4.1 Two national registers, one EU database
Malta's architecture has three layers, in order. Layer one is EUDAMED: actor registration and the SRN, and device / UDI registration as the modules become mandatory — the first four modules becoming mandatory as from 28 May 2026 on the timetable notified by the MMA (source: MMA notice MMA/MDPCED-001/2026). Layer two is the MMA organisation register (MT-MDF02 / GL-MDF02), producing a unique organisation registration number (source: MMA, GL-MDF02/05, February 2026). Layer three is the MMA device notification (MT-MDF05 / GL-MDF05), producing an MMA reference number per device, quoted in later revision and revocation applications (source: MMA, GL-MDF05/06, February 2026). Malta-based manufacturers have a further route: MT-MDF03, the application for device registration to place devices on the EU market, producing a Medical Device Registration Letter from the MMA (source: servizz.gov.mt, MT-MDF03 service page). Do not confuse that EU-market route (EUR 100 per declaration of conformity under Schedule 5) with the MT-MDF05 local-market notification.
4.2 Organisation registration (MT-MDF02) in detail
The organisation registration exists, in the MMA's words, "to increase functionality at all regulatory levels and to allow courteous communication between distributors, importers, manufacturers, authorised representatives, notified bodies and the National Competent Authority" (source: MMA, GL-MDF02/05, February 2026).
Applicants are manufacturers, authorised representatives, importers, distributors and notified bodies, first-time or amending. The mechanics: the form must be completed electronically — handwritten forms will not be accepted; a signed scanned copy with supporting documentation is submitted by email to mdforms.medicinesauthority@gov.mt; for a manufacturer, authorised representative or importer, the EUDAMED SRN is inputted; payment follows GL-MDF07, selecting standard or fast-track service; and on successful review a unique organisation registration number is forwarded to the applicant (source: MMA, GL-MDF02/05, February 2026). The government service page confirms standard service 30 days, fast track 10 days (source: servizz.gov.mt, MT-MDF02, updated 16 June 2026). Until that number exists, the device notification file is incomplete, because MT-MDF05 asks for the organisation registration number (source: MMA, GL-MDF05/06, February 2026). Sequence in Malta is built into the forms.
4.3 Device notification (MT-MDF05) in detail
The notification form may be submitted by economic operators with a registered place of business in Malta and by operators with a registered place of business outside Malta but with no local supplier in Malta; a Malta-based applicant must be registered with the MMA (source: MMA, GL-MDF05/06, February 2026). That second limb matters for direct models: a foreign manufacturer with no Maltese supplier can notify in its own name, while still carrying the EU requirement for an AR if it is non-EU. The form has six components: Section A (introduction — date, applicant, notification type as first-time, amendment or revocation, applicant type, organisation details including registration number and SRN); Section B (manufacturer details, corresponding to the declaration of conformity); Section C (authorised representative details, where the manufacturer is outside the EU); Section D (the device notification sheet); Section E (details of payment); and the data-protection consent, self-declaration and MMA declaration (source: MMA, GL-MDF05/06, February 2026). Section D is the heart of the file: for each device, the source country supplier, the MMA reference number (for revisions and revocations), brand name, generic name, catalogue number, device type, class, and where applicable Basic UDI-DI and UDI-DI — each as written on the declaration of conformity, with brand name, generic name, catalogue number, device type and class mandatory. The documents submitted with the form are the notification sheet, the declaration of conformity and a soft copy of proof of payment; additional documents must be made available on request (source: MMA, GL-MDF05/06, February 2026).
Two fee-shaping definitions deserve attention. Devices bearing the same Basic UDI-DI are considered variants; where it is not available, variants are groups with the same intended purpose and manufacturing process that vary in size, colour or version. The guidance's examples: wrist supports of the same brand with and without a splint are not variants but separate devices requiring an individual fee; syringes of the same brand in different volumes require one notification fee; catheters in different gauges require one fee (source: MMA, GL-MDF05/06, February 2026). A portfolio's Maltese notification cost is therefore a function of its Basic UDI-DI structure, not its SKU count — build the notification sheet from the UDI database, not the price list.
4.4 What registration is not
Neither registration nor notification is a conformity assessment, clinical evaluation or quality audit; the MMA checks completeness, consistency with the declaration of conformity, applicant registration status and payment. Responsibility stays where the MDR puts it — with the manufacturer for conformity and each operator for its own obligations — and the MT-MDF05 self-declaration states that expressly. Any later change to the EU file (manufacturer name, AR, Basic UDI-DI structure, class) is a trigger to consider an amendment notification, priced at EUR 100 under Schedule 5 (source: S.L. 458.46, Schedule 5).