Lithuania Medical Device Registration2026 Field Guide

No Lithuanian CE mark — EU MDR applies directly. Lithuania adds its own layer on top: a VASPVT registration confirmed with registration numbers before a device is placed on the market, a 14-working-day data duty for the higher classes, and a Lithuanian-language rule whose English exception needs the professional recipient's consent. Add a payer whose reimbursable device price lists move only twice a year.

window to submit device data to VASPVT after placing Class IIa, IIb, III and custom-made active implantable devices on the market (Law on the Health System)
14 working days
of the population covered for a core set of services (OECD, Health at a Glance 2025)
99%
of GDP spent on health (OECD, Health at a Glance 2025)
7.6%

Sourced from the State Health Care Accreditation Agency, the National Health Insurance Fund, the Law on the Health System and the OECD, and verified 2026 publications.

Download the guide

Enter your work email and we'll send the PDF straight to your inbox.

Professional review is required before business use. See our privacy policy.

What's inside

Seventeen pages, zero fluff

Every figure in this guide traces to a named source — the State Health Care Accreditation Agency, the National Health Insurance Fund, the Law on the Health System, the OECD and the European Observatory, and verified 2026 publications.

EU MDR + the Lithuanian layer

The Law on the Health System frames the national layer: the State Health Care Accreditation Agency (VASPVT) receives the national registration filings, supervises the device market and holds suspension, withdrawal and recall powers — while the National Health Insurance Fund (VLK) is the payer.

EUDAMED + the VASPVT registration

Manufacturers and devices register in EUDAMED — but the Lithuanian registration survives alongside it: the manufacturer filing to VASPVT, with technical data and conformity evidence, must be confirmed with registration numbers before placement, and Class IIa, IIb and III devices carry a further data submission within 14 working days of placing on the market.

Lithuanian language + the payer list

Instructions for use must be in Lithuanian; English is possible only for devices intended for healthcare professionals, with the recipient's consent, and safety notices are always Lithuanian. Outpatient payment runs through the VLK reimbursable medical aid device lists, whose price lists are updated twice a year.

Complete guide

The guide in full

Every section of the PDF on one page. Use the form above to keep a copy.

Overview

Last updated: November 2026. The EU Medical Device Regulation (MDR, Regulation (EU) 2017/745) and the In Vitro Diagnostic Medical Devices Regulation (IVDR, Regulation (EU) 2017/746) apply directly in Lithuania. The Lithuanian national layer sits mainly in the Law on the Health System of the Republic of Lithuania (Lietuvos Respublikos sveikatos sistemos Istatymas, in this guide the Law on the Health System), whose medical-device provisions — in particular the Article 59 series — set the national registration, notification, language, supervision and enforcement rules that sit on top of the two EU regulations, together with implementing orders of the Minister of Health and the practice of the competent authority. Around them sit the Law on Health Insurance, which frames reimbursement from the Compulsory Health Insurance Fund, the Law on Public Procurement, which frames hospital buying, and the Law on Advertising, which frames device promotion. Always confirm critical steps against current guidance from the State Health Care Accreditation Agency under the Ministry of Health (ValstybinI akreditavimo sveikatos priežiIros veiklai tarnyba prie Sveikatos apsaugos ministerijos, VASPVT) before filing.

Lithuania is the largest Baltic market. There were 2.9 million people living in Lithuania in 2024 (source: OECD, Economic Surveys: Lithuania 2025). Lithuania spends USD 4,259 per capita on health, equal to 7.6% of GDP, compared to an OECD average of 9.3% of GDP, and 99% of the population is covered for a core set of services (source: OECD, Health at a Glance 2025, Lithuania country note). In terms of financial coverage, 67% of spending in Lithuania was covered by mandatory prepayment, lower than the OECD average of 75% (source: OECD, Health at a Glance 2025, Lithuania country note). The State of Health in the EU Country Health Profile 2025 records total health expenditure of 7.3% of GDP in 2023, financed primarily through contributions to a mandatory health insurance scheme administered by the National Health Insurance Fund, with population coverage at about 99% of residents (source: OECD / European Observatory on Health Systems and Policies, Country Health Profile 2025: Lithuania). For a foreign manufacturer, Lithuania is usually the first Baltic step: the largest patient pool of the three states, a single national device authority, a single national payer, and a hospital sector anchored by Vilnius and Kaunas. Its trap is the mirror image of its attraction. Because the market reads as "EU, therefore CE mark, therefore done", entrants miss that Lithuania kept a real national registration step — with confirmation and registration numbers required before placement for the core manufacturer filing, and a separate 14-working-day data submission for the higher classes — plus a Lithuanian-language rule whose English exception is consent-based and narrower than many teams assume.

Regulatory reality check up front: Lithuania is not a separate certification regime from the EU. There is no "Lithuanian CE mark", and VASPVT does not approve medical devices in the sense of granting a medicines-style marketing authorisation. The principal requirements for placing medical devices on the Lithuanian market are set out in the Law on the Health System, and devices placed on the market must comply with the MDR, while in vitro diagnostic devices must comply with the IVDR, including labelling, quality and safety requirements (source: IBA Healthcare and Life Sciences Law Committee, Trade and Distribution of Therapeutic Products Survey — Lithuania, 2026 survey, authored by Ellex Valiunas). Regulatory oversight and market monitoring of medical devices are carried out by VASPVT (source: IBA survey, Lithuania, 2026). What Lithuania adds — and what foreign manufacturers most often underestimate — is a national layer with six specific features:

  1. 1. A national registration step that survives alongside EUDAMED — manufacturers of medical devices, and persons who assemble or sterilise procedure packs or systems and place them on the market in their own name, must submit documents to VASPVT with their registered office address, technical data and evidence of conformity of the devices with the MDR or the IVDR; where the manufacturer is not established in an EEA country, the authorised representative established in Lithuania must submit the relevant data, and confirmation of registration by the competent authority and assignment of registration numbers are required before placement on the market (source: IBA survey, Lithuania, 2026, citing the Law on the Health System).
  2. 2. A second, class-graded data duty with a hard clock — manufacturers, authorised representatives, importers and distributors of Class IIa, IIb and III medical devices and of custom-made active implantable devices must submit their registered office address and data on any devices placed on the Lithuanian market to VASPVT no later than 14 working days from the date of placing those devices on the market (source: IBA survey, Lithuania, 2026, citing the Law on the Health System). Fourteen working days is roughly three calendar weeks; a launch plan that treats the Lithuanian filing as an afterthought will miss it.
  3. 3. No separate national wholesale licence for devices — the wholesale distribution of medicinal products is a licensed pharmaceutical activity in Lithuania, but no separate national wholesale distribution licence is required for medical devices; placing devices on the market is subject to compliance with the MDR and the IVDR (source: IBA survey, Lithuania, 2026). The Lithuanian barrier is registration, language and payer access, not a distributor licence.
  4. 4. A Lithuanian-language rule with a consent-based English exception for professionals — manufacturers, authorised representatives and persons assembling or sterilising procedure kits or systems must provide the instructions for use and maintenance required under the MDR and the IVDR in Lithuanian, or in English if the device is intended for healthcare professionals, with the recipient's consent; local safety notices must be delivered to users in Lithuanian (source: IBA survey, Lithuania, 2026, citing the Law on the Health System). Patient-facing information has no English shortcut.
  5. 5. MDR vigilance clocks, enforced nationally — serious incidents are reported under Article 87 of the MDR (no later than 15 days as the general rule, 2 days for a serious public health threat, 10 days for death or an unanticipated serious deterioration), and VASPVT holds national inspection, suspension, withdrawal and recall powers, with administrative fines for device-law infringements of up to EUR 850, and up to EUR 1,500 for a repeated infringement, for managers of legal entities or other responsible persons (source: IBA survey, Lithuania, 2026, citing the Law on the Health System and the Code of Administrative Offences).
  6. 6. A reimbursement gate that is not run by the device authority — to be reimbursed from the Compulsory Health Insurance Fund budget, medical aid devices must be included in the Lists of Reimbursable Medical Aid Devices and in the Reimbursable Medical Aid Devices Price List, and the price lists are approved and updated twice a year; the Fund is managed by the National Health Insurance Fund under the Ministry of Health (ValstybinI ligoniI kasa, VLK), which prepares the price lists and makes key reimbursement decisions (source: IBA survey, Lithuania, 2026). Registration with VASPVT and listing with VLK are two different doors, opened by two different institutions, on two different clocks.

A note on institutions, because the names recur throughout this guide and are routinely confused. The Ministry of Health (Sveikatos apsaugos ministerija) holds health policy. VASPVT — full name, the State Accreditation Service for Health Care Activities under the Ministry of Health — is the device authority: registration and market supervision of medical devices, and, in its wider remit, licensing and accreditation of healthcare institutions and professionals. The State Medicines Control Agency (ValstybinI vaistI kontrolIs tarnyba, VVKT) is the medicines authority, not the device authority; a device filing sent to VVKT is a filing sent to the wrong agency. VLK, the National Health Insurance Fund, is the payer. The Public Procurement Office oversees procurement law, the Central Public Procurement Information System (CentrinI viešIjI pirkimI informacinI sistema, CVP IS) is where tenders live, and CPO LT is the central purchasing body whose electronic catalogue many institutions must or may buy through. This guide keeps those six roles separate throughout, because most Lithuanian market-access errors are, at root, a filing or a pitch sent to the wrong one of them.

A note on terminology. Lithuanian law uses the term medicinos priemonIs (medicinos prietaisai) — medical devices (medical appliances) — for the MDR and IVDR product universe, and a narrower term, medicinos pagalbos priemonIs (MPP) — medical aid devices — for the reimbursable outpatient category administered through the VLK lists. A device can be a perfectly lawful medicinos priemonI on the Lithuanian market and still not be a reimbursable MPP. Sections 1 to 9 of this guide are about the first status; Section 10 is about the second, and about hospital procurement, which is a third route again.

1. The Regulatory Framework: EU MDR + the Lithuanian National Layer

1.1 The two EU regulations apply directly — Lithuania does not re-run them

The MDR and the IVDR are EU regulations, not directives: they apply in Lithuania in their own terms, without a Lithuanian transposing statute recreating them. Classification, conformity assessment, clinical and performance evaluation, the EU declaration of conformity, CE marking, Unique Device Identification (UDI), economic-operator obligations and vigilance all run on the EU texts, read with the guidance of the Medical Device Coordination Group (MDCG). No Lithuanian authority repeats a notified body's conformity assessment, and no Lithuanian certificate replaces CE marking. A Lithuanian registration confirmation is an administrative registration of a device that is already lawfully CE marked — it is not a second conformity decision, and it must never be described to a board, a distributor or a customer as a Lithuanian "approval" of safety or performance.

The transition architecture is therefore the EU one: the extended MDR transitional provisions under Regulation (EU) 2023/607 for legacy devices, the staggered IVDR transition, and the phased mandatory use of EUDAMED modules under Regulation (EU) 2024/1860, which also introduced the obligation to inform in the event of an interruption or discontinuation of supply of certain devices — an obligation the Lithuanian survey expressly notes as applicable to Lithuanian supply chains, with economic operators expected to pass the information without undue delay to the operators, health institutions and professionals they directly supply (source: IBA survey, Lithuania, 2026). Any date in a launch plan that depends on a transition deadline or on whether a given EUDAMED module is mandatory at the filing date is a date to be verified at the time of filing, because both move.

1.2 The Lithuanian layer: the Law on the Health System and VASPVT

On top of the EU layer, the Law on the Health System carries the national device regime. Its device provisions do four things that matter commercially. First, they designate the supervision model: regulatory oversight and market monitoring of medical devices are carried out by VASPVT (source: IBA survey, Lithuania, 2026). Second, they impose the national registration and data-submission duties described in Sections 4 and 5 — the manufacturer filing with confirmation and registration numbers before placement, and the 14-working-day submission for the higher classes. Third, they impose the national language and traceability duties described in Section 6, including the duty, on the competent authority's request, to provide traceability information across the supply chain. Fourth, they give VASPVT its enforcement toolkit: access to all information, inspection of premises and collection of samples related to the manufacture, safety, supply, import, marketing and use of medical devices; on identifying infringements, VASPVT may temporarily suspend the placing of devices on the market or require their withdrawal or recall (source: IBA survey, Lithuania, 2026, citing the Law on the Health System).

Alongside the Law on the Health System sit: the Law on Health Insurance and the VLK list system for reimbursement (Section 10.2); the Law on Public Procurement, CVP IS and CPO LT for institutional buying (Section 10.1); the Law on Advertising, under which a device advertisement must not use the name, surname or image of a patient or refer to the recommendations of healthcare institutions, healthcare professionals or their professional organisations, and advertising that does not comply with the MDR or the IVDR is prohibited (source: IBA survey, Lithuania, 2026, citing Article 15 of the Law on Advertising); and the Code of Administrative Offences for fines. Clinical investigations add a further Lithuanian gate — the Lithuanian Bioethics Committee and VASPVT both feature in the national pathway for clinical investigations and performance studies (source: Create Lithuania / Invest Lithuania, Regulatory Pathways for Medicines, Medical Devices and Novel Foods, 2025) — the detail of which is study-specific and to be verified for any given protocol.

ElementEU levelLithuanian level
Legal basisMDR (EU) 2017/745; IVDR (EU) 2017/746Law on the Health System (Article 59 series); Minister of Health orders
Conformity assessmentManufacturer + notified body where the class requires itNot repeated in Lithuania
Device authorityEUDAMED; national competent authorities networkVASPVT — registration, market supervision, vigilance
Medicines authorityEuropean Medicines Agency networkVVKT — medicines only, not devices
Actor registrationEUDAMED actor module, Single Registration NumberLithuanian filing data held by VASPVT alongside EUDAMED
Device registrationEUDAMED UDI / device module as modules become mandatoryVASPVT registration with confirmation and registration numbers before placement (manufacturer filing)
Higher-class data dutyEUDAMED device dataRegistered office + device data to VASPVT within 14 working days of placing Class IIa, IIb, III and custom-made active implantable devices on the market
LanguageLanguage determined by the Member State (MDR Annex I)Lithuanian as the rule; English possible for healthcare-professional devices with recipient consent; safety notices in Lithuanian
PayerNone at EU levelVLK — Compulsory Health Insurance Fund; reimbursable medical aid device lists and price list, updated twice a year
ProcurementDirective 2014/24/EU principlesLaw on Public Procurement; CVP IS notices; CPO LT central catalogue

1.3 What VASPVT is — and what it is not

VASPVT was established in 1996 and, in its wider remit, accredits and licenses healthcare institutions and professionals; for devices, its functions are registration, market supervision and patient-safety oversight. Three practical consequences follow. First, VASPVT is a supervisor, not a promoter: it will test whether a filing is complete and whether a device on the market matches its documentation, but it does not help a manufacturer enter the market, and its registration confirmation is not an endorsement that can be quoted in marketing — the Law on Advertising independently bars recommendation-based device advertising. Second, VASPVT's device register and EUDAMED are parallel records: until EUDAMED registration is fully mandatory, Lithuanian-based manufacturers and authorised representatives must register devices with VASPVT and receive written confirmation before placing them on the market (source: Create Lithuania, Regulatory Pathways, 2025, TRL 8 summary). Third, because VASPVT also licenses the institutions that buy devices, its view of a supplier's compliance behaviour is formed across both files; a vigilance or labelling problem is never only a product problem.

1.4 The market in numbers

Lithuania's 2.9 million residents (source: OECD, Economic Surveys: Lithuania 2025, recording 2.9 million people in 2024) are served by a system spending 7.6% of GDP on health (source: OECD, Health at a Glance 2025). The Country Health Profile 2025 adds the structural facts a device plan needs: physician density was 4.6 per 1,000 population in 2023, the highest level since 2000 and above the EU average, while the system faces a looming shortage of nurses and a deferred national target for the nurse-to-doctor ratio (source: OECD / European Observatory, Country Health Profile 2025: Lithuania). Health system restructuring was initiated in 2023 (source: Country Health Profile 2025: Lithuania). Read together, the numbers describe a clinician-rich, budget-disciplined, centralised system: clinical adoption can move fast where a Vilnius or Kaunas centre champions a device, but every euro is contested, the payer's lists update on a fixed twice-yearly rhythm, and public money buys through formal procurement. No reliable, officially published single figure for the total Lithuanian medical device market size was identified for this guide; commercial market-size estimates vary widely by definition and are to be verified before any figure is used in an investment or board document.

2. Device Classification

2.1 The EU rules decide the class — Lithuania applies them unchanged

Medical devices are classified by risk in accordance with the rules in Annex VIII to the MDR into Classes I, IIa, IIb and III; in vitro diagnostic devices are classified in accordance with Annex VIII to the IVDR into Classes A, B, C and D (source: IBA survey, Lithuania, 2026). Lithuania adds no national class and no national reclassification. The class decided under the EU rules drives everything downstream in this guide: whether a notified body must be involved, whether the Lithuanian 14-working-day data duty is triggered, and how heavy the vigilance and post-market documentation set must be.

Class (MDR)Risk profileNotified bodyLithuanian consequence
Class ILowNo, except sterile (Is), measuring (Im) or reusable surgical (Ir) featuresManufacturer registration filing before placement; no 14-working-day higher-class duty
Class IIaMedium-lowYes14-working-day data submission to VASPVT after placing on the market
Class IIbMedium-highYes14-working-day data submission to VASPVT after placing on the market
Class IIIHighYes, with the heaviest scrutiny14-working-day data submission; implant-card and summary of safety and clinical performance duties under the MDR
Custom-made active implantablePatient-specific, highRoute-specificExpressly inside the 14-working-day duty

Class (IVDR) Risk profile Notified body Note Class A Low No, except sterile Class A Entry class for many laboratory products Class B Moderate Yes Default class for many IVDs

Class (IVDR)Risk profileNotified bodyNote
Class CHigherYesIncludes many near-patient and companion-diagnostic-a djacent tests
Class DHighestYes, with EU reference laboratory involvement where applicableHighest-scrutiny IVD class

The notified body number — a four-digit number — must be indicated next to the CE marking where a notified body was involved, including for Class IIa, IIb and III devices, Class B, C and D IVDs, Class I sterile or measuring devices and Class A sterile IVDs (source: IBA survey, Lithuania, 2026). A Lithuanian distributor or hospital tender officer checking a label will look for exactly that number; a CE mark without it, on a class that requires a body, is an immediate credibility failure in a tender evaluation.

2.2 Classification errors that matter in Lithuania

Three classification mistakes recur in Baltic launches. The first is software: software that drives, influences or informs a medical decision is a device in its own right under MDR Rule 11, and its class is set by the significance of the information and the seriousness of the condition — teams that file patient-facing software as Class I wellness product discover the error at the VASPVT registration step, when the technical data must evidence conformity for the class actually claimed. The second is accessories and procedure packs: a person who assembles or sterilises procedure packs or systems and places them on the market in their own name carries their own Lithuanian filing duty (Section 4), so a kitting operation in Vilnius is a regulated actor, not a logistics detail. The third is the Annex XVI borderline: products without an intended medical purpose that the MDR nonetheless regulates — the Lithuanian implementing discussion in 2019 already flagged examples such as non-corrective contact lenses and cosmetic-surgery equipment being drawn into the device regime (source: Sorainen, Life Sciences and Healthcare sector news in Lithuania, January 2020, reporting the Ministry of Health and VASPVT approval of renewed medical equipment regulations based on the MDR and the IVDR). Borderline determinations for a specific product are to be verified with VASPVT before the Lithuanian plan is fixed.

3. The Authorised Representative, Importer and Distributor in Lithuania

3.1 The EU chain applies — with Lithuanian filing points inside it

The MDR and the IVDR allocate obligations by role: the manufacturer (Article 10 of the MDR), the authorised representative (Article 11), the importer (Article 13) and the distributor (Article 14). Lithuania does not rewrite those articles; it attaches national filing and cooperation duties to the same roles. Importers must be established in the EU and may only place devices on the EU market that comply with the MDR or the IVDR, including the obligations on conformity, labelling and documentation, traceability, storage and transport conditions, post-market vigilance, cooperation with authorities and corrective actions set out in Article 13 (source: IBA survey, Lithuania, 2026). Distributors carry the parallel Article 14 verification and traceability duties. None of these roles needs a Lithuanian wholesale licence for devices — and none of them is excused from the Lithuanian registration data duties because of that.

3.2 The authorised representative: the Lithuanian filing falls to the AR when the manufacturer is outside the EEA

Where the manufacturer is not established in an EEA country, the authorised representative established in Lithuania must submit the relevant registration data to VASPVT (source: IBA survey, Lithuania, 2026). Read that sentence carefully before signing an AR agreement for a Baltic launch. A generic EU authorised representative established in another Member State satisfies the MDR mandate, but the Lithuanian text attaches the filing to an authorised representative established in Lithuania for the non-EEA manufacturer case. Whether, for a specific structure, a non-Lithuanian EEA authorised representative's filing is accepted in practice, or a Lithuanian-established representative is required, is a point to be verified with VASPVT before the launch structure is locked — it decides contracts, not just paperwork. The AR's wider MDR duties are unchanged: verify the EU declaration of conformity and technical documentation, keep documentation available for VASPVT, cooperate on preventive and corrective actions, and carry joint liability exposure under Article 11(5) of the MDR where the manufacturer defaults.

3.3 The importer of record

The importer is the EU-established actor who places a third-country device on the Union market. In a Lithuanian launch the importer of record must be identified by name in the plan, because the importer carries: verification that the device is CE marked, that the declaration of conformity was drawn up, that the manufacturer and any authorised representative are identified, and that the device is registered in EUDAMED where required; the duty to indicate its own name and address on the device, its packaging or an accompanying document; storage and transport conditions that do not jeopardise conformity; complaint and vigilance forwarding; and, for Lithuania specifically, a share of the 14-working-day data duty for the higher classes (Section 4.3) and the traceability-on-request duty (Section 6.3). A distributor who in fact first places a third-country device on the EU market is an importer in law, whatever the distribution agreement calls them — the single most common role error in Baltic structures.

3.4 The distributor: no licence, real duties

Lithuanian distributors of devices need no separate national wholesale distribution licence (source: IBA survey, Lithuania, 2026). What they carry instead is a duty set that a serious distributor prices into the margin: verify CE marking, declaration of conformity, labelling and Lithuanian-language information before making a device available; ensure storage and transport conditions preserve conformity; maintain traceability and technical documentation and promptly address non-compliance; and cooperate with the competent authority, including enabling VASPVT to inspect storage premises (source: IBA survey, Lithuania, 2026, citing the Law on the Health System). Sales to end users are not a separately licensed activity under Lithuanian law, and remote (online) sales of medical devices are permitted without a separate licence or authorisation for the website or the seller, provided the MDR and IVDR requirements are met (source: IBA survey, Lithuania, 2026). Reimbursable medical aid devices, however, are issued to residents by pharmacies or other designated persons, and remote sales of reimbursable medical aid devices dispensed on electronic prescriptions follow the remote-sales framework for medicinal products under Minister of Health Order No. V-1491 (source: IBA survey, Lithuania, 2026) — a consumer webshop plan and a reimbursable-device plan are therefore different builds.

Two further distribution rules shape channel design. Parallel import of devices needs no separate national parallel-import licence, but notification is required, and an importer or distributor that relabels or repackages must inform the manufacturer and the competent authority in advance, under Article 16(4) of the MDR and the IVDR (source: IBA survey, Lithuania, 2026). And foreign suppliers may not treat Lithuania as a drop-ship market: any entity placing devices on the EU market must be established in the EU and ensure full MDR and IVDR compliance (source: IBA survey, Lithuania, 2026).

4. The Registration Architecture: EUDAMED + the Lithuanian Registration

4.1 Two records, one device

A device launched in Lithuania lives in two records. The first is EUDAMED, the EU database: actor registration and the Single Registration Number (SRN), UDI and device registration, notified-body and certificate data, and, as modules become mandatory, vigilance and post-market reporting. The second is the Lithuanian record held by VASPVT under the Law on the Health System. Neither record substitutes for the other while the national regime stands: EUDAMED registration does not discharge the Lithuanian filing, and the Lithuanian confirmation does not discharge EUDAMED. Until EUDAMED is mandatory, Lithuanian-based manufacturers and authorised representatives must register devices with VASPVT and receive written confirmation before placing them on the market (source: Create Lithuania, Regulatory Pathways, 2025). The mandatory-use position of each EUDAMED module at the actual filing date is to be verified at that date.

4.2 The manufacturer filing: confirmation and registration numbers before placement

The core Lithuanian rule, stated plainly: manufacturers of medical devices — and persons who assemble or sterilise procedure packs or systems and place them on the market in their own name — must submit documents to VASPVT showing their registered office address, technical data, and evidence of conformity of the devices with the MDR or the IVDR; where the manufacturer is not established in an EEA country, the authorised representative established in Lithuania submits the relevant data; and confirmation of registration by the competent authority and assignment of registration numbers are required before placement on the market (source: IBA survey, Lithuania, 2026, citing the Law on the Health System). Three consequences. First, sequencing: the Lithuanian confirmation sits on the critical path before first sale, not after it — build it into the launch Gantt as a gate, with the conformity evidence complete before the filing is made. Second, scope: the filing is evidence-based (technical data plus conformity evidence), not a bare name-and-address notice, so the technical documentation set must be filing-ready, not merely audit-ready. Third, the current form, submission channel and document checklist for this filing at the date of filing are to be verified directly with VASPVT, because forms and channels are updated by implementing practice rather than by the statute itself.

4.3 The 14-working-day duty for the higher classes

Separately, manufacturers, authorised representatives, importers and distributors of Class IIa, IIb and III devices and of custom-made active implantable devices must submit their registered office address and data on devices placed on the Lithuanian market to VASPVT no later than 14 working days from the date of placing those devices on the market (source: IBA survey, Lithuania, 2026, citing the Law on the Health System). Note the breadth: the duty names all four actor roles, so a contract must assign who files, for which devices, within the window — silence in the distribution agreement means four actors each assuming another filed. Note also the direction of the clock: this duty runs from placement, while the Section 4.2 confirmation must exist before placement. A compliant Lithuanian launch therefore has two dated artefacts — a pre-market registration confirmation with registration numbers, and a post-placement data submission inside 14 working days for the higher classes — and a file that can produce both, with dates, on VASPVT's request.

Secondary commercial summaries of the Lithuanian regime publish differing registration and notification durations. No VASPVT-published processing duration for the registration confirmation was verified for this guide; any specific number of days for VASPVT to issue the confirmation is to be verified with the Agency at the time of filing, and no launch date should be promised against an unverified duration (see Section 7).

4.4 Systems, procedure packs and custom-made devices

Persons who assemble or sterilise procedure packs or systems and place them on the market in their own name are filers in their own right under the manufacturer filing (source: IBA survey, Lithuania, 2026). Custom-made active implantable devices are expressly inside the 14-working-day duty. For custom-made devices more broadly, and for devices manufactured and used within health institutions under Article 5(5) of the MDR, the Lithuanian national position and any additional VASPVT notification practice at the filing date are to be verified for the specific product, because the statute-level sources used for this guide do not settle every category line.

5. Step-by-Step: Placing a Medical Device on the Lithuanian Market

  • Step 1 — Fix the qualification and the class. Confirm the product is a medical device (or an Annex XVI product, or an IVD), fix the MDR or IVDR class under Annex VIII, and record the classification rationale. Every Lithuanian duty in this guide keys off that decision.
  • Step 2 — Complete the EU conformity route. Technical documentation, clinical or performance evaluation, risk management, post-market surveillance planning and, for the class, notified-body assessment; draw up the EU declaration of conformity and affix CE marking, with the notified body number where a body was involved. Nothing in this step is Lithuania-specific, and no Lithuanian authority re-runs it.
  • Step 3 — Build the actor chain. Non-EEA manufacturer: appoint an authorised representative by written mandate, and resolve the Lithuanian filing point in Section 3.2. Identify the importer of record for third-country devices. List every actor who will place or distribute the device in Lithuania — that list drives both Lithuanian filings and the vigilance routing.
  • Step 4 — Register the actors in EUDAMED. Manufacturer, authorised representative and importer obtain SRNs, with validation by the competent authority of each actor's place of establishment. Do not schedule a launch date that assumes same-day validation.
  • Step 5 — Register the devices in EUDAMED. Complete Basic UDI-DI and device data as the applicable module duties require at the filing date, reconciled against the declaration of conformity and the labels.
  • Step 6 — Prepare the Lithuanian documentation set. Lithuanian labelling and instructions as the rule; for healthcare-professional devices, the documented recipient consent if English instructions are to be used (Section 6). Implant cards and patient information in Lithuanian where the device carries them. Safety-notice templates in Lithuanian, ready before they are needed.
  • Step 7 — File the Lithuanian manufacturer registration with VASPVT before placing on the market. Registered office address, technical data and evidence of conformity; obtain the registration confirmation and the registration numbers. No first sale, no first tender delivery and no first hospital evaluation unit before this gate is passed.
  • Step 8 — Run the 14-working-day clock for the higher classes. On placing Class IIa, IIb or III devices or custom-made active implantable devices on the Lithuanian market, submit the registered office address and device data to VASPVT within 14 working days, through the actor the contracts designate as filer, and file the proof of submission with the launch record.
  • Step 9 — Open the payer and procurement routes in parallel, not after. If the device is an outpatient medical aid device, prepare the VLK list application against the twice-yearly price-list rhythm (Section 10.2). If it is a hospital product, map the CVP IS notices and the CPO LT catalogue modules for the category (Section 10.1) and identify the Vilnius and Kaunas buyers whose specifications will decide the category.
  • Step 10 — Switch on post-market operations before the first unit ships. Complaint intake, vigilance assessment against the MDR clocks, field safety corrective action procedure with Lithuanian-language notices, traceability records that can answer a VASPVT request, and the supply-interruption information duty under Regulation (EU) 2024/1860.

6. Language, Labelling and Information Supplied with the Device

6.1 The rule: Lithuanian — with a consent-based English lane for professionals only

Manufacturers, authorised representatives and persons assembling or sterilising procedure kits or systems must provide the instructions for use and maintenance required under the MDR and the IVDR in Lithuanian — or in English if the device is intended for healthcare professionals, with the recipient's consent — must deliver local safety notices to users in Lithuanian, and must, on the competent authority's request, provide traceability information across the supply chain (source: IBA survey, Lithuania, 2026, citing the Law on the Health System). Importers and distributors must supply all manufacturer-provided instructions for use and maintenance in Lithuanian, or in English for healthcare professionals with consent, and provide traceability information to the competent authority on request (source: IBA survey, Lithuania, 2026).

Three operational readings. First, the exception is conditional twice over: professional-only intended users, and the recipient's consent. A hospital's consent is an artefact to obtain and file — ideally in the supply or tender documentation — not an assumption to make because the staff "speak English". Second, the exception does not travel to the patient: information for patients and lay users, implant cards and patient instructions are Lithuanian work. Third, safety notices have no exception at all in the surveyed text: local safety notices go to users in Lithuanian. A field safety notice process that must pause for translation is a process designed to miss its clock; translate the templates at launch, not at the incident.

Secondary summaries sometimes state the Lithuanian rule more absolutely — labelling information in Lithuanian regardless of the user's competence. The statute-based survey position in Section 6.1 is the more precise one, but the gap between the two readings is exactly why the language plan for a specific product — label, instructions, software interface, declaration documents — is to be verified against the current VASPVT practice for that product type before print and release decisions are locked.

6.2 Labelling content

Beyond language, the label and information set must carry the MDR Annex I content: manufacturer and, where applicable, authorised representative and importer identification; UDI carrier; the CE marking with the notified body number where a body was involved (Section 2.1); sterile, single-use, custom-made and other applicable indications; and the instructions for use, which may be omitted for Class I and Class IIa devices only where the device can be used safely without them under the MDR conditions. Electronic instructions for use are governed by the EU eIFU regulation and its conditions; whether a given Lithuanian buyer or category accepts electronic-only supply in practice is a tender-by-tender fact, not a launch assumption.

6.3 Traceability on request

The Lithuanian layer makes traceability an on-demand duty to the authority, layered over the MDR's UDI and supply-chain identification system: on VASPVT's request, the manufacturer-side actors and the importers and distributors must provide traceability information across the supply chain (source: IBA survey, Lithuania, 2026). Practically, each actor must be able to answer, quickly and from records rather than reconstruction: which devices, which UDI-DIs and lots, were supplied to which Lithuanian customers, and from which upstream actor they were received. A distributor who cannot answer that question from its own system is not ready to hold a Lithuanian higher-class line, whatever its sales coverage.

7. Timelines

7.1 The clocks that are actually fixed

Lithuania's device timeline is a stack of clocks from three different legislators, and a credible launch plan separates them instead of quoting one blended "registration time".

ClockDurationSource
Registration confirmation + registration numbersRequired before placement on the market (manufacturer filing)Law on the Health System, via IBA survey 2026
Higher-class data submission (Class IIa, IIb, III; custom-made active implantable)No later than 14 working days from placing on the marketLaw on the Health System, via IBA survey 2026
Serious incident report (general)No later than 15 days after becoming awareMDR Article 87
Serious incident — death or unanticipated serious deteriorationNo later than 10 daysMDR Article 87
Serious public health threatNo later than 2 daysMDR Article 87
Reimbursable price lists (medicines and medical aid devices)Approved / updated twice a yearIBA survey 2026, on the VLK list system
Remote-sales and other medicines-side notification clocksNot applicable to devices as suchSee Section 3.4 — device remote sales need no separate licence

7.2 The clocks that are not fixed in a verified source

No VASPVT-published standard processing time for the Section 4.2 registration confirmation was verified from an official source for this guide, and commercial secondary sources publish figures that differ from each other. The honest planning treatment is: file early, treat the confirmation as a gate of unverified duration, and confirm the expected duration with VASPVT when the filing is made — the duration is to be verified at that point. The same applies to EUDAMED actor-validation durations, which vary by validating authority and workload, and to VLK list-assessment durations for a specific medical aid device application, which depend on the application cycle and the evidence requested. Notified-body conformity assessment remains, for most higher-class entrants, the longest single block in the whole timeline, and it is an EU-level block: Lithuania neither shortens nor lengthens it.

7.3 A realistic sequence

For a CE-marked higher-class device with a complete technical file, the Lithuanian increment over a generic EU launch is: actor-chain and AR structuring (including the Section 3.2 question), Lithuanian language set and consent artefacts, the pre-market VASPVT registration filing and its confirmation, and the launch-day start of the 14-working-day submission clock — run in parallel with, not after, the VLK list application (twice-yearly rhythm) or the first CVP IS / CPO LT procurement moves. For a device still in conformity assessment, the Lithuanian increment is small compared with the notified-body block, and the correct sequence is to prepare the Lithuanian filings during the assessment so the confirmation gate, not the paperwork, is the only Lithuanian wait at the end.

8. Fees and Costs

8.1 Official fees: what could — and could not — be verified

This guide does not publish a Lithuanian device registration fee, because no official fee for the VASPVT medical-device registration or data submission was verified from a primary source during its preparation. The searches run for this guide covered the Lithuanian state fee (valstybIs rinkliava) framework — Government Resolution No. 1458 of 15 December 2000 approving the list of specific state fee amounts, whose consolidated text remains in force — and VASPVT's published service information, without locating a device-registration line item that could be quoted responsibly with an amount. Two errors are therefore equally barred: quoting a euro amount copied from a commercial registration vendor (such figures circulate, they differ, and none traced to the Resolution No. 1458 list for the device procedure), and stating that the Lithuanian registration is free (the absence of a located fee line is not evidence that no fee, or no paid-service charge for a related procedure, applies). The correct statement, and the one to carry into a budget, is: any state fee or paid-service charge for the specific VASPVT procedure planned is to be verified with VASPVT and against the current Resolution No. 1458 list at the time of filing. Clinical-investigation-related charges, certificates and any expert-assessment fees are likewise to be verified procedure by procedure.

8.2 The costs that actually dominate a Lithuanian launch

Cost blockNatureDriver
Notified body conformity assessmentEU-level feeClass, device complexity, body capacity
Authorised representativeCommercial mandate feeWhether a Lithuania-established representative is used (Section 3.2)
Lithuanian translationCommercialInstructions, labels, implant cards, safety-notice templates, tender documents
Regulatory filing labourInternal / adviserTwo Lithuanian filings (pre-market registration; 14-working-day submission) plus EUDAMED upkeep
VLK list applicationInternal / adviser + evidenceHealth-technology and budget-impact evidence for a medical aid device listing
Procurement participationInternal / adviserCVP IS monitoring, CPO LT catalogue qualification, tender documentation in Lithuanian
Post-market operationRecurringVigilance, complaints, traceability systems, supply-interruption duties

The pattern matches the other Baltic states: the state's own charges, whatever their verified level, are a minor line. The money sits in conformity assessment, in the representative and translation structure, in payer evidence, and in the recurring cost of being a well-documented supplier in a procurement-driven system.

9. Vigilance, Post-Market Surveillance and Incident Reporting

9.1 The EU system, with VASPVT as the national supervisor

Post-market surveillance (MDR Articles 83 to 86), vigilance (Articles 87 to 92) and market surveillance (Articles 93 onwards) run on the EU texts in Lithuania, with VASPVT as the authority that supervises the Lithuanian market. The manufacturer's clocks are the MDR clocks: report a serious incident immediately and no later than 15 days after becoming aware of it as the general rule; no later than 2 days for a serious public health threat; no later than 10 days where the incident involves death or an unanticipated serious deterioration in health (source: MDR Article 87). Field safety corrective actions are reported before implementation unless urgency requires otherwise, trend reporting applies to statistically significant increases in non-serious incidents or expected side effects (Article 88), and periodic safety update reports follow the class-based rhythm of Article 86. Regulation (EU) 2024/1860 adds the supply-interruption information duty noted in Sections 1.1 and 5.

9.2 The Lithuanian enforcement reality behind the clocks

VASPVT may access all information, inspect premises and collect samples related to the manufacture, safety, supply, import, marketing and use of medical devices; on identifying infringements it may temporarily suspend the placing of devices on the market, or require withdrawal or recall (source: IBA survey, Lithuania, 2026, citing the Law on the Health System). Under the Code of Administrative Offences, non-compliance with medical-device laws may result in fines of up to EUR 850 for managers of legal entities or other responsible persons, and up to EUR 1,500 for a repeated infringement (source: IBA survey, Lithuania, 2026). The fine levels are modest; the suspension, withdrawal and recall powers are not. A Lithuanian vigilance failure is therefore priced in market access, not in fines — a suspended line loses tenders and catalogue positions that a fine never touches.

9.3 What the distributor and importer must do locally

Importers and distributors must ensure appropriate storage, maintain traceability and technical documentation, promptly address non-compliance and cooperate with the competent authorities, including enabling VASPVT to inspect storage premises (source: IBA survey, Lithuania, 2026, citing the Law on the Health System). In practice the Lithuanian partner is the manufacturer's early-warning system: complaints and incidents surface at the hospital, in Lithuanian, during Lithuanian working hours. The distribution agreement must therefore carry a same-day forwarding duty for potential serious incidents, a Lithuanian-language intake route, and a prohibition on the local partner assessing reportability alone. Users and healthcare institutions feed the same system from their side; local safety notices, when they come, go to users in Lithuanian (Section 6.1).

10. Market Access in Practice: Procurement and Reimbursement

10.1 Hospital and institutional procurement: CVP IS and CPO LT

Public buyers in Lithuania procure under the Law on Public Procurement, and notices are published through the Central Public Procurement Information System, CVP IS (CentrinI viešIjI pirkimI informacinI sistema, at viesiejipirkimai.lt). Above the EU thresholds, notices also appear in Tenders Electronic Daily. The practical system has three layers a device supplier must work simultaneously. The first is the individual hospital tender: specifications written by the Vilnius and Kaunas university hospitals and the larger regional hospitals set the category standard, and a specification written around a competitor's parameter set is the commonest reason a compliant, CE-marked device loses without ever being evaluated on merit — specification monitoring on CVP IS, before notices close, is a sales function, not an administrative one. The second layer is CPO LT, the central purchasing body: suppliers qualify into its electronic catalogue through procurement procedures run as dynamic purchasing systems, with the procedure announced on CVP IS and the resulting catalogue (katalogas.cpo.lt) letting contracting authorities — including institutions of the Lithuanian national health system — order against standardised technical specifications, with the lowest-priced compliant offer winning the order in the catalogue model (sources: CPO LT, supplier instruction documents and participant guidance, cpo.lt and katalogas.cpo.lt). Health-sector modules in the catalogue have included categories such as sterilisation products and medical and other gases (source: CPO LT catalogue supplier instructions). Whether a given device category is bought through a CPO LT module, and the current qualification window for that module, are to be verified on CVP IS and with CPO LT at the time of planning. The third layer is the supplier's own tender-readiness: Lithuanian-language tender documentation, the registration confirmation and registration numbers from Section 4.2 ready to exhibit, and delivery and service capacity inside Lithuania.

10.2 Outpatient reimbursement: the VLK medical aid device lists

The outpatient payer route is separate from both registration and hospital procurement. Medical aid devices (medicinos pagalbos priemonIs, MPP) are reimbursed from the Compulsory Health Insurance Fund (Privalomojo sveikatos draudimo fondas) budget only if included in the Lists of Reimbursable Medical Aid Devices and in the Reimbursable Medical Aid Devices Price List; price lists are approved and updated twice a year (source: IBA survey, Lithuania, 2026). The Fund is managed by the National Health Insurance Fund, VLK, which prepares the price lists, concludes agreements and makes key decisions on reimbursement of medical aid devices (source: IBA survey, Lithuania, 2026). Only insured patients are entitled, devices are prescribed and issued through the established channels — reimbursable medical aid devices are issued to residents by pharmacies or other designated persons (source: IBA survey, Lithuania, 2026) — and VLK's public patient guidance confirms that only devices entered in the reimbursement lists are compensated from the Fund (source: VLK, patient guidance on reimbursable medicines and medical aid devices, ligoniukasa.lrv.lt).

Pricing inside the list system is regulated across the supply chain: base prices and patient co-payments are set under government-approved criteria and grouping rules; reimbursable products are supplied within capped wholesale and retail mark-ups set by the Minister of Health, with VAT applied; and non-reimbursable medical aids, by contrast, are sold at market price (source: IBA survey, Lithuania, 2026, on the pricing provisions). Lithuania also applies external reference pricing principles in its medicines system — prices set by reference to prices in other EU countries (source: Create Lithuania, How to apply for reimbursement in Lithuania, 2026) — so a Lithuanian list price should be modelled as a regional price decision, and the precise reference method, base-price group and co-payment level for a specific medical aid device are to be verified with VLK for the application cycle in question. Suppliers of reimbursed medical aid devices must ensure continuous supply (source: IBA survey, Lithuania, 2026): a listing is a supply commitment, and a stock-out is a compliance event, not only a commercial one.

10.3 The payer landscape in one paragraph

One fund, one manager, two doors. The Compulsory Health Insurance Fund, managed by VLK, pays for services provided by contracted public and private providers and for listed medicines and medical aid devices in the outpatient channel, while devices consumed inside hospital care are bought by the provider through Section 10.1 procurement and paid inside the service price. A launch plan that files a hospital consumable into the outpatient list, or an outpatient medical aid device into a tender-only plan, has chosen the wrong door — and in Lithuania the wrong door costs a full list cycle (the price lists move twice a year) or a full tender cycle, not a week.

11. Common Mistakes

  • Mistake 1 — Treating the CE mark as the end of the Lithuanian work. The Lithuanian registration confirmation and registration numbers are required before placement for the manufacturer filing (Section 4.2). A first sale made before that confirmation exists is a launch built on a missing gate.
  • Mistake 2 — Missing the 14-working-day clock. The higher-class data submission runs from the date of placing on the market, and it names manufacturers, authorised representatives, importers and distributors. If the contracts do not designate the filer, the window closes while four actors assume another filed.
  • Mistake 3 — Sending the device file to the medicines agency. VVKT regulates medicines. Devices belong to VASPVT. The error is common because both agencies sit under the same ministry and both are called "the agency" in casual Baltic advice.
  • Mistake 4 — Assuming a non-Lithuanian EU authorised representative settles the Lithuanian filing. The surveyed Lithuanian text attaches the filing, for a non-EEA manufacturer, to the authorised representative established in Lithuania. Resolve the point with VASPVT before the structure is signed (Section 3.2).
  • Mistake 5 — Using the professional English exception without consent, or for patients. English instructions require a healthcare-professional device and the recipient's consent, filed as an artefact. Patient information and safety notices are Lithuanian (Section 6.1).
  • Mistake 6 — Believing a distributor needs a Lithuanian wholesale licence — or that needing none means needing nothing. There is no device wholesale licence; there are registration data duties, storage and traceability duties, and inspection cooperation duties (Sections 3.4 and 9.3).
  • Mistake 7 — Quoting an unverified registration fee or duration. No official VASPVT device-registration fee amount or standard processing duration was verified for this guide (Sections 7.2 and 8.1). Budgets and launch promises built on vendor-blog figures are built on nothing.
  • Mistake 8 — Confusing registration with reimbursement. A registered medicinos priemonI is not thereby a reimbursable medicinos pagalbos priemonI. The VLK lists and their twice-yearly price-list rhythm are a separate application with its own evidence (Section 10.2).
  • Mistake 9 — Watching tenders but not specifications or the catalogue. In Lithuania the category is decided in CVP IS specifications and in CPO LT catalogue modules before the order is placed. Monitoring only published award notices is monitoring the past.
  • Mistake 10 — Advertising with clinician or patient endorsements. Lithuanian advertising law bars device advertisements that use a patient's name or image or refer to recommendations of healthcare institutions, professionals or their organisations (Section 1.2). Standard global campaign material routinely breaches this.

12. Frequently Asked Questions

Is there a separate Lithuanian approval for medical devices?

No. Conformity assessment and CE marking run under the MDR and the IVDR at EU level, and no Lithuanian authority repeats them. Lithuania adds a national registration layer: a manufacturer filing to VASPVT with registered office address, technical data and conformity evidence, with registration confirmation and registration numbers required before placement on the market (source: IBA survey, Lithuania, 2026).

Who is the Lithuanian authority for medical devices?

VASPVT — the State Health Care Accreditation Agency (State Accreditation Service for Health Care Activities) under the Ministry of Health. It handles device registration and market supervision. The State Medicines Control Agency (VVKT) is the medicines authority, and VLK is the payer; neither registers devices.

Which devices trigger the 14-working-day submission?

Class IIa, IIb and III medical devices and custom-made active implantable devices. Manufacturers, authorised representatives, importers and distributors of those devices must submit their registered office address and data on devices placed on the Lithuanian market to VASPVT no later than 14 working days from the date of placing them on the market (source: IBA survey, Lithuania, 2026).

Does a distributor need a Lithuanian licence for medical devices?

No separate national wholesale distribution licence is required for medical devices (source: IBA survey, Lithuania, 2026). Distributors still carry MDR verification, storage, traceability and cooperation duties, and share the 14-working-day data duty for the higher classes.

Can instructions for use be in English?

Only in the defined lane: the device must be intended for healthcare professionals and the recipient must consent. Otherwise the instructions for use and maintenance must be in Lithuanian, and local safety notices must be delivered in Lithuanian in all cases (source: IBA survey, Lithuania, 2026). Patient-facing information should be planned in Lithuanian without exception.

What does the Lithuanian registration cost?

No official fee amount for the VASPVT device registration was verified from a primary source for this guide, and this guide does not quote one. Verify any state fee or paid-service charge for the specific procedure with VASPVT and against the current state fee list (Government Resolution No. 1458) at the time of filing — see Section 8.1. The dominant launch costs are conformity assessment, representation, translation, payer evidence and procurement participation.

How long does Lithuanian registration take?

The verified fixed points are the sequencing rules, not a processing duration: confirmation before placement, and the 14-working-day submission after placement for the higher classes. No VASPVT-published standard processing duration was verified; confirm the expected duration with the Agency when filing (Section 7.2).

How are medical devices reimbursed in Lithuania?

Hospital devices are bought by providers through public procurement (CVP IS notices; CPO LT central catalogue) and paid inside service prices. Outpatient medical aid devices are reimbursed from the Compulsory Health Insurance Fund only if included in the Lists of Reimbursable Medical Aid Devices and the Reimbursable Medical Aid Devices Price List, which are updated twice a year and prepared by VLK (source: IBA survey, Lithuania, 2026).

Can medical devices be sold online in Lithuania?

Yes. Remote (online) sales of medical devices are permitted without a separate licence or authorisation for the website or the seller, provided the MDR and IVDR requirements are met. Reimbursable medical aid devices dispensed on electronic prescriptions follow the remote-sales framework of Minister of Health Order No. V-1491 (source: IBA survey, Lithuania, 2026).

What happens if a device breaks the rules in Lithuania?

VASPVT can inspect premises, access information and collect samples, and may temporarily suspend placing on the market or require withdrawal or recall. Administrative fines for device-law infringements run up to EUR 850 for responsible persons, and up to EUR 1,500 for a repeated infringement (source: IBA survey, Lithuania, 2026).

Does EUDAMED registration replace the Lithuanian filing?

No. While the national regime stands, the Lithuanian VASPVT registration and the EUDAMED records are parallel duties. Until EUDAMED is mandatory, Lithuanian-based manufacturers and authorised representatives must register with VASPVT and receive written confirmation before placing devices on the market (source: Create Lithuania, Regulatory Pathways, 2025).

Sources & scope

  • vaspvt.gov.lt, State Health Care Accreditation Agency under the Ministry of Health (VASPVT) — medical device registration, market supervision and vigilance under the Law on the Health System.
  • ligoniukasa.lrv.lt, National Health Insurance Fund of Lithuania (VLK) — Compulsory Health Insurance Fund and the reimbursable medical aid device lists and price list.
  • eur-lex.europa.eu, Regulation (EU) 2017/745 (MDR), Regulation (EU) 2023/607.

This guide is market-access intelligence, not regulatory or legal advice. State Health Care Accreditation Agency and National Health Insurance Fund practice, list cycles and any state fee amounts change — professional review is required before business use.

Market Access Strategy Session

Lithuania on your roadmap? Start with the full picture.

A Market Access Strategy Session maps your Lithuania route — EUDAMED registration, the VASPVT registration and 14-working-day submission, Lithuanian labelling, and VLK list and CVP IS / CPO LT procurement access.

Professional review is required before business use.