Last updated: November 2026. The EU Medical Device Regulation (MDR, Regulation (EU) 2017/745) and the In Vitro Diagnostic Regulation (IVDR, Regulation (EU) 2017/746) apply directly in Slovenia, an EU Member State since 1 May 2004 and a euro-area member since 1 January 2007. The Slovenian national layer is built around the Medical Devices Act (Zakon o medicinskih pripomoIkih — ZMedPri), cited by the Agency for Medicinal Products and Medical Devices of the Republic of Slovenia (Javna agencija Republike Slovenije za zdravila in medicinske pripomoIke — JAZMP) in its current legislation list as published in the Official Gazette of the Republic of Slovenia, No. 40/25, with amendments; JAZMP's own current guidance also refers to the Act as ZMedPri-1, and JAZMP continues to publish an English translation of the earlier Act text (Official Gazette RS, No. 98/2009) on its website. Around that core sit the Decree on the implementation of the Regulation (EU) on medical devices and JAZMP's implementing rules (including the Rules on vigilance and the Rules relating to fees, Official Gazette RS, No. 24/19); the Tariff of JAZMP for medical devices (Official Gazette RS, No. 6/26), which carries the Agency's annual fees; the Health Care and Health Insurance Act (ZZVZZ) and the Rules on Mandatory Health Insurance, which govern what the Health Insurance Institute of Slovenia (ZZZS) pays for; and the Public Procurement Act (ZJN-3), which governs how public hospitals buy. Always confirm critical steps against the current guidance of JAZMP and the consolidated Slovenian texts before filing: much Slovenian device law and guidance is published in Slovene only, the Act was recast in 2025, and the consolidated wording in force at the date of filing — not a summary, and not this guide — governs.
Slovenia is a small, concentrated and almost entirely publicly financed market. According to the Statistical Office of the Republic of Slovenia (StatistiIni urad Republike Slovenije — SURS), Slovenia had 2,138,751 residents on 1 July 2026, of whom foreign citizens represented 10.8% (source: SURS, population release for the first half of 2026). Slovenia spends USD 5,527 per capita on health (PPP), less than the OECD average of USD 5,967, equal to 9.9% of GDP against an OECD average of 9.3%, and all of the population is covered for a core set of services, against an OECD average of 98% (source: OECD, Health at a Glance 2025, Slovenia country note). In the most recent EU country profile, health expenditure reached 9.3% of GDP in 2023, below the EU average of 10.0% (source: OECD / European Observatory on Health Systems and Policies, Slovenia: Country Health Profile 2025, data for 2023). Care is delivered through a concentrated public hospital sector — 27 public, state-owned hospitals in the European Observatory's system description, including ten general (regional) hospitals and the two university medical centres in Ljubljana and Maribor, which are by far the largest hospitals in the country — (source: European Observatory on Health Systems and Policies, Health Systems in Transition profile for Slovenia, physical resources). A single purchaser, ZZZS, operates the compulsory health insurance scheme. For a foreign manufacturer, the EU conformity file is necessary but not sufficient: the Slovenian layer decides who must be entered in JAZMP's registers and on what clock, what must be supplied in Slovene, who pays JAZMP's annual fees, and whether a device that may lawfully be sold will also be paid for — through the ZZZS standards and lists for devices supplied to insured persons, and through public procurement under ZJN-3 for devices bought by public hospitals.
Regulatory reality check up front: Slovenia is not a separate certification regime from the EU. There is no "Slovenian CE mark" and JAZMP does not approve medical devices in the way a pre-market authorisation authority approves medicines. The manufacturer carries out the conformity assessment on its own responsibility, involving a notified body for the higher risk classes, and affixes the CE marking; JAZMP keeps the national registers, validates Slovenian actors for EUDAMED, supervises the market, runs materiovigilance, decides classification questions put to it, and issues certificates of free sale for registered devices. What Slovenia adds — and what foreign manufacturers most often underestimate — is a national layer with six specific features:
- 1. Two registration systems, and the national one unlocks the European one — JAZMP maintains a national Register of Medical Devices and a Register of economic operators established in the Republic of Slovenia that carry out activities in the field of medical devices. Registration in that national register of economic operators is necessary for the confirmation of the economic operator in EUDAMED: JAZMP verifies the operator's information in accordance with Article 31 of the MDR, and the operator is then assigned its Single Registration Number (SRN) through EUDAMED (source: JAZMP, EUDAMED information page). A Slovenian importer or distributor that has never engaged with JAZMP's national register can therefore stall the whole chain's EUDAMED validation — the national step is not a legacy formality running beside the European one; for Slovenia-established actors it sits inside it.
- 2. Activity registration before the activity starts, for the whole operator chain — under the Medical Devices Act as summarised in current professional guidance, registration with JAZMP is mandatory before commencing activities in Slovenia for distributors, manufacturers of custom-made devices, reprocessors of single-use devices and healthcare institutions based in Slovenia; registration is also mandatory for importers, manufacturers and authorised representatives with their seat in Slovenia, and a manufacturer or authorised representative with a seat in Slovenia must register the medical device itself with JAZMP before placing it on the market for as long as the pre-EUDAMED national device registration applies, reporting any change to a registered device within eight days of the change (source: Lexology, "Q&A: marketing authorisation for pharmaceuticals and medical devices in Slovenia", summarising the Medical Devices Act registration regime). The interaction of these duties with mandatory EUDAMED registration since 28 May 2026 is to be verified with JAZMP at filing — see Sections 4 and 7.
- 3. An annual fee that reaches operators many entrants never budget for — notified bodies, manufacturers, authorised representatives, importers, distributors, manufacturers of custom-made devices, reprocessors of medical devices, and healthcare institutions that manufacture in-house devices for their own use all pay JAZMP annual fees as specified in the JAZMP Tariff for Medical Devices (Official Gazette RS, No. 6/26) (source: JAZMP, Certificate of Free Sale pages for MDR/IVDR and for legacy-directive devices, each stating the annual-fee population in those terms). The annual fee is not a one-off registration charge and it is not limited to manufacturers: a Slovenian distributor carrying your devices is inside the fee population, and its JAZMP standing is part of your route to market.
- 4. A Slovene-language rule with a narrow professional-use opening — the instructions for use must be written in the Slovene language, legible and understandable for the user; for professional use, the instructions for use can be written in the language understandable for the user, which the European Commission's language-requirements table records as normally meaning English is acceptable (source: European Commission / MDCG, "Language requirements for manufacturers — MDR", Rev. 3, August 2025, Slovenia entry, citing Article 33 of the Medical Devices Act, Official Gazette RS, No. 98/2009). The same table records Slovene across the other information categories for Slovenia — labelling, implant card and patient information, safety and performance information and the EU declaration of conformity among them. JAZMP's clinical-investigation guidance applies the same logic in the current Act: under Article 7(5) of ZMedPri-1, the labelling and instructions for use intended for the lay person, the implant-card information under Article 18(1) MDR, the synopsis of the investigation and the informed-consent documents must be provided in Slovenian (source: JAZMP, "Notification, review and conclusion of a clinical investigation"). Plan Slovene as the default for everything user-facing; treat the professional-use opening as an instructions-for-use point, not as a general English permission — Section 6 works this through.
- 5. A certificate of free sale that depends on a valid registration decision — JAZMP issues certificates only for devices entered in the Register of Medical Devices with a valid registration decision and, for MDR/IVDR devices, recorded in EUDAMED, on the prescribed form (Form 513-05), once registration decisions have become final (source: JAZMP, Certificate of Free Sale pages). Validity tracks risk: one year for Class I (MDR) / Class A (IVDR) devices; for higher-risk devices, as long as the notified body's certificate is valid. Professional guidance records issue within 30 days of a complete application (source: Lexology Q&A, above). An export plan therefore needs the national registration finished and final first.
- 6. Payment by standard, not by a single national device listing — medical devices may be paid for from compulsory health insurance in accordance with ZZVZZ and the Rules on Mandatory Health Insurance, and for each type of medical device a standard is set which provides the price, the material and the duration of use; distributors are obliged to make standard medical devices available without any co-payment, and a patient who needs or chooses a better or more expensive device co-pays the difference (source: Lexology, "Spotlight: medicine and medical device pricing and reimbursement in Slovenia", summarising ZZVZZ and the Rules on Mandatory Health Insurance). ZZZS publishes the operational lists behind that system — the list of medical devices with codes and prices, the list of contractual suppliers, and the lists by supplier (source: ZZZS, "Medical Devices" pages). A device can be fully registered and still have no funded route if it does not fit a standard, a code and a contractual supplier — Section 10 separates that gate from hospital procurement.
A note on institutions, because Slovenia splits responsibilities in a way that confuses first-time entrants: the Ministry of Health (Ministrstvo za zdravje — MZ) holds framework competence — legislation and oversight of the system. JAZMP, the Agency for Medicinal Products and Medical Devices, is the competent authority for devices in operation: the national registers of devices and of economic operators, EUDAMED actor validation for Slovenia-established operators, market surveillance, materiovigilance, classification procedures, clinical-investigation validation and assessment together with the Medical Ethics Committee of the Republic of Slovenia (Komisija za medicinsko etiko — KME RS), and certificates of free sale. ZZZS is the single purchaser for compulsory health insurance — it sets the device standards, keeps the lists of devices and contractual suppliers, and contracts with providers; it is not a device regulator. The certifying notified body may be established anywhere in the EU; whether any Slovenia-established body currently holds MDR or IVDR designation is to be verified in the European Commission's NANDO database when the route is chosen.