The licence is the gate
No SAHPRA product registration exists yet — the Section 22C establishment licence, with your devices and classes on its product list, is the legal right to import, distribute and sell.
Field guide · PDF
SAHPRA licenses establishments, not products. The Section 22C licence, the resident Authorised Representative, and an ISO 13485 timetable that hardens through 2028 — with product registration still to come, by phased call-up.
Sourced from SAHPRA's medical devices page, the SAHPGL-MD-07 licensing guideline, and the 2016 Regulations.
Enter your work email and we'll send the PDF straight to your inbox.
What's inside
Every figure in this guide traces to a named source — SAHPRA's medical devices page and licensing Q&A guideline (SAHPGL-MD-07), the 2016 Regulations in Gazette No. 40480, SAHPRA's ISO 13485 communication MD01-2025/2026_v2, and the gazetted fee schedules.
No SAHPRA product registration exists yet — the Section 22C establishment licence, with your devices and classes on its product list, is the legal right to import, distribute and sell.
Every licensed establishment appoints an Authorised Representative — a natural person resident in South Africa, holding a written mandate, with a CV on file and legal responsibility for compliance.
Verified ISO 13485 certification phases in from 2025 to 2028 — renewals first, then existing licence holders, amendments, and finally every new application.
Complete guide
Every section of the PDF on one page. Use the form above to keep a copy.
Last updated: November 2026. South Africa regulates medical devices and in vitro diagnostic medical devices (IVDs) through the South African Health Products Regulatory Authority (SAHPRA), an entity of the National Department of Health, under the Medicines and Related Substances Act, 1965 (Act No. 101 of 1965) and the Regulations relating to Medical Devices and IVDs, published in Government Gazette No. 40480 on 9 December 2016. The single most important fact for a foreign manufacturer in 2026 is this: SAHPRA licenses establishments today; it does not yet operate a general product-registration system for medical devices. Always confirm the current licence forms, fee gazette and ISO 13485 verification position on SAHPRA's own medical devices page before filing.
South Africa is the largest medical device market in sub-Saharan Africa and the natural entry point for the Southern African region: a population of roughly 63 million (Stats SA mid-year estimate for 2025, as reported in the market sources checked for this guide), a dual health system in which a well-resourced private sector serves the insured minority while the public sector serves the large majority, nine provincial health departments buying alongside national transversal contracts, and a private hospital market dominated by three groups — Netcare, Mediclinic and Life Healthcare. Market sources checked for this guide put the device market at about US$1.12 billion in 2025, with more than 90% of devices imported in the commonly cited estimates; a localisation study reported in October 2026, based on Stats SA supply and use tables, put the import share of total medical appliance supply at 80.8% in 2022. Either way, this is an import market. For a foreign manufacturer, however, South Africa is not a "register the product, then ship" market in the Kenyan or European sense — and planning it as if it were is the most common and most expensive misunderstanding.
Regulatory reality check up front — six features of the South African system that foreign manufacturers most often underestimate:
1. The establishment licence is the market gate today, not product registration. SAHPRA states plainly on its own medical devices page: "No medical device may be manufactured, distributed, imported, exported or sold without a valid SAHPRA medical device establishment licence," and, on the same page, "The registration process for medical devices is still in development." Section 22C(1)(b) of the Medicines Act is the operative provision: the company that manufactures, imports, exports, distributes or wholesales the device in South Africa must be licensed. A CE mark, an FDA clearance or a TGA listing does not replace that licence, and there is currently no SAHPRA product certificate to apply for in place of it.
2. Product registration is coming, in phases, starting from the top of the risk ladder — but no Call-Up Notice had been issued as at November 2025. Section 14 of the Medicines Act prohibits the sale of a device that has not been registered if SAHPRA has declared that device or class registrable. As at November 2025, SAHPRA had issued no such Call-Up Notice for medical devices (Legal500 2025 recap, based on SAHPRA reporting). What exists instead is a Medical Device Registration Feasibility Study, launched with an Expression of Interest in May 2024, testing the proposed registration process on Class C and Class D IVD and non-IVD products for HIV and tuberculosis. In October 2025 SAHPRA reported the study ongoing, with 15 applicants participating and 22 of a targeted 32 products received. SAHPRA expects a preliminary report in the 2025/26 financial year, and the study is described as the preliminary step before a Call-Up Notice for a pilot Class D call-up. Doğrulanacak: whether a Call-Up Notice, a pilot Class D call-up or the feasibility study's preliminary report has been published between that reporting and your filing date — check SAHPRA's medical devices page and stakeholder communications first.
3. The licence is class-specific and product-list-specific. An applicant must list all the medical devices it manufactures, distributes or wholesales, and the licence is granted for the classes listed. Adding a new class, a new product category or a new activity is an amendment, not an informal extension — you cannot lawfully import a Class D implant under a licence that only lists Class B consumables.
4. One narrow exemption exists — and it is narrower than it sounds. A manufacturer, distributor or wholesaler of a non-sterile, non-measuring Class A medical device is exempt from licensing, per SAHPRA position statement 9.106 "Class A medical devices," as stated on SAHPRA's medical devices page. Sterile Class A, measuring Class A, and every Class B, C and D activity sit inside the licensing net. Do not build a launch plan on the exemption without confirming your exact device falls within it.
5. ISO 13485 is moving from declaration to hard prerequisite, on a published phased timetable through 2028. At application, a company declares the status of its quality management system. On renewal, manufacturers and distributors must already provide ISO 13485 evidence. SAHPRA's stakeholder communication MD01-2025/2026_v2 (18 September 2025), "ISO 13485 Certificate as a prerequisite for the approval of a Medical Device Establishment Licence," then phases verification in: renewals from 1 June 2025, existing licence holders submitting certificates for verification from 1 April 2026, licence amendments from 1 June 2027, and all new applications from 1 April 2028 — in each case a valid ISO 13485:2016 certificate in the name and at the address of the South African licensed establishment, issued by a conformity assessment body (CAB) recognised by SAHPRA. A foreign parent's ISO certificate in a different legal name and at a different address does not automatically satisfy a requirement written for the South African establishment.
6. The Authorised Representative is a natural person resident in South Africa — not a company mailbox. Every establishment that applies for a SAHPRA licence must appoint an Authorised Representative who is a natural person based in South Africa, with a written mandate to represent the establishment, and who is responsible for adherence to the law, regulations and guidelines. The representative's curriculum vitae forms part of the application, and the representative initials and signs the application. Choose this person as your regulatory accountable officer in South Africa, not as a formality.
A note on old and recycled information: the web is full of pages that describe a South African "medical device registration" with a single fee, a 6–8 week approval and a product certificate, often recycling Medicines Control Council (MCC)-era forms and directives (including references to EC Directive 93/42/EEC). The MCC no longer exists — SAHPRA replaced it under the Medicines and Related Substances Amendment Act 14 of 2015 — the EU directives referenced have themselves been replaced by the EU MDR and IVDR, and the 6–8 week figure that circulates refers, where it refers to anything current, to establishment licence processing, not to a product registration that does not yet exist. File to the SAHPRA system described below, and treat any source that promises a SAHPRA product registration certificate today as a red flag about that source.
| Instrument | What it does |
|---|---|
| Medicines and Related Substances Act, 1965 (Act No. 101 of 1965), as amended | Primary statute. Section 22C(1)(b) requires a licence to manufacture, import, export, distribute or wholesale medical devices and IVDs. Section 14 provides for registration of devices once SAHPRA declares a device or class registrable — the legal basis for the future call-up system. Section 22D and the fee regulations govern licensing and retention. |
| Medicines and Related Substances Amendment Act 14 of 2015 | Created SAHPRA in place of the Medicines Control Council and expanded the regulatory mandate expressly to medical devices and IVDs — the foundation of the current establishment-licensing regime and of the future registration regime. |
| Regulations relating to Medical Devices and IVDs (Government Gazette No. 40480, 9 December 2016) | The operative device regulations: establishment licensing (notably Regulation 5), classification of devices and IVDs, the register, amendments, clinical investigations, adverse event reporting and vigilance (Regulation 17), advertising, labelling of devices (Regulation 22) and instructions for use (Regulations 23–24), and custom-made devices. |
| Call-up notice, Government Gazette No. 40637, 24 February 2017 | Required manufacturers and distributors to apply for a SAHPRA licence within six months of publication, and wholesalers within twelve months — the start of compulsory establishment licensing. Since 31 March 2020 an acknowledgement letter on submission of an application may no longer be used in lieu of a licence. |
| SAHPRA Guideline on Questions and Answers: Licensing of Medical Device Establishments (SAHPGL-MD-07, version dated 21 May 2025) | The working Q&A guideline for new, amendment and renewal applications: submission format, cover letter, application forms, Authorised Representative CV, Quality Manual / Site Master File, proof of payment, costs, and quality-manual expectations. Use the current version on SAHPRA's website at filing. |
| SAHPRA Guideline for Classification of Medical Devices and IVDs | SAHPRA's classification guidance for non-IVD devices and IVDs, aligned to the Act and the 2016 Regulations — the reference for the A–D rules in Section 2. |
| SAHPRA Communication MD01-2025/2026_v2 (18 September 2025) — ISO 13485 prerequisite | Phases ISO 13485:2016 certification in as a verified prerequisite for establishment licences through April 2028 (Section 9). The certificate must be in the South African establishment's name and address and issued by a SAHPRA-recognised CAB. |
| Regulations Regarding Fees Payable under the Medicines Act (fee gazettes, including the dispensation gazetted 12 February 2025) | Sets licence application, renewal, amendment and annual retention fees. SAHPRA states that the current medicine fee structure for licensing of establishments is adopted for medical device establishments. Figures change by gazette — see Section 7.2 and verify at filing. |
| Preferential Procurement Policy Framework Act (PPPFA) and 2022 Preferential Procurement Regulations; Broad-Based Black Economic Empowerment (B-BBEE) Act | Not device law, but decisive for public selling: preference-point scoring (80/20 and 90/10 systems) in national and provincial tenders, in which the bidder's B-BBEE level earns points (Section 8). |
| National Health Insurance Act (signed May 2024) | Will, over time, centralise public purchasing toward a single government buyer. Implementation timing and device-procurement consequences: doğrulanacak — plan for today's nine-province system, monitor NHI implementation. |
South Africa is not in the European Union, and there is no EU MDR route into this market: CE marking is evidence SAHPRA and South African buyers respect, and it is one of the recognised-jurisdiction approvals SAHPRA says its future registration system will rely on — but it confers no South African market right by itself. SAHPRA became an Affiliate Member of the Medical Device Single Audit Program (MDSAP) in April 2025; MDSAP participation is not mandatory for South African entry, but SAHPRA can request MDSAP audit reports and certificates from participating manufacturers, which makes a clean MDSAP/ISO 13485 file increasingly useful evidence.
South Africa classifies devices and IVDs in four risk classes aligned to the GHTF/IMDRF model. The class drives whether the licensing exemption can apply, what supporting evidence the establishment licence application must carry for the products listed, how demanding tender evaluators will be, and — once call-up begins — the order in which products are likely to be called for registration, highest risk first.
| Class | Risk | Examples (from SAHPRA-aligned summaries) |
|---|---|---|
| A | Low | Bandages and basic wound dressings, non-invasive low-risk devices |
| B | Low–moderate | Powered wheelchairs, electrotherapy devices |
| C | Moderate–high | Diagnostic X-ray equipment, implantable hearing aids |
| D | High | Implantable pacemakers, insulin pumps and other life-sustaining / highest-risk devices |
For IVDs, SAHPRA's classification guidance applies a parallel A–D ladder driven by public-health and personal risk: Class A is no public-health risk or low personal risk; Class B is low public-health risk or moderate personal risk; Class C is moderate public-health risk or high personal risk; Class D is high public-health risk — the category that captures, for example, assays detecting transmissible agents in blood, tissues and organs that can cause serious disease. The IVD rules in SAHPRA's guidance address, in sequence, transmissible-agent detection, blood grouping / tissue typing antigens and antibodies, other defined high-consequence markers, self-testing devices (generally Class C unless an exception applies), control materials, and general reagents and instruments, with a residual rule placing IVDs not covered by an earlier rule in Class B. This is why HIV and tuberculosis IVDs sit at the centre of the registration feasibility study: they are the Class C/D, high-public-health-risk end of the IVD ladder.
Practical classification notes:
1. Duration, invasiveness and activity move a device up. SAHPRA's framework weighs intended use, the probability and severity of harm, and duration of contact — transient (less than 60 minutes), short-term (up to 30 days) and long-term (more than 30 days) — together with invasiveness, sterility, measuring function and whether the device is active. Document your classification rationale against SAHPRA's classification guideline; in a dispute, SAHPRA's determination prevails.
2. Sterile or measuring Class A is not exempt Class A. The licensing exemption is confined to non-sterile, non-measuring Class A. A sterile dressing set or a Class A device with a measuring function pulls the establishment back into licensing.
3. Your EU or GHTF class is a starting point, not the answer. South Africa's classes use the same A–D letters as the GHTF model, which are not the EU MDR's I / IIa / IIb / III. Map the device deliberately; do not copy a European class label into a South African product list.
4. Classification decides your future call-up exposure. SAHPRA has signalled phased, transitional call-up by risk class and condition, starting with the highest risk. A Class D portfolio should be prepared for registration first — dossier, CAB conformity assessment and recognised-jurisdiction approvals assembled now, not when the notice publishes.
South African market entry is a structure, not a single application. Three roles must be understood separately:
1. Authorised Representative — mandatory, natural person, resident in South Africa. SAHPRA's medical devices page states that establishments applying for a licence "must appoint an Authorised Representative who must be a natural person based in South Africa" and that "[t]he representative is responsible for adherence to the law, regulations and guidelines." The licensing Q&A guideline (SAHPGL-MD-07) defines the AR as a natural person resident in the Republic who has the written mandate to represent the manufacturer, importer, distributor, wholesaler, retailer or service provider, and acts on its behalf. The AR's CV is submitted with the application; the AR signs the cover letter and initials/signs the application forms. For a foreign manufacturer, the AR is normally an officer or appointee of your South African distributor or subsidiary — but the mandate, competence and continuity of this individual are regulatory facts, not administrative decoration. If the AR leaves, the establishment must regularise the appointment by amendment; a licence whose AR has quietly departed is a compliance defect.
2. The licensed distributor (importer) — the normal entry vehicle for foreign manufacturers. The distributor licence covers import, export and distribution. Your South African distributor must hold this licence, for the classes of device you will supply, with your devices on its product list, before the first commercial import. Verify the licence itself — number, holder name, address, activities and product classes — do not accept a company profile or a tender history as evidence. One structural warning belongs in every distribution agreement: the licence and the product listing sit with the distributor's establishment. If the relationship ends, your products come off that establishment's list and must be added to a new licensed establishment by amendment before lawful import resumes. Agree notice, handover and listing-transfer mechanics in writing at appointment, not at exit.
3. Wholesaler — a separate, narrower licence. The wholesaler licence covers storage, transportation and delivery; wholesalers buy from licensed manufacturers and distributors and do not import. If your route to market uses a wholesaler layer (common for reaching smaller hospitals, pharmacies and rural facilities), that wholesaler needs its own licence, and your distributor's licence does not cover it.
4. Manufacturer licence — wider than factory production. The manufacturer licence covers manufacture, pack, label, service, import and export. Local assembly, labelling, refurbishment or servicing operations in South Africa can amount to manufacturing activity; a foreign group setting up a South African labelling or service operation should expect to need this licence type, not a distributor licence.
5. No own subsidiary is required. A foreign manufacturer can enter entirely through an independent licensed distributor plus its AR. A South African subsidiary becomes relevant for tender strategy, B-BBEE structuring, local servicing and control — commercial reasons, not a licensing precondition.
Three licence types exist, per SAHPRA: manufacturer (manufacture, pack, label, service, import, export), distributor (import, export, distribute) and wholesaler (storage, transportation, delivery). The application, per SAHPRA and the SAHPGL-MD-07 Q&A guideline, comprises:
Applications are submitted to SAHPRA's medical devices administration (the mdadmin channel named in SAHPRA process materials). Screening confirms completeness, the file is referred for technical evaluation, and the licence issues after approval and fee confirmation. The licence is valid for five years, subject to annual retention fees and to renewal — with ISO 13485 evidence required at renewal for manufacturers and distributors. Amendments (new classes or products, changed premises, changed AR, changed conditions) are formal applications with their own fee, not notifications by email.
SAHPRA's stated plan is a phased Registration Call-Up Plan: devices will be called up sequentially by type and class, prioritising the highest risk, and registration will implement reliance pathways based on verification of registration in Australia, the United States, the European Union, Brazil, Canada or Japan, and/or WHO prequalification of IVDs. The feasibility study running since May 2024 is the proving ground: voluntary, confined to licensed establishments, targeting 32 HIV/TB-related products (IVD tests, condoms and TB-screening X-ray devices among them), split between locally manufactured and imported products, with conformity assessment by accredited CABs rather than direct SAHPRA assessment. As at the October/November 2025 reporting, 15 applicants were participating, 22 products had been received (18 product applications from 12 applicants were reported for Q4 2024/25, categorised as Class C and Class D IVD and non-IVD HIV/TB products), no Call-Up Notice had been issued, and a preliminary report was expected in the 2025/26 financial year, ahead of a possible pilot Class D call-up.
What this means in practice for a 2026 entrant: (a) you cannot obtain, and no consultant can obtain for you, a general SAHPRA product registration today — anyone selling one is selling Gate 1 paperwork under a Gate 2 label; (b) your recognised-jurisdiction approvals (CE marking under EU MDR/IVDR, FDA, TGA, Health Canada, PMDA, ANVISA) are the raw material of the future reliance route — keep them current, keep the technical file that supports them complete and consistent, and keep the South African product list aligned to them; and (c) Class D and high-risk Class C manufacturers should build the registration dossier now, because call-up, when it comes, will come for them first and will run through CAB conformity assessment.
Step 1 — Confirm the product is a device (or IVD) and classify it. Apply the Medicines Act definition and SAHPRA's classification guideline; record the A–D classification rationale, including the IVD rules for diagnostics. Check the narrow Class A exemption honestly: non-sterile and non-measuring, or the exemption does not apply.
Step 2 — Choose the South African structure. Independent licensed distributor, wholesaler layer, local manufacturing/servicing presence, or own subsidiary — and confirm which licence type each party needs (Section 3). If a partner claims to be licensed, obtain and verify the licence and its product-class scope.
Step 3 — Appoint the Authorised Representative. Written mandate to a named natural person resident in South Africa; obtain the AR's CV; agree the AR's duties (SAHPRA correspondence, vigilance, licence maintenance) and the succession/amendment procedure if the person changes.
Step 4 — Build the licence file. Cover letter; Forms 6.21 / 6.22 / 6.26 in Excel and signed PDF; complete product list with classes and, where the form requires it, nomenclature codes (doğrulanacak: current form fields at filing); Free Sale Certificate for Class B products; recognised-jurisdiction approval evidence for Class C and D products; Quality Manual / Site Master File and QMS declaration; ISO 13485 certificate or CAB agreement evidence as the applicable phase requires (Section 9).
Step 5 — File with SAHPRA and pay the gazetted fee. Submit through SAHPRA's medical devices administration channel, attach bank proof of payment for the correct licence-type fee from the current fee gazette (Section 7.2), and keep the full submission set — the file you submitted is the baseline for every later amendment.
Step 6 — Answer screening and evaluation queries completely. Incomplete product lists, mismatched company names or addresses across the ISO certificate, Free Sale Certificate and application, and unsigned forms are the routine causes of delay. Respond as one consistent file, through the AR.
Step 7 — Receive the establishment licence and diarise its life. Five-year validity; annual retention fee cycle (SAHPRA issues retention-fee notifications for the year, and licensing summaries place the annual retention payment in the June cycle — doğrulanacak: the current retention due date and amount in SAHPRA's latest retention-fee communication at filing); renewal with ISO 13485 evidence; amendments before any change to classes, products, premises or AR.
Step 8 — Set up import mechanics. Customs registration and a clearing agent for the importing establishment; correct tariff classification and duty/VAT treatment per HS code (doğrulanacak with SARS / your clearing agent for your exact codes — rates are code-specific); and, for regulated categories beyond SAHPRA (for example radiation-emitting devices and electrical medical equipment), confirm any additional authority or compulsory-specification requirements that apply to your product type before shipment (doğrulanacak per product).
Step 9 — Prepare the tender and private-channel dossiers in parallel. CSD registration, SARS tax compliance status, B-BBEE certificate or affidavit, and the SAHPRA licence evidence that public tenders require (Section 8). Supplier qualification by the private groups runs on the same licence evidence.
Step 10 — Go live with post-market systems before the first sale. Vigilance and adverse-event reporting under Regulation 17, complaint handling, recall execution, distribution records for traceability, and the ISO 13485 verification calendar must be operating from day one (Section 9) — and the regulatory watch for the first Call-Up Notice must have a named owner.
The 2016 Regulations prescribe labelling for medical devices (Regulation 22) and instructions for use for devices and IVDs (Regulations 23 and 24). The labelled information must identify the device and its manufacturer, and the IFU must support safe use of the device as registered/listed with the establishment. Two practical rules follow:
Advertising of devices is also regulated under the 2016 Regulations (Regulation 21, compliance and advertising) — claims in South African marketing must stay within the device's intended purpose and the evidence held. Do not market a device as "SAHPRA registered" while only an establishment licence exists: the licence authorises the establishment and its listed activities; it is not a product approval, and overstating it misleads buyers and tenders alike.
| Step | Typical time | Notes |
|---|---|---|
| Establishment licence application (complete file) | ~6–8 weeks | The processing range quoted in SAHPRA-aligned summaries for a complete application; screening queries and incomplete product lists extend it. It is a licence timeline, not a product-registration timeline. |
| ISO 13485 certification of the South African establishment (if not already held) | Build 4–9 months into the plan | Audit scheduling with a SAHPRA-recognised CAB is the long pole for renewals and, from April 2028, for all new applications; doğrulanacak by CAB capacity at filing |
| Licence amendment (new class/product, premises, AR) | Weeks, file-dependent | Required before trading in the amended scope; doğrulanacak: current SAHPRA service standard at filing |
| Licence renewal | File before expiry of the 5-year term | Renewal requires ISO 13485 evidence for manufacturers and distributors |
| Product registration | Not yet available | Feasibility study ongoing at last reporting (Oct/Nov 2025); preliminary report expected 2025/26 FY; Call-Up Notice to follow — no filing timeline exists until SAHPRA publishes the call-up plan |
The planning error to avoid is arithmetic on the wrong gate: a launch plan that adds "8 weeks for SAHPRA registration" has confused the establishment licence with a product approval and will misstate both the regulatory position and the tender evidence.
SAHPRA states that the current medicine fee structure for licensing of establishments is adopted for medical device establishments, and fees are set by gazette and revised periodically — the fee dispensation gazetted on 12 February 2025 superseded the December 2020 schedule for the categories it covers. The table below shows the two reference points found in the sources checked. Treat every figure as doğrulanacak against the current fee gazette and the SAHPRA invoice at filing.
| Fee | 2020 schedule (Gazette, 22 Dec 2020) | February 2025 dispensation (as reported in SAHPRA fee-review materials and secondary summaries) |
|---|---|---|
| New manufacturer licence (Section 22C(1)(b)) | R25,200 | R26,200 |
| New distributor licence (import, export, distribute) | R15,000 | R15,600 |
| New wholesaler licence | R15,000 | R15,600 |
| Annual retention fee | R4,200 (SAMED 2021 summary) | R4,400 (secondary summary of the 2025 regulations) — doğrulanacak |
| Licence amendment | R5,300 (SAMED 2021 summary) | Doğrulanacak in the current gazette |
| Renewal | Per the licensing fee schedule | SAHPRA fee-review materials show renewal figures above the new-application figures for licence categories — doğrulanacak by licence type at filing |
Any third-party figure quoted in dollars, or a single "registration fee" such as the R84,000 that appears in one regulations-library summary, could not be reconciled to SAHPRA's own fee materials in the sources checked and is not used in this guide — doğrulanacak items must be resolved from the gazette, never from aggregator pages. Beyond SAHPRA fees, budget for: AR and local regulatory support (commercial); ISO 13485 certification and surveillance audits by a recognised CAB for the South African establishment (commercial, and now timetable-critical); CSD/tender administration; customs duty and VAT by HS code; and the internal cost of keeping the product list, labels and foreign approvals aligned. As elsewhere in Africa, the state fee is the smallest line in the launch — the expensive failures are trading outside the licence's listed scope and letting retention or ISO verification lapse.
A licence makes supply lawful; it does not make anyone buy. South Africa is two markets sharing one licence:
1. The public sector — nine provincial systems plus transversal contracts. The national Department of Health and the nine provincial health departments (Gauteng, KwaZulu-Natal, Western Cape, Eastern Cape, Limpopo, Mpumalanga, Free State, North West and Northern Cape) procure under the Public Finance Management Act and National Treasury supply-chain rules. Large, high-volume categories are often bought through transversal (national) contracts against which provinces and hospitals place call-offs; the rest runs as provincial and hospital-level tenders published on the National Treasury eTenders portal (etenders.gov.za) and provincial portals. The mechanics that decide eligibility are unglamorous and absolute: registration on the Central Supplier Database (CSD), a current SARS tax compliance status, company registration (CIPC) evidence, signed standard bidding documents, and — specifically for devices — evidence of a valid SAHPRA medical device establishment licence, which SAHPRA itself notes is required to be eligible to bid for national and provincial tenders. Scoring runs under the PPPFA preference-point systems (80/20 for lower-value, 90/10 for higher-value contracts), in which the bidder's B-BBEE level earns preference points: your local partner's empowerment structure is therefore a commercial variable in every public bid, not a back-office detail. For IVDs and laboratory products, the NHLS runs its own procurement and is a volume channel in its own right.
2. The private sector — three groups set the pace. Netcare, Mediclinic and Life Healthcare together control the large majority of private hospital capacity (market sources put the three groups at roughly 80% of the private hospital market). They procure commercially and qualify suppliers centrally: the SAHPRA licence, ISO 13485 position, service and maintenance capability, and total-cost files are the entry conversation. Timelines are faster than public tenders; the compliance questions are the same.
3. The NHI horizon. The National Health Insurance Act, signed in May 2024, points toward eventual centralisation of public purchasing under a single buyer, which would narrow and standardise the supplier pool. It does not change today's filing or tender mechanics — doğrulanacak: implementation regulations and procurement transition as they publish — but it raises the value of being an established, licensed, ISO-certified supplier before any centralisation begins.
Practical points: public specifications are fixed when a tender publishes, so clinical evaluation and health-economics work must happen before the notice, through the buyer's announced engagement channels; provincial payment discipline varies, so price the payment terms as carefully as the unit price; and never submit a bid listing a device outside your establishment licence's product list and classes — tender evaluation checks the licence, and an out-of-scope listing discovered after award puts the contract itself at risk.
Structure and representative
Establishment licence application (per SAHPGL-MD-07, current version)
Licence lifecycle
Import and tender readiness
Post-market
1. Budgeting for a product registration that does not exist — or paying a consultant for one. The live gate is the establishment licence. Product registration is in a feasibility study. Any "SAHPRA registration certificate in 6–8 weeks" offer is mislabelled licensing paperwork.
2. Assuming CE marking or FDA clearance is a South African market right. They are evidence — for the licence file (Classes C/D), for tenders, and for the future reliance route. The Section 22C licence is the right to trade.
3. Trading outside the product list. The licence covers the classes and devices listed. Importing an unlisted class or model before an amendment is granted is unlicensed trading, however valid the rest of the licence is.
4. Building on the Class A exemption without checking its edges. Non-sterile and non-measuring only. Sterile or measuring Class A, and all of B/C/D, are licensed activities.
5. Treating the Authorised Representative as a formality. The AR must be a named, resident natural person with a written mandate, a filed CV and legal responsibility for compliance. An AR who has left, or who cannot operate vigilance, is a licence defect.
6. Holding ISO 13485 in the wrong name. From the 2025–2028 phases, the certificate SAHPRA verifies is the South African establishment's, at its address, from a recognised CAB. A parent-company certificate does not automatically answer it.
7. Missing retention or renewal. Five-year validity with annual retention is the same trap as elsewhere, with the same consequence: no valid licence, no lawful trade, no tender eligibility.
8. Confusing distributor and wholesaler licences. Wholesalers cannot import. A route to market built on a wholesaler importing directly is unlawful at its first step.
9. Ignoring B-BBEE and CSD until the tender publishes. Preference points and database registration decide public bids before the technical file is opened. Structure the local partnership with tenders in mind from day one.
10. Calling the licence an approval in marketing. "SAHPRA licensed establishment" is the accurate claim. "SAHPRA registered/approved device" is not — yet — a claim anyone can truthfully make for a general device, and buyers' compliance teams know it.
11. No plan for call-up. Class D and high-risk Class C portfolios will be called first. Manufacturers without a CAB-ready technical file and current recognised-jurisdiction approvals when the notice publishes will be at the back of a phased queue.
12. Copying another country's vigilance timelines into the South African SOP. Regulation 17 sets the duties; the operative reporting timeframes must be taken from current SAHPRA guidance (doğrulanacak), not from EU or Kenyan templates.
Who regulates medical devices in South Africa? SAHPRA — the South African Health Products Regulatory Authority, an entity of the National Department of Health — through its Medical Devices Unit, under the Medicines and Related Substances Act 101 of 1965 and the 2016 Regulations relating to Medical Devices and IVDs (Gazette No. 40480).
Do I need to register my medical device with SAHPRA? Not yet — because, for general devices, you cannot: SAHPRA states that the registration process is still in development, and as at November 2025 no Call-Up Notice under Section 14 had been issued. What you must have today is a valid Medical Device Establishment Licence (Section 22C(1)(b)) held by the South African establishment that manufactures, imports, distributes or wholesales your device, with your devices and classes on its product list. Registration will arrive by phased call-up, highest risk first, following the feasibility study now under way.
Do I need a local representative? Yes. Every licensed establishment must appoint an Authorised Representative who is a natural person based in South Africa, holding a written mandate and responsible for adherence to the law, regulations and guidelines. A foreign manufacturer enters through a South African licensed establishment (usually its distributor); it does not need its own subsidiary, but the AR appointment is not optional.
How are devices classified? In four risk classes, A (low) to D (high), aligned to the GHTF/IMDRF model — bandages in A, powered wheelchairs in B, diagnostic X-ray equipment in C, implantable pacemakers in D. IVDs follow a parallel A–D ladder driven by public-health and personal risk, with high-public-health-risk assays (for example blood-borne transmissible-agent detection) in Class D. Note that A–D are not the EU MDR's I–III classes.
Is any device exempt from licensing? One narrow category: establishments dealing only in non-sterile, non-measuring Class A devices are exempt, per SAHPRA position statement 9.106. Sterile or measuring Class A devices, and all Class B, C and D activities, require a licence.
How much does it cost? Per the fee schedules in the sources checked (doğrulanacak against the current gazette at filing): a new distributor or wholesaler licence was R15,000 under the 2020 schedule and R15,600 under the February 2025 dispensation; a manufacturer licence R25,200 rising to R26,200; annual retention was R4,200, reported as R4,400 under the 2025 dispensation; amendments were R5,300 under the earlier schedule. Add AR, ISO 13485 certification, tender and customs costs.
How long does approval take, and how long is a licence valid? A complete establishment licence application is quoted at about 6–8 weeks in SAHPRA-aligned summaries. The licence is valid for five years, subject to annual retention fees, and renewal requires ISO 13485 evidence for manufacturers and distributors. There is no product-registration timeline until SAHPRA publishes the call-up plan.
My device is CE-marked. Does that help? Yes — as evidence, in three places: the licence file for Class C and D products (market approval in a recognised jurisdiction, which includes the EU), tender and private-group qualification, and the future registration system, for which SAHPRA has said it will implement reliance pathways based on registration in Australia, the US, the EU, Brazil, Canada and Japan, and/or WHO prequalification of IVDs. It does not replace the establishment licence.
Can my distributor also handle wholesaling and the AR role? A distributor licence covers import, export and distribution; wholesaling (storage, transport, delivery, onward sale without importing) is a separate licence, though one group may hold both through appropriately licensed establishments. The AR is a named natural person — often from within the distributor — appointed by written mandate, with a CV filed in the licence application.
What changes in 2026–2028 that I should plan for now? Two things. First, ISO 13485 verification: existing licence holders submit certificates for verification from 1 April 2026, amendments require verified certification from 1 June 2027, and all new applications require it from 1 April 2028. Second, product registration: the feasibility study's preliminary report is expected in the 2025/26 financial year, to be followed by a Registration Call-Up Plan and a likely pilot Class D call-up. Keep the technical file, CAB relationships and recognised-jurisdiction approvals ready.
What are my duties after licensing? Pay annual retention; renew at five years with ISO 13485 evidence; amend the licence before changing classes, products, premises or the AR; operate vigilance and adverse-event reporting under Regulation 17 through the AR; maintain complaint, recall and traceability systems; keep labels and marketing within the listed device and its intended purpose; and monitor SAHPRA communications for the call-up that will add product registration to the licence.
Is there a public tender channel I should plan around? Yes — a large one. Nine provincial health departments plus national transversal contracts, published on etenders.gov.za, gated by CSD registration, SARS tax compliance, B-BBEE preference points under the PPPFA, and evidence of a valid SAHPRA establishment licence; the NHLS procures IVDs and laboratory products at scale; and the private market is led by Netcare, Mediclinic and Life Healthcare, which qualify suppliers centrally on the same licence and quality evidence.
Items marked "doğrulanacak" in this guide were not verifiable to a primary-source figure or text at the time of writing and must be confirmed against current SAHPRA guidance, the current fee gazette and SAHPRA's medical devices page before filing or budgeting. Fee figures reproduce the gazetted schedules and SAHPRA fee-review materials cited above and must be confirmed at filing. This guide is general information for market-entry planning, not legal advice.
This guide is market-access intelligence, not regulatory or legal advice. SAHPRA timelines, licence types, and import rules are product-specific — professional review is required before business use.
Market Access Strategy Session
A Market Access Strategy Session maps your SAHPRA route — establishment and licence-type selection, the Authorised Representative appointment, ISO 13485 verification timing, and call-up readiness.
Professional review is required before business use.
Back to ghmap.io Questions? Talk to our team