EU MDR + the Dutch layer
The Medical Devices Act (Wet medische hulpmiddelen) frames the national layer and the IGJ inspectorate supervises it — the IGJ inspects and enforces; it does not certify or approve products.
Field guide · PDF
No Dutch CE mark — EU MDR applies directly. What the Netherlands adds: an inspectorate with real enforcement teeth (the IGJ), a NOTIS-to-EUDAMED migration with a hard 28 November 2026 deadline, Dutch-language labelling, and hospitals that buy inside insurer-negotiated budgets.
Sourced from Farmatec/CIBG, the IGJ, business.gov.nl, and verified 2026 publications.
Enter your work email and we'll send the PDF straight to your inbox.
What's inside
Every figure in this guide traces to a named source — Farmatec/CIBG, the IGJ, business.gov.nl, and verified 2026 publications.
The Medical Devices Act (Wet medische hulpmiddelen) frames the national layer and the IGJ inspectorate supervises it — the IGJ inspects and enforces; it does not certify or approve products.
Farmatec's NOTIS notification gives way to mandatory EUDAMED registration and NOTIS data does not transfer automatically — pre-May 2026 devices must be registered by 28 November 2026, while custom-made devices stay in NOTIS.
There is no national device reimbursement list: hospital devices are paid inside DBC/zorgproduct budgets negotiated with insurers, and outpatient aids move through insurer-contracted suppliers.
Complete guide
Every section of the PDF on one page. Use the form above to keep a copy.
Last updated: October 2026. The EU Medical Device Regulation (MDR 2017/745) and the In Vitro Diagnostic Regulation (IVDR 2017/746) apply directly in the Netherlands. The Dutch national layer consists mainly of the Medical Devices Act (Wet medische hulpmiddelen, in force 26 May 2021), the Ministerial Regulation on Medical Devices (Regeling medische hulpmiddelen), supervision by the Health and Youth Care Inspectorate (Inspectie Gezondheidszorg en Jeugd, IGJ), and the notification system run by Farmatec (part of CIBG, an agency of the Ministry of Health, Welfare and Sport, VWS) — historically NOTIS, and since 28 May 2026 primarily EUDAMED. Always confirm critical steps against current Farmatec, IGJ and business.gov.nl guidance before filing.
The Netherlands is a small country with an outsized medical technology footprint. It is a major European logistics and distribution hub, home to large university medical centres (Amsterdam UMC, Erasmus MC, UMC Utrecht, Radboudumc, UMCG, LUMC, Maastricht UMC+), a dense medtech cluster around Eindhoven, Enschede, Leiden and Amsterdam, and — unusually for its size — four MDR-designated notified bodies. Commercial market-size estimates vary materially by product scope, so treat any single figure as indicative, not official. What is not in dispute is the structure of demand: a mandatory health-insurance system (Zorgverzekeringswet, Zvw) with competing private insurers, hospitals that buy largely within diagnosis-treatment combination budgets, and a strong culture of health-technology assessment led by Zorginstituut Nederland (ZIN, the National Health Care Institute) and tariff regulation by the Nederlandse Zorgautoriteit (NZa, Dutch Healthcare Authority).
Regulatory reality check up front: the Netherlands is not a separate certification regime from the EU. There is no "Dutch CE mark" and no Dutch pre-market product approval on top of CE marking. What the Netherlands adds — and what foreign manufacturers most often underestimate — is a national layer that sits around and after CE marking:
This guide covers the EU layer briefly (it is the same as in Germany, France, Italy or Spain) and the Dutch layer in depth, including the 2026 NOTIS-to-EUDAMED transition. That transition is the most time-sensitive point in Dutch device regulation today: file in the wrong system, or assume a NOTIS notification carries over automatically, and you will miss a registration duty that is not automatic.
The Wet medische hulpmiddelen (Medical Devices Act), which entered into force on 26 May 2021 and replaced the old Wet op de medische hulpmiddelen, regulates only the matters the MDR and IVDR leave to member states. In substance it:
Two structural contrasts with southern Europe matter for planning. First, the Netherlands has no AEMPS-style prior operating licence and no Italian-style establishment authorisation for a Dutch importer or distributor of medical devices as such. A Dutch establishment needs a Chamber of Commerce (Kamer van Koophandel, KvK) registration and must meet its MDR Article 13/14 duties, but there is no separate Dutch "importer licence" to obtain before trading. Second, Dutch competent-authority work is split: VWS is the competent/designating authority, Farmatec/CIBG runs notifications and export declarations, and the IGJ supervises and enforces. Sending a conformity question to the IGJ as if it were a certification body is a category error.
| Body | Role |
|---|---|
| Ministry of Health, Welfare and Sport (VWS) | Competent authority in the formal sense; designates Dutch notified bodies (on joint assessment with the IGJ, the European Commission and other member states); owns the Wet and Regeling medische hulpmiddelen. |
| IGJ — Inspectie Gezondheidszorg en Jeugd (Health and Youth Care Inspectorate) | Day-to-day supervisor for medical devices and IVDs: market surveillance, post-market surveillance supervision, serious-incident and FSCA intake, inspections, enforcement and fines. Not a certification body. |
| Farmatec (part of CIBG, VWS) | Runs device notifications (historically NOTIS; now the Dutch interface for the EUDAMED transition and for custom-made devices in NOTIS), and issues export declarations / Free Sales Certificates (€62 per declaration — see Section 8). Access to NOTIS uses eHerkenning (the Dutch business login). |
| CCMO — Centrale Commissie Mensgebonden Onderzoek (Central Committee on Research Involving Human Subjects) | Competent body for clinical investigations and performance studies: validation via its national office (Landelijk Bureau) for MDR Article 62 / 74.2 investigations, allocation to an accredited medical research ethics committee (MREC/METC) or review by the CCMO itself; submissions run through the national Research Portal. |
| Accredited MRECs / METCs | Review clinical investigations (and, directly, Article 74.1 PMCF and Article 82 investigations), alongside the WMO (Medical Research Involving Human Subjects Act) framework where it applies. |
| Zorginstituut Nederland (ZIN) | Assesses whether care — including care involving devices — meets the statutory criterion of "established medical science and medical practice" for the basic insured package; advises the Minister. Not a device regulator, but decisive for reimbursement of new forms of care. |
| NZa — Nederlandse Zorgautoriteit | Maintains the hospital payment system (zorgproducten / DBC structure, procedure codes — zorgactiviteiten), sets maximum tariffs for roughly 30% of DRGs and supervises insurers' and providers' market conduct; the rest of hospital tariffs are negotiated between hospitals and insurers. |
| Dutch notified bodies | The IGJ lists four MDR-designated bodies in the Netherlands: BSI Group The Netherlands B.V. (NB 2797), DEKRA Certification B.V. (NB 0344), Kiwa Dare B.V. (NB 1912) and Scarlet NB B.V. Any EU-designated notified body can be used for the Dutch market; it does not have to be Dutch. |
| LIR / MEBI / RIVM | LIR (Landelijk Implantaten Register) is the National Implant Registry; MEBI (Meldpunt en Expertisecentrum Bijwerkingen Implantaten, hosted with RIVM involvement) receives reports of adverse effects of implants from patients and professionals. |
The Netherlands uses the MDR classification system directly — four risk classes, 22 classification rules in Annex VIII MDR:
| Class | Risk Level | Conformity Assessment Route | Examples (per business.gov.nl / MDR practice) |
|---|---|---|---|
| Class I | Low risk | Self-declaration (Annex II+III technical documentation; EU Declaration of Conformity; no notified body — except Class Is [sterile], Im [measuring function], Ir [reusable surgical instruments]) | Adhesive bandages, spectacles, wheelchairs, non-sterile examination gloves |
| Class IIa | Low–medium risk | Notified body required (Annex IX chapters I+III, or Annex II+III + Annex XI) | Dental fillings, surgical clamps, hearing aids, ultrasound devices |
| Class IIb | Medium–high risk | Notified body required | Anaesthesia machines, infusion pumps, ventilators, surgical lasers |
| Class III | High risk | Notified body required, strictest scrutiny (Annex IX, including clinical evaluation consultation for certain devices) | Cardiac catheters, devices in direct contact with the heart, central circulatory system or central nervous system, pacemakers, drug-eluting stents |
For IVDs, the IVDR classes A–D apply (A lowest risk — specimen containers, prepared selective media; B medium — pregnancy tests, cholesterol self-tests; C medium-high — blood-glucose self-tests, genetic testing; D highest — hepatitis B blood-donor screening, HIV diagnostic blood tests), with notified-body involvement for Classes B, C and D and for sterile Class A devices.
Classification matters in the Netherlands for three practical reasons beyond the conformity route itself. First, it historically determined the national notification duty: under Sections 24/25 of the Dutch framework as described by the IGJ, Dutch manufacturers and authorised representatives of non-EU manufacturers established in the Netherlands were obliged to register (notify) certain risk classes nationally — in practice the Farmatec/NOTIS notification centred on Class I devices and IVDs placed on the European market by a Dutch-established manufacturer or AR, precisely because higher classes were already visible through notified-body certificates. Second, classification drives the Dutch clinical-investigation route: invasive Class IIa/IIb/III investigations for conformity purposes can only be reviewed under the Article 62/74.2 track with CCMO validation (Section 5). Third, classification drives post-market reporting: PSUR duties apply from Class IIa upwards, and implantables trigger the LIR and implant-card layer (Section 10).
Special MDR categories also apply in the Netherlands: custom-made devices (Annex XIII statement instead of CE marking; no UDI; Dutch registration in NOTIS remains mandatory — Section 4), systems and procedure packs (Article 22 MDR statement), in-house devices manufactured and used only within a health institution under Article 5(5) MDR (business.gov.nl confirms the lighter regime where the Article 5(5) conditions are met), and Annex XVI products without a medical purpose (non-corrective contact lenses, dermal fillers and similar products listed in Annex XVI).
An incorrect classification is not a clerical error: it invalidates the conformity route, can mean a missing notified body, and can result in IGJ enforcement. Resolve borderline questions (device vs. medicinal product vs. cosmetic vs. biocidal product) early — for medicines-borderline questions the Dutch Medicines Evaluation Board (CBG/MEB) is also part of the landscape, and MEB consultation fees for certain device-borderline procedures have been reported in regulatory round-ups;
The Netherlands imposes no separate Dutch representative requirement on top of the MDR for a foreign manufacturer that simply sells into the Netherlands through others. Keep the EU roles and the Dutch establishment facts separate.
PRRC: The Person Responsible for Regulatory Compliance (Art. 15 MDR) requirement applies unchanged: a relevant degree plus one year of professional experience, or four years of experience without a degree; the AR must also have a PRRC permanently and continuously at its disposal. Micro and small manufacturers may have the PRRC permanently and continuously at their disposal without employing them.
No physical office is required to sell into the Netherlands. A foreign manufacturer does not need a Dutch office, branch or warehouse merely to register and sell devices in the Netherlands. It needs: an EU AR (if non-EU), an importer/distributor chain that meets Articles 13–14, correct labelling language, EUDAMED registration and — where the manufacturer or AR is Dutch-established — the remaining NOTIS duties for custom-made devices. A Dutch office becomes regulatory-relevant mainly if that establishment itself is the manufacturer, AR, importer or a reprocessor, in which case the corresponding MDR duties (and, for investigations and implants, the Dutch systems in Sections 5 and 10) attach to it directly.
This is the section most generic "sell into the EU" guides get wrong for the Netherlands in 2026, because the Netherlands changed systems mid-year — and, unlike Spain, did not replace its national register with a new national marketing register.
The old system (until 27 May 2026). Farmatec ran NOTIS, the Dutch notification system for medical devices, accessed with eHerkenning. Its core users were Dutch-established manufacturers and ARs notifying Class I devices and IVDs, and manufacturers (and, where applicable, ARs) of custom-made devices, who must also maintain and submit the list of types of custom-made devices they provide. A NOTIS notification was an administrative notification, not an approval: Farmatec's stated processing time, once the information and required documents were complete, was a maximum of 90 days for a notification application, and a maximum of 3 months for an amendment or deregistration.
What changed on 28 May 2026. With the first four EUDAMED modules becoming mandatory EU-wide, Farmatec states the rule plainly: as of 28 May 2026 it is mandatory to register medical devices through EUDAMED (UDI/Device registration). The transitional periods are:
What stays in NOTIS. Two things. First, custom-made devices: in the Netherlands, registration is mandatory for manufacturers (and, where applicable, the AR) of custom-made devices under Article 2 of the Regeling medische hulpmiddelen, and that registration is completed in NOTIS, not as an ordinary EUDAMED device registration; custom-made devices bear no CE marking and require no UDI. Second, housekeeping of the old register: confirmed product notifications that have not yet been processed in EUDAMED can still be changed online in NOTIS (product updates — amendment/WIJZ, conversion, deregistration/AFM) up to and including 28 November 2026, logging in with eHerkenning; the processing time for a deregistration or amendment is a maximum of 3 months.
What the Netherlands did NOT add. There is no Dutch equivalent of Spain's RECOPS or Italy's Repertorio: no separate national marketing register for ordinary CE-marked devices on top of EUDAMED, and no per-product national filing fee for ordinary devices that has been verified in current Farmatec guidance.
Practical consequence for planning. Budget for one registration system in 2026 and after — EUDAMED (EU-wide, mandatory, no fee) — plus a migration project if you hold legacy NOTIS notifications (re-register in EUDAMED by 28 November 2026; finish any NOTIS amendments/deregistrations by the same date), plus a continuing NOTIS duty if you make custom-made devices. Do not file a fresh ordinary NOTIS notification for a device first marketed after 27 May 2026, and do not assume NOTIS data migrates automatically — it does not.
Step 1 — Confirm the product is a medical device and classify it under Annex VIII MDR (or IVDR Annex VIII for IVDs). Resolve borderline questions early; for device–drug combinations, expect the notified body to run the ancillary-substance consultation procedure (in the Netherlands, the CBG/MEB is part of that landscape).
Step 2 — Establish the quality management system aligned to MDR Art. 10(9), normally evidenced by ISO 13485 certification. Notified bodies audit the QMS as part of conformity assessment; it is a prerequisite, not a parallel task. Business.gov.nl summarises the Dutch expectation in the same terms: the QMS must cover risk management, clinical evaluation, the post-market surveillance system and serious-incident reporting processes.
Step 3 — Compile the technical documentation (Annex II) and clinical evaluation (Annex XIV), including the post-market surveillance plan and PMCF plan where applicable. For IVDs, compile the performance evaluation under the IVDR.
Step 4 — Complete conformity assessment with a notified body (all classes above I, and Is/Im/Ir; IVD Classes B, C, D and sterile A). The notified body does not need to be Dutch — any body designated for your device codes in NANDO qualifies — but the Netherlands is one of the few member states where choosing a local body (BSI Netherlands, DEKRA Netherlands, Kiwa Dare, Scarlet NB) is a realistic option across many scopes, which can simplify audit logistics and language. Verify scope in NANDO, not on the body's marketing page.
Step 5 — Draw up the EU Declaration of Conformity (Annex IV) and affix the CE marking (with the notified body's number where a body was involved). There is no separate Dutch marketing authorisation for an ordinary device: the Dutch system is CE marking plus registration/notification, policed after the fact by the IGJ — not a Dutch product approval.
Step 6 — Assign UDI (Art. 27 MDR: Basic UDI-DI, UDI-DI, UDI-PI) and prepare UDI carriers for labels, on the staggered MDR timetable by class. Custom-made devices are exempt from UDI. Importers and distributors in the Netherlands must check that the device has a UDI — and Dutch hospitals check it again at goods-in.
Step 7 — Appoint the EU Authorized Representative (if you are established outside the EU) and execute the Art. 11 mandate — before registrations, because the AR is a registrant in EUDAMED. Decide deliberately whether the AR sits in the Netherlands (with the Dutch NOTIS consequences for custom-made devices described in Section 3) or elsewhere in the EU.
Step 8 — Decide the Dutch establishment question. Will any company established in the Netherlands be the manufacturer, AR, importer, distributor or reprocessor? If yes, that establishment needs KvK registration, its own EUDAMED actor registration/SRN and working Article 10/11/13/14 systems — but no Dutch device operating licence (Section 3). If a Dutch health institution will manufacture devices for its own exclusive use, test the Article 5(5) in-house conditions instead of assuming a full manufacturer route.
Step 9 — Register the actors and the device in EUDAMED. Obtain the SRN for manufacturer, AR and importer through actor registration (validated by the competent authority of the member state of establishment — for Dutch establishments, the Dutch validation route via Farmatec/VWS), then register the device/UDI data. Mandatory since 28 May 2026 for the first four modules; new devices must be registered before placement on the market. If the device was placed before 28 May 2026 and notified in NOTIS, the EUDAMED deadline is 28 November 2026.
Step 10 — Register custom-made devices in NOTIS (only if applicable). Dutch manufacturers (and, where applicable, ARs) of custom-made devices register in NOTIS under Article 2 of the Regeling, keep the Annex XIII statement instead of a Declaration of Conformity/CE marking, and maintain the list of types of custom-made devices provided. Farmatec's maximum processing time for a complete notification application is 90 days.
Step 11 — Prepare the Dutch labelling. Translate labels and IFU into Dutch as the default (Section 6 on language and the professional-use English nuance). For Class I and IIa devices, an IFU is not required if the device can be used safely without one — but the labelling duties remain.
Step 12 — If you will run a clinical investigation in the Netherlands, file through the CCMO route before starting (see the box below). Do not treat a hospital's local approval as a substitute for CCMO/MREC review.
Step 13 — Set up vigilance, post-market surveillance, implant and supply-continuity workflows before first sale: PMS system and reporting calendar (Section 10), a serious-incident reporting workflow to the IGJ, LIR/MEBI workflows if you supply implants, and the Article 10a supply-interruption workflow (IGJ + national reporting point + Zorg Inkoop Netwerk Nederland).
Step 14 — Build the commercial route (separate from registration). Registration makes sale lawful; it does not create a buyer. For the public sector: monitor tenders on TenderNed and hospital/regional purchasing cooperatives, and expect the Aanbestedingswet 2012 (Dutch Public Procurement Act) framework for above-threshold contracts. For hospital devices, work out whether your device is paid inside an existing zorgproduct/DBC budget or needs a coding/tariff conversation with the NZa and insurers (Section 9). For outpatient aids (hulpmiddelenzorg), the buyer is effectively the health insurer's contracted-supplier system, not the individual hospital.
> Clinical investigations in the Netherlands (CCMO route). All clinical investigations with medical devices must be submitted digitally via the national Research Portal. The CCMO is the authorised agency for registering clinical investigations with medical devices — for non-CE-marked devices and for CE-marked devices investigated outside their intended use alike. For investigations for conformity purposes (MDR Article 62 and Article 74.2), the research file is first validated by the CCMO's national office (Landelijk Bureau, CCMO_LB); after a positive validation decision, the CCMO_LB forwards the file to an accredited MREC or the CCMO for review. Invasive Class IIa/IIb/III investigations under Articles 62/74.2 can only be reviewed by an accredited MREC or the CCMO. Post-market clinical follow-up investigations (Article 74.1) and other investigations (Article 82) may be submitted directly to an accredited MREC — CCMO_LB validation is not necessary. Investigations no longer need to be reported separately to the IGJ; the CCMO handles that as part of validation. Sponsors prepare a standard research file, including an Investigational Medical Device Dossier (IMDD) (a light IMDD may suffice for certain investigator-initiated low-risk studies without manufacturer involvement). Review fees for MDR Article 62/74.2 and IVDR files follow the national CCMO fee schedule (a single fee per trial, redistributed by the competent authority); individual MRECs publish their own tariffs for WMO and local-feasibility reviews — one university METC's published 2025 tariff, for example, lists €4,000 (single-centre commercial WMO) / €5,500 (multi-centre commercial WMO), €500 for a WMO amendment and €1,200 for local approval, with MDR 62/74.2 files charged at the national CCMO rates.
What does NOT exist in the Netherlands: an IGJ "approval certificate" for an ordinary CE-marked device, a Dutch importer licence, and a Dutch national product register on top of EUDAMED. The proof of lawful Dutch market access is: CE certificate + EU Declaration of Conformity + EUDAMED registrations (+ a NOTIS registration for custom-made devices, where applicable). Anyone selling you a Dutch "product approval" beyond this is selling you a filing service, not an approval.
The Dutch language rule is the point where summaries most often mislead, because two true statements sit side by side:
Practical rule for planning: budget and produce full Dutch labels and IFU for anything that reaches a patient, a lay user, an implant card, a field safety notice or a Dutch authority. English-only supporting information may be defensible for narrowly professional-use elements under the Regeling's conditions — but treat that as a documented, device-specific position you verify against the current Regeling text and Commission table at filing, not as a general "the Netherlands accepts English" licence. Hospitals' procurement and vigilance teams in the Netherlands routinely reject English-only IFUs at vendor qualification even where a legal argument exists. Conversely, do not over-translate by assumption either: confirm which Annex I Section 23 information elements you are translating, for which user group, before commissioning translation of the entire technical file — the duty attaches to the information supplied with the device, not to your whole dossier.
Electronic IFU (eIFU) follows the EU eIFU regulation framework; the language duty is unchanged by the medium.
| Activity | Typical timeline |
|---|---|
| CE certification via notified body (Class IIa, new submission) | 12–18 months as a realistic planning figure; notified-body capacity remains the critical path EU-wide |
| CE certification (Class III) | 18–24+ months, including the clinical evaluation consultation procedure where it applies |
| Class I self-declaration route | 2–4 months once technical documentation and Dutch labelling are complete |
| EUDAMED actor registration / SRN issuance | Weeks, after competent-authority verification. Mandatory from 28 May 2026 for the first four modules |
| EUDAMED device registration (new device, from 28 May 2026) | Must be completed before placing the first unit on the market; allow weeks for data preparation and validation, not days |
| Migration of pre-28 May 2026 devices from NOTIS to EUDAMED | Deadline 28 November 2026 — a fixed date, not a processing time; start the data project early, NOTIS data does not transfer automatically |
| NOTIS notification (custom-made devices; and historically Class I by Dutch-established operators) | Farmatec: maximum 90 days once the information and required documents are complete |
| NOTIS amendment (WIJZ) / conversion / deregistration (AFM) | Maximum 3 months; online changes to confirmed notifications possible only until 28 November 2026 |
| Export declaration / Free Sales Certificate (Farmatec) | Within 4 weeks if the application includes all information and all required documents |
| Clinical investigation (MDR Art. 62/74.2) | MDR Art. 66–74 timetable (validation ~10 days; assessment report within 45 days, extendable), routed in the Netherlands via CCMO_LB validation then MREC/CCMO review, plus local feasibility approval at each centre; Art. 74.1/82 files go directly to an accredited MREC |
| Hospital tender / insurer-contract cycles | Not a registration step — but allow 6–18 months from registration to first meaningful public-sector revenue; Dutch hospitals and purchasing cooperatives run their own calendars, and insurer contracting for outpatient aids is annual |
Transition deadlines for legacy (MDD/AIMDD) devices (Regulation (EU) 2023/607) — the same EU-wide dates apply in the Netherlands:
Eligibility conditions: continued directive compliance, no significant change in design or intended purpose, no unacceptable risk, an MDR-compliant QMS in place by 26 May 2024, a formal notified-body application by 26 May 2024, and a written notified-body agreement by 26 September 2024. IVDR legacy deadlines (Regulation (EU) 2024/1860) are staggered by class: Class D by end-2027, Class C by end-2028, Class B and sterile Class A by end-2029.
The Netherlands' national filing fees for devices are low to non-existent for ordinary products; the money is spent on certification, clinical investigations, and Dutch-language and hospital-qualification work.
Currency: all official fees in euros (€).
Legal market access is only the first gate in the Netherlands. The second gate is how a system of competing insurers and largely autonomous hospitals pays, and it works differently for hospital devices and for outpatient aids.
Hospital devices: paid inside the treatment, not from a device list. The Netherlands has no single national reimbursement list covering most hospital medical devices (contrast France's LPPR). Specialist medical care is paid through the zorgproduct / DBC system: all specialist care (inpatient, day-case, ambulatory) within the same major disease category and period (for most services 120 days; for surgical procedures 42 days) is covered by a single DRG-type payment, determined by the combination of a diagnosis code and a procedure code (zorgactiviteit). The coding nomenclature and payment structure are maintained by the NZa and released annually; around 30% of DRGs have maximum tariffs set nationally by the NZa, and tariffs for the remainder are negotiated between individual hospitals and insurers. Services outside the DRG payment are handled as "other care products" (overige zorgproducten, OZPs) in defined categories. Practical consequence: for most hospital devices the commercial question is not "get reimbursed" but "get specified by clinicians, fit (or improve) the hospital's zorgproduct economics, and win supply contracts at a price the hospital can absorb inside its negotiated tariffs." Where a genuinely new form of care is involved, ZIN's assessment of whether it meets the "established science and practice" criterion for the basic package, and any NZa coding/tariff step, become the gating items — both are evidence projects, not filings.
Outpatient aids (hulpmiddelenzorg): the insurer is the buyer. Devices supplied for use at home — diabetes, incontinence, stoma and wound-care aids are the classic categories — are generally arranged through health insurers' contracted suppliers: patients order from a contracted supplier and the aid is delivered to the home. The NZa actively supervises insurers' statutory duty of care (zorgplicht) for these aids, including whether acute provision works in practice. Practical consequence: a manufacturer of outpatient aids does not sell its way in hospital by hospital; it must be ranged by the suppliers and distribution channels that hold insurer contracts, and service levels in acute situations are part of the product.
Interactions with healthcare professionals — a Dutch legal trap for foreign teams. The statutory inducement prohibition (gunstbetoon) in the Wet medische hulpmiddelen, the GMH code (advertising to professionals, hospitality, service agreements) and Transparency Register Healthcare disclosure of financial relationships above the applicable thresholds govern speaker fees, consultancy, sponsorship and hospitality. The IGJ has inspected hospital departments (notably cardiology) over opaque payments routed via physicians' personal entities or foundations, and found insufficient internal controls — without, in those published cases, immediately imposing fines, but with explicit warnings of enforcement on repeat findings. Structure every Dutch HCP payment through the institution where required, document the service and fair-market value, and register disclosable relationships. A US- or UK-style speaker programme transplanted unchanged into the Netherlands is a compliance incident waiting for a date. should be confirmed before you set internal limits.)
Advertising. Device advertising to the public and to professionals is governed by the Wet medische hulpmiddelen inducement rules, the GMH code for professional-facing promotion, and the specific code for public advertising of self-care devices in pharmaceutical form; there is no pre-clearance licensing system for device advertising to professionals. The KOAG pre-review route that exists for medicines advertising to the public is not available for medical devices.
Do I need a Dutch authorized representative? No. You need an EU Authorized Representative established in any member state (MDR Art. 11) if you are established outside the EU. Choosing an AR established in the Netherlands is a commercial and logistical decision — it can help with Dutch notified bodies and distributors, but it also historically carried the Dutch NOTIS notification duty for Class I devices and still carries NOTIS duties for custom-made devices. Decide it deliberately (Section 3).
Is there a Dutch registration on top of CE marking? Yes — an administrative one, not an approval: EUDAMED registration (mandatory EU-wide from 28 May 2026; before placement for new devices, by 28 November 2026 for devices already on the market). The Netherlands has no additional national marketing register for ordinary devices on top of EUDAMED. The exception is custom-made devices, which are registered in NOTIS on a continuing basis.
I notified my device in NOTIS years ago. Am I done? No. Farmatec and business.gov.nl are explicit: a NOTIS-notified device that is not yet in EUDAMED must still be registered in EUDAMED, by 28 November 2026 if it was placed on the market before 28 May 2026. Until that date you can still amend or deregister the NOTIS notification online with eHerkenning; after it, EUDAMED is the system of record.
Does Class I need Dutch registration? In EUDAMED: yes, like every device. In NOTIS: the historical Dutch notification duty centred on Dutch-established manufacturers and ARs for Class I devices; from 28 May 2026 the registration duty for ordinary Class I devices is discharged in EUDAMED. Custom-made devices — a separate category, often confused with Class I — remain in NOTIS in all cases.
Do I need a Dutch importer licence or a Dutch office? Neither, for ordinary devices. There is no Dutch prior operating licence for device importers (contrast Spain), and no office is required merely to sell into the Netherlands. A Dutch importer needs KvK registration, EUDAMED actor registration and Article 13 systems; verify those, and verify that any consultant charging for a "licence" can name its legal basis.
How much does Dutch registration cost? EUDAMED: €0. No per-product NOTIS fee for ordinary notifications has been verified in current Farmatec guidance. A Farmatec export declaration / Free Sales Certificate costs €62 per declaration + €62 administration per invoice. Real costs are notified-body fees (roughly €10,000–€50,000+ by class and complexity, market-reported), EU AR services, CCMO/MREC fees if you run investigations in the Netherlands, and Dutch translations.
How long does the whole process take? For a new Class IIa device starting from no CE marking: 12–18 months for notified-body certification dominates everything else; EUDAMED registration is weeks of data work once the certificate and UDI data exist, and must be finished before first placement. For an already CE-marked device entering the Netherlands: actor/device registration in EUDAMED, Dutch labelling and importer set-up are typically a 1–3 month project — plus, commercially, 6–18 months to first meaningful hospital or insurer-channel revenue.
Can I use a Dutch notified body if I am not Dutch? Yes. Notified bodies serve the whole EU market. The Netherlands designates four MDR bodies — BSI Group The Netherlands (2797), DEKRA Certification (0344), Kiwa Dare (1912) and Scarlet NB — but any EU-designated body with the right scope codes in NANDO can certify for the Dutch market. Choose on scope, queue and sector experience, not on flag.
Do labels and IFUs have to be in Dutch? As the default, yes — the Dutch government portal states that label and IFU information must be in Dutch. The Regeling allows Dutch or English for certain professional-use information under conditions, without prior approval — a narrow, document-specific position to verify per device, not a general English licence. Patient-facing information, implant cards and field safety notices: Dutch.
How are hospital devices paid for if there is no reimbursement list? Inside the treatment payment. Specialist care is paid via zorgproducten/DBC tariffs maintained by the NZa (about 30% nationally capped, the rest negotiated between hospitals and insurers), so your device must fit a hospital's product economics and be specified by clinicians. Genuinely new care may need a ZIN "established science and practice" assessment and NZa coding work. Outpatient aids are a different channel: insurers' contracted suppliers.
What is the single most common Dutch mistake in 2026? Treating the NOTIS-to-EUDAMED transition as automatic. It is not: NOTIS data does not migrate, the EUDAMED deadlines (before placement / 28 November 2026) are fixed, NOTIS amendments close on the same November date, and custom-made devices stay behind in NOTIS — four facts that must all be in the same project plan.
Sources consulted for this guide (October 2026): Farmatec (english.farmatec.nl) — Notifications page (EUDAMED mandatory from 28 May 2026; NOTIS transition deadlines; custom-made registration in NOTIS; 90-day / 3-month processing) and Free Sales Certificate page (€62 + €62; 4-week processing); business.gov.nl — "Requirements for medical devices" (IGJ as supervisory body; classes; EUDAMED and NOTIS transition; UDI; CE marking and notified-body classes; Dutch labelling; CCMO submission; PMS/PSUR; implant card, LIR within 1 month, MEBI; supply-interruption reporting at least 6 months ahead via the national reporting point and Zorg Inkoop Netwerk Nederland; reprocessing and in-house rules); IGJ (english.igj.nl) — registration of a medical device or IVD (Dutch-established manufacturers/ARs, Sections 24/25), role of the notified body (four Dutch designated bodies: BSI Netherlands, DEKRA Certification, Kiwa Dare, Scarlet NB; VWS designation), supervision of post-market surveillance; European Commission language-requirements table for the MDR (Netherlands — Regeling medische hulpmiddelen Art. 1: Dutch / Dutch or English by information category); CCMO and Amsterdam UMC research guidance (Research Portal submission, CCMO_LB validation for Art. 62/74.2, accredited MREC review, Art. 74.1/82 direct to MREC, IMDD, national CCMO fees) and a published university METC tariff (METC UMCG, 2025); NZa/ZIN market-access summaries (zorgproduct/DBC structure, ~30% national maximum tariffs, OZP categories, ZIN package advice) and Aanbestedingswet 2012 / TenderNed procurement guidance; Dutch legal commentary on the Wet medische hulpmiddelen (IGJ fines up to €870,000 / 10% of turnover; statutory gunstbetoon prohibition; GMH code and Transparency Register Healthcare; IGJ cardiology inspections). Items marked "to be verified" (to be verified) were not confirmed against a primary source at the time of writing and must be checked before filing.
This guide is market-access intelligence, not regulatory or legal advice. Farmatec and IGJ practice and the EUDAMED transition change — professional review is required before business use.