The MD-14 → MD-15 route
MD-14 filed on SUGAM leads to the MD-15 license — perpetual validity, 6–9 month typical timelines.
Field guide · PDF
The MD-14 to MD-15 import licence, the mandatory Indian Authorised Agent, the predicate fork, and the five-year retention rule — for a 1.4-billion-patient market.
Sourced from CDSCO and the Medical Devices Rules, 2017 — checked November 2026.
Enter your work email and we'll send the PDF straight to your inbox.
What's inside
Every figure in this guide traces to a named source — CDSCO, the Medical Devices Rules 2017, and the sources named in the guide.
MD-14 filed on SUGAM leads to the MD-15 license — perpetual validity, 6–9 month typical timelines.
Foreign manufacturers cannot file directly — the agent files, answers queries, and handles vigilance.
Second Schedule fees by class, the no-predicate route (MD-26/MD-27), 15/30-day materiovigilance, and the mistakes that reset applications.
Complete guide
Every section of the PDF on one page. Use the form above to keep a copy.
Last updated: November 2026. India regulates medical devices — including in vitro diagnostics (IVDs) — nationally under the Drugs and Cosmetics Act, 1940 and the Medical Devices Rules, 2017 (MDR 2017). The national regulatory authority is the Central Drugs Standard Control Organisation (CDSCO), headed by the Drugs Controller General of India (DCGI) under the Ministry of Health and Family Welfare (MoHFW). India is not part of the EU MDR system: a CE mark is supporting evidence, never an Indian authorisation. Since notification S.O. 648(E) of 11 February 2020, all medical devices are regulated under MDR 2017, with mandatory licensing phased in for Class A and B devices from October 2022 and for Class C and D devices from October 2023. Always confirm the current classification list, fee schedule and portal workflow on CDSCO's own sites — cdsco.gov.in, the SUGAM portal and the CDSCO Medical Devices Online portal (cdscomdonline.gov.in) — and, for higher-risk and single-window filings, the National Single Window System (NSWS) before filing.
India is Asia's fourth-largest medical device market after Japan, China and South Korea, and one of the world's top 20. The India Brand Equity Foundation (IBEF), citing Department of Pharmaceuticals data, valued the industry at about Rs. 1,02,660 crore (about US$12 billion) in FY24. Government and industry sources checked for this guide put imports at roughly 70–80% of domestic demand — about US$8.6–8.8 billion in FY2024-25, concentrated in imaging, electro-medical equipment, IVDs, implants and high-end consumables — against exports of about US$4.0–4.1 billion in the same year. That import dependence is precisely why the import-licence gate matters: for a foreign manufacturer, India is not a distributor-led free sale market. You cannot legally import, stock or sell a regulated device without a CDSCO licence held by an Indian entity, and the licence is tied to a specific overseas manufacturing site and specific devices.
Regulatory reality check up front — seven features of the Indian system that foreign manufacturers most often underestimate:
A note on old information: pre-2020 sources describe only a short list of "notified" devices, a three-year registration certificate and Form 40 filing. All three are out of date for a 2026 filing. The current system covers all devices, uses MDR 2017 forms, and grants perpetual licences with five-year retention.
| Instrument | What it does |
|---|---|
| Drugs and Cosmetics Act, 1940 | Parent statute under which MDR 2017 is made; offences and penalties for unlicensed manufacture, import or sale sit here. |
| Medical Devices Rules, 2017 (effective 1 January 2018) | The operative device rules: classification (Chapter II, Rule 4, First Schedule), manufacturing licensing (Chapter V), import licensing (Chapter VI, Rules 34–39), clinical investigation (Chapter VII), labelling (Chapter VI / Rule 46 onwards), testing laboratories, Notified Bodies and post-approval changes. Amended repeatedly since 2020 to bring all devices into scope. |
| Notification S.O. 648(E), 11 February 2020 | Extended the device definition so that *all* medical devices fall under MDR 2017, ending the notified-devices-only era. |
| Second Schedule to MDR 2017 | Government fee schedule: per-site and per-device fees for manufacturing and import (Section 7.2). |
| Fifth Schedule / QMS requirements | Quality management system requirements aligned to ISO 13485; Plant Master File and Device Master File are the core dossier expressions of this system. |
| National Medical Devices Policy, 2023 | Industrial policy (Department of Pharmaceuticals): domestic manufacturing, PLI incentives, medical device parks, export promotion. Policy, not a licensing instrument — it shapes incentives, not approvals. |
| Production Linked Incentive (PLI) Scheme for Medical Devices | Rs. 3,420 crore outlay, FY23–FY27 in the sources checked: 5% incentive on incremental sales for defined categories (cancer care / radiotherapy, radiology and imaging, anaesthetics and cardio-respiratory including renal care, and implants). Relevant if you localise manufacturing; irrelevant to an import licence. |
| Legal Metrology Act, 2009 and Packaged Commodities Rules | Consumer-pack labelling layer that sits alongside MDR labelling for imported packaged devices: importer name and address, maximum retail price (MRP) inclusive of taxes, country of origin, customer-care details. |
| Drugs (Prices Control) Order (DPCO) / NPPA notifications | Ceiling prices for scheduled / notified devices. Applies to a defined list only — check, do not assume. |
India also operates the Materiovigilance Programme of India (MvPI), launched 6 July 2015 and coordinated by the Indian Pharmacopoeia Commission (IPC) as National Coordination Centre, and a network of registered Notified Bodies (accredited through the National Accreditation Board for Certification Bodies) that audit Class A and B manufacturing sites, while CDSCO Medical Device Officers and authorised auditors handle Class C and D scrutiny.
| Route | Application | Licence / output | Authority |
|---|---|---|---|
| Class A non-sterile, non-measuring — manufacture or import | Online registration on CDSCO MD portal (Chapter IIIB self-certification framework, G.S.R. 777(E) in the sources checked) | System-generated registration number — no licence | CDSCO portal |
| Manufacture Class A (sterile/measuring) and Class B | MD-3 (loan licence MD-4) | MD-5 (loan MD-6) | State Licensing Authority |
| Manufacture Class C and D | MD-7 (loan licence MD-8) | MD-9 (loan MD-10) | Central Licensing Authority |
| Import, all classes | MD-14 | MD-15 | Central Licensing Authority |
| Sale, stock, exhibit or distribution | MD-41 | MD-42 registration certificate | State Licensing Authority |
| Test licence — manufacture for test / evaluation / clinical investigation | MD-12 | MD-13 | CLA / SLA by class |
| Test licence — import for test / evaluation / clinical investigation | MD-16 | MD-17 | Central Licensing Authority |
| Clinical investigation of investigational device | MD-22 | MD-23 permission | Central Licensing Authority |
| Clinical performance evaluation of new IVD | MD-24 | MD-25 permission | Central Licensing Authority |
| Device with no predicate — permission to import / manufacture | MD-26 | MD-27 permission | Central Licensing Authority |
India classifies devices into four risk classes under Rule 4 of MDR 2017, aligned to GHTF/IMDRF principles:
| Class | Risk | Examples (from CDSCO-aligned sources checked) |
|---|---|---|
| A | Low | Surgical dressings, examination gloves (non-sterile), tongue depressors, surgical masks, nasopharyngeal catheters, general culture media and manual staining solutions (IVD) |
| B | Low–moderate | Hypodermic needles, IV cannulae, suction units, blood pressure monitors, oxygen concentrators, insulin pen needles; blood glucose monitors and pregnancy test kits (IVD) |
| C | Moderate–high | Haemodialysis catheters, ventilators, bone screws and orthopaedic implants, breast implants, surgical lasers, X-ray equipment; hepatitis B/C kits and PSA assays (IVD) |
| D | High | Coronary stents, pacemakers and implantable defibrillators, heart valves, deep-brain stimulation devices; HIV diagnostic kits and blood-grouping reagents (IVD) |
Practical classification notes:
A foreign manufacturer cannot file, import or distribute in India on its own. Three local functions must be in place:
Register on the CDSCO Medical Devices Online portal, self-certify compliance with MDR 2017 and the Essential Principles, and obtain the system-generated registration number. There is no MD-15 and no per-device government fee in the summaries checked for this sub-route. This is the only genuinely light route in India — and it applies to a narrow slice of Class A. Sterile or measuring Class A devices leave this route entirely.
The standard foreign-manufacturer route: the IAA files Form MD-14 on the SUGAM / MD Online system with the Plant Master File (PMF), Device Master File (DMF), ISO 13485 certificate, Free Sale Certificate and the full technical file (Section 10), pays the Second Schedule fees in US dollars per overseas site and per distinct device, answers CDSCO scrutiny queries, and — for higher-risk devices — supports any site inspection or testing CDSCO directs. CDSCO grants the import licence in Form MD-15.
Rule 36 gives CDSCO up to nine months from application to grant or reject an MD-15; CDSCO's Citizen Charter (2025) target cited in the sources checked is 270 working days. Well-prepared notified/predicate files are commonly reported at 6–9 months; Class C and D imports at 9–12 months. Treat any promise of a few weeks for a Class B–D import as a red flag.
A predicate device (Rule 3(zm) in diagnostic summaries) is a device already legally marketed / licensed in India with similar intended use, materials, design and technology. If none exists:
Class A and B devices typically do not require clinical investigation. Class C and D devices with a predicate and strong foreign data often proceed on clinical evaluation; Class C and D devices without a predicate should plan for the investigation question from day one.
If you manufacture in India instead of importing: Class A (sterile/measuring) and B apply to the SLA in MD-3 for MD-5 (Notified Body audit — for Class B, pre-grant audit within 90 days in the summaries checked; for Class A, audit within 120 days after grant); Class C and D apply to the CLA in MD-7 for MD-9 (mandatory pre-licensing inspection). Fees are in rupees and far lower than import fees (Section 7.2) — one reason the PLI / Make in India route is commercially attractive for high-volume devices. A loan licence (MD-4 → MD-6, MD-8 → MD-10) covers manufacture on another licensee's premises.
Step 1 — Confirm the product is a device (or IVD) and get its Indian class. Apply the MDR 2017 definition (Rule 3(z)) and the First Schedule rules; check CDSCO's published classification list for your exact device type. If unlisted or novel, request classification from the CLA (SUGAM Risk Classification Module for novel uses). Document the rationale.
Step 2 — Run the predicate and reference-market check. Search whether a substantially equivalent device is licensed in India, and assemble your foreign approval set (FDA, CE under EU MDR, Health Canada, TGA, PMDA, MHRA) and marketing history. This one check decides standard vs. MD-26 route, clinical expectations and realistic timeline.
Step 3 — Appoint the Indian Authorised Agent. Execute the Power of Attorney (apostille / embassy authentication), verify the agent's wholesale / manufacturing / MD-42 position, IEC and portal account, and sign a written agreement covering licence holding, query response times, retention payment, vigilance, recall execution and exit/handover.
Step 4 — Build the PMF and DMF. Plant Master File: site layout, organisation, QMS, manufacturing and QC operations for the actual overseas site named in the application. Device Master File: device description and variants, intended use, design history, materials, risk management (ISO 14971), Essential Principles checklist, verification and validation, biocompatibility, sterilisation validation, shelf-life / stability, software validation where applicable, clinical evaluation, draft Indian labelling and IFU. Add ISO 13485 certificate covering that site and scope, and the Free Sale Certificate from the country of origin / competent authority.
Step 5 — Prepare Indian labelling before filing, not after. MDR labelling plus Legal Metrology importer declarations (Section 6). The labels filed must be the labels shipped.
Step 6 — File Form MD-14 online and pay the fees. The IAA files on SUGAM / CDSCO MD Online (via NSWS where the integrated route applies), uploads the dossier into the designated slots, and pays the Second Schedule fees per site and per device in USD. Keep the fee receipt and the filed dossier as a frozen copy.
Step 7 — Clear scrutiny, queries, testing and inspection. Respond to CDSCO deficiency / query letters completely within the stipulated window (commonly 30–45 days in portal practice summaries — to be verified: window stated on your query letter). For Class C and D, expect deeper technical scrutiny, possible MDTL testing and possible overseas-site inspection coordination. Partial answers restart cycles.
Step 8 — Receive Form MD-15. Check every field on grant: manufacturer legal name, site address, agent name, device names, models/variants and class. Errors on the licence become customs and tender problems. Diarise the five-year retention date on the day the licence arrives.
Step 9 — Complete the selling gate. Confirm MD-42 coverage and wholesale licensing for every entity/premises that will stock and distribute, agree customs classification and duty treatment with your clearing agent, and set up the import labelling / over-stickering process so no unit is sold before Indian declarations are applied.
Step 10 — Go live with post-market systems before the first commercial shipment. Materiovigilance reporting, complaint handling, FSCA/recall procedure, distribution traceability and change control (Section 9) must be operating, with named responsibilities split between manufacturer, IAA and distributors. Hold a test licence (MD-16 → MD-17) separately if you need pre-licence units for evaluation, exhibition or clinical work — an MD-15 is not a sample licence, and a test licence is not a sales licence.
Dossier filing and labelling are in English — English is the working language of CDSCO filing, and Hindi or other Indian languages are not a general MDR labelling condition for professional devices. to be verified: any state- or product-specific consumer-information expectation for lay / home-use devices at filing.
The Indian label set is two layers that must both be satisfied:
UDI — plan for it, do not claim it is fully live. Rule 46 of MDR 2017 contains the unique identification provision. The original 1 January 2022 mandatory date was delayed, and the sources checked in November 2026 consistently describe detailed UDI database, format and phased enforcement as still being finalised by CDSCO, with the strongest expectations on Class C and D devices (UDI-DI + UDI-PI in human-readable and machine-readable form). to be verified: current UDI notification, scope and date at filing. The prudent build is GS1-compatible DI/PI artwork readiness without printing a compliance claim CDSCO has not yet enforced.
Label, model, indication or site changes after licensing are post-approval changes under Rule 38 — major changes need prior CLA approval. Shipping a globally updated label to India before the Indian change is approved is a compliance breach, not a marketing update.
| Step | Typical time | Notes |
|---|---|---|
| Classification request (unlisted / novel device) | Allow 1–3 months in the plan (*to be verified*) | Precedes a clean MD-14 filing |
| Class A non-sterile, non-measuring registration | Days to a few weeks | Self-certification route; audit exposure follows |
| Import MD-15 — Class A (licensed) and B | ~6–9 months | Standard predicate file, complete dossier |
| Import MD-15 — Class C and D | ~9–12 months | Deeper scrutiny; testing / inspection can extend |
| Statutory CDSCO window for MD-15 (Rule 36) | Up to 9 months to grant or reject | Citizen Charter target cited: 270 working days |
| Domestic manufacture — Class A and B (MD-5) | ~3–6 months | Includes Notified Body audit cycle |
| Domestic manufacture — Class C and D (MD-9) | ~6–12 months | Mandatory pre-licensing inspection |
| No-predicate route (MD-26 / expert committee ± Indian clinical investigation) | Add 6–18+ months to the licence clock (*planning figure, to be verified per device*) | Investigation duration is protocol-driven |
| Test licence (MD-17 import for evaluation / clinical) | Shorter than MD-15; confirm on portal at filing | Never a sales route |
Timelines are scrutiny estimates, not guarantees; dossier completeness and query response speed move them in both directions.
Import licence (Form MD-14 → MD-15) — paid in US dollars, per overseas manufacturing site and per distinct device:
| Device class | Per overseas site | Per distinct device |
|---|---|---|
| Class A (medical device) | US$1,000 | US$50 |
| Class B (medical device) | US$2,000 | US$1,000 |
| Class C and D (medical device) | US$3,000 | US$1,500 |
| Class A and B (IVD) | US$1,000 | US$10 |
| Class C and D (IVD) | US$3,000 | US$500 |
Class A non-sterile, non-measuring registration carries no MD-14 fee in the summaries checked.
Domestic manufacturing licence — paid in rupees:
| Route | Per site | Per device |
|---|---|---|
| Class A and B manufacture (MD-3 → MD-5) | Rs. 5,000 | Rs. 500 |
| Class C and D manufacture (MD-7 → MD-9) | Rs. 50,000 | Rs. 1,000 |
Retention fee: the retention amount payable every five years to keep a perpetual licence alive follows the Second Schedule for the licence type — to be verified: exact retention figure for your licence on the portal invoice at the retention date; diarise the date, not just the amount. Late retention: 2% per month up to 90 days (Rule 37), then deemed cancellation.
Beyond government fees, budget for: IAA and regulatory support fees (commercial, and the real cost driver for most foreign makers); Notified Body / inspection and MDTL testing charges where directed; Power of Attorney apostille / legalisation; customs duty and GST by HS classification (to be verified with your clearing agent for your exact code — concessional entries exist for defined medical categories and are code-specific, never assume them); Legal Metrology over-labelling; and GeM / tender participation costs. A worked example: a single-site Class B import with three distinct devices costs US$2,000 + (3 × US$1,000) = US$5,000 in government fees alone, before any agent, testing or freight cost.
Registration makes a device legal in India; it does not make it bought. The buying landscape has three channels:
Practical points: public specifications are fixed when a tender publishes, so clinical evaluation placement, health-economics files and reference-site development must happen before the tender, through announced engagement channels and lawful evaluation placements (test licence route). Never market a device as "CDSCO approved" beyond what the licence says — CDSCO licenses manufacture/import for sale of the listed devices; it does not endorse efficacy claims beyond the licensed intended use.
Agent and applicant
Classification and predicate
Technical dossier (filed with Form MD-14)
Registration lifecycle
Post-market
Who registers medical devices in India? The Central Drugs Standard Control Organisation (CDSCO), headed by the Drugs Controller General of India (DCGI) under the Ministry of Health and Family Welfare. The DCGI as Central Licensing Authority grants all import licences (MD-15) and Class C/D manufacturing licences; State Licensing Authorities grant Class A/B manufacturing licences and MD-42 sale/distribution certificates.
Do I need a local representative? Yes, if you are a foreign manufacturer. You must appoint an Indian Authorised Agent — an Indian entity holding a manufacturing licence, a wholesale licence or a Form MD-42 certificate, with an IEC and a CDSCO portal account — under an apostilled or embassy-authenticated Power of Attorney (Rules 3(f) and 34). The agent files Form MD-14 and holds the MD-15. You do not need your own Indian subsidiary, though a subsidiary is the strongest agent structure.
How are devices classified? In four risk classes, A (low) to D (high), under Rule 4 and the First Schedule to MDR 2017 — dressings and masks in A; needles, BP monitors and oxygen concentrators in B; ventilators, implants and X-ray systems in C; stents, pacemakers, heart valves and HIV kits in D. The Central Licensing Authority makes the final determination and publishes classification lists; novel uses go through the SUGAM Risk Classification Module.
My device is CE-marked. How fast can I register? CE marking strengthens the file but does not shorten the statutory route by itself: plan about 6–9 months for a Class A (licensed) / B import and 9–12 months for Class C/D, within CDSCO's nine-month Rule 36 decision window plus query time. CE / FDA status matters most for the no-predicate question — two years of marketing in the US, UK, Australia, Canada or Japan (and, to be verified, the EU) can support waiver of an Indian clinical investigation under the Rule 63 framework.
How much does registration cost? Government import fees (Second Schedule, USD): Class A US$1,000 per site + US$50 per device; Class B US$2,000 + US$1,000; Class C/D US$3,000 + US$1,500; IVDs pay US$10 (A/B) or US$500 (C/D) per device at the same per-site fees. Domestic manufacturing fees are Rs. 5,000 + Rs. 500 (A/B) and Rs. 50,000 + Rs. 1,000 (C/D). Add agent, testing, inspection, customs and labelling costs — the government fee is rarely the largest line.
How long is a licence valid? In perpetuity, provided the retention fee is paid every five years (Rule 37). Late payment: 2% per month for up to 90 days; after that the licence is deemed cancelled.
Can my distributor also be my authorised agent and importer? Yes — one Indian company may hold all three roles, but it must actually hold each licence (MD-15 as agent, MD-42 / wholesale for distribution) and the MD-15 will sit with that company. Agree exit and handover terms before filing; changing agent is not a simple transfer.
What if no similar device is registered in India? You take the no-predicate route: Form MD-26 to the CLA, Subject Expert Committee review, permission in Form MD-27, and potentially an Indian clinical investigation (MD-22 → MD-23; MD-24 → MD-25 for a new IVD's performance evaluation) unless the CLA waives investigation on the strength of two years' reference-market history and your foreign clinical data.
Does India require UDI? Rule 46 of MDR 2017 provides for unique device identification, but the original January 2022 date was delayed and detailed phased enforcement was still being finalised in the sources checked in November 2026, with Class C and D the expected first scope. Build GS1-compatible DI/PI readiness; to be verified: current notification and date at filing.
What are my duties after registration? Five-year retention; materiovigilance — suspected unexpected serious adverse events and recalls to the licensing authority within 15 days, and MvPI reporting within 15 days (serious) / 30 days (non-serious with serious-harm potential, including near misses); PMS and complaint trending; FSCA/recall execution with lot traceability; prior CLA approval for major changes (Rule 38); current ISO 13485 / PMF / DMF; and MD-42-compliant distribution.
Is there a public tender channel I should plan around? Yes. Government buyers purchase through the Government e-Marketplace (GeM) and institutional/state tenders, private chains buy commercially, and NPPA price ceilings apply to notified categories. Registration is the entry ticket to all three; GeM catalogue correctness and the price-control check decide whether the ticket is usable.