Step 1 — Confirm the product is a device, and get the class right. Apply the MDR Article 2 definition and Annex VIII rules, including Rule 11 for software and the Annex XVI list for products without a medical purpose. If qualification or classification is genuinely doubtful, NOMA is the authority that decides for Norway — a wrong self-assessment discovered in a tender or an inspection is far more expensive than an early borderline analysis.
Step 2 — Build the conformity file once, for the EEA, in English. Quality management system (in practice EN ISO 13485), risk management (EN ISO 14971), technical documentation under MDR Annexes II–III, clinical evaluation under Article 61 and Annex XIV, and post-market surveillance planning under Articles 83–86. Norway requires the documentation used in conformity assessment procedures — including technical documentation and audit, assessment and inspection reports — to be in English, so the EEA file you build for your notified body is the file Norway expects; do not machine-translate it into Norwegian (see Section 6 for what must be Norwegian instead).
Step 3 — Choose and contract a notified body if the class requires one. Classes IIa, IIb and III, sterile / measuring / reusable Class I, sterile Class A IVDs and Classes B–D IVDs need a notified body designated for the device category. DNV Product Assurance AS is the Norway-designated option under both Regulations; any EEA body works. Confirm designation scope against Commission Implementing Regulation (EU) 2017/2185 codes, apply once (parallel applications to several bodies for the same device are prohibited), and budget for queue time — see Section 7.
Step 4 — Register as an actor in EUDAMED and obtain the SRN. Non-EEA manufacturers first appoint the EEA authorised representative; the AR must exist in EUDAMED before the manufacturer can complete registration. Norwegian-established actors are validated by NOMA. Do this before the notified-body application where the procedure requires it, and in all cases before placing the device on the market.
Step 5 — Complete the Norwegian-language package. Translate the label and the instructions for use into Norwegian, prepare the Norwegian implant card and patient information where Article 18 MDR applies, localise user-facing software information into Norwegian, and prepare Norwegian field safety notice templates. If Norwegian is genuinely impossible for a defined professional product, prepare the section 15 exemption application instead — with the user-group evidence and the safe-use argument NOMA requires (see Section 6) — and do not ship before it is granted. Allow translation and in-country review time in the launch plan; this step, not registration, is the usual critical path for Norway.
Step 6 — Draw up the EU Declaration of Conformity and affix the CE mark. The Declaration of Conformity may be in English or Norwegian for Norway. Affix the CE mark with the notified body's number where a body was involved. There is no additional Norwegian mark.
Step 7 — Register the device in EUDAMED and assign UDI. Complete UDI/Device module registration before the first unit is placed on the Norwegian market, and apply UDI carriers to the label and higher packaging levels on the MDR timetable for the class.
Step 8 — Set up the Norwegian supply chain roles in writing. Importer and distributor agreements should allocate, expressly: verification duties under Articles 13–14, Norwegian-language complaint intake, vigilance reporting support and deadlines, field safety corrective action execution in Norway (including Norwegian-language field safety notices), Article 10a shortage notifications, the ten/fifteen-year documentation and traceability retention, and who deals with NOMA in Norwegian if NOMA writes in Norwegian. Customs and import VAT handling should be allocated too — Norway is in the EEA, not the EU customs union.
Step 9 — Stand up vigilance and post-market surveillance for Norway. Register the internal route by which a Norwegian incident reaches the manufacturer's vigilance function fast enough to meet MDR Article 87 deadlines to NOMA (2 / 10 / 15 days — Section 9), brief Norwegian distributors and hospital customers that healthcare facilities in Norway have their own mandatory reporting duty, and prepare the Manufacturer Incident Report (MIR) workflow to NOMA's medical device vigilance unit.
Step 10 — Plan the commercial gate in parallel. Identify whether the device sells into existing Sykehusinnkjøp HF framework agreements or needs a new tender cycle, whether the hospital trusts buy it inside activity-based (DRG-type) financing, whether it is a genuinely new method that belongs in the Nye metoder pipeline, or whether it is an aid supplied through the municipal / NAV assistive-technology channel. Regulatory completion and commercial readiness are different dates in Norway; plan both (Section 10).