Romania Medical Device Registration2026 Field Guide

No Romanian CE mark — EU MDR applies directly, and Romania has no notified body of its own. Romania adds an activity authorisation on top: importers, distributors and service providers need an ANMDMR operating permit (aviz de funcționare) before they start, alongside mandatory EUDAMED from 28 May 2026 and a national database that binds only Romania-established representatives and manufacturers. Add a Romanian-language rule, prior approval for public advertising, and a payment system split between CNAS and SICAP tenders.

operating permit validity (aviz)
3 yrs
of GDP spent on health
5.8%
hospital beds per 1,000 population (2023)
7.3

Sourced from the ANMDMR, Law No. 95/2006, Order No. 566/2020 and the OECD, and verified 2026 publications.

Download the guide

Enter your work email and we'll send the PDF straight to your inbox.

Professional review is required before business use. See our privacy policy.

What's inside

Thirty pages, zero fluff

Every figure in this guide traces to a named source — ANMDMR, Regulation (EU) 2017/745 and Regulation (EU) 2017/746, Law No. 95/2006, Emergency Ordinances No. 46/2021 and No. 137/2022, Ministerial Orders No. 566/2020, No. 3539/2022 and No. 3467/2022, and verified 2026 legal and trade publications.

EU MDR + the Romanian layer

Law No. 95/2006 and Order No. 566/2020 carry the national layer: the ANMDMR, created in its current form by Law No. 134/2019, authorises the activities behind the market — import, distribution, installation and maintenance — and surveils devices in operation. A CE file alone authorises none of them.

EUDAMED + record + operating permit

EUDAMED actor registration is mandatory from 28 May 2026 — and Romania's national database is narrower than most guides say: it binds authorised representatives established in Romania and Romanian manufacturers, not representatives established elsewhere in the EU. The operating permit (aviz de funcționare) is the gate that stops shipments.

Romanian language + tender access

Labels and instructions must reach the end user in Romanian, with a narrow Romanian-or-English opening for professional users. Public advertising of devices needs ANMDMR approval before it runs — a tariffed visa — while hospital revenue is decided in SICAP tenders and CNAS programme budgets.

Complete guide

The guide in full

Every section of the PDF on one page. Use the form above to keep a copy.

Overview

Last updated: November 2026. Romania is a full EU member state, so the EU Medical Device Regulation (MDR, Regulation (EU) 2017/745) and the In Vitro Diagnostic Regulation (IVDR, Regulation (EU) 2017/746) apply directly in Romania. The Romanian national layer is built around the National Agency for Medicines and Medical Devices of Romania — ANMDMR in Romanian, NAMMDR in the agency's English materials — the institutional ordinances that make the MDR and IVDR work in Romania, and the Romanian rules on national-database registration, operating permits for importers and distributors, language, advertising, vigilance and public procurement that sit on top of the MDR.

Always confirm critical steps, the current NAMMDR forms channel and the current indexed fee amounts against NAMMDR guidance in force at filing before you file. Where this guide cannot verify a current figure from a primary source, it says so and marks it to be verified.

Romania is a market of roughly nineteen million people — the second-largest market in south-eastern Europe and one of the larger hospital systems in the EU's east — served by a public hospital network organised under the Ministry of Health and county authorities, a single dominant social health insurance system administered by the National Health Insurance House (Casa Națională de Asigurări de Sănătate, CNAS) through county health insurance houses, and a growing private hospital and clinic sector concentrated in Bucharest, Cluj-Napoca, Timișoara, Iași and the other regional centres. For a foreign manufacturer, Romania is not entered on population alone. It is entered because three systems have to line up at the same time: the EU conformity system that puts the CE mark on the device, the Romanian registration and operating-permit system administered by NAMMDR that makes the device and the operator lawfully present on the Romanian market, and the Romanian payment system — CNAS reimbursement for defined devices and programmes, and public procurement through the electronic SICAP platform (the system still widely called SEAP) for devices supplied into public hospitals — that decides whether anyone actually pays for it. A device can be perfectly CE-marked and still not sell lawfully or commercially in Romania if the national-database position has not been worked through, if the Romanian importer or distributor does not hold the right operating permit, if the Romanian-language labelling set is missing, or if the procurement and reimbursement logic has never been planned.

Regulatory reality check up front: Romania is not a separate certification regime from the EU. There is no "Romanian CE mark" and NAMMDR does not grant a pre-market approval for ordinary medical devices in the way a non-EU regulator grants a registration certificate. NAMMDR is the Romanian competent authority and market surveillance authority for medicines and medical devices, operating under the Ministry of Health and reorganised in its current form by Law No.

134/2019. It validates Romanian registrations in EUDAMED, maintains the Romanian national database of medical devices, issues operating permits (aviz de funcționare) for import, distribution, installation and maintenance activities, authorises clinical investigations, approves advertising material, handles vigilance and field safety corrective actions, supervises devices in use in Romanian healthcare units, and enforces the sanctions chapter of Law No. 95/2006. The manufacturer remains responsible for compliance; a conforming device bears the CE mark and can then be traded across the EU/EEA. What Romania adds — and what foreign manufacturers most often underestimate — is a national layer with seven specific features:

1. Romania kept a national database alongside EUDAMED — and who must register in it is narrower and stranger than most guides say. Under Minister of Health Order No. 3539/2022 — the methodological rules for placing medical devices on the market and registering economic operators in EUDAMED and in the national database — registration in the Romanian national database is required, in the cases NAMMDR itself publishes, for authorised representatives established in Romania when they place on the market Class I, IIa, IIb and III devices (including sterile and measuring devices), systems and procedure packs under MDR Article 22, and active implantable devices, and for Romanian manufacturers, including Romanian manufacturers of custom-made devices placing devices on the market under their own name (see also Order No. 2845/2022 on custom-made devices). NAMMDR states expressly that authorised representatives established in other EU Member States are NOT registered in the Romanian national database, and that free sale certificates for export are issued only to manufacturers and authorised representatives established in Romania. Do not copy a generic "register every device nationally in every EU country" checklist into Romania, and do not copy the opposite error either: if your EU authorised representative is established in Romania, or you manufacture in Romania, the national-database duty is real, tariffed, and enforced. Doğrulanacak: the exact current scope of national-database registration for your operator type and device family at filing — NAMMDR has been aligning national practice with mandatory EUDAMED during 2026, and the division of work between the two systems must be confirmed for your category.

2. The operating permit is the Romanian layer that actually stops shipments. Under Minister of Health Order No. 566/2020, approving the methodological rules for Title XX of Law No.

95/2006, three activities in Romania are subject to approval control: (a) import of medical devices, (b) distribution of medical devices, and (c) installation and/or maintenance of medical devices. A Romanian importer or distributor without a valid operating permit covering the activity, the work point and the manufacturers it handles is not a lawful Romanian channel, however good its sales team is. The permit is issued, renewed and modified by NAMMDR on a documented file (application Annex 1 or Annex 5, document lists in Annexes 2, 3 or 6, the manufacturers table in Form F6, staff and PRRC forms), is reported in the secondary literature as valid for three years if the conditions on which it was issued remain unchanged, and must be modified when manufacturers, activities or work points change. From 2026, NAMMDR also requires an operator seeking an import permit (or a modification adding import) to already hold an importer Single Registration Number (SRN) from EUDAMED actor registration. Vet a Romanian partner on its permit first — scope, manufacturers listed, work points, expiry — and only then on price.

3. EUDAMED became mandatory during 2026 — Romania follows the EU date, not a Romanian one. By Commission Decision (EU) 2025/2371 of 26 November 2025, the first four EUDAMED modules — actor registration, UDI/device registration, notified bodies and certificates, and market surveillance — were declared functional, and the related obligations apply from 28 May 2026. NAMMDR validates EUDAMED actor registrations for Romanian-established manufacturers, authorised representatives, importers and system/procedure pack producers, charges a published tariff for validation/verification of EUDAMED registration, and may validate, reject or request corrections. A device first placed on the market under the mandatory regime must be registered in EUDAMED before first placement; devices already on the market follow the transitional deadlines in the MDR as amended. Receiving an SRN does not register the device: actor registration and device registration are separate acts.

4. Romania has no MDR or IVDR notified body of its own. NAMMDR states this expressly: currently Romania has no MDR-designated or IVDR-designated notified body. A Romanian manufacturer — and any manufacturer whose file a Romanian partner is managing — must therefore contract a notified body designated in another EU/EEA state whose NANDO scope covers the device and the conformity route. That means foreign-language notified body queues, foreign audit travel and longer, less controllable timelines are a structural feature of the Romanian route for Class IIa and above, not a local failure. There is no rule requiring a Romanian body for the Romanian market, because there is none to require.

5. Romanian is the language of the file, with a narrow professional-use English path that needs approval. The European Commission's MDR language table for Romania (Revision 3, August 2025), citing Emergency Government Ordinance No. 46/2021 and Order No. 3539/2022, records Romanian for labelling and instructions for use, with Romanian-or-English possibilities in defined cases — in particular for devices intended exclusively for professional users, on the basis of the healthcare professional's written consent/request and, in the cases the table records, the competent authority's approval. Patient and lay-user information has no such shortcut. A file that passed in an English-tolerant market will fail in Romania on language alone if the Romanian set is missing, exactly as in Greece, Poland or Portugal. Field safety notices reaching Romanian users must be handled in Romanian in the cases the table records.

6. Advertising to the public is a prior-approval activity in Romania, and it is tariffed. Under Emergency Government Ordinance No. 46/2021, promotion of medical devices in Romania covers medical-representative visits, samples, sponsorship of promotional and scientific events and the supply of information to professionals and the public. Only devices that can be used without medical intervention may be advertised to the public, public-facing advertising material requires NAMMDR approval, and NAMMDR's tariff list prices approval of advertising material and of educational material as separate, time-limited visas (see Section 8). Healthcare-professional material is not prior-approved in the same way, but NAMMDR may check it for compliance, and misleading claims collide with MDR Article 7 as well as Romanian law. Never launch a Romanian consumer campaign on the strength of an approval obtained in another Member State.

7. For public-hospital sales, Romania is an electronic-procurement market with a payer, not a single buyer. Public hospitals and other contracting authorities buy through SICAP, Romania's electronic public procurement system (successor branding to SEAP), under the Romanian public procurement framework supervised by the National Public Procurement Agency (ANAP), which has published a dedicated guide for the procurement of medical devices. Reimbursement from public health funds, in the cases where a device is reimbursed outside or alongside hospital payment, runs through agreements with the health insurance houses and at set prices and conditions. There is therefore no single Romanian door: there is the individual hospital's SICAP procedure, centralised procurement where it applies to your category, the private hospital groups buying commercially, and the CNAS agreement layer where your device is a reimbursed device.

Doğrulanacak: whether your category is bought centrally, reimbursed by CNAS, or both, at entry — this is category-specific and changes by programme and by year.

A note on institutions: the Ministry of Health (Ministerul Sănătății) sets policy and issues the ministerial orders that form most of the Romanian device layer. NAMMDR / ANMDMR, headquartered in Bucharest (Aviator Maior Ștefan Sănătescu Street No. 48, Sector 1), is the competent authority, the market surveillance authority, the EUDAMED validating authority for Romanian operators, the operator of the national database and the issuer of operating permits, free sale certificates and advertising approvals. CNAS and the county health insurance houses are the payer layer. ANAP supervises public procurement policy; SICAP is the transactional platform. The National Trade Register Office (ONRC) is where Romanian companies — including your distributor and any Romanian manufacturer — are registered, and NAMMDR's 2025 notices require Romanian manufacturers notifying devices to evidence ONRC registration and the CAEN activity code specific to the device type.

1. The Regulatory Framework: EU MDR/IVDR (Directly Applicable) + Romanian National Layer

1.1 The EU layer (applies directly in Romania — no transposition step)

  • Regulation (EU) 2017/745 (MDR) — applicable since 26 May 2021: classification (Article 51 and Annex VIII, 22 rules), General Safety and Performance Requirements (GSPR, Annex I), conformity assessment (Article 52), CE marking (Article 20), UDI (Articles 27–28), EUDAMED registration (Articles 29–31), clinical evaluation and investigations (Articles 61–82), post-market surveillance and vigilance (Articles 83–92), the ban on misleading claims (Article 7), and the supply-interruption notification duty in Article 10a (applicable since 10 January 2025, introduced by Regulation (EU) 2024/1860).
  • Regulation (EU) 2017/746 (IVDR) — applicable since 26 May 2022, with staggered transition for legacy IVDs by class (to end-2027, end-2028 and end-2029 depending on class, subject to the Article 110 conditions).
  • Regulation (EU) 2023/607, as amended by Regulation (EU) 2024/1860 — extended MDR transition for eligible legacy devices to 31 December 2027 for Class III and Class IIb implantable devices (with stated exceptions) and to 31 December 2028 for other Class IIb, Class IIa and Class I devices that require a notified body, subject to the Article 120 conditions, including a timely formal application to a notified body and a signed written agreement. A legacy device that does not meet those conditions may not be placed on the Romanian market on the strength of an old directive certificate.
  • Annex XVI products without a medical purpose — the MDR common specifications (Commission Implementing Regulation (EU) 2022/2346) apply to listed non-medical products such as coloured contact lenses, cosmetic implants and fillers, liposuction equipment and certain lasers and light-based equipment.
  • EUDAMED mandatory modules — following Commission Decision (EU) 2025/2371, use of the Actor Registration, UDI/Device Registration, Notified Bodies & Certificates and Market Surveillance modules became mandatory EU-wide from 28 May 2026, including in Romania. NAMMDR validates Romanian actor applications and Romanian-linked registrations, and charges the published validation tariff in Section 8.

1.2 The Romanian layer

InstrumentWhat it does
Law No. 95/2006 on healthcare reform, republished, Title XX — Medical DevicesThe parent Romanian health law for devices: the framework NAMMDR's methodological rules implement, including the market surveillance and sanctions chapters (Articles 934–936 are the sanctions provisions NAMMDR cites for devices). Read in its current republished, amended form — never from an old consolidated copy.
Law No. 134/2019Reorganises NAMMDR (the merger of the former medicines agency and the former medical devices technical office into the current agency) and is cited as a legal basis throughout NAMMDR's current orders and tariffs.
Emergency Government Ordinance (EGO) No. 46/2021Establishes the institutional framework and measures for applying the MDR in Romania: designates the competent authority, sets the Romanian language rules relied on in the Commission language tables, the advertising and promotion regime for devices, free sale certificates, vigilance centralisation and the Romanian derogation (Article 59 MDR) machinery.
Emergency Government Ordinance No. 137/2022 (Official Gazette No. 995 of 13 October 2022)The IVDR counterpart to EGO 46/2021: designates NAMMDR as competent and market surveillance authority for IVDs, and sets the Romanian rules on free sale certificates (IVDR Article 55), central registration of serious incident and field safety reporting, and performance studies. Note the Romanian IVDR framework arrived after the IVDR application date — secondary Romanian IVD practice was still settling in the following years.
Minister of Health Order No. 3539/2022The core methodological rules for placing medical devices on the market and registering economic operators in EUDAMED and in the national database, and for derogations from conformity assessment procedures. This is the order NAMMDR's national-database, authorised-representative and language notices rely on. Doğrulanacak: the order's exact Official Gazette publication date and its current amended form at filing — Romanian sources cite it both by its November 2022 adoption and its December 2022 publication.
Minister of Health Order No. 2845 of 29 September 2022Methodological rules, under Article 13 of EGO 46/2021, for the registration of custom-made devices placed on the market under their own name by manufacturers based in Romania.
Minister of Health Order No. 566/2020Approves the methodological rules for applying Title XX of Law 95/2006 to the approval of activities in the field of medical devices: import, distribution, installation and/or maintenance. Contains the operating permit procedure, the document lists (Annexes 1–6), the evaluation and invoicing timetable (the invoice is issued within a maximum of 90 days of application under Article 10(1), as NAMMDR publishes it), and the temporary/emergency approval route described in the secondary literature.
Minister of Health Order No. 3467 of 17 November 2022 (Official Gazette Part I, No. 1123 of 22 November 2022)Approves the tariffs for NAMMDR's medical device activities — the source of every fee figure in Section 8 of this guide. Tariffs are in lei, are payable in advance unless the governing rules provide otherwise, are indexed annually by 31 January by the previous year's average inflation rate (from January 2024), and are not refunded for rejected applications. The figures in Section 8 are therefore the 2022 base tariffs: to be verified — the indexed amount actually payable in the filing year, on NAMMDR's current Forms and fees page, before payment.
Minister of Health Order No. 330 of 13 February 2023Methodological rules for clinical evaluation and clinical investigations with medical devices in Romania, replacing the 2006 procedure (Order No. 792/2006). Read together with MDR Articles 62–82 and the ethics committee layer for any Romanian investigation site.
Minister of Health Order No. 2219 of 14 July 2022Control by periodic verification of medical devices in use, performance evaluation of second-hand devices put into service, and issuance of approvals for use of devices in healthcare units and in pre-hospital emergency intervention vehicles (MIAMUP). This is the Romanian "devices in use" layer hospitals and service providers live under after installation.
Minister of Health Order No. 650/2025 (published 27 February 2025)Procedural rules for applying MDR Article 5(5) to devices manufactured and used only within healthcare institutions in Romania, including notification of such devices to NAMMDR on form F.MD/IVD_IS, charged as a tariffed NAMMDR activity.
Romanian public procurement framework and the ANAP Guide for Procurement of Medical DevicesHow public hospitals buy: electronic procedures in SICAP, ANAP policy supervision, and the joint ANAP–AmCham best-practice guide for medical device procurement (planning, technical specifications referencing ISO/IEC/EN standards, and evaluation indicators). Doğrulanacak: the current procurement law thresholds, direct-award limits and any centralised category list at tender — thresholds are revised periodically and must never be taken from a 2021 summary.
CNAS / health insurance house reimbursement frameworkThe payer layer for devices reimbursed from public health funds: reimbursement is performed on the basis of agreements with the health insurance houses, for devices meeting specific requirements and at set prices, as the CEE/SEE medical devices guides summarise the Romanian position. Doğrulanacak: whether your device is a CNAS-reimbursed device at all, and if so the current programme, price and agreement conditions, at entry.

1.3 Who does what

  • Manufacturer — full MDR Article 10 duties. If established in Romania: EUDAMED actor registration validated by NAMMDR, national-database registration of the devices it places on the market in the cases Order 3539/2022 covers, ONRC company registration with the CAEN code specific to the device type evidenced in the notification file (per NAMMDR's 2025 notice to Romanian manufacturers), and Romanian-language labelling and IFU. If established elsewhere in the EU and supplying Romania: no second Romanian marketing authorisation, but the Romanian language, channel-permit and payment layers in this guide still have to be worked through with the Romanian chain. If established outside the EU: an EU authorised representative is mandatory (see Section 3).
  • Authorised representative (AR) — mandatory for manufacturers established outside the EU (MDR Article 11), established anywhere in the EU. Romania adds the specific rule in point 1 above: only an AR established in Romania registers in the Romanian national database (and only a Romania-established AR or manufacturer can obtain a Romanian free sale certificate). An AR established in Germany or Ireland who covers Romania for MDR purposes does not thereby acquire a Romanian national-database file — and does not need one on NAMMDR's published position.
  • Importer — MDR Article 13 duties; the first EU operator making a third-country device available.

A Romanian importer must additionally hold a NAMMDR operating permit for import under Order 566/2020 and, from the 2026 EUDAMED regime, must hold an importer SRN before the import permit is issued or modified. It verifies CE marking, labelling, UDI and the EU Declaration of Conformity, and carries Romanian traceability and complaint duties.

  • Distributor — MDR Article 14 duties, plus the Romanian operating permit for distribution (and for installation/maintenance if it services devices), with the manufacturers it distributes listed in its permit file. This is the operator NAMMDR, CNAS, ANAP-supervised buyers and the hospitals will actually deal with day to day. Vet a Romanian distributor on its permit scope and expiry, its SICAP record, its warehousing and traceability procedures and its service capability — before signing, not after the first hospital asks for the permit.
  • Custom-made device manufacturers based in Romania — register under Order 2845/2022 and the MDR Annex XIII statement procedure rather than CE marking. NAMMDR registers information on Romanian custom-made manufacturers, their headquarters and the devices they place on the market, and holds vigilance information in the national database. Do not label a mass-produced device adapted to a professional user's requirements as "custom-made" to use this route.
  • Health institutions that manufacture and use devices in-house (MDR Article 5(5)) — subject to the MDR conditions for in-house manufacture and use, and in Romania to the notification procedure in Order 650/2025.
  • Servicing organisations — installation and/or maintenance is itself a permitted activity in Romania. A foreign manufacturer's Romanian service partner needs its own operating permit covering installation/maintenance for the device groups it services (Form F7 groups, trained personnel, measuring and monitoring equipment in Forms F8–F10), and devices in use are subject to periodic verification under Order 2219/2022.

2. Device Classification

Romania applies the MDR classification rules (Annex VIII) unchanged — there is no Romanian reclassification layer. NAMMDR's own classification guidance states that medical devices are classified into four risk classes — I, IIa, IIb and III — taking into account the intended purpose and inherent risks, that the manufacturer of a plain Class I device is responsible for its own conformity assessment and CE marking, that manufacturers of Class I sterile, Class I measuring, IIa, IIb and III devices must involve a notified body, and that classification should follow MDCG 2021-24, the Commission's guidance on classification of medical devices. Classification is by intended purpose and inherent risk across 22 rules in four groups (non-invasive Rules 1–4, invasive Rules 5–8, active Rules 9–13, special Rules 14–22). Where more than one rule applies, the rule resulting in the higher class prevails. Accessories are classified in their own right, software is classified under Rule 11, and the classification rationale must be documented in the technical documentation.

ClassTypical examplesConformity route in brief
I (incl. Is sterile, Im measuring, Ir reusable surgical)Non-invasive low-risk devices, simple dressings, non-sterile examination gloves, wheelchairs, simple instrumentsSelf-declaration for plain Class I (Article 52(7)); notified body involvement limited to the sterile, measuring or reusable-surgical aspect for Is/Im/Ir
IIaShort-term invasive devices, hearing aids, dental materials, many monitoring and diagnostic devices, most decision-support software as the starting point under Rule 11Notified body (Annex IX QMS + technical-documentation assessment on at least one representative device per category, or Annex X + XI route); typical end-to-end timelines in current EU practice of around 12–18 months
IIbLong-term invasive devices, infusion pumps, ventilators, most implants not in Class III, active therapeutic devices with potentially hazardous energy exchangeNotified body, with technical-documentation assessment per generic device group; commonly around 12–18 months, longer for implantables
IIIHeart valves, implantable pacemakers and neurological devices, absorbable implants, breast implants, devices incorporating an ancillary medicinal substance, defined surgical meshesNotified body with full technical-documentation / design examination; Summary of Safety and Clinical Performance (SSCP) publicly accessible via EUDAMED; expert-panel or clinical-evaluation consultation and, for ancillary medicinal substances, competent-authority/EMA consultation where applicable; commonly 18–24 months
IVDs (IVDR)Classes A–D under IVDR Annex VIII (Rules 1–7)A: self-declare (notified body for sterile A); B–D: notified body, with the heaviest scrutiny, EU reference laboratory involvement and common-specification compliance for Class D

Three Romanian practical notes:

1. If you need a classification answer from Romania, it is a paid NAMMDR service. Order 3467/2022 tariffs a standard classification of a product (one product per request), contesting a classification, and a complex classification of a borderline product as three separate activities (see Section 8). Budget for the borderline route for substance-based, combination and software-borderline products rather than assuming a free informal opinion.

2. Do not import an old MDD class into a Romanian filing. Software (Rule 11), substance-based devices, nanomaterial devices and reusable surgical instruments in particular moved up under the MDR. NAMMDR, a (foreign) notified body, a CNAS checker or a hospital buyer checking your EUDAMED record will work from the MDR class — and any Romanian national-database record must match it.

3. There is no Romanian notified body to ask. See point 4 of the reality check: every Romanian Class IIa+ file goes to a notified body designated in another Member State. Obtain that body's own current lead time in writing before promising a Romanian launch date, and check its NANDO scope for your device codes at application.

Software deserves a separate line for Romania: MDR Rule 11 puts most software that provides information used for diagnostic or therapeutic decisions at Class IIa or higher, and higher where the decision could cause death or serious deterioration of health. Classify the claim you actually make in the Romanian IFU — the Romanian text is the text Romanian users, NAMMDR and a Romanian court will read.

3. The EU Authorised Representative, Importer and Distributor in Romania

A non-EU manufacturer (for example a Turkish, US, UK or Chinese maker — note that Great Britain is a third country for this purpose, while Northern Ireland follows EU goods rules under the Windsor Framework) selling into Romania needs the standard MDR chain, with Romanian specifics at each link:

1. EU Authorised Representative — appointed by written mandate (MDR Article 11), established anywhere in the EU. NAMMDR's published clarification (28 February 2023) restates the rule precisely: only manufacturers outside the European Union are required to appoint a single authorised representative to place the device on the EU market. The AR verifies that the EU Declaration of Conformity and technical documentation have been drawn up, keeps documentation available for NAMMDR and other competent authorities for at least 10 years after the last device is placed on the market (15 years for implantables), cooperates on corrective actions, and is jointly liable for defective devices on the same basis as the manufacturer where the manufacturer has not complied with its obligations. Romanian twist: if the AR is established in Romania, it must register in the NAMMDR national database for the device types listed in point 1 of the reality check, and NAMMDR is its validating authority in EUDAMED. If the AR is established in another Member State, it is not registered in the Romanian national database.

2. Importer — the first EU operator making a third-country device available, with Article 13 verification (CE marking, labelling, UDI, EU Declaration of Conformity, manufacturer and AR identification), storage and transport conditions that do not jeopardise compliance, and complaint-register and traceability duties. A Romanian-established importer registers in EUDAMED (validated by NAMMDR), must hold an importer SRN for its operating permit, and must hold the Order 566/2020 operating permit for import covering the manufacturers it imports.

3. Distributor — Article 14 verification and traceability duties, plus the Romanian layer: a valid operating permit for distribution (and for installation/maintenance where applicable), the manufacturers it handles listed alphabetically in its permit file (Form F6) with the exact names from the declarations of conformity, a named Person Responsible for Regulatory Compliance (PRRC) and/or warehouse manager (Form F9), documented work procedures for the permitted activities, and the ability to bid in SICAP in Romanian, to execute a field safety corrective action in Romanian without delay, and to support periodic verification of devices in use. Distributor selection in Romania is a regulatory act, not only a commercial one.

4. Servicing partner — if your device needs installation, maintenance, calibration, safety checks or repair in Romania, put a permitted Romanian service chain in place before the first tender. Installation/maintenance is a separately permitted activity with its own document list (Annex 3), device-group list (Form F7), staff list (Form F8) and measuring/monitoring equipment list (Form F10). Romanian hospital tenders routinely score or require local servicing, spare-parts and response-time capability, and a foreign manufacturer with no Romanian service answer loses on that ground alone, however strong the CE file is.

5. Relabelling / repackaging trap (MDR Article 16) — a distributor or importer that translates, relabels or repackages a device for Romania can step into manufacturer-level duties for that activity, including notified-body certification of its quality system for the activity where required. Romanian translation of labelling and IFU should therefore be done under the manufacturer's control wherever possible, with the translator working from the controlled master file — not improvised by the local partner.

6. PRRC — the Person Responsible for Regulatory Compliance (MDR Article 15) requirement is EU-wide, and it appears expressly in the Romanian permit file as Form F9 (decision appointing the PRRC and/or warehouse manager). Micro and small enterprises may rely on an external PRRC who is permanently and continuously at their disposal.

7. Customs and tax border — Romania is inside the EU customs union and uses the leu (RON) — it is not in the euro area, so Romanian pricing, NAMMDR tariffs, tender values and CNAS prices are denominated and paid in lei, and exchange-rate movement is a real line in a Romanian business plan. A shipment from a German or Italian warehouse to a Bucharest hospital is an intra-EU supply, not an import. Goods arriving from outside the EU are imported through Romanian customs, with import VAT and any duty accounted for at import by commodity code. Romanian VAT changed materially in 2025: by Law No. 141/2025 the standard VAT rate rose from 19% to 21% from 1 August 2025, and the former 5% and 9% reduced rates were consolidated into a single reduced rate of 11% for defined categories (medicines for human use are among the categories reported at the reduced rate); the former VAT exemption for supplies of medical equipment to non-profit entities was repealed in the same package. Doğrulanacak: the current VAT treatment for your exact device category and supply route, at pricing and at shipment — device VAT treatment in Romania is category- and customer-specific and changed in 2025, so never price from a pre-August-2025 source.

Free movement point: a device that is CE-marked and lawfully marketed in another EU country can be brought into Romania with no additional Romanian marketing authorisation. The Romanian work is therefore not a second certification; it is (a) the EUDAMED registration and, where your operator type triggers it, the Romanian national-database registration, (b) a Romanian channel holding the correct Order 566/2020 operating permits, (c) the Romanian-language set, (d) advertising approval for any public-facing campaign, and (e) the CNAS / SICAP work that makes the device payable and biddable in Romania.

4. The Romanian Registration Architecture: EUDAMED + National Database + Operating Permit

Romania currently has three regulatory layers that foreign teams routinely conflate. Keep them separate in your launch plan.

4.1 Layer 1 — EUDAMED (EU-wide, mandatory from 28 May 2026)

1. Actor registration — manufacturer, AR, importer and system/procedure pack producers register and receive a Single Registration Number (SRN) after validation by the competent authority of their place of establishment — NAMMDR for Romanian-established operators. The non-EU manufacturer links to its AR in EUDAMED; that linking can only be completed once the AR has its SRN. NAMMDR checks the data entered, and may validate, reject or request corrections. A separate payment form is filed for each EUDAMED registration application, submitted to NAMMDR headquarters in Bucharest or electronically to its registry address, after which NAMMDR invoices the validation tariff (220 lei at the 2022 base tariff — see Section 8).

2. Device registration — in the UDI/Device module, at Basic UDI-DI / UDI-DI level with the European Medical Device Nomenclature (EMDN). Receiving an SRN does not register the device: actor registration and device registration are separate acts, and both must be complete before first placement under the mandatory regime.

3. Certificates and vigilance modules — notified body certificates are recorded in EUDAMED, and market surveillance and vigilance reporting increasingly run through EUDAMED alongside the national NAMMDR channel. Keep both channels in your vigilance procedure until NAMMDR confirms a national channel is closed for your report type. Note that EGO 46/2021 and EGO 137/2022 both provide for NAMMDR to ensure, centrally at national level, the registration of manufacturers' reports of serious incidents and field safety corrective actions — the Romanian central reporting duty survives alongside EUDAMED.

4.2 Layer 2 — NAMMDR National Database (Romanian-specific, operator-dependent)

Under Order 3539/2022, the national database is the Romanian register NAMMDR maintains in parallel with EUDAMED. On NAMMDR's published position:

  • Who registers: authorised representatives established in Romania, for Class I, IIa, IIb and III devices (including sterile and/or measuring devices), systems and procedure packs under MDR Article 22, and active implantable devices; and Romanian manufacturers, including Romanian custom-made manufacturers under Order 2845/2022, who provide information on their headquarters and a description of the devices placed on the market. NAMMDR also registers vigilance information in the national database.
  • Who does not: authorised representatives established in other Member States are expressly not registered in the Romanian national database. A foreign EU manufacturer selling into Romania through a non-Romanian AR and a Romanian distributor should not assume a national-database device filing in its own name — confirm the position for your chain at filing (to be verified note in point 1 of the reality check applies).
  • The output is a certificate: Order 3467/2022 tariffs the "issuance of a certificate of registration of medical devices into the national database (maximum 5 products)" as a distinct activity, and tariff-lines also cover registration of changes to database information (maximum 5 items) and searching the national database. Treat the certificate as your completion evidence — file it with the launch records, not in an email inbox.
  • Free sale certificates are a Romanian-established-operator product: NAMMDR issues free sale certificates for export only to manufacturers headquartered in Romania or their authorised representatives headquartered in Romania (EGO 137/2022 states the same rule for IVDs under IVDR Article 55). A non-Romanian AR cannot obtain a Romanian free sale certificate for use in a third-country filing; plan third-country dossiers around the certificate of the country where the manufacturer or AR is actually established.
  • Never write "approved by NAMMDR" in Romanian marketing material. Registration in the national database is an identification and surveillance process — not a NAMMDR assessment of safety or performance, and not an endorsement. The claim collides with MDR Article 7 (misleading claims) and the Romanian advertising layer.

4.3 Layer 3 — Operating Permit (aviz de funcționare) for the Romanian channel

Under Order 566/2020 and Law 95/2006 Title XX, the activities subject to approval control are import, distribution, and installation and/or maintenance of medical devices. Practical consequences from NAMMDR's published operating-permit guidance:

  • The file is headed by Annex 1 (issuance or renewal) or Annex 5 (modification), with the document list for the activity — Annex 2 for import/distribution, Annex 3 for installation/maintenance, Annex 6 for temporary approval of import/distribution — ticked off in the column for economic operators. Documents must be complete, signed, dated, with registration number or "in accordance with the original" certification as applicable, in A4, legible, with no missing pages. Annexes 2, 3 and 6 and Forms F6 and F7 are also submitted in editable format (Word or Excel) on digital media (USB stick) — a strikingly analogue detail in an otherwise electronic system, and a common cause of a file being treated as incomplete.
  • One file can cover several activities or work points, with common documents submitted once. The manufacturers to be sold are listed alphabetically in the Manufacturers Table (Form F6), with the exact name from the declaration of conformity, and one declaration of conformity per manufacturer is submitted, with the corresponding conformity certificate according to the device's risk class.
  • Tariffs for the permit are invoiced after application: under Article 10(1) of Order 566/2020 as NAMMDR publishes it, the invoice is issued within a maximum of 90 days of submission, and failure to pay the fiscal invoice within its stated term leads to the filing (closure) of the file. This is the opposite order from most NAMMDR tariffs, which are payable in advance — do not apply one payment logic to the other.
  • If evaluation finds non-compliant or missing documents, NAMMDR requests additions. After two re-evaluations following non-tariff addition requests, any further re-evaluation carries a new tariff of 50% of the initial tariff. Front-load completeness; the third look is not free.
  • Permit tariffs scale with the size of the file (number of manufacturers for import/distribution; number of device groups for installation/maintenance), with discounts built into Order 3467/2022 for additional activities (50% from the second activity), additional work points (50% from the second, 25% from the eleventh) and a single charge for the import activity. See Section 8.
  • The permit must be kept current: adding a manufacturer, adding an activity, adding a work point, changing the operator's name or headquarters address are all tariffed modification events. A distributor that starts distributing your brand before you are added to its permit is distributing outside its permit, even if your device is perfectly CE-marked.

4.4 Payment layers (not marketing authorisations)

CNAS reimbursement and SICAP procurement are covered in Section 11. Neither is a marketing authorisation, and neither substitutes for Layers 1–3. A device with a CNAS agreement but no compliant Romanian channel is as unsellable as a device with a perfect channel and no payer.

5. Step-by-Step: Placing a Device on the Romanian Market

The sequence below assumes a foreign (non-Romanian) manufacturer. Romanian manufacturers start at Step 2 and carry the national-database duty in their own name.

Step 1 — Qualify and classify the device under the MDR/IVDR. Confirm the product is a medical device (or an Annex XVI listed product, or an IVD), document the Annex VIII classification rationale, and resolve borderline questions early — in Romania a formal classification, including a complex/ borderline classification, is a tariffed NAMMDR service if you need the authority's answer (Section 8). Do not let a distributor classify the device for you in its permit file with a different class from your technical documentation.

Step 2 — Complete EU conformity assessment. Build the technical documentation (MDR Annexes II–III), the clinical evaluation, risk management, and post-market surveillance system; complete the conformity route for the class (self-declaration for plain Class I; a notified body designated in another Member State for everything above it — there is no Romanian notified body); draw up the EU Declaration of Conformity; affix CE marking; assign Basic UDI-DI and UDI-DI and prepare the UDI carriers.

Step 3 — Appoint the EU authorised representative (non-EU manufacturers) and decide where it sits. The AR's seat decides the Romanian national-database position: a Romania-established AR registers in the national database and can obtain Romanian free sale certificates; an AR established elsewhere does neither. Make that choice deliberately, in writing, as part of the mandate — it is a Romanian market-structure decision, not an administrative detail.

Step 4 — Register actors in EUDAMED and obtain SRNs. Manufacturer, AR and (Romanian) importer each complete actor registration; Romanian-established actors are validated by NAMMDR on payment of the validation tariff per application. Complete the manufacturer–AR linking. A Romanian operator that will import must have its importer SRN before its operating permit is issued or modified to add import.

Step 5 — Register the devices in EUDAMED. Complete UDI/device registration at Basic UDI-DI and UDI-DI level with EMDN coding, and select the markets where the device is placed, keeping that list current as markets are added. Both actor and device registration must be complete before first placement under the mandatory regime that applies from 28 May 2026.

Step 6 — Complete the Romanian national-database step, if your chain triggers it. If the AR is established in Romania, or the manufacturer is Romanian, file the national-database registration under Order 3539/2022 (Order 2845/2022 for Romanian custom-made manufacturers), pay the registration-certificate tariff, and obtain and file the certificate of registration (maximum 5 products per certificate at the tariffed unit). Keep the record current: changes to database information are a separate tariffed registration.

Step 7 — Put the Romanian channel on a valid operating permit. Confirm — by inspecting the permit, not by asking the partner — that the Romanian importer's permit covers import, the distributor's permit covers distribution, and the service partner's permit covers installation/maintenance for your device groups; that your company is listed in the permit's manufacturers table (Form F6) with the exact name from your declaration of conformity; and that the permit is in date for every work point that will hold or service stock. If you are not yet listed, the partner files a modification (Annex 5) to add manufacturers — a tariffed event scaled by the number of manufacturers added — before the first shipment, not after it.

Step 8 — Produce the Romanian-language set. Labels and instructions for use in Romanian; implant cards, EU declarations, field safety notices and user interfaces in Romanian or in the English variant only where the Commission language table for Romania actually allows it for your user type (professional-only, with the recorded consents/approvals). Do the translation under manufacturer control (MDR Article 16 trap, Section 3). Verify the Romanian IFU says the same thing as the classification rationale and the clinical evaluation — the Romanian text is the operative text in Romania.

Step 9 — Set up vigilance, PMS and in-use support before the first sale. Written procedures for serious incident reporting and field safety corrective actions through both the EUDAMED and the NAMMDR central channels; complaint handling and traceability at importer and distributor level (one-up/one-down, UDI-based); Romanian-language field safety notice templates; and, for installed equipment, the periodic-verification and approval-for-use regime of Order 2219/2022 built into the service offer.

Step 10 — Clear the commercial gates. Price in lei, with the post-August-2025 VAT position for your category confirmed; decide the route to market per segment — SICAP tenders (individual hospital or centralised), private hospital groups, retail/pharmacy where applicable, and CNAS agreement if the device is reimbursed; obtain NAMMDR approval for any public-facing advertising material before the campaign runs; and brief the Romanian partner that "NAMMDR registered" is never to be rendered as "NAMMDR approved" in any Romanian material.

Step 11 — Place the device on the market and maintain the file. First placement only after Steps 2–8 are evidenced. Then maintain: EUDAMED and (where applicable) national-database updates, permit modifications as the chain changes, permit renewal in good time before expiry, PSUR/PMS reporting, vigilance timelines, and tender/CNAS record-keeping. Romanian files decay at renewals and modifications, not at first filing — diarise them at launch.

6. Document Checklist

6.1 Manufacturer core file (EU file, used at every Romanian step)

  • [ ] Qualification and classification rationale (MDR Annex VIII rule(s), MDCG 2021-24 reasoning)
  • [ ] Technical documentation (Annexes II–III) and GSPR checklist (Annex I)
  • [ ] Clinical evaluation report / IVD performance evaluation, and PMCF/PMPF plan
  • [ ] Risk management file, usability, software lifecycle and cybersecurity documentation where applicable
  • [ ] Notified body certificates (for Is/Im/Ir and Class IIa+) — from a body designated in another Member State; there is no Romanian body
  • [ ] EU Declaration of Conformity
  • [ ] Labels, packaging artwork and instructions for use — master file and Romanian set
  • [ ] UDI data set (Basic UDI-DI, UDI-DI, EMDN code) and UDI carrier specimens
  • [ ] PMS plan, PMS report / PSUR as applicable, vigilance procedures and Romanian field safety notice templates
  • [ ] AR mandate (non-EU manufacturers), PRRC identification, insurance cover for MDR Article 10(16) financial responsibility

6.2 Romanian registration file

  • [ ] EUDAMED actor registration evidence and SRN for each Romanian-established operator; payment form per EUDAMED application and NAMMDR validation invoice/receipt
  • [ ] EUDAMED device registration evidence (UDI/Device module)
  • [ ] If AR established in Romania or manufacturer Romanian: national-database registration file under Order 3539/2022 and the certificate of registration into the national database (maximum 5 products per certificate)
  • [ ] Romanian manufacturers additionally (per NAMMDR's 2025 notice): ONRC registration certificate and evidence of authorisation of device manufacturing showing the CAEN code (rev.

3) specific to the notified device type

  • [ ] Romanian custom-made manufacturers: file under Order 2845/2022 and the MDR Annex XIII statement set
  • [ ] Free sale certificate — only if obtainable (Romanian-established manufacturer or AR) and only where a third-country dossier actually requires a Romanian certificate

6.3 Romanian channel file (operating permit — inspect the partner's, file your own if you establish in Romania)

  • [ ] Operating permit (aviz de funcționare) for import (importer), distribution (distributor), installation/maintenance (service partner), in date, covering each work point
  • [ ] Application form Annex 1 (issuance/renewal) or Annex 5 (modification); document list Annex 2 (import/distribution), Annex 3 (installation/maintenance) or Annex 6 (temporary approval)
  • [ ] Manufacturers Table Form F6 — manufacturers listed alphabetically, exact declaration-of-conformity names, your company listed before first shipment
  • [ ] One EU Declaration of Conformity per manufacturer, with the conformity certificate corresponding to the device risk class
  • [ ] Form F7 device categories/groups (installation/maintenance), Form F8 staff list, Form F9 PRRC / warehouse manager appointment decision, Form F10 measuring and monitoring equipment list, Form F11 technical inspection bulletin template where applicable
  • [ ] Documented work procedures for the permitted activities; warehouse and traceability procedures
  • [ ] Editable-format copies (Word/Excel) of Annexes 2/3/6 and Forms F6/F7 on digital media, as NAMMDR's guidance requires
  • [ ] Importer SRN evidence (for the import permit, from the 2026 regime)

7. Timelines

StepRealistic / published timelineNotes
Plain Class I self-declaration routeWeeks, manufacturer-controlledNo notified body; the Romanian steps (language set, channel permit position) usually take longer than the conformity step
Notified body conformity (IIa / IIb / III)Commonly ~12–18 months (IIa/IIb) and ~18–24 months (III) in current EU practiceNo Romanian notified body exists — queue, audit and certification all happen in another Member State; obtain the body's lead time in writing
EUDAMED actor validation by NAMMDRNo single published statutory period verified for this guide — to be verified at filingNAMMDR may validate, reject or request corrections; one payment form per application
National-database registration certificateNo single published statutory period verified for this guide — to be verified at filingTariffed per certificate (maximum 5 products)
Operating permit evaluation (Order 566/2020)Reported in the Chambers secondary summary as an evaluation of up to 120 days, ending in a positive or negative evaluation report; NAMMDR issues the tariff invoice within a maximum of 90 days of application (Article 10(1))Permit validity is reported as 3 years if conditions remain unchanged; a temporary approval route for emergency conditions is reported with evaluation within a maximum of 7 days and 6 months' validity. Doğrulanacak: the current renewal filing deadline in Order 566/2020 as amended at renewal — the Chambers summary reports that a renewal not filed in good time restarts the procedure from the beginning; confirm the current text before relying on any specific month figure.
Advertising approval (public-facing material)NAMMDR approval is a tariffed visa valid 6 months, extendable at the applicant's proposal to a maximum of 1 year (Order 3467/2022 notes) timelineThe 2021 secondary literature reported a 30-day assessment expectation for public advertising material under EGO 46/2021 — to be verified at campaign filing
Clinical investigation authorisationTariffed by device class (Section 8); procedure under MDR Articles 62–82 and Order 330/2023Ethics opinion required in parallel; no single Romanian statutory period is stated in this guide — to be verified at protocol filing
Legacy-device transitionTo 31 Dec 2027 (Class III / IIb implantable, with exceptions) / 31 Dec 2028 (other IIb, IIa, Is/Im/Ir)Only if the Article 120 conditions (timely notified body application + signed agreement) are met

Planning rule for Romania: build the launch plan backwards from the slowest of three clocks — the foreign notified body's queue, the Romanian channel's operating-permit position (a modification to add your company takes permit-procedure time, not sales time), and the first realistic SICAP/ CNAS cycle for your category. In practice the channel permit and the payer cycle, not the CE mark, set the Romanian revenue date.

8. Fees and Tariffs

All NAMMDR figures below are the base tariffs in Annex 1 to Minister of Health Order No.

3467/2022, in lei (RON), as published in the Official Gazette Part I, No. 1123 of 22 November 2022.

They are indexed annually by 31 January by the previous year's average inflation rate, starting January 2024 — so the amount actually payable in 2026 is higher than the base figure.

Doğrulanacak: the indexed tariff in force on NAMMDR's Forms and fees page at payment, for every line you use. General tariff rules from the same order: tariffs are paid in lei, in advance before registration of the application unless the governing rules provide otherwise (operating permits are the exception — invoiced after application, Section 4.3); an underpaid tariff must be completed within 30 days of notification or the request is rejected; tariffs for rejected applications are not refunded; withdrawal before NAMMDR's completion request retains 10%, after it 50%, after the final document is drawn up 100%.

8.1 Regulatory activities (Order 3467/2022, Section 1)

ActivityBase tariff (lei)
Issuance of a customs notice (per product)120
Grant of an out-of-scope notice (per product)120
Issuance of a free sale certificate (1–10 items)1,240
Standard classification of a product (one product per request)2,840
Contesting a classification2,840
Complex classification of a borderline product4,740
Issuance of a certificate of registration of medical devices into the national database (maximum 5 products)1,310
Assessment of the notification file for devices manufactured and used in healthcare facilities (per product) — the Article 5(5) in-house notification1,310
Assessment of a notification file for a change to in-house devices350
Registration of changes to information in the medical devices database (maximum 5 items)350
Search of the national database220
Validation/verification of registration into EUDAMED220

8.2 Operating permit activities (Order 3467/2022, Section 5)

ActivityBase tariff (lei)
Assessment for issuance/ renewal/modification adding import or distribution, by number of manufacturers: 1–10 / 11– 30 / 31–50 / 51–100 / over 1001,480 / 1,730 / 1,980 / 2,230 / 2,620
Assessment for issuance/ renewal/modification adding installation and/ or maintenance, by number of device groups: 1–1,660 / 1,900 / 2,150 / 2,320

Activity Base tariff (lei)

10 / 11–30 / 31–50 / over Assessment of modification by adding a work point Assessment of change of operator name / non-profit headquarters address change Assessment of modification adding manufacturers (as importer/distributor/EU authorised representative): 1–10 / 11–30 / over 30 manufacturers 490 / 1,040 / 1,760 Assessment of deletion of manufacturers / deletion of a work point Built-in discounts and surcharges (Article 6 of the order): the tariff is per activity and per work point; non-profit registered offices are not charged; from the second activity, 50% of the first activity's tariff; from the second work point with the same activity, 50%, and from the eleventh work point, 25%; the import activity is charged only once; re-assessment after an unfavourable report is 50% of the basic tariff; and after two re-evaluations following addition requests, each further re-evaluation is a new 50% tariff.

8.3 Clinical investigations and performance studies (Order 3467/2022, Section 3 — selected)

Activity Base tariff (lei)

Assessment for authorisation of a clinical investigation — active implantable, Class III, IIb invasive and similar Annex XVI devices 20,450 Assessment to validate an investigation application — Class IIa and IIb non-invasive and similar devices 18,030 Assessment to validate an investigation application 8,770

ActivityBase tariff (lei)
— Class I and IIa non-invasive and similar devices
Substantial/technical amendments to an approved investigation (and substantial IVD performance-study amendments)5,930
Administrative amendments (and non-substantial IVD amendments)2,620
Resubmission after withdrawal or rejection6,780
Assessment and authorisation of performance evaluation for IVDs — self-testing and Classes C and D13,710
Assessment and authorisation of performance evaluation for IVDs — Classes A and B11,200
Authorisation of IVD performance studies using leftover samples (IVDR Article 58(2))6,080
PMCF notification (MDR) / PMPF notification (IVDR) assessment1,880

8.4 Advertising (Order 3467/2022, Section 4)

ActivityBase tariff (lei)
Approval of advertising material for medical devices (visa valid 6 months; applicant may propose up to 1 year)2,720
Approval of educational material for medical devices (visa valid 6 months; applicant may propose up to 1 year)1,730

8.5 What is NOT in the tariff list

Notified body fees (payable to a foreign body, in its currency and on its schedule — usually the largest single cost in a IIa+ file), Romanian translation of the labelling/IFU set, legal and regulatory consultancy, SICAP participation costs, CNAS agreement administration, warehouse and service-network costs, and the internal cost of maintaining a Romanian-speaking regulatory contact. Budget the NAMMDR tariffs as the smallest, most predictable line in the Romanian budget — and the only line whose base figures can be read off a published table.

9. Language, Labelling and UDI in Romania

  • Labels and IFU: Romanian. The Commission's MDR language table for Romania (Revision 3, August 2025), citing EGO 46/2021 (Article 3) and Order 3539/2022, records Romanian for the label and instructions for use. Build the Romanian set as a controlled translation of the master file, with the same intended purpose, indications, contraindications and warnings — a Romanian IFU that widens or narrows the claim changes the device's Romanian legal position, and its classification under Rule 11 for software.
  • Defined English paths — professional use only, and conditional. The same table records Romanian-or-English possibilities in defined cases: for the implant card and for professional-user information in the cases it specifies, English may be accepted on the basis of the healthcare professional's written consent or request and, where the table records it, the competent authority's approval. The EU Declaration of Conformity is recorded as Romanian for Romanian manufacturers and Romanian or English for foreign manufacturers in the IVDR table's terms; conformity-assessment documents are recorded in English and Romanian; field safety notices are recorded in Romanian, or Romanian or English based on the written consent of the healthcare professional in the cases specified. Read the current table row for your document type before relying on any English variant — the exceptions are document-specific, not a general professional-use waiver, and none of them covers patient or lay-user information.
  • UDI is EU-wide, execution is Romanian. UDI assignment, carriers and EUDAMED registration follow MDR Articles 27–31 with no Romanian variant. The Romanian execution point is verification: the Romanian importer and distributor must verify UDI and labelling as part of their Article 13/14 checks and their permit-file duties, and hospitals' in-use controls under NAMMDR's market surveillance programme check label and IFU compliance against Annex I, Chapter III, point 23 of the MDR (point 20 of the IVDR) on the devices actually in use and in storage.
  • Advertising language and claims are part of labelling discipline. A Romanian advertisement that claims more than the Romanian IFU supports is both an MDR Article 7 problem and a Romanian advertising-approval problem. Keep one controlled claim set, in Romanian, across IFU, tender dossiers and campaigns.

10. Clinical Investigations in Romania

Romania applies MDR Articles 62–82 directly, with the Romanian procedural layer in Order 330/2023 (clinical evaluation and clinical investigations, replacing the 2006 Order 792/2006 procedure) and the tariff scale in Section 8.3. Practical points:

  • The sponsor files for authorisation/validation according to the device class and invasiveness — the tariff itself signals the Romanian triage: full authorisation assessment for active implantable, Class III, IIb invasive and Annex XVI-type devices at the top tariff; validation assessments for IIa/ IIb non-invasive and for I/IIa non-invasive devices at lower tariffs; separate tariffs for substantial and administrative amendments, for resubmission after withdrawal or rejection, and for PMCF notifications.
  • An ethics opinion is required in parallel with the competent-authority procedure — build both clocks into the protocol timeline, and confirm the current committee and submission channel at protocol filing (to be verified).
  • IVD performance studies follow the IVDR with Romanian tariffs split by class (C/D and self-testing at the higher tariff; A/B lower; leftover-sample studies under Article 58(2) as a separate, lower-tariff route).
  • Devices manufactured and used only within a Romanian healthcare institution are not a shortcut around investigation rules: they sit under MDR Article 5(5) and the Romanian notification procedure in Order 650/2025, with their own tariffed notification and change-notification files.
  • NAMMDR's approval-for-use regime for devices equipping pre-hospital emergency vehicles (MIAMUP) under Order 2219/2022 is an in-use equipment assessment, not a marketing authorisation and not an authorisation of the vehicle — NAMMDR states this expressly. Do not cite a MIAMUP approval as product approval in any dossier.

11. Reimbursement and Public Procurement: CNAS and SICAP

Regulatory registration makes a device lawful in Romania; these two systems make it paid for.

11.1 CNAS reimbursement

Romania's social health insurance is administered by CNAS through county health insurance houses. For medical devices, the position summarised in the CEE/SEE comparative guides is that reimbursement from public health funds is performed only on the basis of agreements with the health insurance houses, for devices that meet specific requirements and at set prices. In practice that means:

  • Not every CE-marked device is a reimbursed device. Reimbursement is category- and programme-specific — defined device categories, national health programmes and insured-patient entitlements — and a device sold outside those categories is paid by the hospital's own budget (through procurement), by a private payer, or by the patient.
  • Where reimbursement applies, the commercial terms are set by the framework — agreement, requirements and price — not negotiated freely per sale. Build the Romanian price architecture from the reimbursed price backwards, in lei, with the VAT position for the category confirmed.
  • Doğrulanacak at entry, for your exact category: whether the device is reimbursed at all, under which programme or list, at what set price, through which county houses, and with what agreement and reporting duties. These are set and revised by Romanian decisions and must never be taken from a secondary summary or another country's analogy.

11.2 Public procurement through SICAP

Public hospitals and other contracting authorities buy under the Romanian public procurement framework, supervised as policy by ANAP and transacted electronically in SICAP (the electronic public procurement system, successor to the SEAP branding still used colloquially and in older documents). ANAP and AmCham Romania have published a joint Guide for Procurement of Medical Devices — planning, technical-specification and evaluation best practice — which is the best public window into how Romanian buyers are told to buy devices. Research on Romanian device procurement (2019–2024 SICAP files) shows technical specifications dominated by ISO, IEC and EN standards — ISO 9001, ISO/IEC 60601 for electrical safety, sterilisation and laboratory standards by category — so a tender dossier that cannot answer in standards language loses before price is discussed.

Practical consequences for a foreign manufacturer:

  • Your bidder is your Romanian channel. The SICAP bidder will normally be the Romanian importer/distributor holding the operating permit — another reason the permit file (Section 4.3) is a commercial document. Confirm the bidder's permit lists your company before the tender, not during clarification.
  • Specifications are written around standards and the incumbent. Monitor SICAP for your category continuously, challenge discriminatory specifications through the Romanian remedies channels within their short deadlines where the specification is written around a competitor's proprietary features, and keep a Romanian-language technical file ready to answer specification points line by line.
  • Centralised vs hospital-by-hospital buying must be mapped per category. Some categories are bought centrally (Ministry of Health / centralised procurement bodies); the rest are bought by individual public hospitals and county structures as contracting authorities. Doğrulanacak: the current centralised category list and the current procurement thresholds and direct-award limits at tender — Romanian thresholds have been revised repeatedly (a 2021 summary is not a 2026 source) and the applicable procedure turns on them.
  • Private sector is a separate, faster channel. Private hospital groups and clinic chains in Bucharest and the regional cities buy commercially, without SICAP, but they apply the same regulatory checks — CE marking, EUDAMED position, Romanian IFU, permitted distributor, service capability. A private-sector first sale is often the fastest Romanian reference site for a later public tender.

11.3 Pricing discipline

Romania is a lei market with a 2025 VAT change (standard rate 21% from 1 August 2025; single reduced rate 11% for defined categories) and set-price reimbursement in the categories where CNAS pays. Price once, in writing, per channel — reimbursed, public tender, private — with the VAT treatment, the permit-holder's margin, service and periodic-verification costs, and lei/euro exchange exposure all inside the number. A euro price list converted at the day's rate is not a Romanian pricing policy.

12. Post-Market Surveillance, Vigilance and Devices in Use

  • PMS system (MDR Articles 83–86): PMS plan, PMS report for Class I, PSUR for IIa and above, trend reporting — EU-wide duties with no Romanian variant in substance, executed in Romania through the chain you actually have: the Romanian importer/distributor's complaint intake is your Romanian PMS front door, so its procedures are part of your PMS system, audited as such.
  • Vigilance (MDR Articles 87–92): serious incidents and field safety corrective actions are reported within the MDR deadlines (2 days for a serious public health threat, 10 days for death or unanticipated serious deterioration, 15 days otherwise). Romania adds the central national registration layer: EGO 46/2021 (MDR) and EGO 137/2022 (IVDR, Article 7) provide for NAMMDR to ensure, centrally at national level, the registration of manufacturers' reports of serious incidents and field safety corrective actions. Run both the EUDAMED and the NAMMDR channel in your procedure until NAMMDR confirms a channel is closed for your report type, and issue the Romanian field safety notice in the language the Commission table requires for your users.
  • Market surveillance (MDR Article 93; Regulation (EU) 2019/1020): NAMMDR, as market surveillance authority, may carry out announced and unannounced inspections, on-site and physical product checks without prior notice, and may enter premises, areas and means of transport used for the operator's activity to identify non-compliance and obtain evidence. Its published control programme includes controls in hospital units on devices in use — checking, among other things, whether the hospital has appointed a person responsible for records of devices in use and as NAMMDR contact, and whether labels and IFUs of devices in use and in storage comply with the MDR/IVDR. Your Romanian users' inspection readiness is therefore part of your post-market system: a hospital that cannot produce your Romanian IFU in an inspection becomes your problem.
  • Devices in use (Order 2219/2022): devices in service in Romanian healthcare units are subject to control by periodic verification, performance evaluation of second-hand devices put into service, and approvals for use — executed through the permitted installation/maintenance channel.

Contracts for installed equipment should allocate periodic-verification duties, costs and scheduling expressly; an unverified device in use is a surveillance finding waiting for an inspector.

  • Supply interruption (MDR Article 10a): manufacturers must notify anticipated interruption or discontinuation of supply where reasonably foreseeable serious harm could result — applicable since 10 January 2025 — through the Romanian competent-authority channel as part of the EU procedure.
  • Sanctions: device sanctions sit in Law 95/2006, Title XX, Chapter 4 (Articles 934–936, as NAMMDR cites them), alongside the MDR corrective powers — withdrawal, recall, prohibition on making available, and suspension or withdrawal of the importer's/distributor's authorisations in the cases the Romanian framework provides. The operating permit is the enforcement handle: a channel that loses its permit stops your Romanian sales lawfully overnight, which is why permit compliance is monitored, not assumed.
  • Record retention: technical documentation and declarations kept available for at least 10 years after the last device is placed on the market (15 years for implantables) — by the manufacturer and, within its mandate, the AR — with the Romanian importer/distributor traceability records aligned to the same horizon.

13. Common Mistakes in Romania

1. Assuming every EU device needs a Romanian national-database filing — or that none does. The duty turns on the operator: Romania-established ARs (for the listed device types) and Romanian manufacturers register; ARs established in other Member States do not. Both errors waste money or create an infringement; resolve your chain's position in writing at launch.

2. Choosing the AR's seat without seeing the Romanian consequences. An AR established in Romania brings national-database registration and access to Romanian free sale certificates; an AR elsewhere brings neither. Choosing on price alone and discovering the free-sale consequence during a third-country filing is the classic version of this mistake.

3. Signing a Romanian distributor without inspecting its operating permit. If the permit does not cover the activity, the work point and — in its Form F6 manufacturers table — your exact company name, the channel is not lawful for your devices yet. A modification to add manufacturers takes permit-procedure time and a tariff; shipments before it are outside the permit.

4. Treating the permit invoice as a filing fee. Permit tariffs are invoiced after application (within a maximum of 90 days under Article 10(1) of Order 566/2020) and an unpaid invoice closes the file — while most other NAMMDR tariffs are prepaid and unrefunded if rejected. Teams that apply one logic to both lose files both ways.

5. Quoting 2022 tariffs as 2026 prices. Order 3467/2022 tariffs are indexed annually for inflation from January 2024. A budget built on the base table without the indexed figures underfunds every Romanian filing.

6. Shipping with an English-only IFU. Romanian labelling and IFU are the rule; the English paths are narrow, document-specific and conditional on professional-user consent/approval. Patient-facing English-only material has no path at all.

7. Letting the local partner translate and relabel informally. That is the MDR Article 16 trap: the partner can step into manufacturer duties for the relabelled device. Translate under manufacturer control, from the controlled master.

8. Planning notified body time as if Romania had a body. It has none. Every IIa+ file queues abroad; a Romanian launch date promised before the foreign body's written lead time exists is a guess presented as a plan.

9. Launching consumer advertising on another country's approval. Public-facing device advertising in Romania needs NAMMDR approval, is limited to devices usable without medical intervention, and the visa runs 6 months (maximum 1 year). An unapproved campaign risks the material, the visa record and the permit-holder's standing.

10. Pricing in euro, or on pre-August-2025 VAT. Romania is a lei market; the standard VAT rate rose to 21% on 1 August 2025 and the reduced-rate structure was rebuilt at the same time. Euro-converted price lists and old VAT assumptions both break the first Romanian invoice.

11. Confusing a CNAS agreement, a SICAP award and a registration. They are three different systems (payer, procurement, regulation). Winning one does not complete the others, and "registered" must never be marketed as "approved" or "endorsed" by NAMMDR.

12. Ignoring devices-in-use duties after installation. Periodic verification under Order 2219/2022, hospital in-use inspections and the MIAMUP approval-for-use regime all operate after the sale. A service contract silent on verification produces surveillance findings in your reference hospitals.

14. Frequently Asked Questions

Does Romania require a separate national approval for CE-marked devices? No. There is no Romanian pre-market approval and no Romanian CE mark. Conformity is EU-wide under the MDR/ IVDR. Romania adds registration layers (EUDAMED validation, and national-database registration for the operator types Order 3539/2022 covers), operating permits for the Romanian channel, Romanian language, advertising approval for public material, and the CNAS/SICAP payment systems.

Who is the Romanian competent authority? The National Agency for Medicines and Medical Devices of Romania — ANMDMR in Romanian, NAMMDR in its English materials — under the Ministry of Health, headquartered in Bucharest. It is the competent authority and market surveillance authority for devices (EGO 46/2021) and for IVDs (EGO 137/2022), validates Romanian EUDAMED actor registrations, maintains the national database, and issues operating permits, free sale certificates and advertising approvals.

Must my devices be registered in the Romanian national database? It depends on your chain.

Authorised representatives established in Romania register for Class I, IIa, IIb and III devices (including sterile/measuring), Article 22 systems and procedure packs, and active implantable devices;

Romanian manufacturers register their devices, with a specific regime for Romanian custom-made manufacturers. Authorised representatives established in other Member States are expressly not registered in the Romanian national database, on NAMMDR's published position. Confirm your chain's position at filing — NAMMDR practice has been aligning with mandatory EUDAMED during 2026.

Do I need a Romanian authorised representative? No — you need an EU authorised representative if you are a manufacturer outside the EU, and it may be established in any Member State. Establishing it in Romania triggers Romanian national-database registration and gives access to Romanian free sale certificates; establishing it elsewhere does neither. Only non-EU manufacturers need an AR at all.

Does my Romanian distributor need a licence? Yes, in substance: import, distribution and installation/maintenance of medical devices in Romania are activities subject to approval control under Order 566/2020, exercised through a NAMMDR operating permit (aviz de funcționare) covering the activity, the work points and the manufacturers listed in the permit file. Verify the permit — scope, Form F6 listing of your company, expiry — before signing or shipping.

How long is an operating permit valid, and when is it renewed? The secondary literature reports a validity of 3 years if the conditions of issuance remain unchanged, an evaluation of up to 120 days, and a restart of the procedure if renewal is not filed in good time before expiry.

Doğrulanacak: the exact current renewal deadline and validity text in Order 566/2020 as amended, at filing/renewal.

Is there a Romanian notified body? No. NAMMDR states that Romania currently has no MDR-designated or IVDR-designated notified body. Use a body designated in another EU/EEA state whose NANDO scope covers your device; there is no requirement (and no possibility) to use a Romanian body.

What language must labelling and the IFU be in? Romanian. Narrow English variants exist for defined document types and professional-only use, conditional on the healthcare professional's written consent/request and, where specified, the competent authority's approval, per the Commission's MDR/IVDR language tables for Romania. Patient and lay information must be in Romanian.

What does NAMMDR charge? Published tariffs under Order 3467/2022 (2022 base, indexed annually for inflation from January 2024): EUDAMED validation 220 lei per application; national-database registration certificate 1,310 lei (maximum 5 products); free sale certificate 1,240 lei (1–10 items); standard classification 2,840 lei and borderline classification 4,740 lei; operating-permit assessments from 1,480 lei (import/distribution, 1–10 manufacturers) scaled by manufacturers or device groups; advertising approval 2,720 lei; clinical investigation assessments from 8,770 to 20,450 lei by class. Always confirm the indexed amount payable in the filing year.

Can I advertise my device to the Romanian public? Only if the device can be used without medical intervention, and only with NAMMDR-approved material (a tariffed visa valid 6 months, extendable to a maximum of 1 year). Professional-directed material is not prior-approved in the same way but is subject to NAMMDR compliance checks and to MDR Article 7.

How are devices bought by Romanian public hospitals? Through electronic public procurement in SICAP under the Romanian procurement framework (ANAP supervises policy and co-published the medical device procurement guide), hospital by hospital or through centralised procurement where the category is centralised. Where a device is reimbursed from public health funds, reimbursement runs through agreements with the health insurance houses (CNAS system) at set prices and conditions. Map your category to the right channel before entry.

What currency and VAT should I plan for? Lei (RON) — Romania is not in the euro area. Standard VAT is 21% from 1 August 2025 (Law 141/2025), with a single reduced rate of 11% for defined categories; the VAT treatment of your exact device category and customer type must be confirmed at pricing (to be verified).

I manufacture in Romania. What is different? You register in EUDAMED as a Romanian manufacturer (NAMMDR-validated), register your devices in the national database under Order 3539/2022, evidence ONRC registration and the CAEN code specific to your device type in the notification file (NAMMDR's 2025 notice), follow Order 2845/2022 if your devices are custom-made, and you — or your Romania-established AR — are the operators who can obtain Romanian free sale certificates for export dossiers.

Where should a foreign manufacturer start in practice? In this order: (1) MDR classification and conformity route with a foreign notified body, if needed; (2) the AR-seat decision, taken with its Romanian consequences in view; (3) EUDAMED actor and device registration; (4) inspection of the Romanian channel's operating permit and your Form F6 listing; (5) the Romanian-language set; (6) the CNAS/SICAP map for your category. In ghmap.io terms, Romania rewards the feasibility-first sequence — doctor and hospital acceptance tested through a permitted, payable channel — because a CE mark without a permitted Romanian distributor and a payer route produces no Romanian revenue, while a permitted channel with real hospital demand makes every regulatory step above a formality with a tariff attached.

Sources & scope

  • anm.ro, National Agency for Medicines and Medical Devices of Romania (ANMDMR) — operating permits, EUDAMED validation, device tariffs and market surveillance.
  • cnas.ro, National Health Insurance House (CNAS) — provider contracting under the Framework Contract and settlement of listed outpatient medical devices.
  • eur-lex.europa.eu, Regulation (EU) 2017/745 (MDR), Regulation (EU) 2017/746 (IVDR), Regulation (EU) 2023/607.

This guide is market-access intelligence, not regulatory or legal advice. ANMDMR and CNAS practice, tariff annexes and Framework Contract norms change — professional review is required before business use.