EU MDR + the Polish layer
The Act of 7 April 2022 frames the national layer and the URPL supervises it — market surveillance, inspections, vigilance, and fines of up to PLN 5,000,000 for placing a non-compliant device on the market.
Field guide · PDF
No Polish CE mark — EU MDR applies directly. What Poland adds catches foreign teams out: a URPL with real enforcement teeth, national distributor and user lists that survived EUDAMED, Polish-language rules, Europe's strictest device advertising regime, and payment through the NFZ and the JGP system.
Sourced from the URPL, the Act of 7 April 2022 on Medical Devices, and verified 2026 publications.
Enter your work email and we'll send the PDF straight to your inbox.
What's inside
Every figure in this guide traces to a named source — the URPL, the 2022 Act, the Ministry of Health and NFZ, and verified 2026 publications.
The Act of 7 April 2022 frames the national layer and the URPL supervises it — market surveillance, inspections, vigilance, and fines of up to PLN 5,000,000 for placing a non-compliant device on the market.
URPL distributor and professional-user lists have run since 1 July 2023 next to EUDAMED. Labels and instructions for lay users are Polish, and since 2023 public advertising may not use a healthcare professional's image or promote professional devices to the public.
There is no simple listing route: most hospital devices are paid inside the JGP (DRG) system, hospital-network budgets or add-on categories, with AOTMiT health technology assessment built into reimbursement decisions.
Complete guide
Every section of the PDF on one page. Use the form above to keep a copy.
Last updated: October 2026. The EU Medical Device Regulation (MDR 2017/745) and the In Vitro Diagnostic Regulation (IVDR 2017/746) apply directly in Poland. The Polish national layer consists mainly of the Act of 7 April 2022 on Medical Devices (in force from 26 May 2022, with the last provisions from 1 July 2023), supervision by the Office for Registration of Medicinal Products, Medical Devices and Biocidal Products — URPLWM, commonly URPL — Polish-language rules in Article 12 of that Act, a strict national advertising regime in Articles 54–61, the distributor and user lists launched on 1 July 2023, and a separate reimbursement and hospital-payment system run through the Ministry of Health, the National Health Fund (NFZ) and the Agency for Health Technology Assessment and Tariff System (AOTMiT). Always confirm critical steps against current URPL, Ministry of Health, NFZ and AOTMiT guidance before filing.
Poland is Central Europe's largest medical device market by population — about 38 million inhabitants — and one of its most consequential. Independent market reporting cited for this guide valued the Polish medical devices market at around USD 4.5 billion on a five-year historical basis, driven by an ageing population, rising healthcare expenditure, a large public hospital sector and a strong domestic manufacturing base concentrated around Warsaw, Kraków and Wrocław. Poland is also a manufacturing and distribution hub for Central and Eastern Europe: a device that is set up correctly for Poland — Polish labelling, a compliant distributor chain, a workable NFZ payment story — is far better placed for the wider CEE region than a device that is only CE-marked on paper.
Regulatory reality check up front: Poland is not a separate certification regime from the EU. There is no "Polish CE mark" and no Polish pre-market product approval on top of CE marking for an ordinary device. What Poland adds — and what foreign manufacturers most often underestimate — is a national layer that is, in two areas, stricter than in most member states:
This guide covers the EU layer briefly (it is the same as in Germany, France, Italy, Spain, the Netherlands or Belgium) and the Polish layer in depth, including the 2022 Act, the 1 July 2023 lists, the Polish-language and advertising rules that catch foreign teams most often, and the NFZ/JGP payment reality that decides whether a registered device actually sells.
Poland implemented the MDR/IVDR late. Sector analyses (Wolf Theiss, Medical Devices in CEE & SEE) note that the adapting Act came into force in 2022 — a year later than it should have — creating a period of regulatory uncertainty. Do not rely on summaries based on the old law.
Two structural points matter for planning. First, Poland has very few Polish-designated notified bodies and none for the full IVDR scope in the analyses consulted: PCBC (Polskie Centrum Badań i Certyfikacji, NB 1434) and TÜV NORD Polska (NB 2274) are listed in NANDO for MDR scopes, and Wolf Theiss notes that for IVDR certifications only 12 notified bodies were available in Europe and none was based in Poland at the time of its guide, with not all MDR conformity routes (for example Annex X type examination) available in Poland. A manufacturer may use any EU-designated notified body for the Polish market — and for many device codes, in practice, must. Verify scope in NANDO, not on a body's marketing page. Second, Polish entrepreneurs continue to report notified-body bottlenecks — longer waits and higher certification costs, especially for SMEs — even though no significant device shortages were reported in Poland in the sector guide consulted.
| Body | Role |
|---|---|
| URPL / URPLWM — Office for Registration of Medicinal Products, Medical Devices and Biocidal Products (Urząd Rejestracji Produktów Leczniczych, Wyrobów Medycznych i Produktów Biobójczych), Warsaw (Al. Jerozolimskie 181C, 02-222 Warszawa) | Polish competent authority for medical devices and IVDs: implements MDR/IVDR, designates and supervises Polish notified bodies, market surveillance and inspections, vigilance, clinical investigations, distributor and user lists, custom-made activity registration, advertising supervision, Free Sale / classification opinions in defined cases (including for customs and tax authorities, e.g. on VAT classification), and administrative fines. |
| Ministry of Health (Ministerstwo Zdrowia — MZ) | Owns health legislation, the advertising implementing regulation, highly specialised services paid directly by the Ministry, and formal reimbursement decisions. Triggers most AOTMiT HTA work by request. |
| NFZ — National Health Fund (Narodowy Fundusz Zdrowia) | The single public payer. Contracts hospitals, runs the JGP (Polish DRG) payment logic in practice, publishes procedure (ICD-9-PL) and diagnosis (ICD-10-PL) coding that determines DRG grouping, and pays network and non-network hospitals differently (Section 9). |
| AOTMiT — Agency for Health Technology Assessment and Tariff System (Agencja Oceny Technologii Medycznych i Taryfikacji) | Maintains the DRG tariff system and conducts HTA as an integral part of reimbursement decision-making, on request of the Minister of Health or NFZ. Not a filing office a manufacturer can simply apply to for a routine review. |
| Chief Pharmaceutical Inspectorate (GIF) / Voivodeship Pharmaceutical Inspectors (WIF) | Supervise, to a limited extent, device trade when devices are sold through pharmaceutical wholesalers and pharmacies. |
| PCBC S.A. (NB 1434) and TÜV NORD Polska (NB 2274) | Polish-designated MDR notified bodies listed in secondary sources consulted. PCBC's own regulations describe EU QMS certificates (Annex IX Ch. I), EU technical documentation assessment certificates (Annex IX Ch. II) and EU product quality assurance certificates (Annex XI Part A), valid for a maximum of five years, issued in Polish and English. Any EU notified body may be used. |
| POLMED | Polish Chamber of Commerce for Medical Devices; sector association and owner of the Polish ethical business practice code. |
Poland uses the MDR classification system directly — four risk classes, 22 classification rules in Annex VIII MDR (Article 51 MDR), based on the manufacturer's intended purpose. If more than one rule applies, the rule resulting in the highest class applies. MDCG 2021-24 is the classification guidance notified bodies and the URPL work to.
| Class | Risk Level | Conformity Assessment Route | Examples |
|---|---|---|---|
| Class I | Low risk | Self-declaration (Annex II + III technical documentation; EU Declaration of Conformity; no notified body — except Class Is [sterile], Im [measuring function], Ir [reusable surgical instruments], where a notified body is involved only for the sterile, measuring or reprocessing aspect) | Non-sterile examination gloves, adhesive dressings, spectacles, wheelchairs |
| Class IIa | Low–medium risk | Notified body required (Annex IX QMS + sampled technical documentation, or Annex XI route) | Dental fillings, surgical clamps, hearing aids, diagnostic ultrasound |
| Class IIb | Medium–high risk | Notified body required (Annex IX, or Annex X type examination + Annex XI; technical documentation assessed for each implantable device, with defined exceptions) | Anaesthesia machines, infusion pumps, ventilators, surgical lasers, blood bags |
| Class III | High risk | Notified body required, strictest scrutiny (Annex IX including assessment of technical documentation for each device, or Annex X + Annex XI; clinical evaluation consultation / expert panel for specified devices) | Pacemakers, heart valves, devices in direct contact with the heart, central circulatory or central nervous system |
For IVDs, the IVDR classes A–D apply (A lowest risk; B; C; D highest risk — e.g. high-risk blood-donor screening), with notified-body involvement for Classes B, C and D and for sterile Class A devices. Software is classified under Rule 11: software that provides information used for diagnostic or therapeutic decisions is at least Class IIa, and can be IIb or III depending on the seriousness of possible deterioration — a frequent up-classification trap for digital health products entering Poland. The Ministry of Health has initiated an e-health application certification programme intended to clarify the landscape for software as a medical device (SaMD), but sector sources describe the SaMD / AI reimbursement and regulatory pathway as still evolving.
Classification disputes between a manufacturer and a notified body are resolved in Poland by the President of the URPL by administrative decision — a Polish procedural point worth knowing before a dispute arises, not after. The URPL also retains a classification / qualification opinion role in defined national cases, for example when customs authorities seek confirmation whether a product is a medical device, typically to determine the applicable VAT rate.
Special MDR categories also apply in Poland: custom-made devices (Annex XIII statement instead of ordinary CE marking route; Polish activity registration since 1 July 2023 — Section 4), systems and procedure packs (Article 22 MDR), in-house devices manufactured and used only within a health institution under Article 5(5) MDR, and Annex XVI products without a medical purpose (non-corrective contact lenses, dermal fillers and similar). An incorrect classification is not a clerical error: it invalidates the conformity route, can mean a missing notified body, and in Poland can be fined as placing a non-compliant device on the market.
Poland imposes no separate Polish representative requirement on top of the MDR for a foreign manufacturer that simply sells into Poland through others. Keep the EU roles and the Polish establishment facts separate — the Polish list duties attach to whoever is established in Poland, in whatever role, and to whoever first introduces a device into Poland.
PRRC: The Person Responsible for Regulatory Compliance (Art. 15 MDR) requirement applies unchanged: a relevant degree plus one year of professional experience, or four years of experience without a degree; the AR must also have a PRRC permanently and continuously at its disposal. Micro and small manufacturers may have the PRRC permanently and continuously at their disposal without employing them.
No physical office is required to sell into Poland. A foreign manufacturer does not need a Polish office, branch or warehouse merely to register and sell devices in Poland. It needs: an EU AR (if non-EU), an importer / distributor chain that meets Articles 13–14, Polish labelling for lay-facing information, EUDAMED registration and — where an operator in its chain is Polish-established or first introduces the device into Poland — that operator's URPL list registration. A Polish office becomes regulatory-relevant mainly if that establishment itself is the manufacturer, AR, importer, distributor or custom-made operator, in which case the corresponding MDR duties plus the Polish list, language and advertising duties attach to it directly.
This is the section most generic "sell into the EU" guides get wrong for Poland, because Poland kept national lists when the EU system became mandatory — and because older summaries still describe the repealed 2010 Act.
The old rule you will still find online — do not build on it. Under the Act of 20 May 2010, a manufacturer or authorised representative domiciled in Poland had to notify the President of the URPL at least 14 days before first placing a device on the Polish market, and Polish distributors / importers had to notify without delay, and no later than seven days after first introduction. That Act was repealed by the 2022 Act. The seven-day concept survives — but as the distributor-list deadline in Section 3, in a different legal form, on a different platform. Any consultant, checklist or internal SOP that still cites a "14-day URPL product notification" as the current Polish route must be re-verified against the consolidated 2022 Act and the live URPL platform before use.
Non-compliance with the new list / registration obligations can attract a financial penalty of up to PLN 200,000 (Baker McKenzie summary of the 2022 Act sanctions).
What changed on 28 May 2026 (EUDAMED). Poland applies the same EU timetable as every other member state:
What Poland did NOT add. There is no Polish equivalent of Italy's Repertorio per-product registration fee for ordinary devices, and no Polish "product approval certificate" on top of CE marking. The Polish extras are: the distributor and user lists, custom-made activity registration, Polish-language supply, the strict advertising regime, and — commercially decisive but legally separate — the NFZ / JGP payment system. A URPL list registration is an administrative registration, not an approval: it does not certify conformity, and the URPL does not "approve" an ordinary CE-marked product through it. Conversely, the absence of a required list registration is a compliance defect the URPL can fine even for a perfectly CE-marked device.
Practical consequence for planning. Budget for two registration systems in 2026 and after — EUDAMED (EU-wide, mandatory, no fee) plus the URPL lists for every Polish first-introducer / Polish-established operator in your chain — plus Polish labelling, plus, if you sell to hospitals, a JGP / NFZ payment project that runs on its own logic (Section 9). Do not assume an EUDAMED registration replaces the URPL distributor list, and do not assume a URPL list entry replaces EUDAMED. In Poland, in 2026, both statements are false.
Step 1 — Confirm the product is a medical device and classify it under Annex VIII MDR (or IVDR Annex VIII for IVDs). Resolve borderline questions (device vs. medicinal product vs. cosmetic vs. biocidal product) early; where classification is disputed with a notified body, the President of the URPL decides by administrative decision. For device–drug combinations, expect the notified body to run the ancillary-substance consultation procedure.
Step 2 — Establish the quality management system aligned to MDR Art. 10(9), normally evidenced by ISO 13485 certification. Notified bodies audit the QMS as part of conformity assessment; it is a prerequisite, not a parallel task.
Step 3 — Compile the technical documentation (Annex II) and clinical evaluation (Annex XIV), including the post-market surveillance plan and PMCF plan where applicable. For IVDs, compile the performance evaluation under the IVDR. Documentation for market surveillance in Poland must be producible in English or Polish — non-EU suppliers in particular should ensure the file that travels with the device into the EU is complete in one of those languages (Wolf Theiss).
Step 4 — Complete conformity assessment with a notified body (all classes above I, and Is / Im / Ir; IVD Classes B, C, D and sterile A). The notified body does not need to be Polish — any body designated for your device codes in NANDO qualifies. Poland's own options are PCBC (NB 1434) and TÜV NORD Polska (NB 2274) for the MDR scopes listed in NANDO; PCBC does not offer every route (for example Annex X in the sector guide), and no Polish body was listed for IVDR in the analyses consulted — plan IVD certification outside Poland. Verify scope in NANDO and obtain written timelines: notified-body capacity is the critical path, in Poland as across the EU.
Step 5 — Draw up the EU Declaration of Conformity (Annex IV) and affix the CE marking (with the notified body's number where a body was involved). There is no separate Polish marketing authorisation for an ordinary device: the Polish system is CE marking plus registrations / lists, policed by the URPL — not a Polish product approval.
Step 6 — Assign UDI (Art. 27 MDR: Basic UDI-DI, UDI-DI, UDI-PI) and prepare UDI carriers for labels, on the staggered MDR timetable by class. Your Polish distributor and any directly-importing hospital will need the Basic UDI-DI for the URPL list within seven days of first introduction — supply it before the goods ship, not after they land.
Step 7 — Appoint the EU Authorized Representative (if you are established outside the EU) and execute the Art. 11 mandate — before registrations, because the AR is a registrant in EUDAMED. Decide deliberately whether the AR sits in Poland (with the Polish list / URPL-facing consequences in Section 3) or elsewhere in the EU.
Step 8 — Decide the Polish establishment and first-introduction question. Who first introduces each device into Poland — your Polish distributor, a Polish importer, a directly-importing hospital, or your own Polish establishment? That actor carries the Article 21 / 22 list duty. Put the seven-day data flow (Basic UDI-DI, manufacturer name / address, type and trade name) into the distribution agreement, with a named owner on both sides. If any company established in Poland will manufacture, represent or import custom-made devices, complete the URPL activity registration before placing the device on the market.
Step 9 — Register the actors and the device in EUDAMED. Obtain the SRN for manufacturer, AR and importer through actor registration (validated by the competent authority of the member state of establishment — for Polish establishments, the URPL), then register the device / UDI data. Mandatory since 28 May 2026 for the first four modules; new devices must be registered before placement on the market.
Step 10 — Complete the URPL list registrations for the Polish chain. Distributor list (first introduction, seven days), user list (directly-importing professional users), and custom-made activity registration (before market placement) as applicable (Sections 3–4). Confirm the live URPL platform form for your operator type at filing — the platforms launched on 1 July 2023 and their forms and submission channels (online platform, ePUAP trusted profile, qualified e-signature, in person, mail) are the authoritative workflow, not third-party summaries.
Step 11 — Prepare the Polish labelling. Translate labels and IFU into Polish as the default for lay users; build the professional-user English position, if you intend to rely on it, as a documented, device-specific decision under Article 12, with patient-facing information in Polish regardless (Section 6). Check the user interface language as a separate item — it has its own Polish rule.
Step 12 — If you will run a clinical investigation or performance study in Poland, file through the URPL / ethics route before starting (see the box below). Do not treat a hospital's local approval as a substitute.
Step 13 — Set up vigilance, post-market surveillance and supply-continuity workflows before first sale: PMS system and reporting calendar (Section 11), a serious-incident reporting workflow, field safety corrective action readiness, and the Article 10a supply-interruption workflow (notify the competent authority and affected actors at least six months in advance, absent exceptional circumstances).
Step 14 — Build the reimbursement and commercial route (separate from registration). Registration makes sale lawful; it does not create a payer. Work out, before you price the Polish launch, whether the device is paid inside a JGP group, inside a network hospital's adjusted budget, as an add-on / separately-financed service, directly by the Ministry of Health as a highly specialised service, or through the outpatient Reimbursement Act list (Section 9). For hospital sales generally: Polish public hospitals buy under public procurement law — expect tenders where, in the sector summaries consulted, price is still the primary determinant for bundled devices, with formal ESPD-type documentation and Polish-language tender documents.
> Clinical investigations and performance studies in Poland (URPL + ethics committee). The 2022 Act sets the Polish rules and procedures for conducting a clinical investigation of a medical device and performance testing of an IVD, within the MDR Articles 62–82 / IVDR Articles 58–82 framework: sponsor notification / authorisation through the URPL via EUDAMED / national route as applicable at filing, plus a favourable ethics committee opinion, before the investigation starts in Poland. Patient and participant information and consent documents must be in Polish (Commission language table: Polish for documents intended for study participants and patient information). Substantial modifications, safety reporting (serious adverse events / device deficiencies) and annual safety reporting follow the MDR / Polish timetable. The URPL is empowered to inspect trial sponsors.
What does NOT exist in Poland: a URPL "approval certificate" for an ordinary CE-marked device, a Polish CE mark, and a per-product national marketing register on top of EUDAMED of the Italian Repertorio type. The proof of lawful Polish market access is: CE certificate + EU Declaration of Conformity + EUDAMED registrations + URPL list registration for the Polish first-introducer in the chain (+ a JGP / reimbursement position if you want the public system to pay). Anyone selling you a Polish "product approval" beyond this is selling you a filing service, not an approval.
Poland's language rule is written into Article 12 of the 2022 Act, and it is one of the two Polish rules (with advertising) that most often forces a late, expensive rework of a pan-EU pack.
Practical rule for planning: budget and produce full Polish labels and IFU for anything that reaches a patient, a lay user, an implant card, a field safety notice or a Polish authority. English-only supply is defensible only for genuinely professional-only devices, as a documented Article 12 position, with patient information in Polish and the interface question answered explicitly. Polish hospitals and distributors will reject English-only patient material at vendor qualification even where a legal argument exists for professional material. Secondary sources cite label-related fines in Poland — for example up to PLN 250,000 for missing importer details — and incorrect labelling is listed among the triggers the URPL treats as non-compliance.
Electronic IFU (eIFU) follows the EU eIFU regulation framework; the Polish language duty is unchanged by the medium.
| Activity | Typical timeline |
|---|---|
| CE certification via notified body (Class IIa, new submission) | 12–18 months as a realistic planning figure; notified-body capacity remains the critical path EU-wide, and Polish-designated capacity is limited — most manufacturers certify outside Poland |
| CE certification (Class III) | 18–24+ months, including the clinical evaluation consultation procedure where it applies |
| Class I self-declaration route | 2–4 months once technical documentation and Polish labelling are complete. One secondary source (OMC Medical) cites 4–6 weeks for the Class I registration process after submission |
| EUDAMED actor registration / SRN issuance | Weeks, after competent-authority verification (for Polish establishments, URPL validation). Mandatory from 28 May 2026 for the first four modules |
| EUDAMED device registration (new device, from 28 May 2026) | Must be completed before placing the first unit on the market; allow weeks for data preparation and validation, not days |
| Registration of pre-28 May 2026 devices in EUDAMED | By late November 2026 ; a URPL list entry does not discharge this duty |
| URPL distributor list (first introduction into Poland) | Data submitted within 7 days from the date of introducing the device into Poland — this is a deadline after introduction, not a pre-approval wait. Build the data flow before the goods ship |
| URPL custom-made activity registration (Polish-established operators) | Before placing the device on the market; allow weeks for account, form signature (hand / ePUAP trusted profile / qualified e-signature) and submission |
| Clinical investigation | MDR timetable (validation and assessment phases) routed in Poland as URPL + ethics committee review, plus local feasibility at each centre |
| Reimbursement — outpatient list route (Reimbursement Act) | Statutory procedure is described in legal summaries as among the longest in Europe, although it should be finished within 180 days; HTA (AOTMiT), Economic Commission negotiation on official selling price and payment level, and ministerial decision are sequential steps, not parallel ones |
| Reimbursement — hospital JGP route | No manufacturer "application timeline" in the ordinary case: payment follows the existing JGP group / budget / add-on structure, or it does not. A genuinely new procedure code or tariff is an evidence and policy project — allow 12–24 months as a planning range |
| Hospital tender cycles | Not a registration step — but allow 6–18 months from registration to first meaningful public-sector revenue; Polish hospitals buy under public procurement law on their own calendars |
Transition deadlines for legacy (MDD / AIMDD) devices (Regulation (EU) 2023/607) — the same EU-wide dates apply in Poland:
Eligibility conditions: continued directive compliance, no significant change in design or intended purpose, no unacceptable risk, an MDR-compliant QMS in place by 26 May 2024, a formal notified-body application by 26 May 2024, and a written notified-body agreement by 26 September 2024. IVDR legacy deadlines (Regulation (EU) 2024/1860) are staggered by class: Class D by end-2027, Class C by end-2028, Class B and sterile Class A by end-2029.
Poland's national filing fees for ordinary devices are modest to non-existent; the real Polish costs are certification, Polish translation (including the interface), advertising-compliance rework if you get it wrong, and the reimbursement / tender evidence work. The financial risk that dwarfs all filing fees is the penalty scale.
Currency: official Polish fees and all penalties are in Polish złoty (PLN). Budget in PLN, not euros, for URPL-facing items.
Legal market access is only the first gate in Poland. The second gate is how the public system pays, and the third — unique in its strictness in Europe — is what you may say about the device once you sell it.
MTR Consulting's market-access summary describes Poland with one key theme: reimbursement is payment — via the DRG-adjusted budget and fee-for-service (add-on) mechanisms. There is no separate "funding approval" step, because the coverage decision is made in parallel with the creation of the procedure code, and there are no stand-alone HTA activities, because HTA is an integral part of reimbursement decision-making.
In practice (MTR / Wolf Theiss summaries):
A separate route exists for devices reimbursed outside hospital procedures, under the Reimbursement Act of 12 May 2011. In the Global Legal Insights summary: a device may be reimbursed if it is authorised / marketed, available on the Polish market, and has a GTIN or equivalent identification code. The procedure is initiated by an application to the Minister of Health by an authorised applicant (for devices, a business entity as defined in the MDR), submitted electronically only through the Reimbursement List System (SOLR) since 1 January 2018, with the mandatory attachments in Articles 24–25 of the Act. After formal examination, AOTMiT and its President may be involved (HTA: clinical effectiveness, economic analysis, health-system impact), then negotiations with the Negotiation Team / Economic Commission on the official selling price, payment level and reimbursed indications, then the ministerial decision. Legal summaries describe the Polish procedure as still one of the longest in Europe, although it should finish within 180 days. Do not promise a board an outpatient listing date on the basis of the 180-day figure alone.
There is no specific reimbursement framework for IVD tests in Poland (MTR). Most IVD tests are funded on a global budget principle as part of laboratory funding. A limited number of expensive IVD tests are reimbursed fee-for-service using add-on payment categories in the DRG system. Practical consequence: an IVD's Polish business case is usually a laboratory-budget and tender case, not a per-test reimbursement case, unless the test is one of the expensive add-on categories.
Poland regulated device advertising more broadly than the EU (CMS, Taylor Wessing, Wolf Theiss). The regime has applied since 1 January 2023, with a transition to 30 June 2023 for advertising already disseminated — long expired. It is supervised by the President of the URPL. In substance, for advertising to the public:
Penalty: advertising in breach can be fined up to PLN 2,000,000. The URPL can also order cessation / correction measures for misleading advertising under Article 7 MDR / IVDR. Every Polish campaign asset — including global social content geo-targeted at Poland, influencer content, and clinic co-marketing — needs a Polish advertising review before publication. This is not a translation task; global creative built around a clinician testimonial is unlawful for the Polish public audience in its original form.
Post-market duties are MDR duties, enforced in Poland by the URPL with Polish inspection powers and Polish fines.
Do I need a separate Polish approval on top of CE marking? No. There is no Polish product approval or Polish CE mark for an ordinary device. You need CE marking, EUDAMED actor and device registration, the URPL list registration that applies to the Polish first-introducer in your chain, Polish labelling, and — commercially — a JGP / reimbursement position.
Who is the Polish regulator? The President of the Office for Registration of Medicinal Products, Medical Devices and Biocidal Products (URPL), based in Warsaw. The URPL implements the MDR/IVDR, supervises Polish notified bodies, runs market surveillance, vigilance, clinical investigations, the distributor and user lists, and advertising supervision, and imposes administrative fines.
Does my EU Authorized Representative have to be in Poland? No. Any EU-established AR covers Poland under MDR Article 11. A Polish-established AR creates additional Polish operator duties in practice (and, for custom-made devices, an activity-registration duty) — choose the AR's seat deliberately.
What is the seven-day rule in Poland? A Polish-based distributor that first introduces a device, system or procedure pack into Poland must submit its data — Basic UDI-DI, manufacturer name and address, type and trade name — to the URPL List of Distributors within seven days of introduction. It comes from Article 21 of the 2022 Act, in force for the lists from 1 July 2023.
Do I still need to notify the URPL 14 days before placing a device on the market? That was the rule under the repealed Act of 20 May 2010, still widely reproduced online. Do not rely on it as the current route. Confirm at filing whether any category-specific notification still applies to your device — the current general system is EUDAMED plus the URPL lists described in Section 4.
Can I supply the device in English only? Only, in defined circumstances, for genuinely professional-only devices under Article 12 — and even then patient-specific information must be in Polish, and the interface rule must be answered. Lay-user labelling, IFU, implant cards, field safety notices and patient documents must be in Polish.
Can I use my global campaign in Poland? Not without a Polish advertising review. Since 1 January 2023, public advertising may not show or suggest a healthcare professional, may not promote professional-only devices to the public, may not appeal directly to children, must be lay-understandable, must carry the required device / manufacturer / advertiser identification and warnings, and ad templates must be kept for two years. Breach: up to PLN 2,000,000.
How are hospital devices paid for in Poland? Mainly inside the JGP (Polish DRG) system: network hospitals (63% in 2022) through activity-adjusted budgets, some services through separate DRG case-by-case or fee-for-service add-on categories, highly specialised services directly by the Ministry of Health, and non-network hospitals through NFZ contract tenders paid case-by-case. The procedure code (ICD-9-PL) plus diagnosis code (ICD-10-PL) determines the group; AOTMiT maintains tariffs.
How long does Polish registration take? CE certification is the long pole: 12–18 months for a new Class IIa, 18–24+ months for Class III, as planning figures. EUDAMED and URPL list work is weeks — but the distributor list has a seven-day deadline after first introduction, so it must be prepared before shipment, not started after it.
What are the biggest fines in Poland? Up to PLN 5,000,000 for placing a non-compliant device on the market, up to PLN 2,000,000 for non-compliant advertising, and up to PLN 200,000 for distributor / user list and custom-made registration failures — all under the 2022 Act, imposed by the President of the URPL.
Can a Polish notified body certify my device? For some MDR scopes, yes — PCBC (NB 1434) and TÜV NORD Polska (NB 2274) are Polish-designated. For many codes and for IVDR, you will in practice certify with a notified body outside Poland, which is fully valid for the Polish market. Verify scope in NANDO at application.
Does ghmap.io help with Poland? ghmap.io works feasibility-first — Doctor → Hospital → Distributor → Manufacturer: device samples are tested by hospitals and doctors in the target market, and distributor engagement follows clinical acceptance and hospital demand, not the other way round. For Poland that means validating clinical fit and the JGP / tender payment story with Polish hospitals first, then structuring the distributor chain — including the Article 21 list duty and Polish labelling — around proven demand. No prices are published in this guide; engagement is scoped per project.
This guide is market-access intelligence, not regulatory or legal advice. URPL practice, NFZ contracting and AOTMiT processes change — professional review is required before business use.
Market Access Strategy Session
A Market Access Strategy Session maps your Poland route — URPL list duties, Polish labelling and advertising compliance, and an NFZ/JGP payment strategy.