:

ClassRiskExamples (from CDSCO-aligned sources checked)
ALowSurgical dressings, examination gloves (non-sterile), tongue depressors, surgical masks, nasopharyngeal catheters, general culture media and manual staining solutions (IVD)
BLow–moderateHypodermic needles, IV cannulae, suction units, blood pressure monitors, oxygen concentrators, insulin pen needles; blood glucose monitors and pregnancy test kits (IVD)
CModerate–highHaemodialysis catheters, ventilators, bone screws and orthopaedic implants, breast implants, surgical lasers, X-ray equipment; hepatitis B/C kits and PSA assays (IVD)
DHighCoronary stents, pacemakers and implantable defibrillators, heart valves, deep-brain stimulation devices; HIV diagnostic kits and blood-grouping reagents (IVD)

Practical classification notes:

  1. 1. The CLA decides. Classification is done by the Central Licensing Authority under Rules 3 and 4(3) against Part I of the First Schedule and published in CDSCO's dynamic classification lists. If your device is not listed, submit device details, intended use, technical specifications, labelling/IFU and the regulatory status of the same or similar product in other countries and request a classification — since November 2025 through the SUGAM Risk Classification Module for novel uses.
  2. 2. Sterile or measuring changes a Class A device's route. A Class A device that is sterile or has a measuring function needs a licence (MD-5 for domestic manufacture; MD-15 for import). Only Class A devices that are both non-sterile and non-measuring take the registration-number-only route, with post-licence / post-registration audit exposure.
  3. 3. Software is a device. Standalone medical device software (SaMD / MDSW), including screening and diagnostic software, is classified and licensed under MDR 2017 like other devices. CDSCO released draft guidance on software as a medical device in late 2025 / October 2025 in the sources checked — to be verified: final SaMD guidance status and any AI-specific filing expectations at filing.
  4. 4. IVDs are in the same A–D ladder but carry their own per-device import fees (Section 7.2) and, for new IVDs, a clinical performance evaluation pathway (MD-24 → MD-25) rather than a device clinical investigation.
  5. 5. Misclassification is not a shortcut. Filing a Class C implant as Class B surfaces at scrutiny or audit and resets the application. Document the classification rationale, the rule applied and the predicate comparison in the Device Master File up front.

3. The Indian Authorised Agent, Importer and Distributor

A foreign manufacturer cannot file, import or distribute in India on its own. Three local functions must be in place:

  1. 1. Indian Authorised Agent (IAA) — mandatory for foreign manufacturers. Under Rule 3(f), the authorised agent is the Indian entity appointed by the overseas manufacturer through a Power of Attorney. Under Rule 34, the agent must hold a manufacturing licence, a wholesale licence (commonly Forms 20B/21B under the Drugs Rules in agent materials), or a registration certificate in Form MD-42 for sale and distribution. The agent must also be an India-registered business with an Importer Exporter Code (IEC) from the Directorate General of Foreign Trade and an account on CDSCO's online licensing portal. The Power of Attorney must be apostilled if executed in a Hague Convention country, or authenticated by the Indian Embassy / consulate elsewhere.
  2. 2. The agent holds the MD-15 — and the risk that comes with it. The MD-15 is issued to the Indian agent for the named overseas manufacturer, site and devices. The licence cannot simply be "transferred" to a new distributor: changing agent in practice means a fresh licence strategy. Choose the agent as a long-term regulatory holder, not as your first sales distributor by default. The commercially cleanest structures in India are (a) your own Indian subsidiary as agent, or (b) an independent regulatory agent that is separate from distribution, so a distributor change does not orphan the licence. If your distributor is the agent, agree licence continuity, document handover and exit terms in writing before filing.
  3. 3. Importer and distribution are licensed separately. Holding an MD-15 authorises import of the listed devices; stocking, exhibiting and distributing for sale requires the MD-42 certificate (Form MD-41, SLA) for the entity and premises doing the selling, plus the wholesale licence position the agent/importer relies on. Verify each licence for the exact entity and premises — "we have CDSCO" without the form numbers is not verification.
  4. 4. Agent responsibilities after approval. The IAA files and maintains the MD-14/MD-15, answers CDSCO queries, handles post-approval changes (Rule 38: major changes need prior CLA approval, decided within 60 days in the sources checked; other changes follow the notification / fresh-application windows in Rules 38–39 — to be verified: current change categorisation for your change at filing), pays the five-year retention fee, coordinates recalls and Field Safety Corrective Actions, and is CDSCO's and customs' first point of contact on safety and compliance.
  5. 5. One company can wear several hats — if licensed for each. A single Indian company may act as IAA, importer and distributor, but it must actually hold each licence and be named in each capacity. Diligence means copies of the MD-15, the MD-42 and the wholesale/manufacturing licence, checked against the portal record.

4. Registration Routes: Standard, Predicate and No-Predicate

4.1 Class A non-sterile, non-measuring — registration only

Register on the CDSCO Medical Devices Online portal, self-certify compliance with MDR 2017 and the Essential Principles, and obtain the system-generated registration number. There is no MD-15 and no per-device government fee in the summaries checked for this sub-route. This is the only genuinely light route in India — and it applies to a narrow slice of Class A. Sterile or measuring Class A devices leave this route entirely.

4.2 All other imports — Form MD-14 to Form MD-15

The standard foreign-manufacturer route: the IAA files Form MD-14 on the SUGAM / MD Online system with the Plant Master File (PMF), Device Master File (DMF), ISO 13485 certificate, Free Sale Certificate and the full technical file (Section 10), pays the Second Schedule fees in US dollars per overseas site and per distinct device, answers CDSCO scrutiny queries, and — for higher-risk devices — supports any site inspection or testing CDSCO directs. CDSCO grants the import licence in Form MD-15.

Rule 36 gives CDSCO up to nine months from application to grant or reject an MD-15; CDSCO's Citizen Charter (2025) target cited in the sources checked is 270 working days. Well-prepared notified/predicate files are commonly reported at 6–9 months; Class C and D imports at 9–12 months. Treat any promise of a few weeks for a Class B–D import as a red flag.

4.3 No predicate — Form MD-26 to Form MD-27, expert committee and possible Indian clinical investigation

A predicate device (Rule 3(zm) in diagnostic summaries) is a device already legally marketed / licensed in India with similar intended use, materials, design and technology. If none exists:

  • File Form MD-26 to the CLA for permission to import or manufacture a device without a predicate; permission, if granted, issues in Form MD-27.
  • CDSCO refers the technical data to a Subject / Special Expert Committee, which opines on marketing on the existing data or on conducting a clinical investigation in India.
  • If an Indian clinical investigation is required, apply in Form MD-22 for permission in Form MD-23, with Ethics Committee approval, CTRI registration, Good Clinical Practice compliance and ICMR-aligned protocol design. For a new IVD, the parallel instrument is clinical performance evaluation (MD-24 → MD-25).
  • Waiver possibility (Rule 63 framework): clinical investigation may be waived where the device has been legally marketed for two years in a reference market — the sources checked consistently list the United States, United Kingdom, Australia, Canada and Japan (one 2026 summary reports extension of Rule 63 recognition to EU approvals — to be verified: current Rule 63 country list at filing, it is amendment-prone) — and the CLA is satisfied with the safety and performance data. A waiver is a CLA decision, not an entitlement: file the two-year marketing proof and the foreign clinical file as if you expect to be challenged.

Class A and B devices typically do not require clinical investigation. Class C and D devices with a predicate and strong foreign data often proceed on clinical evaluation; Class C and D devices without a predicate should plan for the investigation question from day one.

4.4 Domestic manufacturing — for completeness

If you manufacture in India instead of importing: Class A (sterile/measuring) and B apply to the SLA in MD-3 for MD-5 (Notified Body audit — for Class B, pre-grant audit within 90 days in the summaries checked; for Class A, audit within 120 days after grant); Class C and D apply to the CLA in MD-7 for MD-9 (mandatory pre-licensing inspection). Fees are in rupees and far lower than import fees (Section 7.2) — one reason the PLI / Make in India route is commercially attractive for high-volume devices. A loan licence (MD-4 → MD-6, MD-8 → MD-10) covers manufacture on another licensee's premises.

5. Step-by-Step: Placing a Medical Device on the Indian Market (Import Route)

Step 1 — Confirm the product is a device (or IVD) and get its Indian class. Apply the MDR 2017 definition (Rule 3(z)) and the First Schedule rules; check CDSCO's published classification list for your exact device type. If unlisted or novel, request classification from the CLA (SUGAM Risk Classification Module for novel uses). Document the rationale.

Step 2 — Run the predicate and reference-market check. Search whether a substantially equivalent device is licensed in India, and assemble your foreign approval set (FDA, CE under EU MDR, Health Canada, TGA, PMDA, MHRA) and marketing history. This one check decides standard vs. MD-26 route, clinical expectations and realistic timeline.

Step 3 — Appoint the Indian Authorised Agent. Execute the Power of Attorney (apostille / embassy authentication), verify the agent's wholesale / manufacturing / MD-42 position, IEC and portal account, and sign a written agreement covering licence holding, query response times, retention payment, vigilance, recall execution and exit/handover.

Step 4 — Build the PMF and DMF. Plant Master File: site layout, organisation, QMS, manufacturing and QC operations for the actual overseas site named in the application. Device Master File: device description and variants, intended use, design history, materials, risk management (ISO 14971), Essential Principles checklist, verification and validation, biocompatibility, sterilisation validation, shelf-life / stability, software validation where applicable, clinical evaluation, draft Indian labelling and IFU. Add ISO 13485 certificate covering that site and scope, and the Free Sale Certificate from the country of origin / competent authority.

Step 5 — Prepare Indian labelling before filing, not after. MDR labelling plus Legal Metrology importer declarations (Section 6). The labels filed must be the labels shipped.

Step 6 — File Form MD-14 online and pay the fees. The IAA files on SUGAM / CDSCO MD Online (via NSWS where the integrated route applies), uploads the dossier into the designated slots, and pays the Second Schedule fees per site and per device in USD. Keep the fee receipt and the filed dossier as a frozen copy.

Step 7 — Clear scrutiny, queries, testing and inspection. Respond to CDSCO deficiency / query letters completely within the stipulated window (commonly 30–45 days in portal practice summaries — to be verified: window stated on your query letter). For Class C and D, expect deeper technical scrutiny, possible MDTL testing and possible overseas-site inspection coordination. Partial answers restart cycles.

Step 8 — Receive Form MD-15. Check every field on grant: manufacturer legal name, site address, agent name, device names, models/variants and class. Errors on the licence become customs and tender problems. Diarise the five-year retention date on the day the licence arrives.

Step 9 — Complete the selling gate. Confirm MD-42 coverage and wholesale licensing for every entity/premises that will stock and distribute, agree customs classification and duty treatment with your clearing agent, and set up the import labelling / over-stickering process so no unit is sold before Indian declarations are applied.

Step 10 — Go live with post-market systems before the first commercial shipment. Materiovigilance reporting, complaint handling, FSCA/recall procedure, distribution traceability and change control (Section 9) must be operating, with named responsibilities split between manufacturer, IAA and distributors. Hold a test licence (MD-16 → MD-17) separately if you need pre-licence units for evaluation, exhibition or clinical work — an MD-15 is not a sample licence, and a test licence is not a sales licence.

6. Labelling and Language

Dossier filing and labelling are in English — English is the working language of CDSCO filing, and Hindi or other Indian languages are not a general MDR labelling condition for professional devices. to be verified: any state- or product-specific consumer-information expectation for lay / home-use devices at filing.

The Indian label set is two layers that must both be satisfied:

  1. 1. MDR 2017 labelling (Chapter VI, Rule 46 onwards). Device name, manufacturer name and address, manufacturing site, lot/batch or serial number, manufacturing and expiry dates where applicable, storage conditions, warnings, symbols, and the import licence number and IAA/importer identification required for imported devices. The label and IFU filed in the DMF must match the marketed label.
  2. 2. Legal Metrology (Packaged Commodities) declarations for imports. Importer name and complete address in India, MRP inclusive of all taxes, net quantity where applicable, country of origin, month/year of import, and consumer-care name, address, phone and email. In practice this is applied by compliant over-labelling / stickering in a bonded or approved process before retail sale — build it into landed cost and lead time, do not treat it as a port afterthought.

UDI — plan for it, do not claim it is fully live. Rule 46 of MDR 2017 contains the unique identification provision. The original 1 January 2022 mandatory date was delayed, and the sources checked in November 2026 consistently describe detailed UDI database, format and phased enforcement as still being finalised by CDSCO, with the strongest expectations on Class C and D devices (UDI-DI + UDI-PI in human-readable and machine-readable form). to be verified: current UDI notification, scope and date at filing. The prudent build is GS1-compatible DI/PI artwork readiness without printing a compliance claim CDSCO has not yet enforced.

Label, model, indication or site changes after licensing are post-approval changes under Rule 38 — major changes need prior CLA approval. Shipping a globally updated label to India before the Indian change is approved is a compliance breach, not a marketing update.

7. Timelines and Fees

7.1 Timelines

StepTypical timeNotes
Classification request (unlisted / novel device)Allow 1–3 months in the plan (*to be verified*)Precedes a clean MD-14 filing
Class A non-sterile, non-measuring registrationDays to a few weeksSelf-certification route; audit exposure follows
Import MD-15 — Class A (licensed) and B~6–9 monthsStandard predicate file, complete dossier
Import MD-15 — Class C and D~9–12 monthsDeeper scrutiny; testing / inspection can extend
Statutory CDSCO window for MD-15 (Rule 36)Up to 9 months to grant or rejectCitizen Charter target cited: 270 working days
Domestic manufacture — Class A and B (MD-5)~3–6 monthsIncludes Notified Body audit cycle
Domestic manufacture — Class C and D (MD-9)~6–12 monthsMandatory pre-licensing inspection
No-predicate route (MD-26 / expert committee ± Indian clinical investigation)Add 6–18+ months to the licence clock (*planning figure, to be verified per device*)Investigation duration is protocol-driven
Test licence (MD-17 import for evaluation / clinical)Shorter than MD-15; confirm on portal at filingNever a sales route

Timelines are scrutiny estimates, not guarantees; dossier completeness and query response speed move them in both directions.

7.2 Fees (Second Schedule to MDR 2017)

Import licence (Form MD-14 → MD-15) — paid in US dollars, per overseas manufacturing site and per distinct device:

Device classPer overseas sitePer distinct device
Class A (medical device)US$1,000US$50
Class B (medical device)US$2,000US$1,000
Class C and D (medical device)US$3,000US$1,500
Class A and B (IVD)US$1,000US$10
Class C and D (IVD)US$3,000US$500

Class A non-sterile, non-measuring registration carries no MD-14 fee in the summaries checked.

Domestic manufacturing licence — paid in rupees:

RoutePer sitePer device
Class A and B manufacture (MD-3 → MD-5)Rs. 5,000Rs. 500
Class C and D manufacture (MD-7 → MD-9)Rs. 50,000Rs. 1,000

Retention fee: the retention amount payable every five years to keep a perpetual licence alive follows the Second Schedule for the licence type — to be verified: exact retention figure for your licence on the portal invoice at the retention date; diarise the date, not just the amount. Late retention: 2% per month up to 90 days (Rule 37), then deemed cancellation.

Beyond government fees, budget for: IAA and regulatory support fees (commercial, and the real cost driver for most foreign makers); Notified Body / inspection and MDTL testing charges where directed; Power of Attorney apostille / legalisation; customs duty and GST by HS classification (to be verified with your clearing agent for your exact code — concessional entries exist for defined medical categories and are code-specific, never assume them); Legal Metrology over-labelling; and GeM / tender participation costs. A worked example: a single-site Class B import with three distinct devices costs US$2,000 + (3 × US$1,000) = US$5,000 in government fees alone, before any agent, testing or freight cost.

8. Selling After Registration: GeM, Tenders, Price Control and the Private Sector

Registration makes a device legal in India; it does not make it bought. The buying landscape has three channels:

  1. 1. Public procurement — GeM and tenders. Central and state government hospitals and institutions buy heavily through the Government e-Marketplace (GeM, the national public procurement portal under the Department of Commerce) and through institutional / state medical corporation tenders. Supplier onboarding, product catalogue listing with correct licence numbers, and local-content / Make in India preference rules under public procurement orders all affect competitiveness. A device not correctly catalogued on GeM with its MD-15 details is invisible to a large part of public demand.
  2. 2. Price control — check before you price. The NPPA caps prices for devices notified under the DPCO; coronary stents and knee implants are the long-standing capped categories, and trade-margin rationalisation has been applied to selected device groups in recent years. to be verified: current capped list and any trade-margin notification covering your device at pricing. Pricing India off a global list without this check has forced mid-launch repricing for foreign makers before.
  3. 3. Private hospitals and chains. India's large private sector — national chains, regional groups and teaching hospitals — buys commercially, qualifies suppliers on licence, QMS and clinical evidence, and typically moves faster than public tenders. The first qualification question is still the same: MD-15 / licence copy, please. Ayushman Bharat (PM-JAY) package rates shape procedure economics in empanelled hospitals and therefore device price sensitivity, even though the scheme buys procedures, not devices directly.

Practical points: public specifications are fixed when a tender publishes, so clinical evaluation placement, health-economics files and reference-site development must happen before the tender, through announced engagement channels and lawful evaluation placements (test licence route). Never market a device as "CDSCO approved" beyond what the licence says — CDSCO licenses manufacture/import for sale of the listed devices; it does not endorse efficacy claims beyond the licensed intended use.

9. After Placing on the Market: Retention, Materiovigilance and Change Control

  • Retention every five years. Pay the retention fee before the five-year date (Rule 37). Late fee 2% per month for up to 90 days; thereafter the licence is deemed cancelled and the device can no longer be legally imported or sold on it.
  • Materiovigilance — two reporting doors, learn both. (a) Licence-holder duty under MDR 2017: report suspected unexpected serious adverse events, and any action taken including recalls, to the concerned licensing authority within 15 days of the event coming to your notice. (b) MvPI system (IPC as National Coordination Centre): under the updated hospital guidance issued with WHO India in September 2026 (Version 1.3 in the sources checked), serious adverse events involving death or serious deterioration in health are reported to MvPI within 15 days, and non-serious events with the potential to cause serious harm — including near-miss malfunctions detected before use — within 30 days. Reports flow through Medical Device Adverse Event Monitoring Centres and the ADRMS portal; IPC evaluates, issues safety alerts (37 alerts and 59 CDSCO recommendations from 47,614 cases reviewed to September 2025 in IPC reporting checked) and forwards cases to CDSCO. Your IAA must operate both doors, not just forward hospital mail.
  • Post-market surveillance. Maintain PMS: complaint trending, proactive incident review, distribution traceability by lot/serial, periodic review of foreign field actions, and cooperation with CDSCO market surveillance and MDTL sampling.
  • Field Safety Corrective Actions and recalls. Execute FSCAs and recalls in India under CDSCO oversight, including foreign-origin FSCAs affecting Indian-supplied units, with consignee-level traceability. Report the action under the 15-day licence-holder duty above.
  • Post-approval changes (Rules 38–39). Major changes — device design, intended use, manufacturing site, labelling with regulatory impact — require prior CLA approval (60-day decision window cited in the sources checked). Agent change is a strategic event, not an administrative one (Section 3). Build an India check into your global change-control form.
  • QMS continuity. Keep ISO 13485 certification, PMF and DMF current for the licensed site; a lapsed QMS certificate undermines the licence it supports at retention, inspection or variation.
  • Distribution control. Every stockist in the chain must be within the MD-42 / wholesale licensing perimeter, store and transport under labelled conditions, and keep records sufficient for a lot-level recall.

10. Document Checklist (India Import Launch File)

Agent and applicant

  • ☐ Power of Attorney from the overseas manufacturer to the IAA — apostilled (Hague) or Indian Embassy-authenticated
  • ☐ IAA's wholesale licence / manufacturing licence / Form MD-42 certificate (copy verified against portal)
  • ☐ IAA's IEC, Indian incorporation documents and SUGAM / MD Online portal account confirmation
  • ☐ Written agent agreement: licence holding, query SLAs, retention payment, vigilance, FSCA/recall, exit and document handover

Classification and predicate

  • ☐ CDSCO classification-list extract for the device type, or CLA classification communication
  • ☐ Classification rationale (First Schedule rule applied) in the DMF
  • ☐ Predicate comparison (intended use, materials, design, technology) or MD-26 no-predicate strategy

Technical dossier (filed with Form MD-14)

  • ☐ Plant Master File (PMF) for the actual overseas manufacturing site
  • ☐ Device Master File (DMF): description, variants, design, materials, specifications
  • ☐ Essential Principles checklist and declaration of conformity
  • ☐ Risk management file (ISO 14971)
  • ☐ Design verification and validation; biocompatibility; sterilisation validation; shelf-life / stability
  • ☐ Software validation and cybersecurity file where applicable
  • ☐ Clinical evaluation report (and foreign clinical / marketing-history file for any Rule 63 waiver request)
  • ☐ ISO 13485 certificate naming the site and scope in the application
  • ☐ Free Sale Certificate from the country of origin / competent authority
  • ☐ Foreign approvals (FDA / CE / Health Canada / TGA / PMDA / MHRA) for the identical device and use
  • ☐ Indian labels and IFU as filed (MDR + Legal Metrology layers)

Registration lifecycle

  • ☐ Form MD-14 filing receipt and Second Schedule fee receipt (per site + per device, USD)
  • ☐ Form MD-15 licence — every field checked on grant; five-year retention date diarised
  • ☐ MD-42 / wholesale coverage for each stocking / distributing entity and premises

Post-market

  • ☐ Vigilance SOP covering both doors: 15-day licensing-authority report and 15 / 30-day MvPI timelines, with named IAA responsibilities
  • ☐ Complaint handling, FSCA/recall procedure and lot/serial traceability records
  • ☐ Change-control rule: no India shipment on an unapproved Rule 38 change

11. Common Mistakes in India

  1. 1. Treating CE marking or FDA clearance as an Indian authorisation. They are evidence inside the MD-14 file and support for a Rule 63 waiver — the MD-15 from CDSCO is still required before import or sale.
  2. 2. Filing without an agent strategy. The IAA holds your MD-15. Handing the licence to a sales distributor by default, without exit and handover terms, makes a later distributor change a re-licensing event.
  3. 3. Quoting the Class A timeline for a Class C device. Registration-number Class A takes days; a Class C/D import takes 9–12 months, and a no-predicate device materially longer. Launch dates promised on the wrong route destroy partner credibility.
  4. 4. Discovering the predicate question at scrutiny. No predicate means MD-26, expert committee and possible Indian clinical investigation. Run the predicate search before the commercial plan, not after the query letter.
  5. 5. Inconsistent dossiers. Different device names, models, sites or intended uses across the CE file, ISO 13485 certificate, Free Sale Certificate, PMF and DMF are the most common self-inflicted query round.
  6. 6. Forgetting the second labelling layer. MDR labelling alone does not satisfy Legal Metrology importer declarations and MRP. Units sold without compliant importer over-labelling are enforcement exposure at retail inspection, not at the port alone.
  7. 7. Missing the five-year retention. Perpetual validity invites complacency; retention is a hard diary event with 2% monthly late fee for only 90 days, then deemed cancellation.
  8. 8. Pricing before the NPPA check. A capped or margin-rationalised category reprices your whole channel model. Check the DPCO / NPPA position before signing distributor margins.
  9. 9. Shipping samples on a sales licence — or selling on a test licence. Evaluation, exhibition and clinical units belong on the test licence (MD-16 → MD-17); commercial units belong on MD-15. Mixing the two is a licensing offence in both directions.
  10. 10. Shipping label or site changes without a Rule 38 approval. Global change control that does not include an India gate ships breaches automatically.
  11. 11. Treating software as unregulated IT. Diagnostic, screening and treatment software is a classified device in India; draft SaMD guidance is tightening, not loosening, the file expectations.
  12. 12. Ignoring state-level distribution licensing. The MD-15 is central; MD-42 sale/distribution coverage is state-level. Both gates must close before the first sale in each stocking location.

12. FAQ

Who registers medical devices in India? The Central Drugs Standard Control Organisation (CDSCO), headed by the Drugs Controller General of India (DCGI) under the Ministry of Health and Family Welfare. The DCGI as Central Licensing Authority grants all import licences (MD-15) and Class C/D manufacturing licences; State Licensing Authorities grant Class A/B manufacturing licences and MD-42 sale/distribution certificates.

Do I need a local representative? Yes, if you are a foreign manufacturer. You must appoint an Indian Authorised Agent — an Indian entity holding a manufacturing licence, a wholesale licence or a Form MD-42 certificate, with an IEC and a CDSCO portal account — under an apostilled or embassy-authenticated Power of Attorney (Rules 3(f) and 34). The agent files Form MD-14 and holds the MD-15. You do not need your own Indian subsidiary, though a subsidiary is the strongest agent structure.

How are devices classified? In four risk classes, A (low) to D (high), under Rule 4 and the First Schedule to MDR 2017 — dressings and masks in A; needles, BP monitors and oxygen concentrators in B; ventilators, implants and X-ray systems in C; stents, pacemakers, heart valves and HIV kits in D. The Central Licensing Authority makes the final determination and publishes classification lists; novel uses go through the SUGAM Risk Classification Module.

My device is CE-marked. How fast can I register? CE marking strengthens the file but does not shorten the statutory route by itself: plan about 6–9 months for a Class A (licensed) / B import and 9–12 months for Class C/D, within CDSCO's nine-month Rule 36 decision window plus query time. CE / FDA status matters most for the no-predicate question — two years of marketing in the US, UK, Australia, Canada or Japan (and, to be verified, the EU) can support waiver of an Indian clinical investigation under the Rule 63 framework.

How much does registration cost? Government import fees (Second Schedule, USD): Class A US$1,000 per site + US$50 per device; Class B US$2,000 + US$1,000; Class C/D US$3,000 + US$1,500; IVDs pay US$10 (A/B) or US$500 (C/D) per device at the same per-site fees. Domestic manufacturing fees are Rs. 5,000 + Rs. 500 (A/B) and Rs. 50,000 + Rs. 1,000 (C/D). Add agent, testing, inspection, customs and labelling costs — the government fee is rarely the largest line.

How long is a licence valid? In perpetuity, provided the retention fee is paid every five years (Rule 37). Late payment: 2% per month for up to 90 days; after that the licence is deemed cancelled.

Can my distributor also be my authorised agent and importer? Yes — one Indian company may hold all three roles, but it must actually hold each licence (MD-15 as agent, MD-42 / wholesale for distribution) and the MD-15 will sit with that company. Agree exit and handover terms before filing; changing agent is not a simple transfer.

What if no similar device is registered in India? You take the no-predicate route: Form MD-26 to the CLA, Subject Expert Committee review, permission in Form MD-27, and potentially an Indian clinical investigation (MD-22 → MD-23; MD-24 → MD-25 for a new IVD's performance evaluation) unless the CLA waives investigation on the strength of two years' reference-market history and your foreign clinical data.

Does India require UDI? Rule 46 of MDR 2017 provides for unique device identification, but the original January 2022 date was delayed and detailed phased enforcement was still being finalised in the sources checked in November 2026, with Class C and D the expected first scope. Build GS1-compatible DI/PI readiness; to be verified: current notification and date at filing.

What are my duties after registration? Five-year retention; materiovigilance — suspected unexpected serious adverse events and recalls to the licensing authority within 15 days, and MvPI reporting within 15 days (serious) / 30 days (non-serious with serious-harm potential, including near misses); PMS and complaint trending; FSCA/recall execution with lot traceability; prior CLA approval for major changes (Rule 38); current ISO 13485 / PMF / DMF; and MD-42-compliant distribution.

Is there a public tender channel I should plan around? Yes. Government buyers purchase through the Government e-Marketplace (GeM) and institutional/state tenders, private chains buy commercially, and NPPA price ceilings apply to notified categories. Registration is the entry ticket to all three; GeM catalogue correctness and the price-control check decide whether the ticket is usable.

Sources (checked November 2026)

  • Central Drugs Standard Control Organisation (CDSCO), Medical Device & Diagnostics — cdsco.gov.in (all devices under Drugs and Cosmetics Act 1940 and MDR 2017; A–D risk classification per First Schedule Part I; form routes MD-14→MD-15, MD-7/MD-8→MD-9/MD-10, MD-12→MD-13, MD-16→MD-17, MD-22→MD-23, MD-24→MD-25, MD-26 no-predicate permission; Notified Body MD-1→MD-2).
  • Medical Devices Rules, 2017 (CDSCO consolidated text) — cdsco.gov.in (Rule 3 definitions including authorised agent and predicate; Rule 4 classification; Rules 34–39 import licensing, Rule 36 nine-month decision window, Rule 37 five-year retention with 2% monthly late fee to 90 days; Rule 46 UDI provision; Second Schedule fees).
  • CDSCO FAQ on Medical Devices Rules (2024) and Citizen Charter (2025), as summarised in the consultant sources below (portal routes cdscomdonline.gov.in / SUGAM / NSWS; 270-working-day MD-15 target).
  • Shen Clinical Services, Medical Device Registration in India and Indian Authorized Agent — shenclinicalservices.com (form matrix; IAA eligibility under Rules 3(f) and 34–39 — manufacturing / wholesale / MD-42 holding, Power of Attorney apostille/embassy authentication; import fees USD 1,000/50, 2,000/1,000, 3,000/1,500 and IVD per-device USD 10 / 500; Rule 38 major-change 60-day decision; perpetual validity with five-year retention).
  • Morulaa, India Medical Device Registration & Import License Guide and Indian Authorized Agent (IAA) Services — morulaa.com (6–9 month notified-device timeline; MD-14 filing by IAA on SUGAM; fee table by class; November 2025 SUGAM Risk Classification Module; October 2023 all-devices licensing milestone; Rule 63 no-predicate pathway, MD-26→MD-27, expert committee, two-year reference-market waiver including the reported EU extension).
  • NKB Regovanta, How to Register a Medical Device with CDSCO in India — nkbregovanta.com (phased licensing: A/B October 2022, C/D October 2023; manufacturing fees Rs. 5,000/500 and Rs. 50,000/1,000; import timelines 6–9 and 9–12 months; 2026 Rule 63 / EU note and draft SaMD guidance).
  • EfilingCompany, CDSCO Registration: Forms, Fees & Process (2026) — efilingcompany.com (Class A non-sterile non-measuring registration-number route; SLA vs CLA split; MD-41→MD-42 sale/distribution certificate; portal split cdscoonline / cdscomdonline).
  • Medical Device Registration (MDR portal guides), UDI in India: Current Status & Compliance Roadmap — medicaldeviceregistration.com (Rule 46 UDI provision; January 2022 date delayed; phased implementation and database still being finalised; DI/PI in AIDC and human-readable form expected).
  • Indian Pharmacopoeia Commission / MvPI reporting, via Medical Buyer, PharmaBiz and hospital-guidance summaries (September–October 2026) — medicalbuyer.co.in / pharmabiz.com (MvPI launched 2015, IPC as National Coordination Centre; 47,614 MDAEs reviewed to September 2025, 37 safety alerts, 59 CDSCO recommendations; 775 monitoring centres; updated hospital guidance v1.3 with WHO India, September 2026: serious events to MvPI within 15 days, non-serious with serious-harm potential within 30 days, suspected unexpected serious events to the licensing authority within 15 days; ADRMS portal and helpline 1800-180-3024).
  • India Brand Equity Foundation (IBEF), Medical Devices Industry Analysis (May 2026) — ibef.org (market about US$12 billion in FY24, fourth in Asia, top-20 globally; PLI Rs. 3,420 crore; exports and FDI data).
  • Department of Pharmaceuticals / PIB releases via Redisry and Medical Buyer summaries (November–December 2025 / 2026) — redisry.com / medicalbuyer.co.in (device exports US$2.9 billion FY2021-22 to US$4.1 billion FY2024-25; imports about US$8.6–8.8 billion FY2024-25, 70–80% import dependence; National Medical Devices Policy 2023 implementation; GeM as the national public procurement portal for government buyers).

Sources & scope

  • Central Drugs Standard Control Organisation (CDSCO), Medical Device & Diagnostics — cdsco.gov.in.
  • Medical Devices Rules, 2017 (consolidated text) — classification, import licensing, retention, Second Schedule fees.
  • CDSCO FAQ on Medical Devices Rules (2024) and Citizen Charter (2025).
  • Indian Pharmacopoeia Commission / Materiovigilance Programme of India (MvPI) reporting, 2025–2026.
  • India Brand Equity Foundation (IBEF), Medical Devices Industry Analysis, 2026.
  • Consultant practice summaries checked against the sources above: Shen Clinical Services, Morulaa, NKB Regovanta, EfilingCompany.

This guide is market-access intelligence, not regulatory or legal advice. CDSCO pathways and fee schedules are product-specific — professional review is required before business use.